|
Class 2 Device Recall Natus neoBLUE LED Phototherapy system |
|
Date Initiated by Firm |
April 11, 2016 |
Create Date |
June 15, 2016 |
Recall Status1 |
Terminated 3 on December 14, 2016 |
Recall Number |
Z-2001-2016 |
Recall Event ID |
73169 |
510(K)Number |
K022196
|
Product Classification |
Unit, neonatal phototherapy - Product Code LBI
|
Product |
Natus neoBLUE Phototherapy system, Catalogue Part Number: 010066. Top Assembly Part Numbers: 001103, 001314, 002110, 040906, 040904, 040756, 040758, 040658, 040680.
Product Usage: Used to provide treatment for neonatal hyperbilirubinemia
|
Code Information |
Serial numbers of units distributed in American Samoa, Guam, Puerto Rico and in the US: .22938, .23282, .23283, .23289, .23291, .23302, .23307, .23471, .23473, .23478, .23479, .23480, .23481, .23483, .23538, .23697, .23708, .23887, .23888, .23895, .23916, .23939, .23968, .23969, .23972, .23983, .23993, .23994, .23998, 040756-1, 040756-2, 1, 10090, 10507, 10559, 10592, 10595, 10597, 10598, 10613, 10633, 10634, 10635, 10636, 10637, 10638, 10639, 10640, 10642, 10643, 10645, 10646, 10659, 10660, 10661, 10662, 10663, 10666, 10667, 10668, 10671, 10678, 10679, 10680, 10681, 10682, 10683, 10684, 10684, 10685, 10686, 10687, 10688, 10701, 10702, 10703, 10704, 10705, 10706, 10707, 10708, 10709, 10710, 10711, 10712, 10713, 10714, 10715, 10716, 10717, 10718, 10719, 10725, 10727, 10728, 10730, 10731, 10732, 10734, 10739, 10740, 10743, 10744, 10759, 10760, 10761, 10762, 10763, 10764, 10765, 10766, 10767, 10768, 10771, 10773, 10774, 10775, 10776, 10777, 10778, 10779, 10781, 10782, 10783, 10784, 10785, 10786, 10787, 10788, 10789, 10791, 10793, 10794, 10795, 10796, 10797, 10799, 10800, 10824, 10825, 10826, 10841, 10842, 10843, 10844, 10846, 10847, 10848, 10849, 10850, 10851, 10852, 10853, 10854, 10855, 10856, 10857, 10858, 10859, 10860, 10863, 10869, 10870, 10871, 10872, 10873, 10875, 10877, 10878, 10879, 10880, 10881, 10882, 10883, 10884, 10885, 10886, 10887, 10888, 10889, 10895, 10896, 10897, 10898, 10899, 10905, 10906, 10907, 10908, 10909, 10910, 10911, 10912, 10913, 10914, 10915, 10916, 10917, 10918, 10919, 10920, 10921, 10922, 10923, 10924, 10927, 10928, 10929, 10930, 10931, 10936, 10937, 10938, 10939, 10962, 10963, 10964, 10965, 10966, 10967, 10968, 10969, 10970, 10971, 10972, 10973, 10974, 10975, 10976, 10977, 10978, 10979, 10980, 10998, 10999, 11000, 11001, 11002, 11003, 11004, 11006, 11007, 11008, 11009, 11013, 11020, 11021, 11022, 11023, 11024, 11025, 11026, 11028, 11029, 11030, 11031, 11032, 11033, 11034, 11035, 11036, 11037, 11038, 11039, 11040, 11041, 11042, 11043, 11044, 11045, 11069, 11070, 11071, 11072, 11073, 11074, 11075, 11076, 11078, 11079, 11080, 11081, 11082, 11083, 11084, 11086, 11087, 11088, 11089, 11090, 11091, 11092, 11093, 11094, 11095, 11102, 11103, 11104, 11105, 11106, 11107, 11108, 11109, 11110, 11112, 11113, 11115, 11116, 11117, 11118, 11119, 11120, 11121, 11122, 11123, 11124, 11125, 11126, 11127, 11128, 11129, 11130, 11131, 11132, 11133, 12001, 12002, 12003, 12005, 12006, 12007, 12008, 12009, 12010, 12017, 12021, 12022, 12023, 12024, 12025, 12027, 12028, 12029, 12030, 12048, 12049, 12050, 12051, 12052, 12053, 12054, 12055, 12056, 12057, 12058, 12059, 12060, 12061, 12062, 12063, 12064, 12065, 12066, 12067, 12068, 12069, 12070, 12071, 12072, 12073, 12074, 12075, 12076, 12077, 12078, 12079, 12080, 12081, 12082, 12083, 12084, 12085, 12088, 12089, 12091, 12094, 12095, 12096, 12097, 12098, 12099, 12100, 12101, 12102, 12103, 12104, 12106, 12107, 12108, 12109, 12110, 12111, 12112, 12113, 12114, 12115, 12116, 12117, 12118, 12119, 12120, 12121, 12122, 12123, 12124, 12125, 12126, 12127, 12128, 12129, 12131, 12132, 12133, 12134, 12135, 12136, 12138, 12142, 12143, 12145, 12146, 12147, 12148, 12149, 12151, 12152, 12153, 12154, 12155, 12156, 12157, 12158, 12159, 12160, 12161, 12162, 12163, 12164, 12165, 12166, 12167, 12168, 12169, 12170, 12172, 12173, 12174, 12175, 12176, 12177, 12178, 12179, 12180, 12181, 12182, 12183, 12184, 12185, 12186, 12187, 12188, 12190, 12191, 12192, 12195, 12200, 12204, 12205, 12206, 12207, 12208, 12209, 12211, 12212, 12213, 12226, 12227, 12228, 12229, 12230, 12231, 12232, 12233, 12234, 12235, 12236, 12237, 12238, 12239, 12240, 12241, 12242, 12243, 12244, 12257, 12276, 12277, 12278, 12290, 12291, 12292, 12293, 12294, 12295, 12296, 12297, 12298, 12299, 12300, 12301, 12302, 12303, 12305, 12306, 12307, 12308, 12309, 12310, 12311, 12312, 12313, 12314, 12327, 12328, 12329, 12330, 12331, 12332, 12333, 12334, 12335, 12336, 12340, 12341, 12342, 12343, 12344, 12345, 12346, 12347, 12348, 12349, 12351, 12352.12353, 12354, 12355, 12356, 12357, 12358, 12359, 12360, 12361, 12362, 12363, 12364, 12365, 12366, 12367, 12368, 12369, 12370, 12371, 12372, 12373, 12374, 12375, 12376, 12377, 12378, 12379, 12380, 12381, 12382, 12383, 12384, 12385, 12386, 12387, 12388, 12389, 12390, 12391, 12392, 12393, 12394, 12395, 12396, 12397, 12398, 12399, 12400, 12401, 12403, 12406, 12410, 12412, 12413, 12414, 12415, 12417, 12418, 12419, 12420, 12421, 12422, 12423, 12424, 12425, 12426, 12427, 12428, 12429, 12430, 12431, 12432, 12433, 12434, 12435, 12436, 12437, 12438, 12439, 12440, 12441, 12442, 12443, 12444, 12445, 12446, 12447, 12448, 12449, 12450, 12451, 12452, 12453, 12454, 12455, 12456, 12457, 12458, 12459, 12461, 12462, 12463, 12464, 12465, 12501, 12502, 12503, 12504, 12505, 12506, 12507, 12508, 12509, 12510, 12511, 12512, 12513, 12514, 12516, 12517, 12518, 12519, 12520, 12521, 12522, 12524, 12525, 12526, 12527, 12528, 12529, 12530, 12531, 12532, 12533, 12534, 12535, 12536, 12537, 12538, 12539, 12540, 12541, 12542, 12543, 12544, 12545, 12547, 12548, 12549, 12550, 12551, 12552, 12553, 12554, 12555, 12556, 12557, 12558, 12559, 12569, 12570, 12571, 12584, 12586, 12587, 12588, 12590, 12591, 12593, 12594, 12595, 12601, 12602, 12603, 12604, 12605, 12606, 12607, 12608, 12609, 12610, 12611, 12612, 12613, 12614, 12615, 12616, 12617, 12618, 12619, 12620, 12621, 12622, 12623, 12625, 12627, 12628, 12631, 12632, 12633, 12634, 12635, 12637, 12638, 12639, 12640, 12642, 12643, 12645, 12646, 12647, 12648, 12649, 12650, 12651, 12652, 12653, 12654, 12655, 12656, 12657, 12658, 12659, 12660, 12661, 12662, 12664, 12665, 12667, 12668, 12669, 12670, 12671, 12672, 12673, 12674, 12675, 12676, 12677, 12678, 12679, 12680, 12681, 12682, 12683, 12684, 12685, 12686, 12688, 12689, 12690, 12721, 12722, 12723, 12724, 12725, 12726, 12727, 12728, 12729, 12730, 12731, 12732, 12733, 12734, 12735, 12736, 12737, 12738, 12739, 12740, 12741, 12742, 12743, 12744, 12745, 12746, 12747, 12748, 12749, 12750, 12751, 12752, 12753, 12754, 12755, 12756, 12757, 12758, 12759, 12760, 12761, 12762, 12763, 12764, 12765, 12766, 12767, 12768, 12769, 12770, 12772, 12773, 12774, 12775, 12776, 12777, 12778, 12779, 12780, 12781, 12782, 12783, 12784, 12785, 12786, 12787, 12788, 12789, 12790, 12791, 12792, 12793, 12794, 12795, 12796, 12797, 12797, 12798, 12799, 12800, 12801, 12802, 12803, 12804, 12805, 12806, 12807, 12808, 12809, 12810, 12811, 12812, 12814, 12815, 12816, 12817, 12818, 12819, 12820, 12821, 12822, 12828, 12829, 12830, 12831, 12832, 12833, 12834, 12835, 12836, 12837, 12838, 12839, 12840, 12841, 12842, 12843, 12844, 12845, 12846, 12847, 12848, 12849, 12850, 12851, 12852, 12853, 12854, 12855, 12856, 12857, 12858, 12859, 12860, 12861, 12863, 12864, 