Class 2 Device Recall Sterile, CV0654A Pediatric Heart surgical kit |
|
Date Initiated by Firm |
April 22, 2015 |
Date Posted |
October 07, 2015 |
Recall Status1 |
Terminated 3 on March 28, 2017 |
Recall Number |
Z-0100-2016 |
Recall Event ID |
71904 |
Product Classification |
General and plastic surgery - Product Code FSY
|
Product |
regard Item Number 800565001, Sterile, CV0654A - Pediatric Heart - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a surgical procedure |
Code Information |
Lot numbers 19532, 23789, 24399, and 24478. |
Recalling Firm/ Manufacturer |
Resource Optimization & Innovation Llc 2909 N Neergard Ave Springfield MO 65803
|
For Additional Information Contact |
Robert W. Callahan 314-364-6561
|
Manufacturer Reason for Recall |
The kits contain an equipment glove component which may contain splits or holes compromising the sterility.
|
FDA Determined Cause 2 |
Process control |
Action |
ROi initiated a recall of the custom kits via e-mail to their customers on April 22, 2015. The e-mail included their recall letter dated April 17, 2015. and the supplier's recall letter dated April 16, 2015. The customer was requested to discard the recalled component from their procedure tray at the time of set up and replace it with sterile stock. Customers were instructed to contact any ROi custom pack solutions manager with any concerns. The recalling firm issued a second e-mail notification on 10/19/2015 to the recall coordinator of the distributor. The notice informed them they had been receiving uncorrected product and reminded them to discard the defective component at the time the kit is set up for surgery. Any future distribution will be labeled with an alert to remove the defective component until the kits contain acceptable light handle covers.
For questions regarding this recall call 314-364-6561. |
Quantity in Commerce |
57 kits |
Distribution |
Distribution was made to LA, MS, and TX. |