|
Class 2 Device Recall BrightView |
|
Date Initiated by Firm |
December 17, 2015 |
Create Date |
February 26, 2016 |
Recall Status1 |
Terminated 3 on July 02, 2018 |
Recall Number |
Z-1020-2016 |
Recall Event ID |
73010 |
510(K)Number |
K062298
|
Product Classification |
System, tomography, computed, emission - Product Code KPS
|
Product |
BrightView are gamma cameras; Designed for single or dual detector nuclear imaging. |
Code Information |
Model # 882480; Serial #'s: 4000002, 4000002.75, 4000003, 4000004, 4000006, 4000007, 4000008, 4000009, 4000010, 4000011, 4000013, 4000030, 4000031, 4000032, 4000033, 4000034, 4000035, 4000036, 4000037, 4000038, 4000039, 4000040, 4000041, 4000043, 4000044, 4000045, 4000046, 4000047, 4000048, 4000049, 4000050, 4000051, 4000052, 4000053, 4000054, 4000055, 4000056, 4000057, 4000058, 4000059, 4000060, 4000061, 4000062, 4000064, 4000065, 4000066, 4000068, 4000070, 4000071, 4000072, 4000073, 4000074, 4000075, 4000076, 4000077, 4000078, 4000079, 4000080, 4000081, 4000082, 4000083, 4000086, 4000087, 4000088, 4000089, 4000090, 4000091, 4000093, 4000094, 4000095, 4000096, 4000099, 4000100, 4000102, 4000103, 4000104, 4000105, 4000106, 4000107, 4000108, 4000109, 4000110, 4000111, 4000112, 4000113, 4000114, 4000115, 4000116, 4000117, 4000118, 4000119, 4000120, 4000121, 4000122, 4000123, 4000124, 4000125, 4000126, 4000127, 4000128, 4000129, 4000130, 4000132, 4000133, 4000134, 4000135, 4000136, 4000137, 4000138, 4000139, 4000140, 4000141, 4000142, 4000143, 4000144, 4000145, 4000146, 4000148, 4000149, 4000151, 4000152, 4000153, 4000154, 4000155, 4000156, 4000157, 4000158, 4000159, 4000160, 4000161, 4000162, 4000163, 4000164, 4000165, 4000166, 4000167, 4000168, 4000169, 4000170, 4000171, 4000173, 4000174, 4000175, 4000176, 4000177, 4000178, 4000179, 4000180, 4000181, 4000182, 4000183, 4000184, 4000185, 4000186, 4000187, 4000188, 4000189, 4000190, 4000191, 4000192, 4000193, 4000194, 4000196, 4000197, 4000198, 4000199, 4000200, 4000202, 4000203, 4000204, 4000205, 4000206, 4000207, 4000208, 4000209, 4000211, 4000212, 4000214, 4000215, 4000216, 4000218, 4000219, 4000220, 4000221, 4000222, 4000223, 4000224, 4000225, 4000226, 4000227, 4000228, 4000229, 4000230, 4000231, 4000232, 4000233, 4000234, 4000235, 4000236, 4000237, 4000238, 4000239, 4000240, 4000241, 4000242, 4000243, 4000244, 4000245, 4000246, 4000247, 4000248, 4000249, 4000250, 4000251, 4000253, 4000254, 4000255, 4000256, 4000257, 4000258, 4000259, 4000260, 4000261, 4000262, 4000263, 4000264, 4000265, 4000266, 4000267, 4000268, 4000269, 4000270, 4000271, 4000272, 4000273, 4000274, 4000275, 4000276, 4000277, 4000278, 4000279, 4000280, 4000281, 4000282, 4000283, 4000284, 4000285, 4000286, 4000287, 4000288, 4000289, 4000290, 4000291, 4000292, 4000293, 4000294, 4000295, 4000296, 4000297, 4000298, 4000299, 4000300, 4000301, 4000302, 4000303, 4000304, 4000305, 4000306, 4000307, 4000308, 4000309, 4000310, 4000311, 4000312, 4000313, 4000314, 4000315, 4000316, 4000317, 4000318, 4000319, 4000320, 4000321, 4000322, 4000323, 4000324, 4000325, 4000326, 4000327, 4000328, 4000329, 4000330, 4000331, 4000332, 4000333, 4000334, 4000335, 4000336, 4000337, 4000338, 4000339, 4000340, 4000341, 4000342, 4000343, 4000344, 4000345, 4000346, 4000347, 4000348, 4000349, 4000351, 4000352, 4000353, 4000354, 4000355, 4000356, 4000357, 4000358, 4000359, 4000360, 4000361, 4000362, 4000363, 4000364, 4000366, 4000367, 4000368, 4000369, 4000370, 4000371, 4000372, 4000373, 4000374, 4000375, 4000376, 4000377, 4000378, 4000379, 4000380, 4000381, 4000382, 4000383, 4000384, 4000385, 4000386, 4000387, 4000388, 4000389, 4000390, 4000391, 4000392, 4000394, 4000395, 4000396, 4000397, 4000398, 4000399, 4000400, 4000401, 4000402, 4000403, 4000404, 4000405, 4000406, 4000407, 4000408, 4000409, 4000410, 4000411, 4000412, 4000414, 4000415, 4000416, 4000417, 4000418, 4000419, 4000420, 4000421, 4000422, 4000423, 4000424, 4000425, 4000426, 4000428, 4000429, 4000430, 4000431, 4000432, 4000433, 4000434, 4000435, 4000436, 4000437, 4000438, 4000439, 