12865, 12866, 12867, 12878, 12880, 12881, 12882, 12883, 12884, 12885, 12886, 12887, 12888, 12889, 12890, 12892, 12893, 12895, 12896, 12897, 12898, 12899, 12900, 12901, 12902, 12903, 12904, 12905, 12906, 12907, 12908, 12909, 12910, 12911, 12912, 12913, 12939, 12940, 12941, 12942, 12943, 12944, 12945, 12946, 12947, 12948, 12949, 12950, 12951, 12954, 12955, 12956, 12957, 12958, 12959, 12960, 12961, 12962, 12963, 12964, 12965, 12966, 12967, 12968, 12969, 12970, 12971, 12972, 12973, 12974, 12975, 12976, 12977, 12978, 12981, 12982, 12983, 12984, 12989, 12990, 12991, 12992, 12993, 12994, 12995, 12996, 12997, 12998, 13058, 13059, 13061, 13062, 13063, 13066, 13067, 13069, 13070, 13071, 13072, 13073, 13074, 13075, 13076, 13077, 13078, 13079, 13080, 13081, 13082, 13083, 13084, 13085, 13086, 13087, 13088, 13089, 13090, 13091, 13093, 13095, 13096, 13097, 13098, 13099, 13100, 13101, 13102, 13103, 13104, 13105, 13106, 13107, 13110, 13111, 13112, 13113, 13114, 13115, 13116, 13117, 13118, 13119, 13120, 13121, 13124, 13125, 13126, 13127, 13128, 13129, 13130, 13131, 13132, 13133, 13134, 13135, 13136, 13137, 13138, 13139, 13140, 13141, 13142, 13143, 13144, 13145, 13146, 13147, 13148, 13149, 13150, 13151, 13152, 13153, 13154, 13155, 13156, 13157, 13158, 13159, 13160, 13161, 13162, 13163, 13164, 13165, 13167, 13169, 13170, 13171, 13172, 13173, 13174, 13175, 13176, 13177, 13201, 13202, 13203, 13204, 13205, 13206, 13207, 13208, 13209, 13210, 13211, 13212, 13213, 13214, 13215, 13216, 13217, 13218, 13219, 13220, 13222, 13223, 13224, 13225, 13226, 13227, 13228, 13229, 13230, 13231, 13232, 13233, 13234, 13235, 13236, 13237, 13238, 13239, 13240, 13241, 13242, 13243, 13244, 13245, 13246, 13247, 13248, 13249, 13250, 13251, 13252, 13253, 13254, 13255, 13256, 13257, 13258, 13261, 13262, 13268, 13269, 13270, 13271, 13272, 13273, 13274, 13276, 13281, 13282, 13283, 13284, 13285, 13295, 13296, 13297, 13298, 13299, 13300, 13302, 13303, 13304, 13305, 13306, 13307, 13308, 13309, 13310, 13311, 13312, 13313, 13314, 13315, 13316, 13317, 13318, 13319, 13320, 13321, 13322, 13323, 13324, 13325, 13326, 13328, 13329, 13330, 13331, 13332, 13333, 13334, 13335, 13336, 13337, 13338, 13339, 13340, 13341, 13342, 13343, 13344, 13345, 13346, 13347, 13348, 13349, 13350, 13351, 13352, 13353, 13354, 13355, 13356, 13357, 13358, 13359, 13360, 13361, 13362, 13363, 13364, 13365, 13366, 13367, 13368, 13369, 13370, 13371, 13372, 13374, 13376, 13377, 13378, 13379, 13380, 13381, 13382, 13383, 13384, 13385, 13386, 13387, 13388, 13389, 13390, 13391, 13392, 13394, 13395, 13396, 13397, 13398, 13399, 13400, 13401, 13402, 13403, 13404, 13405, 13406, 13407, 13408, 13409, 13410, 13411, 13412, 13413, 13414, 13415, 13416, 13417, 13418, 13419, 13420, 13421, 13422, 13423, 13424, 13425, 13426, 13427, 13428, 13429, 13430, 13431, 13432, 13433, 13434, 13435, 13436, 13437, 13439, 13441, 13442, 13444, 13445, 13446, 13447, 13448, 13449, 13450, 13451, 13452, 13453, 13454, 13455, 13456, 13457, 13458, 13459, 13460, 13461, 13462, 13463, 13464, 13465, 13466, 13467, 13468, 13469, 13470, 13471, 13472, 13473, 13474, 13475, 13476, 13477, 13478, 13479, 1348, 13480, 13481, 13482, 13483, 13484, 13485, 13487, 13488, 13489, 13490, 13491, 13492, 13493, 13494, 13495, 13496, 13801, 13802, 13803, 13804, 13805, 13806, 13807, 13808, 13809, 13810, 13816, 13817, 13818, 13819, 13820, 13823, 13828, 13829, 13830, 13831, 13832, 13833, 13834, 13835, 13836, 13837, 13838, 13840, 13841, 13842, 13843, 13844, 13845, 13846, 13847, 13848, 13849, 13850, 13851, 13852, 13853, 13855, 13860, 13861, 13862, 13866, 13868, 13870, 13871, 13872, 13873, 13874, 13877, 13878, 13879, 13880, 13881, 13882, 13883, 13884, 13885, 13887, 13898, 13899, 13900, 13901, 13902, 13903, 13904, 13905, 13906, 13907, 13908, 13909, 13911, 13912, 13913, 13914, 13915, 13916, 13917, 13918, 13919, 13920, 13921, 13922, 13923, 13934, 13935, 13959, 13960, 13961, 13962, 13963, 13964, 13965, 13966, 13967, 13968, 13971, 13972, 13973, 13974, 13975, 14021, 14024, 14025, 14026, 14027, 14028, 14029, 14030, 14031, 14032, 14033, 14034, 14035, 14036, 14037, 14038, 14039, 14040, 14041, 14042, 14043, 14044, 14045, 14046, 14047, 14048, 14049, 14050, 14051, 14052, 14053, 14054, 14055, 14056, 14057, 14058, 14058A, 14059, 14060, 14061, 14062, 14063, 14064, 14065, 14066, 14067, 14068, 14069, 14070, 14071, 14072, 14073, 14074, 14075, 14076, 14077, 14078, 14079, 14080, 14081, 14082, 14083, 14084, 14085, 14086, 14087, 14088, 14089, 14090, 14106, 14107, 14108, 14109, 14110, 14111, 14113, 14114, 14115, 14117, 14118, 14119, 14120, 14121, 14122, 14123, 14124, 14125, 14126, 14127, 14128, 14129, 14130, 14131, 14132, 14133, 14134, 14135, 14136, 14137, 14138, 14139, 14140, 14141, 14142, 14143, 14144, 14145, 14146, 14147, 14148, 14149, 14150, 14151, 14152, 14153, 14154, 14155, 14156, 14157, 14158, 14159, 14160, 14161, 14162, 14163, 14164, 14165, 14166, 14168, 14169, 14174, 14175, 14176, 14177, 14178, 14179, 14180, 14211, 14212, 14213, 14214, 14215, 14216, 14217, 14218, 14219, 14220, 14221, 14222, 14224, 14225, 14226, 14227, 14228, 14229, 14230, 14231, 14232, 14233, 14234, 14235, 14236, 14237, 14238, 14239, 14240, 14242, 14243, 14245, 14246, 14247, 14251, 14252, 14339, 14340, 14341, 14342, 14344, 14345, 14346, 14347, 14348, 14349, 14350, 14351, 14352, 14353, 14354, 14355, 14356, 14357, 14360, 14361, 14362, 14363, 14364, 14365, 14366, 14367, 14368, 14371, 14372, 14373, 14374, 14375, 14376, 14377, 14381, 14382, 14387, 14388, 14389, 14390, 14391, 14392, 14393, 14394, 14395, 14396, 14397, 14398, 14399, 14401, 14402, 14403, 14404, 14405, 14406, 14407, 14408, 14409, 14410, 14412, 14413, 14414, 14415, 14416, 14417, 14418, 14420, 14422, 14423, 14424, 14425, 14426, 14427, 14428, 14429, 14430, 14431, 14442, 14443, 14444, 14445, 14446, 14447, 14449, 14450, 14451, 14452, 14453, 14454, 14455, 14456, 14457, 14458, 14459, 14460, 14461, 14462, 14463, 14466, 14467, 14468, 14473, 14474, 14501, 14507, 14508, 14510, 14511, 14512, 14513, 14514, 14515, 14516, 14517, 14518, 14519, 14520, 14522, 14523, 14524, 14525, 14526, 14527, 14528, 14529, 14530, 14531, 14532, 14534, 14535, 14536, 14537, 14538, 14539, 14540, 14541, 14542, 14543, 14544, 14545, 14546, 14547, 14548, 14549, 14550, 14551, 14552, 14553, 14554, 14555, 14558, 14559, 14560, 14561, 14564, 14565, 14567, 14568, 14569, 14571, 14572, 14573, 14574, 14576, 14577, 14578, 14579, 14580, 14581, 14594, 14599, 14601, 14602, 14603, 14604, 14605, 14606, 14607, 14608, 14609, 14610, 14611, 14612, 14613, 14614, 14615, 14616, 14617, 14618, 14619, 14620, 14621, 14622, 14623, 14624, 14625, 14626, 14627, 14633, 14642, 14644, 14646, 14650, 14652, 14657, 14659, 14667, 14668, 14674, 14675, 14676, 14677, 14679, 14681, 14682, 14683, 14684, 14685, 14686, 14687, 14688, 14689, 14690, 14691, 14692, 14693, 14695, 14696, 14697, 14698, 14699, 14700, 14704, 14705, 14706, 14707, 14708, 14709, 14710, 14712, 14713, 14714, 14715, 14716, 14717, 14718, 14719, 14720, 14721, 14722, 14723, 14724, 14725, 14726, 14727, 14728, 14729, 14730, 14731, 14732, 14733, 14734, 14735, 14736, 14737, 14738, 14739, 14740, 14741, 14742, 14743, 14744, 14745, 14746, 14747, 14748, 14749, 14750, 14751, 14753, 14754, 14755, 14756, 14757, 14758, 14759, 14760, 14761, 14762, 14763, 14764, 14765, 14766, 14767, 14768, 14789, 14790, 14791, 14792, 14793, 14794, 14795, 14796, 14797, 14798, 14799, 14800, 14801, 