Class 2 Device Recall LAPCHOLECYSTECTOMY SURGICAL PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall LAPCHOLECYSTECTOMY SURGICAL PACK see related information
Date Initiated by Firm July 10, 2015
Date Posted September 03, 2015
Recall Status1 Terminated 3 on April 12, 2017
Recall Number Z-2688-2015
Recall Event ID 71721
Product Classification laparoscopy kit - Product Code FDE
Product LAP-CHOLECYSTECTOMY SURGICAL PACK
Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
.
Code Information Code:900576A Lots: 140915210 exp. 5/31/15 140915495 exp. 10/31/15 141216351 exp. 12/31/15 141216447 exp. 12/31/15 150116947 exp. 3/31/16 150518827 exp. 5/31/16 150518911 exp. 5/31/16 150519007 exp. 5/31/16 150619186 exp. 6/30/16 150619186 exp. 6/30/16 
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
Manufacturer Reason
for Recall
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
FDA Determined
Cause 2
Packaging
Action Customed issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" notification letter dated July 10, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customer.
Quantity in Commerce 320
Distribution Distributed Only in Puerto Rico.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.