4000440, 4000441, 4000442, 4000443, 4000444, 4000445, 4000446, 4000447, 4000448, 4000449, 4000450, 4000451, 4000452, 4000453, 4000454, 4000455, 4000456, 4000459, 4000460, 4000461, 4000462, 4000463, 4000464, 4000465, 4000466, 4000467, 4000468, 4000469, 4000470, 4000471, 4000473, 4000474, 4000475, 4000476, 4000477, 4000478, 4000479, 4000480, 4000481, 4000482, 4000483, 4000484, 4000485, 4000486, 4000487.4000488, 4000489, 4000490, 4000491, 4000492, 4000493, 4000494, 4000495, 4000496, 4000497, 4000498, 4000499, 4000500, 4000501, 4000502, 4000503, 4000504, 4000505, 4000506, 4000507, 4000510, 4000511, 4000513, 4000516, 4000517, 4000518, 4000519, 4000520, 4000521, 4000522, 4000523, 4000524, 4000525, 4000526, 4000527, 4000528, 4000529, 4000530, 4000531, 4000532, 4000533, 4000534, 4000535, 4000536, 4000537, 4000538, 4000539, 4000540, 4000541, 4000542, 4000543, 4000544, 4000545, 4000546, 4000547, 4000549, 4000550, 4000551, 4000552, 4000553, 4000554, 4000555, 4000556, 4000557, 4000558, 4000559, 4000560, 4000561, 4000562, 4000563, 4000564, 4000565, 4000566, 4000567, 4000568, 4000569, 4000571, 4000572, 4000573, 4000574, 4000575, 4000576, 4000577, 4000578, 4000580, 4000581, 4000582, 4000583, 4000584, 4000585, 4000586, 4000587, 4000588, 4000589, 4000590, 4000591, 4000592, 4000593, 4000594, 4000595, 4000596, 4000597, 4000598, 4000599, 4000600, 4000601, 4000602, 4000603, 4000604, 4000605, 4000606, 4000607, 4000608, 4000609, 4000610, 4000611, 4000612, 4000613, 4000614, 4000615, 4000616, 4000617, 4000618, 4000619, 4000620, 4000621, 4000622, 4000623, 4000624, 4000625, 4000626, 4000627, 4000628, 4000629, 4000630, 4000631, 4000632, 4000633, 4000634, 4000635, 4000636, 4000637, 4000638, 4000639, 4000640, 4000641, 4000642, 4000643, 4000644, 4000645, 4000646, 4000647, 4000648, 4000649, 4000650, 4000651, 4000652, 4000653, 4000654, 4000655, 4000656, 4000657, 4000658, 4000659, 4000660, 4000661, 4000662, 4000663, 4000664, 4000665, 4000666, 4000667, 4000668, 4000669, 4000670, 4000671, 4000672, 4000673, 4000674, 4000675, 4000676, 4000677, 4000678, 4000679, 4000680, 4000681, 4000682, 4000683, 4000684, 4000685, 4000686, 4000687, 4000688, 4000689, 4000690, 4000691, 4000692, 4000693, 4000694, 4000695, 4000696, 4000697, 4000698, 4000699, 4340004, 4340005, 40000092, 40000101, 4000703001, 40007030003, 40007050005, 40007060008, 4000057-OLD, 4000098A, 4000151-1, 4000301-1, 4000472A, 4000484A, 10422, 400054, 4000029, 4000063, 4000067, 4000150, 4000195, 4000201, 4000210, 4000217, 4000252, 4000508, 4000509, 40007060006, 40007060007, 4000012, 4000213 & 4000413. |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Rd Cleveland OH 44143-2131
|
For Additional Information Contact |
Mr. Daniel R. Brown 440-483-2015
|
Manufacturer Reason for Recall |
The firm discovered four issues with the BrightView Family systems.
Issue 1:
During a cardiac non-Auto Body Contouring (ABC) scan, using Cardiac High Resolution (CHR) collimators, the collimator may come in contact with the bottom edge of the patient pallet. If this occurs, collision sensors would pause the scan.
Issue 2:
During a scan, the gantry motion may halt prematurely. This cause
|
FDA Determined Cause 2 |
Software design |
Action |
On 12/22/2015, the firm sent Customer Information Letters to their consignees.
The letter identified the product, problem and actions to be taken by the customers. |
Quantity in Commerce |
677 units |
Distribution |
Worldwide Distribution-US (nationwide) including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, VT, WA, WI, WV & WY, and in DC and PR and in the countries of Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Cayman Islands, China, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Gabon, Germany, Greece, India, Indonesia, Iran, Ireland, Italy, Japan, Korea, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Palestinian, Panama, Philippines, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom & Viet Nam. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KPS and Original Applicant = ADAC LABORATORIES
|
|
|
|