14802, 14803, 14804, 14805, 14806, 14807, 14808, 14812, 14813, 14814, 14815, 14816, 14817, 14818, 14819, 14820, 14821, 14822, 14823, 14824, 14825, 14826, 14827, 14828, 14829, 14830, 14831, 14832, 14833, 14834, 14835, 14836, 14837, 14838, 14839, 14840, 14868, 14869, 14870, 14871, 14872, 14873, 14875, 14876, 14877, 14878, 14879, 14880, 14881, 14882, 14884, 14885, 14886, 14887, 14888, 14889, 14890, 14891, 14892, 14893, 14894, 14895, 14896, 14897, 14898, 14899, 14900, 14901, 14902, 14903, 14904, 14905, 14906, 14907, 14908, 14909, 14910, 14911, 14913, 14914, 14915, 14916, 14917, 14918, 14919, 14920, 14921, 14922, 14923, 14924, 14925, 14926, 14927, 14928, 14929, 14930, 14931, 14932, 14933, 14934, 14935, 14936, 14937, 14938, 14939, 14940, 14941, 14942, 14943, 14944, 14945, 14946, 14947, 14948, 14949, 14950, 14951, 14952, 14953, 14954, 14955, 14956, 14957, 14958, 14959, 14960, 14961, 14962, 14963, 14964, 14965, 14966, 14967, 14968, 14970, 14971, 14972, 14974, 14975, 14976, 14977, 14978, 14979, 14980, 14981, 14982, 14983, 14984, 14985, 14986, 14987, 14988, 14989, 14990, 14991, 14992, 14993, 14994, 14995, 14996, 14997, 14998, 14999, 15000, 15001, 15002, 15003, 15004, 15005, 15006, 15007, 15008, 15009, 15010, 15011, 15012, 15013, 15014, 15015, 15016, 15017, 15018, 15019, 15020, 15021, 15022, 15023, 15024, 15025, 15027, 15028, 15029, 15030, 15061, 15062, 15063, 15064, 15065, 15066, 15067, 15068, 15069, 15070, 15071, 15072, 15073, 15074, 15075, 15076, 15077, 15078, 15079, 15080, 15081, 15082, 15085, 15086, 15168, 15208, 15210, 15211, 15214, 15215, 15216, 15217, 15219, 15220, 15221, 15222, 15223, 15224, 15225, 15226, 15227, 15228, 15229, 15230, 15231, 15232, 15233, 15234, 15235, 15236, 15237, 15238, 15242, 15245, 15253, 15271, 15272, 15273, 15274, 15275, 15276, 15277, 15278, 15279, 15280, 15281, 15282, 15283, 15284, 15285, 15286, 15287, 15288, 15289, 15290, 15291, 15292, 15293, 15294, 15295, 15296, 15297, 15298, 15300, 15301, 15302, 15303, 15304, 15305, 15312, 15313, 15314, 15315, 15316, 15317, 15318, 15319, 15320, 15321, 15322, 15323, 15324, 15325, 15326, 15327, 15328, 15329, 15330, 15331, 15332, 15333, 15334, 15336, 15337, 15338, 15339, 15340, 15341, 15342, 15343, 15344, 15345, 15346, 15347, 15348, 15349, 15350, 15353, 15372, 15373, 15374, 15375, 15376, 15378, 15379, 15381, 15382, 15383, 15384, 15385, 15386, 15387, 15388, 15389, 15390, 15392, 15393, 15394, 15395, 15396, 15398, 15399, 15400, 15405, 15406, 15407, 15408, 15409, 15410, 15411, 15412, 15413, 15416, 15417, 15418, 15419, 15420, 15421, 15422, 15423, 15424, 15425, 15426, 15427, 15428, 15429, 15430, 15431, 15432, 15433, 15434, 15435, 15436, 15437, 15438, 15439, 15440, 15441, 15443, 15444, 15445, 15446, 15497, 15498, 15499, 15500, 15501, 15502, 15503, 15504, 15505, 15506, 15507, 15508, 15509, 15510, 15511, 15513, 15514, 15516, 15517, 15518, 15519, 15520, 15521, 15522, 15524, 15525, 15526, 15527, 15528, 15530, 15531, 15532, 15533, 15534, 15535, 15536, 15537, 15538, 15539, 15540, 15541, 15542, 15543, 15544, 15545, 15546, 15547, 15548, 15550, 15601, 15602, 15603, 15604, 15605, 15606, 15607, 15608, 15609, 15610, 15611, 15612, 15613, 15614, 15615, 15616, 15617, 15618, 15619, 15620, 15621, 15622, 15623, 15624, 15625, 15626, 15627, 15628, 15629, 15630, 15631, 15632, 15633, 15646, 15647, 15648, 15649, 15650, 15651, 15652, 15653, 15654, 15655, 15656, 15657, 15658, 15659, 15660, 15661, 15662, 15663, 15664, 15665, 15666, 15667, 15668, 15669, 15670, 15671, 15672, 15673, 15674, 15675, 15676, 15677, 15678, 15679, 15680, 15681, 15682, 15711, 15712, 15713, 15714, 15715, 15716, 15717, 15718, 15719, 15720, 15721, 16501, 16502, 16503, 16504, 16506, 16507, 16508, 16509, 16510, 16511, 16512, 16513, 16514, 16515, 16517, 16518, 16519, 16520, 16605, 16616, 16617, 16618, 16619, 16620, 16621, 16622, 16624, 16627, 16628, 16629, 16631, 16632, 16633, 16634, 16635, 16636, 16637, 16638, 16640, 16641, 16642, 16643, 16644, 16646, 16647, 16648, 16649, 16650, 16651, 16652, 16653, 16654, 16655, 16656, 16657, 16658, 16659, 16660, 16661, 16662, 16663, 16664, 16665, 16666, 16667, 16668, 16669, 16685, 16686, 16687, 16688, 16689, 16690, 16691, 16692, 16693, 16694, 16695, 16696, 16697, 16698, 16699, 16700, 16701, 16702, 16703, 16704, 16705, 16706, 16707, 16708, 16709, 16710, 16711, 16712, 16713, 16714, 16715, 16716, 16717, 16718, 16719, 16720, 16721, 16722, 16724, 16725, 16726, 16732, 16733, 16734, 16735, 16736, 16737, 16738, 16739, 16740, 16741, 16742, 16743, 16744, 16745, 16746, 16747, 16748, 16749, 16750, 16751, 16752, 16753, 16754, 16755, 16758, 16759, 16760, 16761, 16762, 16763, 16764, 16765, 16766, 16767, 16768, 16769, 16770, 16771, 16772, 16773, 16774, 16775, 16776, 16777, 16778, 16779, 16780, 16781, 16782, 16783, 16784, 16785, 16786, 16807, 16808, 16809, 16810, 16812, 16813, 16814, 16815, 16816, 16817, 16818, 16819, 16820, 16821, 16822, 16823, 16824, 16825, 16826, 16827, 16828, 16833, 16834, 16835, 16836, 16837, 16838, 16839, 16840, 16843, 16846, 16847, 16848, 16849, 16850, 16851, 16852, 16853, 16855, 16856, 16857, 16858, 16859, 16860, 16861, 16862, 16863, 16864, 16865, 16866, 16867, 16868, 16869, 16870, 16871, 16872, 16873, 16874, 16875, 16876, 16877, 16878, 16879, 16880, 16881, 16882, 16883, 16884, 16885, 16886, 16887, 16888, 16939, 16940, 16941, 16942, 16943, 16951, 16952, 16953, 16954, 16955, 16956, 16957, 16958, 16959, 16960, 16961, 16962, 16963, 16964, 16965, 16966, 16967, 16968, 16969, 16971, 16972, 16973, 16974, 16975, 16976, 16977, 16979, 16980, 16981, 16982, 16983, 16984, 16985, 16986, 16987, 16988, 16989, 16990, 16991, 16992, 16993, 16994, 16995, 16996, 16997, 16998, 16999, 17000, 17001, 17002, 17003, 17004, 17005, 17006, 17007, 17008, 17009, 17010, 17011, 17012, 17013, 17014, 17055, 17056, 17057, 17058, 17059, 17060, 17061, 17062, 17073, 17074, 17076, 17077, 17079, 17080, 17081, 17082, 17083, 17084, 17085, 17086, 17087, 17088, 17089, 17090, 17091, 17092, 17093, 17094, 17095, 17096, 17097, 17098, 17099, 17100, 17101, 17105, 17156, 17157, 17158, 17159, 17160, 17161, 17162, 17163, 17164, 17165, 17166, 17167, 17168, 17169, 17170, 17171, 17172, 17173, 17174, 17175, 17176, 17177, 17178, 17179, 17180, 17181, 17182, 17183, 17184, 17185, 17186, 17187, 17188, 17190, 17191, 17192, 17193, 17194, 17195, 17196, 17197, 17198, 17199, 17200, 17201, 17202, 17203, 17204, 17205, 17206, 17207, 17208, 17209, 17210, 17211, 17212, 17221, 17222, 17333, 17334, 17335, 17336, 17337, 17338, 17339, 17340, 17341, 17342, 17343, 17344, 17345, 17346, 17347, 17348, 17350, 17351, 17352, 17353, 17354, 17355, 17356, 17357, 17358, 17359, 17360, 17372, 17373, 17374, 17375, 17376, 17377, 17379, 17380, 17381, 17382, 17383, 17385, 17386, 17387, 17388, 17389, 17390, 17391, 17392, 17393, 17394, 17395, 17396, 17404, 17405, 17418, 17431, 17433, 17434, 17435, 17436, 17446, 17448, 17449, 17450, 17451, 17452, 17453, 17454, 17455, 17456, 17457, 17458, 17459, 17460, 17461, 17462, 17463, 17464, 17465, 17466, 17467, 17468, 17469, 17480, 17484, 17485, 17487, 17488, 17489, 17490, 17491, 17492, 17493, 17494, 17495, 17496, 17497, 17498, 17499, 17500, 17501, 17502, 17503, 17504, 17505, 17506, 17507, 17508, 17509, 17510, 17511, 17514, 17515, 17516, 17517, 17519, 17520, 17524, 17525, 17526, 17527, 17528, 17529, 17530, 17531, 17532, 17538, 17539, 17540, 17541, 17542, 17543, 17544, 17545, 17546, 17547, 17548, 17549, 17550, 17551, 