 Class 2 Device Recall EXTREMITY PACKsee related information
Date Initiated by FirmJuly 10, 2015
Date PostedSeptember 03, 2015
Recall Status1 Terminated 3 on April 12, 2017
Recall NumberZ-2647-2015
Recall Event ID 71721
Product Classification Orthopedic tray - Product Code OJH
ProductEXTREMITY PACK 4/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. .
Code Information Code:9002615 Lots: 140915322 exp. 10/31/15 141015728 exp. 11/30/15 141015728 exp. 11/30/15 141216308 exp. 12/31/15 150116556 exp. 1/1/16 150518786 exp. 5/31/16 
FEI Number 2648727
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
Manufacturer Reason
for Recall
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
FDA Determined
Cause 2
Packaging
ActionCustomed issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" notification letter dated July 10, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customer.
Quantity in Commerce285
DistributionDistributed Only in Puerto Rico.