| | Class 2 Device Recall Dimension Vista System (MG) |  |
| Date Initiated by Firm | September 30, 2015 |
| Date Posted | November 25, 2015 |
| Recall Status1 |
Terminated 3 on September 01, 2016 |
| Recall Number | Z-0339-2016 |
| Recall Event ID |
72399 |
| 510(K)Number | K061655 |
| Product Classification |
Photometric method, magnesium - Product Code JGJ
|
| Product | Dimension Vista System (MG) ; Flex reagent cartridge Catalog number K3057, SMN 10445158. For the quantitative measurement of magnesium in human serum, plasma and urine on the Dimension Vista System. |
| Code Information |
Lot Number 15063BA |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 500 Gbc Dr Ms 514 PO BOX 6101 Newark DE 19702-2466
|
| For Additional Information Contact | 800-441-9250 |
Manufacturer Reason for Recall | Erroneous low results on a small number of reagent wells. There is the potential for under-recovery on certain wells for quality control (QC) and patient sample results. The under-recovery ranged from -0.3 mg/dL [0.12 mmol/L] to -1.6 mg/dL [0.66 mmol/L]; equivalent to a maximum negative bias of -88%. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | An Urgent Medical Device Recall dated September 2015 was provided to all Dimension Vista MG customers who received the affected lot to notify them of the issue, the potential risk to health, and actions to be taken by customers.
1. An Urgent Medical Device Recall was provided to all customers who received the affected lot to notify them of the following:
Siemens has received customer complaints for Dimension Vista MG lot 15063BA, regarding sporadic depressed serum/plasma and QC results.
Discontinue use and discard their remaining inventory of the affected lot.
Review the letter with their Medical Director. Siemens will replace any unused inventory of the affected lot. |
| Quantity in Commerce | 3601 |
| Distribution | Distributed US (nationwide) including DC and the states of IL, CA, GA, LA, TX, TN, FL, SC, IA, TN, DE, NJ, WA, MS, KY, NY, WI, OH, OH, MO, UT, KS, VA, AR, WV, MA, PA, AL, CO, IN, MT, NM, ME, AZ, NM, MO, and NC, and the country of Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = JGJ
|
|
|
|