17552, 17553, 17554, 17555, 17568, 17569, 17570, 17573, 17574, 17575, 17576, 17577, 17578, 17579, 17580, 17581, 17583, 17584, 17585, 17586, 17587, 17588, 17589, 17590, 17591, 17592, 17593, 17594, 17595, 17597, 17598, 17599, 17600, 17601, 17648, 17649, 17650, 17651, 17652, 17653, 17654, 17655, 17657, 17658, 17659, 17660, 17661, 17662, 17663, 17664, 17665, 17666, 17667, 17668, 17669, 17670, 17672, 17673, 17674, 17675, 17676, 17677, 17678, 17679, 17680, 17681, 17682, 17683, 17684, 17685, 17686, 17687, 17688, 17689, 17690, 17691, 17692, 17693, 17694, 17695, 17696, 17697, 17697, 17698, 17699, 17700, 17702, 17703, 17704, 17705, 17707, 17795, 17796, 17797, 17798, 17799, 17800, 17801, 17802, 17803, 17804, 17805, 17806, 17807, 17808, 17809, 17810, 17812, 17813, 17814, 17815, 17816, 17818, 17819, 17820, 17821, 17822, 17823, 17824, 17825, 17826, 17829, 17830, 17831, 17832, 17833, 17834, 17835, 17836, 17837, 17838, 17840, 17841, 17842, 17843, 17844, 17866, 17867, 17868, 17869, 17870, 17871, 17872, 17873, 17874, 17875, 17876, 17877, 17878, 17879, 17880, 17895, 17896, 17897, 17898, 17899, 17900, 17901, 17902, 17903, 17904, 17905, 17906, 17907, 18002, 18003, 18004, 18005, 18006, 18007, 18008, 18009, 18010, 18011, 18012, 18013, 18015, 18016, 18017, 18018, 18019, 18020, 18021, 18022, 18023, 18040, 18041, 18042, 18043, 18044, 18045, 18046, 18047, 18048, 18049, 18050, 18051, 18052, 18053, 18054, 18055, 18056, 18057, 18058, 18059, 18060, 18061, 18062, 18063, 18064, 18065, 18066, 18067, 18068, 18069, 18070, 18071, 18072, 18073, 18074, 18075, 18076, 18077, 18078, 18084, 18086, 18087, 18088, 18089, 18090, 18091, 18092, 18118, 18119, 18120, 18121, 18122, 18133, 18134, 18302, 18303, 18304, 18305, 18306, 18307, 18308, 18309, 18310, 18311, 18312, 18313, 18314, 18315, 18316, 18317, 18318, 18319, 18320, 18328, 18329, 18330, 18331, 18332, 18333, 18334, 18335, 18336, 18337, 18338, 18339, 18340, 18341, 18342, 18343, 18344, 18345, 18346, 18347, 18348, 18349, 18350, 18351, 18352, 18353, 18355, 18356, 18357, 18358, 18359, 18360, 18361, 18362, 18363, 18364, 18365, 18366, 18367, 18368, 18369, 18370, 18371, 18372, 18373, 18374, 18375, 18376, 18377, 18378, 18379, 18380, 18381, 18382, 18383, 18384, 18385, 18386, 18387, 18388, 18389, 18390, 18392, 18405, 18407, 18408, 18409, 18410, 18411, 18413, 18414, 18415, 18416, 18417, 18418, 18419, 18420, 18421, 18422, 18423, 18424, 18425, 18426, 18427, 18428, 18429, 18430, 18431, 18432, 18433, 18434, 18664, 18665, 18666, 18667, 18668, 18669, 18670, 18671, 18672, 18673, 18674, 18675, 18676, 18677, 18678, 18679, 18680, 18681, 18682, 18683, 18684, 18685, 18686, 18687, 18688, 18689, 18690, 18691, 18692, 18693, 18694, 18695, 18696, 18697, 18698, 18699, 18700, 18701, 18702, 18703, 18704, 18705, 18706, 18707, 18708, 18709, 18710, 18711, 18712, 18713, 18714, 18718, 18719, 18720, 18725, 18726, 18728, 18729, 18730, 18731, 18732, 18733, 18734, 18735, 18736, 18737, 18738, 18739, 18740, 18741, 18742, 18743, 18744, 18745, 18746, 18747, 18763, 18764, 18765, 18766, 18767, 18768, 18769, 18770, 18771, 18772, 18773, 18774, 18775, 18776, 18777, 18778, 18779, 18780, 18781, 18782, 18783, 18784, 18785, 18786, 18787, 18788, 18789, 18790, 18791, 18792, 18793, 18794, 18795, 18796, 18797, 18798, 18799, 18800, 18801, 18802, 18803, 18804, 18805, 18806, 18807, 18808, 18809, 18810, 18811, 18812, 18813, 18814, 18815, 18816, 18837, 18838, 18839, 18840, 18841, 18842, 18843, 18844, 18845, 18846, 18847, 18848, 18849, 18850, 18851, 18852, 18853, 18854, 18855, 18856, 18857, 18858, 18859, 18860, 18861, 18862, 18863, 18864, 18866, 18873, 18874, 18875, 18876, 18877, 18878, 18879, 18880, 18881, 18884, 18885, 18886, 18887, 18888, 18889, 18893, 18910, 18911, 18912, 18913, 18914, 18919, 18920, 18921, 18922, 18923, 18924, 18925, 18926, 18927, 18928, 18929, 18930, 18931, 18932, 18933, 18934, 18935, 18936, 18937, 18938, 18939, 18942, 18943, 18944, 18945, 18951, 18954, 18958, 18959, 18960, 18961, 18962, 18963, 18964, 18965, 18966, 18967, 18968, 18969, 18970, 18971, 18972, 18973, 18974, 18975, 18976, 18977, 18978, 18979, 18980, 18981, 18982, 18983, 18984, 18985, 18986, 18987, 18988, 18989, 18990, 18991, 18992, 18993, 18994, 18995, 18996, 18997, 18998, 18999, 19000, 19001, 19002, 19003, 19004, 19005, 19006, 19007, 19008, 19009, 19010, 19011, 19017, 19018, 19019, 19020, 19021, 19022, 19029, 19030, 19031, 19032, 19033, 19035, 19036, 19044, 19048, 19049, 19050, 19051, 19052, 19068, 19069, 19070, 19088, 19100, 19101, 19102, 19103, 19104, 19105, 19106, 19107, 19108, 19109, 19110, 19111, 19112, 19113, 19114, 19115, 19116, 19117, 19118, 19119, 19120, 19121, 19122, 19123, 19124, 19126, 19127, 19128, 19129, 19130, 19131, 19132, 19133, 19134, 19136, 19137, 19138, 19139, 19141, 19142, 19143, 19146, 19147, 19150, 19319, 19320, 19321, 19322, 19323, 19324, 19325, 19326, 19327, 19328, 19329, 19330, 19331, 19332, 19333, 19334, 19335, 19336, 19337, 19338, 19339, 19340, 19341, 19342, 19343, 19344, 19345, 19361, 19366, 19367, 19368, 19369, 19370, 19373, 19374, 19375, 19376, 19377, 19378, 19379, 19380, 19381, 19382, 19383, 19384, 19385, 19386, 19387, 19388, 19389, 19390, 19391, 19392, 19393, 19394, 19395, 19396, 19397, 19398, 19399, 19400, 19401, 19402, 19403, 19404, 19405, 19406, 19409, 19410, 19411, 19412, 19413, 19414, 19415, 19416, 19417, 19418, 19419, 19420, 19421, 19422, 19423, 20637, 20638, 20641, 20642, 20643, 20644, 20645, 20646, 20647, 20648, 20649, 20650, 20651, 20652, 20653, 20654, 20655, 20656, 20657, 20658, 20659, 20660, 20661, 20662, 20663, 20664, 20665, 20666, 20667, 20668, 20669, 20670, 20671, 20672, 20673, 20674, 20675, 20676, 20677, 20710, 20711, 20712, 20713, 20714, 20715, 20716, 20717, 20718, 20719, 20720, 20721, 20722, 20723, 20724, 20725, 20726, 20727, 20728, 20729, 20746, 20747, 20749, 20750, 20751, 20752, 20761, 20762, 20763, 20764, 20765, 20766, 20767, 20768, 20769, 20770, 20771, 20772, 20773, 20774, 20775, 20776, 20777, 20778, 20779, 20780, 20781, 20782, 20783, 20784, 20785, 20786, 20787, 20790, 20791, 20792, 20793, 20794, 20795, 20796, 20797, 20798, 20799, 20800, 20801, 20802, 20803, 20804, 20805, 20806, 20807, 20808, 20809, 20810, 20811, 20812, 20813, 20814, 20815, 20816, 20817, 20818, 20819, 20820, 20821, 20822, 20823, 20824, 20825, 20826, 20827, 20828, 20829, 20830, 20831, 20832, 20833, 20834, 20835, 20836, 20837, 20838, 20839, 20840, 20841, 20852, 20855, 20857, 20858, 20859, 20860, 20861, 20862, 20863, 20864, 20865, 20866, 20867, 20868, 20869, 20870, 20871, 20872, 20873, 20874, 20875, 20876, 20877, 20878, 20879, 20880, 20881, 20882, 20883, 20884, 20885, 20886, 20887, 20888, 20889, 20890, 20891, 20892, 20893, 20894, 20895, 20897, 20898, 20905, 20912, 20916, 20917, 20918, 20922, 20923, 20924, 20925, 20926, 20927, 20928, 20929, 20930, 20931, 20932, 20933, 20934, 20935, 20936, 20937, 20946, 20947, 20949, 20950, 20951, 20952, 20953, 20954, 20955, 20956, 20957, 20958, 20959, 20960, 20961, 20962, 20963, 20964, 20965, 20966, 20967, 20968, 20969, 20970, 20971, 20972, 20973, 20974, 20975, 20976, 20977, 20978, 20979, 20980, 20981, 20982, 20983, 20984, 20985, 20986, 20987, 20988, 20989, 20990, 20991, 20992, 20993, 20994, 20995, 21001, 21002, 21003, 21004, 21005, 21006, 21007, 21008, 21015, 21016, 21017, 21018, 21019, 21020, 21021, 21022, 21023, 21024, 21025, 21026, 21027, 21028, 21029, 21030, 21031, 21032, 21033, 21034, 21035, 21036, 21037, 21038, 21039, 21040, 21041, 21042, 21043, 21044, 21045, 21046, 21047, 21048, 21052, 21053, 21056, 21057, 21058, 21059, 21060, 21061, 21062, 21063, 21068, 21076, 21092, 21093, 21094, 21095, 21096, 21097, 21098, 21100, 21101, 21102, 21103, 21104, 21105, 21106, 21107, 21108, 21109, 21110, 21111, 21112, 21113, 21114, 21115, 21116, 21117, 21118, 21119, 21120, 21121, 21122, 21123, 21124, 21125, 21126, 21127, 21128, 21129, 21140, 21141, 21148, 21149, 21150, 21151, 21152, 21153, 21154, 21155, 21156, 21157, 21158, 21159, 21160, 21161, 21162, 21163, 21164, 21165, 21166, 21167, 21168, 21169, 21170, 21171, 21172, 21191, 21249, 21250, 21251, 21252, 21253, 21254, 21257, 21258, 21259, 21260, 21261, 21262, 21263, 21264, 21266, 21267, 21268, 21269, 21270, 21271, 21272, 21273, 21274, 21275, 21276, 21277, 21278, 21281, 21282, 21283, 21284, 21285, 21286, 21287, 21288, 21289, 21290, 21291, 21292, 21293, 21294, 21295, 21296, 21297, 21298, 21299, 21300, 21301, 21302, 21303, 21304, 21310, 21313, 21314, 21315, 21316, 21345, 21346, 21347, 21348, 21354, 21356, 21363, 21364, 21367, 21371, 21372, 21373, 21374, 21407, 21408, 21409, 21410, 21411, 21412, 21413, 21414, 21415, 21416, 21417, 21418, 21634, 21635, 21636, 21637, 21638, 21639, 21640, 21641, 21642, 21643, 21644, 21645, 21646, 21647, 21648, 21649, 21650, 21651, 21653, 21654, 21655, 21656, 21657, 21658, 21659, 21660, 21661, 21662, 21664, 21665, 21666, 21667, 21668, 21669, 21670, 21671, 21672, 21673, 21674, 21676, 21677, 21678, 21679, 21680, 21681, 21684, 21685, 21686, 21687, 21688, 21689, 21690, 21691, 21692, 21693, 21696, 21738, 21739, 21740, 21741, 21742, 21743, 21744, 21745, 21746, 21747, 21783, 21784, 21785, 21786, 21787, 21790, 21791, 21792, 21793, 21794, 21795, 21796, 21797, 21798, 21799, 21800, 21801, 21802, 21803, 21804, 21805, 21806, 21807, 21808, 21809, 21810, 21811, 21812, 21813, 21814, 21815, 21816, 21817, 21818, 21819, 21820, 21821, 21822, 21823, 21824, 21825, 21826, 21827, 21828, 21829, 21830, 21831, 21832, 21833, 21834, 21835, 21836, 21837, 21838, 21839, 21840, 21841, 21842, 21843, 21844, 21845, 21846, 21847, 21848, 21849, 21850, 21851, 21852, 21868, 21870, 21871, 21872, 21873, 21874, 21875, 21876, 21877, 21878, 21879, 21880, 21881, 21882, 21883, 21884, 21885, 21886, 21887, 21888, 21889, 21890, 21892, 21893, 21894, 21895, 21896, 21897, 21898, 21899, 21900, 21901, 21902, 21903, 21904, 21905, 21906, 21907, 21908, 21909, 21910, 21911, 21912, 21913, 21914, 21915, 21951, 21952, 21953, 21954, 21955, 21956, 21957, 21958, 21959, 21971, 21972, 21973, 21974, 21975, 21976, 21977, 21978, 21979, 21980, 21987, 21989, 21990, 21991, 21992, 21993, 21994, 21995, 21996, 21997, 21998, 21999, 22000, 22001, 22002, 22003, 22004, 22005, 22006, 22007, 22008, 22009, 22010, 22011, 22012, 22013, 22029, 22030, 22031, 22032, 22033, 22034, 22035, 22036, 22037, 22038, 22042, 22043, 22044, 22045, 22046, 22047, 22048, 22049, 22050, 22051, 22052, 22053, 22054, 22055, 22057, 22058, 22059, 22061, 22062, 22063, 22064, 22070, 22071, 22072, 22073, 22074, 22075, 22076, 22077, 22081, 22082, 22083, 22084, 22085, 22086, 22087, 22088, 22089, 22090, 22091, 22092, 22093, 22094, 22095, 22099, 22100, 22101, 22102, 22103, 22104, 22105, 22106, 22107, 22108, 22109, 22110, 22111, 22112, 22113, 22114, 22115, 22116, 22117, 22118, 22119, 22120.22121, 22122, 22123, 22126, 22127, 22128, 22129, 22131, 22132, 22133, 22134, 22135, 22136, 22137, 22138, 22139, 22140, 22141, 22142, 22143, 22144, 22153, 22154, 22155, 22156, 22157, 22158, 22159, 22160, 22161, 22162, 22163, 22164, 22165, 22166, 22167, 22168, 22169, 22170, 22171, 22172, 22173, 22174, 22175, 22176, 22177, 22180, 22181, 22183, 22184, 22187, 22188, 22189, 22192, 22193, 22194, 22195, 22196, 22197, 22198, 22199, 22200, 22201, 22202, 22203, 22204, 22205, 22206, 22207, 22208, 22209, 22210, 22211, 22212, 22213, 22214, 22215, 22216, 22218, 22219, 22220, 22221, 22223, 22224, 22225, 22226, 22227, 22228, 22233, 22234, 22235, 22236, 22237, 22238, 22254, 22255, 22256, 22257, 22258, 22259, 22260, 22261, 22262, 22263, 22264, 22265, 22266, 22267, 22268, 22269, 22270, 22271, 22272, 22273, 22274, 22276, 22277, 22278, 22279, 22280, 22281, 22282, 22283, 22284, 22285, 22286, 22287, 22288, 22289, 22290, 22291, 22292, 22293, 22294, 22295, 22296, 22297, 22299, 22300, 22301, 22302, 22303, 22304, 22305, 22306, 22307, 22308, 22309, 22310, 22311, 22312, 22313, 22314, 22315, 22316, 22317, 22318, 22319, 22320, 22321, 22322, 22323, 22324, 22325, 22326, 22327, 22328, 22329, 22330, 22331, 22332, 22333, 22334, 22335, 22336, 22337, 22338, 22339, 22340, 22341, 22342, 22343, 22344, 22345, 22346, 22347, 22348, 22349, 22350, 22351, 22352, 22353, 22354, 22355, 22356, 22357, 22358, 22359, 22361, 22362, 22363, 22372, 22373, 22374, 22375, 22379, 22380, 22381, 22383, 22384, 22385, 22386, 22387, 22388, 22389, 22390, 22391, 22393, 22394, 22395, 22396, 22397, 22398, 22399, 22400, 22440, 22441, 22442, 22444, 22445, 22446, 22447, 22448, 22449, 22450, 22451, 22452, 22453, 22454, 22455, 22456, 22457, 22458, 22459, 22460, 22461, 22464, 22465, 22467, 22468, 22469, 22470, 22471, 22472, 22475, 22476, 22477, 22478, 22479, 22480, 22481, 22482, 22483, 22484, 22485, 22486, 22487, 22488, 22489, 22490, 22493, 22494, 22495, 22496, 22497, 22498, 22499, 22500, 22501, 22502, 22503, 22505, 22506, 22511, 22513, 22514, 22515, 22516, 22517, 22518, 22519, 22520, 22521, 22522, 22524, 22525, 22527, 22528, 22529, 22530, 22531, 22532, 22533, 22534, 22535, 22536, 22537, 22538, 22539, 22540, 22541, 22542, 22543, 22544, 22545, 22547, 22548, 22549, 22552, 22565, 22568, 22569, 22570, 22571, 22573, 22574, 22575, 22576, 22577, 22578, 22579, 22580, 22581, 22582, 22583, 22584, 22586, 22587, 22588, 22589, 22590, 22591, 22592, 22593, 22594, 22595, 22596, 22598, 22599, 22600, 22607, 22608, 22609, 22610, 22611, 22612, 22613, 22628, 22629, 22630, 22631, 22632, 22633, 22634, 22635, 22636, 22637, 22638, 22639, 22640, 22641, 22642, 22643, 22644, 22645, 22646, 22647, 22648, 22650, 22651, 22652, 22653, 22654, 22655, 22656, 22657, 22658, 22659, 22660, 22661, 22662, 22663, 22664, 22665, 22667, 22668, 22669, 22671, 22673, 22674, 22675, 22676, 22677, 22678, 22679, 22680, 22683, 22684, 22685, 22686, 22687, 22688, 22689, 22690, 22691, 22704, 22705, 22706, 22707, 22712, 22814, 22815, 22816, 22817, 22818, 22819, 22820, 22821, 22822, 22823, 22824, 22825, 22826, 22827, 22828, 22829, 22830, 22831, 22833, 22834, 22835, 22836, 22837, 22838, 22839, 22840, 22841, 22842, 22856, 22857, 22858, 22859, 22860, 22865, 22874, 22875, 22876, 22898, 22899, 22900, 22901, 22902, 22917, 22918, 22919, 22920, 22921, 22922, 22923, 22924, 22925, 22926, 22927, 22928, 22929, 22930, 22931, 22932, 22933, 23010, 23011, 23012, 23013, 23014, 23015, 23016, 23017, 23018, 23019, 23020, 23021, 23022, 23023, 23024, 23025, 23026, 23027, 23028, 23029, 23030, 23031, 23041, 23042, 23043, 23044, 23045, 23046, 23047, 23048, 23049, 23051, 23052, 23053, 23054, 23055, 23056, 23057, 23058, 23059, 23060, 23061, 23062, 23063, 23064, 23065, 23067, 23068, 23069, 23070, 23071, 23072, 23073, 23074, 23075, 23076, 23077, 23078, 23079, 23080, 23081, 23082, 23083, 23084, 23085, 23086, 23087, 23088, 23089, 23090, 23091, 23092, 23093, 23131, 23132, 23133, 23134, 23135, 23136, 23137, 23138, 23139, 23140, 23141, 23142, 23143, 23144, 23145, 23146, 23147, 23148, 23149, 23150, 23151, 23152, 23153, 23154, 23155, 23156, 23157, 23158, 23159, 23160, 23161, 23162, 23163, 23164, 23165, 23166, 23167, 23168, 23169, 23170, 23171, 23172, 23173, 23174, 23175, 23176, 23187, 23188, 23189, 23190, 23191, 23192, 23193, 23194, 23195, 23196, 23197, 23198, 23255, 23256, 23257, 23258, 23259, 23260, 23261, 23262, 23263, 23264, 23265, 23266, 23267, 23268, 23269, 23270, 23271, 23272, 23273, 23274, 23275, 23276, 23277, 23278, 23279, 23280, 23281, 23287, 23293, 23294, 23295, 23296, 23297, 23298, 23303, 23305, 23309, 23310, 23311, 23314, 23315, 23316, 23317, 23318, 23320, 23321, 23322, 23324, 23325, 23326, 23327, 23328, 23331, 23332, 23334, 23335, 23336, 23337, 23341, 23342, 23343, 23344, 23345, 23361, 23362, 23363, 23364, 23365, 23366, 23367, 23368, 23369, 23370, 23371, 23372, 23373, 23374, 23375, 23376, 23377, 23378, 23379, 23380, 23381, 23382, 23383, 23384, 23385, 23386, 23387, 23394, 23395, 23396, 23397, 23398, 23399, 23401, 23402, 23403, 23404, 23405, 23406, 23407, 23409, 23411, 23412, 23449, 23450, 23451, 23452, 23453, 23454, 23455, 23456, 23457, 23458, 23459, 23460, 23461, 23462, 23463, 23464, 23465, 23467, 23468, 23469, 23485, 23486, 23487, 23488, 23489, 23490, 23491, 23492, 23493, 23494, 23495, 23496, 23497, 23498, 23499, 23500, 23501, 23502, 23503, 23504, 23505, 23506, 23507, 23508, 23509, 23510, 23511, 23512, 23513, 23514, 23515, 23516, 23517, 23518, 23519, 23520, 23521, 23522, 23523, 23524, 23525, 23526, 23527, 23528, 23529, 23530, 23531, 23532, 23533, 23534, 23535, 23536, 23537, 23540, 23541, 23542, 23543, 23544, 23545, 23546, 23547, 23548, 23549, 23550, 23551, 23552, 23553, 23554, 23555, 23556, 23557, 23558, 23559, 23560, 23561, 23562, 23563, 23564, 23565, 23566, 23567, 23568, 23569, 23570, 23571, 23572, 23573, 23574, 23575, 23576, 23577, 23578, 23579, 23580, 23581, 23582, 23583, 23584, 23585, 23586, 23587, 23588, 23589, 23590, 23591, 23592, 23593, 23594, 23595, 23596, 23597, 23598, 23599, 23600, 23602, 23603, 23604, 23605, 23606, 23607, 23608, 23609, 23610, 23611, 23612, 23613, 23614, 23615, 23616, 23617, 23618, 23619, 23620, 23621, 23622, 23623, 23624, 23625, 23626, 23627, 23628, 23629, 23630, 23631, 23632, 23633, 23634, 23635, 23636, 23637, 23638, 23639, 23640, 23641, 23642, 23643, 23644, 23645, 23646, 23647, 23648, 23649, 23650, 23651, 23652, 23653, 23654, 23655, 23656, 23657, 23658, 23659, 23660, 23661, 23663, 23664, 23665, 23666, 23667, 23668, 23669, 23670, 23671, 23672, 23673, 23674, 23675, 23676, 23677, 23678, 23679, 23680, 23681, 23682, 23701, 23705, 23710, 23711, 23712, 23713, 23714, 23715, 23716, 23717, 23718, 23719, 23720, 23721, 23722, 23723, 23724, 23725, 23726, 23727, 23728, 23729, 23730, 23731, 23791, 23792, 23793, 23794, 23795, 23796, 23797, 23798, 23799, 23800, 23801, 23803, 23804, 23805, 23806, 23807, 23808, 23809, 23810, 23811, 23812, 23813, 23814, 23815, 23816, 23817, 23818, 23819, 23820, 23821, 23822, 23823, 23824, 23825, 23827, 23828, 23829, 23860, 23861, 23862, 23863, 23864, 23865, 23866, 23867, 23868, 23869, 23870, 23871, 23875, 23876, 23877, 23878, 23879, 23883, 23889, 23892, 23894, 23898, 23899, 23900, 23901, 23902, 23903, 23904, 23907, 23910, 23911, 23912, 23915, 23917, 23918, 23919, 23920, 23922, 23923, 23924, 23926, 23927, 23928, 23929, 23930, 23931, 23932, 23933, 23934, 23935, 23936, 23937, 23938, 23940, 23941, 23942, 23943, 23944, 23945, 23967, 23970, 23973, 23974, 23976, 23977, 23979, 23980, 23981, 23987, 23988, 23991, 24000, 24002, 24041, 24042, 24043, 24044, 24045, 24046, 24047, 24048, 24049, 24050, 24051, 24052, 24053, 24054, 24055, 24056, 24057, 24058, 24059, 24060, 24061, 24062, 24063, 24064, 24065, 24066, 24067, 24068, 24069, 24070, 24071, 24072, 24073, 24074, 24075, 24076, 24077, 24078, 24079, 24080, 24081, 24082, 24083, 24084, 24085, 24086, 24087, 24088, 24089, 24090, 24091, 24092, 24093, 24094, 24095, 24096, 24097, 24098, 24099, 24100, 24101, 24102, 24103, 24104, 24105, 24106, 24107, 24108, 24109, 24110, 24111, 24137, 24138, 24139, 24141, 24142, 24144, 24145, 24146, 24148, 24150, 24151, 24153, 24154, 24155, 24156, 24157, 24158, 24159, 24163, 24164, 24165, 24166, 24167, 24168, 24169, 24170, 24171, 24172, 24173, 24174, 24175, 24176, 24177, 24178, 24179, 24180, 24181, 24182, 24183, 24184, 24185, 24189, 24190, 24196, 24197, 24203, 24204, 24206, 24210, 24211, 24212, 24215, 24216, 24220, 24224, 24225, 24226, 24227, 24228, 24229, 24230, 24231, 24232, 24233, 24234, 24235, 24236, 24238, 24239, 24240, 24242, 24244, 24245, 24248, 24249, 24250, 24251, 24252, 24253, 24254, 24256, 24257, 24258, 24259, 24260, 24262, 24263, 24265, 24266, 24267, 24268, 24269, 24272, 24275, 24276, 24278, 24281, 24283, 24284, 24314, 24316, 24317, 24320, 24325, 24419, 24420, 24422, 24425, 24429, 24430, 24431, 24446, 24447, 24456, 24457, 24458, 24461, 24463, 24464, 24465, 24467, 24470, 24473, 24475, 24476, 24477, 24478, 24479, 24482, 24483, 24484, 24485, 24487, 24488, 24500, 24501, 24510, 24512, 24519, 24521, 24523, 24524, 24527, 24528, 24530, 24533, 24536, 24537, 24539, 24542, 24545, 24546, 24548, 24551, 24554, 24555, 24556, 24557, 24560, 24564, 24565, 24566, 24569, 24572, 24575, 24578, 24581, 24584, 24585, 24586, 24587, 24590, 24591, 24592, 24593, 24594, 24595, 24596, 24599, 24600, 24601, 24602, 24603, 24604, 24605, 24608, 24609, 24611, 24613, 24614, 24617, 24618, 24619, 24620, 24621, 24622, 24623, 24626, 24627, 24628, 24629, 24632, 24635, 24636, 24637, 24638, 24639, 24640, 24641, 24643, 24644, 24646, 24647, 24650, 24652, 24653, 24655, 24656, 24658, 24659, 24661, 24662, 24663, 24665, 24667, 24668, 24686, 24689, 24691, 24692, 24693, 24694, 24824, 24827, 24828, 24830, 24833, 24857, 24860, 24861, 24863, 24865, 24866, 24869, 24872, 24873, 24874, 24875, 24878, 24881, 24882, 24883, 24884, 24885, 24887, 24888, 24890, 24891, 24892, 24893, 24896, 24897, 24898, 24899, 24968, 24969, 24971, 24974, 24977, 24986, 24987, 24988, 24989, 24991, 24992, 24995, 24996, 24998, 24999, 25001, 25004, 25005, 25007, 25008, 25010, 25013, 25014, 25016, 25017, 25019, 25022, 25023, 25025, 25028, 25031, 25034, 25035, 25070, 25073, 25082, 25085, 25088, 25091, 25094, 25095, 25097, 25100, 25103, 25104, 25106, 25107, 25109, 25112, 25115, 25118, 25121, 25122, 25124, 25125, 25127, 25130, 25133, 25134, 25136, 25137, 25138, 25139, 25140, 25141, 25142, 25143, 25144, 25145, 25146, 25147, 25149, 25150, 25151, 25173, 25174, 25175, 25176, 25177, 25178, 25179, 25180, 25181, 25182, 25183, 25184, 25185, 25186, 25187, 25188, 25189, 25190, 25191, 25192, 25193, 25195, 25196, 25197, 25248, 25293, 25294, 25295, 25296, 25297, 25299, 25300, 25301, 25302, 25303, 25304, 25305, 25306, 25307, 25308, 25309, 25310, 25311, 25312, 25313, 25314, 25315, 25316, 25317, 25318, 25319, 25320, 25321, 25322, 25323, 25324, 25325, 25326, 25327, 25328, 25329, 25330, 25331, 25332, 25333, 25334, 25335, 25336, 25337, 25338, 25339, 25340, 25395, 25396, 25397, 25398, 25399, 25400, 25401, 25402, 25403, 25404, 25405, 25406, 25407, 25418, 25419, 25420, 25421, 25422, 25423, 25424, 25425, 25427, 25434, 25435, 25436, 25437, 25438, 25439, 25440, 25441, 25442, 25443, 25444, 25445, 25452, 25453, 25454, 25455, 25456, 25457, 25458, 25459, 25460, 25461, 25462, 25463, 25464, 25465, 25466, 25467, 25468, 25469, 25470, 25471, 25472, 25473, 25474, 25475, 25476, 25477, 25478, 25479, 25480, 25484, 25485, 25486, 25487, 25489, 25490, 25491, 25492, 25553, 25554, 25555, 25556, 25557, 25560, 25561, 25562, 25563, 25565, 25566, 25567, 25568, 25569, 25570, 25571, 25572, 25573, 25574, 25575, 25576, 25579, 25580, 25581, 25582, 25584, 25585, 25586, 25627, 25628, 25629, 25630, 25631, 25632, 25633, 25634, 25635, 25636, 25637, 25638, 25639, 25640, 25641, 25642, 25643, 25644, 25661, 25662, 25663, 25674, 25675, 25676, 25677, 25678, 25679, 25680, 25681, 25682, 25683, 25684, 25685, 25686, 25687, 25688, 25689, 25690, 25692, 25693, 25694, 25696, 25697, 25698, 25699, 25700, 25701, 25702, 25703, 25704, 25705, 25706, 25707, 25708, 25709, 25710, 25711, 25712, 25713, 25714, 25715, 25716, 25717, 25718, 25719, 25720, 25721, 25722, 25723, 25724, 25725, 25727, 25778, 25779, 25780, 25781, 25782, 25783, 25784, 25785, 25786, 25787, 25788, 25789, 25790, 25791, 25792, 25793, 25794, 25798, 25799, 25800, 25801, 25802, 25803, 25804, 25805, 25806, 25808, 25809, 25810, 25811, 25812, 25813, 25814, 25815, 25816, 25817, 25818, 25819, 25820, 25821, 25822, 25823, 25824, 25825, 25827, 25829, 25830, 25831, 25832, 25833, 25834, 25835, 25836, 25837, 25838, 25839, 25840, 25841, 25842, 25843, 25844, 25845, 25846, 25847, 25848, 25849, 25850, 25851, 25852, 25854, 25855, 25856, 25857, 25858, 25859, 25860, 25861, 25862, 25863, 25864, 25865, 25878, 25879, 25880, 25881, 25882, 25892, 25893, 25894, 25895, 25896, 25897, 25898, 25899, 25900, 25901, 25902, 25903, 25904, 25905, 25906, 25907, 25908, 25909, 25910, 25911, 25912, 25913, 25914, 25915, 25918, 25919, 25920, 25921, 25922, 25923, 25924, 25925, 25926, 25927, 25928, 25929, 25930, 25931, 25932, 25933, 25935, 25936, 25937, 25938, 25939, 25940, 25941, 25944, 25945, 25956, 25957, 25958, 25960, 25961, 25962, 25963, 25964, 25965, 25966, 25967, 25968, 25969, 25970, 25971, 25972, 26024, 26064, 26065, 26067, 26068, 26078, 26079, 26091, 26092, 26093, 26094, 26095, 26096, 26097, 26098, 26104, 26105, 26106, 26107, 26108, 26109, 26110, 26111, 26112, 26113, 26114, 26115, 26117, 26121, 26122, 26123, 26124, 26125, 26126, 26127, 26128, 26129, 26130, 26131, 26132, 26133, 26134, 26135, 26136, 26140, 26144, 26145, 26147, 26148, 26149, 26150, 26152, 26158, 26159, 26168, 26169, 26170, 26171, 26319, 26320, 26321, 26322, 26324, 26325, 26326, 26327, 26328, 26329, 26330, 26331, 26332, 26334, 26335, 26336, 26337, 26338, 26341, 26343, 26347, 26348, 26349, 26350, 26351, 26352, 26353, 26355, 26356, 26358, 26359, 26360, 26361, 26362, 26363, 26364, 26365, 26366, 26367, 26368, 26369, 26370, 26371, 26372, 26373, 26374, 26375, 26376, 26378, 26379, 26380, 26381, 26382, 26383, 26384, 26385, 26386, 26388, 26389, 26393, 26394, 26395, 26410, 26411, 26412, 26413, 26414, 26415, 26416, 26417, 26418, 26419, 26420, 26422, 26423, 26424, 26425, 26426, 26427, 26428, 26429, 26431, 26432, 26433, 26434, 26441, 26442, 26443, 26444, 26445, 26446, 26447, 26448, 26449, 26451, 26452, 26454, 26455, 26456, 26457, 26458, 26459, 26460, 26461, 26462, 26463, 26464, 26465, 26467, 26468, 26469, 26470, 26471, 26472, 26473, 26474, 26475, 26476, 26477, 26478, 26479, 26480, 26491, 26492, 26493, 26494, 26495, 26496, 26497, 26498, 26499, 26500, 26501, 26502, 26503, 26504, 26505, 26506, 26507, 26508, 26509, 26510, 26511, 26512, 26513, 26514, 26515, 26516, 26518, 26519, 26520, 26527, 26528, 26529, 26530, 26531, 26532, 26533, 26534, 26535, 26536, 26537, 26538, 26539, 26540, 26541, 26542, 26543, 26544, 26545, 26546, 26557, 26558, 26659, 26660, 26661, 26662, 26663, 26664, 26665, 26666, 26667, 26668, 26669, 26670, 26671, 26672, 26673, 26674, 26677, 26678, 26679, 26680, 26681, 26682, 26683, 26684, 26685, 26686, 26687, 26688, 26689, 26690, 26691, 26692, 26693, 26694, 26695, 26696, 26697, 26698, 26699, 26700, 26702, 26703, 26704, 26705, 26706, 26707, 26708, 26709, 26710, 26711, 26712, 26713, 26714, 26715, 26716, 26717, 26718, 26719, 26720, 26721, 26722, 26723, 26724, 26725, 26726, 26727, 26728, 26729, 26730, 26731, 26732, 26733, 26734, 26735, 26736, 26737, 26738, 26739, 26740, 26741, 26742, 26743, 26744, 26745, 26746, 26747, 26748, 26749, 26750, 26751, 26752, 26753, 26754, 26755, 26756, 26757, 26773, 26774, 26775, 26776, 26777, 26778, 26779, 26780, 26784, 26785, 26786, 26787, 26788, 26789, 26790, 26791, 26793, 26794, 26795, 26796, 26797, 26798, 26799, 26800, 26801, 26802, 26803, 26804, 26805, 26806, 26807, 26808, 26809, 26810, 26811, 26812, 26813, 26814, 26815, 26816, 26817, 26818, 26819, 26820, 26821, 26822, 26823, 26824, 26825, 26826, 26827, 26828, 26829, 26830, 26831, 26832, 26833, 26834, 26835, 26836, 26837, 26838, 26839, 26840, 26841, 26842, 26843, 26844, 26845, 26846, 26847, 26848, 26850, 26851, 26852, 26853, 26854, 26855, 26866, 26867, 26868, 26869, 26870, 26871, 26872, 26873, 26874, 26875, 26876, 26877, 26878, 26879, 26880, 26881, 26882, 26883, 26884, 26885, 26886, 26887, 26888, 26889, 26890, 26891, 26892, 26893, 26894, 26895, 26896, 26897, 26899, 26900, 26901, 26902, 26903, 26904, 26905, 26906, 26907, 26908, 26909, 26910, 26911, 26912, 26913, 26914, 26915, 26916, 26917, 26918, 26919, 26920, 26921, 26922, 26923, 26924, 26925, 26926, 26927, 26928, 26929, 26930, 26931, 26932, 26933, 26934, 26935, 26936, 26937, 26943, 26944, 26945, 26946, 26947, 26948, 26949, 26950, 26951, 26952, 26953, 26954, 26955, 26956, 26957, 26958, 26959, 26960, 26961, 26962, 26963, 26964, 26965, 26966, 26967, 26968, 26969, 26970, 26971, 26972, 26973, 26974, 26975, 26976, 26977, 27028, 27029, 27030, 27031, 27033, 27034, 27035, 27036, 27037, 27038, 27039, 27050, 27052, 27053, 27055, 27056, 27057, 27058, 27059, 27060, 27061, 27062, 27063, 27110, 27111, 27112, 27113, 27114, 27115, 27116, 27117, 27118, 27119, 27120, 27121, 27122, 27123, 27124, 27125, 27126, 27127, 27128, 27129, 27130, 27131, 27138, 27139, 27140, 27141, 27142, 27143, 27144, 27145, 27146, 27147, 27198, 27199, 27200, 27201, 27202, 27203, 27204, 27205, 27206, 27207, 27208, 27209, 27210, 27211, 27212, 27213, 27214, 27215, 27228, 27229, 27230, 27231, 27232, 27233, 27234, 27235, 27237, 27238, 27239, 27240, 27241, 27242, 27243, 27244, 27245, 27246, 27247, 27248, 27249, 27250, 27251, 27252, 27255, 27256, 27257, 27258, 27259, 27260, 27276, 27277, 27278, 27279, 27280, 27281, 27282, 27283, 27284, 27285, 27286, 27287, 27288, 27289, 27290, 27291, 27292, 27293, 27294, 27295, 27296, 27297, 27298, 27299, 27300, 27301, 27302, 27303, 27304, 27305, 27306, 27307, 27308, 27309, 27310, 27311, 27312, 27313, 27314, 27315, 27416, 27417, 27418, 27419, 27420, 27421, 27422, 27423, 27424, 27425, 27426, 27427, 27428, 27429, 27430, 27431, 27432, 27433, 27434, 27435, 27436, 27437, 27438, 27439, 27440, 27441, 27442, 27443, 27444, 27445, 27446, 27447, 27448, 27449, 27450, 27451, 27452, 27453, 27454, 27455, 27456, 27457, 27458, 27459, 27460, 27461, 27462, 27463, 27464, 27465, 27476, 27477, 27478, 27479, 27480, 27481, 27482, 27483, 27484, 27485, 27486, 27487, 27488, 27489, 27490, 27491, 27493, 27494, 27495, 27496, 27497, 27498, 27499, 27500, 27501, 27502, 27503, 27504, 27513, 27514, 27515, 27516, 27517, 27518, 27519, 27520, 27521, 27522, 27523, 27524, 27525, 27526, 27527, 27528, 27529, 27530, 27531, 27532, 27533, 27534, 27535, 27536, 27537, 27538, 27539, 27540, 27541, 27547, 27548, 27549, 27550, 27551, 27552, 27553, 27554, 27555, 27556, 27557, 27558, 27559, 27560, 27561, 27562, 27563, 27564, 27565, 27566, 27567, 27568, 27569, 27570, 27571, 27572, 27573, 27574, 27575, 27580, 27581, 27582, 27583, 27584, 27585, 27586, 27587, 27588, 27589, 27590, 27591, 27592, 27594, 27595, 27596, 27597, 27598, 27599, 27616, 27617, 27618, 27619, 27620, 27621, 27622, 27623, 27624, 27625, 27626, 27627, 27628, 27629, 27630, 27631, 27632, 27639, 27640, 27641, 27642, 27643, 27644, 27645, 27646, 27647, 27648, 27649, 27652, 27653, 27654, 27655, 27656, 27657, 27658, 27659, 27660, 27661, 27662, 27663, 27664, 27665, 27666, 27667, 27668, 27669, 27670, 27671, 27687, 27688, 27689, 27690, 27691, 27692, 27693, 27694, 27695, 27696, 27697, 27698, 27699, 27700, 27701, 27702, 27703, 27704, 27705, 27706, 27707, 27708, 27709, 27710, 27712, 27713, 27714, 27715, 27716, 27717, 27718, 27719, 27720, 27721, 27722, 27723, 27724, 27725, 27726, 27727, 27728, 27729, 27730, 27731, 27732, 27733, 27734, 27735, 27736, 27737, 27738, 27739, 27740, 27741, 27742, 27743, 27744, 27745, 27746, 27747, 27748, 27749, 27750, 27751, 27752, 27753, 27754, 27755, 27756, 27757, 27758, 27759, 27760, 27761, 27762, 27763, 27764, 27765, 27766, 27767, 27768, 27769, 27869, 27870, 27921, 27922, 27923, 27924, 27925, 27926, 27927, 27928, 27929, 27932, 27933, 27934, 27935, 27936, 27937, 27939, 27940, 27941, 27942, 27943, 27944, 27946, 27947, 27948, 27949, 27950, 27951, 27952, 27953, 27954, 27955, 27956, 27957, 27958, 27959, 27960, 27961, 27962, 27963, 28029, 28030, 28031, 28032, 28033, 28034, 28035, 28036, 28037, 28038, 28039, 28040, 28041, 28042, 28043, 28044, 28045, 28046, 28047, 28048, 28049, 28050, 28051, 28102, 28103, 28104, 28105, 28106, 28107, 28108, 28109, 28110, 28111, 28112, 28113, 28114, 28115, 28116, 28117, 28118, 28119, 28120, 28121, 28122, 28123, 28124, 28125, 28126, 28127, 28128, 28129, 28130, 28131, 28132, 28133, 28134, 28135, 28136, 28137, 28138, 28139, 28140, 28141, 28142, 28143, 28148, 28149, 28150, 28151, 28152, 28153, 28154, 28155, 28156, 28157, 28162, 28163, 28164, 28165, 28166, 28167, 28168, 28169, 28170, 28171, 28172, 28173, 28174, 28175, 28176, 28177, 28178, 28186, 28187, 28188, 28189, 28194, 28195, 28196, 28197, 28198, 28199, 28200, 28201, 28202, 28203, 28204, 28205, 28206, 28207, 28208, 28209, 28210, 28211, 28228, 28229, 28230, 28231, 28232, 28233, 28234, 28235, 28236, 28237, 28238, 28239, 28241, 28242, 28243, 28244, 28245, 28246, 28247, 28248, 28252, 28253, 28254, 28255, 28256, 28257, 28258, 28259, 28260, 28261, 28262, 28263, 28264, 28265, 28266, 28267, 28268, 28269, 28270, 28271, 28272, 28273, 28274, 28275, 28276, 28277, 28278, 28279, 28283, 28284, 28285, 28286, 28287, 28288, 28289, 28290, 28291, 28292, 28293, 28294, 28295, 28296, 28297, 28298, 28361, 28362, 28363, 28364, 28365, 28366, 28367, 28368, 28369, 28370, 28371, 28372, 28373, 28374, 28375, 28376, 28377, 28378, 28379, 28383, 28384, 28385, 28386, 28387, 28388, 28389, 28390, 28406, 28407, 28408, 28409, 28410, 28411, 28412, 28413, 28414, 28415, 28416, 28417, 28418, 28419, 28420, 28421, 28422, 28425, 28426, 28427, 28428, 28429, 28430, 28433, 28434, 28435, 28436, 28437, 28438, 28464, 28465, 28466, 28467, 28468, 28469, 28470, 28471, 28472, 28473, 28474, 28475, 28482, 28483, 28484, 28485, 28486, 28487, 28488, 28489, 28490, 28491, 28492, 28493, 28494, 28495, 28496, 28497, 28498, 28499, 28500, 28501, 28513, 28514, 28515, 28517, 28518, 28519, 28520, 28521, 28522, 28523, 28524, 28525, 28526, 28527, 28528, 28529, 28530, 28531, 28532, 28533, 28534, 28535, 28536, 28537, 28538, 28539, 28540, 28541, 28542, 28543, 28544, 28545, 28546, 28547, 28552, 28553, 28554, 28555, 28556, 28557, 28558, 28559, 28560, 28561, 28562, 28563, 28564, 28567, 28568, 28569, 28570, 28571, 28572, 28573, 28574, 28575, 28576, 28577, 28578, 28579, 28580, 28581, 28582, 28583, 28584, 28585, 28586, 28587, 28588, 28589, 28590, 28591, 28592, 28593, 28594, 28595, 28596, 28597, 28598, 28599, 28600, 28601, 28602, 28603, 28604, 28605, 28606, 28607, 28608, 28609, 28610, 28611, 28612, 28613, 28614, 28615, 28616, 28617, 28618, 28619, 28620, 28621, 28622, 28623, 28624, 28625, 28626, 28627, 28628, 28629, 28630, 28631, 28632, 28633, 28634, 28635, 28659, 28660, 28661, 28663, 28677, 28678, 28681, 28682, 28683, 50099, 50102, 50103, 50104, 50105, 50106, 50107, 50108, 50109, 50110, 50111, 50112, 50113, 50114, 50115, 50116, 50117, 50118, 50119, 50120, 50121, 50122, 50123, 50164, 50166, 50168, 50169, 50170, 50171, 50172, 50173, 50174, 50175, 50176, 50177, 50178, 50179, 50180, 50181, 50182, 50183, 50184, 50185, 50186, 50187, 50188, 50192, and 50193. ******************************************************************************************* Serial numbers of units distributed OUTSIDE the US: .23323, .23329, .23330, .23333, .23340, .23408, .23986, 10673, 10674, 10675, 10676, 10838, 10839, 10840, 10890, 10904, 10932, 11077, 11096, 11097, 11098, 11099, 11101, 12011, 12019, 12210, 12214, 12215, 12216, 12217, 12218, 12219, 12220, 12221, 12222, 12223, 12224, 12225, 12280, 12281, 12282, 12283, 12284, 12285, 12286, 12287, 12288, 12289, 12350, 12408, 12409, 12641, 12980, 13068, 13190, 13438, 13440, 13812, 13813, 13814, 13856, 13857, 13858, 13859, 13863, 13864, 13865, 13867, 13875, 13876, 13888, 13889, 13890, 13891, 13892, 13893, 13894, 13895, 13896, 13897, 13924, 13925, 13926, 13927, 13928, 13929, 13930, 13931, 13933, 13937, 13938, 13939, 13940, 13941, 13942, 13943, 13946, 13947, 13948, 13949, 13950, 13951, 13952, 13953, 13954, 13955, 13956, 13957, 13958, 13969, 13970, 14022, 14023, 14167, 14170, 14171, 14172, 14173, 14241, 14253, 14254, 14255, 14256, 14257, 14258, 14259, 14260, 14261, 14262, 14263, 14264, 14265, 14266, 14267, 14268, 14269, 14301, 14326, 14328, 14329, 14330, 14331, 14332, 14333, 14334, 14335, 14336, 14337, 14338, 14358, 14359, 14369, 14370, 14379, 14383, 14384, 14419, 14421, 14432, 14433, 14434, 14435, 14436, 14437, 14438, 14439, 14440, 14441, 14448, 14469, 14470, 14471, 14472, 14475, 14506, 14509, 14562, 14563, 14566, 14570, 14575, 14628, 14629, 14630, 14631, 14632, 14634, 14635, 14636, 14637, 14638, 14639, 14640, 14641, 14643, 14645, 14663, 14671, 14680, 14701, 14702, 14703, 14711, 14769, 14770, 14771, 14772, 14773, 14774, 14775, 14776, 14777, 14778, 14779, 14780, 14781, 14782, 14783, 14784, 14785, 14786, 14787, 14788, 14809, 14811, 14842, 14843, 14844, 14845, 14846, 14847, 14848, 14849, 14859, 14860, 14861, 14862, 14863, 14864, 14865, 14866, 14867, 15031, 15041, 15042, 15043, 15044, 15045, 15046, 15047, 15048, 15049, 15051, 15052, 15053, 15054, 15055, 15056, 15057, 15058, 15059, 15060, 15156, 15213, 15306, 15307, 15308, 15309, 15310, 15311, 15351, 15523, 15549, 15557, 15558, 15559, 15560, 15561, 15562, 15563, 15564, 15565, 15566, 15567, 15568, 15569, 15570, 15571, 15572, 15573, 15574, 15575, 15576, 15577, 15578, 15579, 15580, 15581, 15582, 15583, 15584, 15585, 15586, 15587, 15588, 15590, 15591, 15592, 15593, 15594, 15595, 15596, 15597, 15598, 15599, 15600, 15634, 15635, 15636, 15637, 15638, 15639, 15640, 15641, 15642, 15643, 15644, 15645, 15683, 15684, 15685, 15686, 15687, 15688, 15689, 15690, 15691, 15692, 15693, 15694, 15695, 15696, 15697, 15698, 15699, 15700, 15701, 15702, 15703, 15704, 15705, 15706, 15707, 15708, 15709, 15710, 15722, 15723, 15724, 15725, 15726, 15727, 15728, 15729, 15730, 15731, 15732, 15733, 15734, 15738, 15739, 15740, 15741, 15742, 15743, 15744, 15745, 15746, 15747, 15748, 15749, 15750, 15751, 15752, 15753, 15754, 15755, 15756, 15757, 15758, 15759, 15760, 15761, 15762, 15763, 15764, 15765, 15766, 15767, 15768, 15770, 15771, 15772, 15773, 15774, 15775, 15777, 15778, 15779, 15780, 15781, 15782, 15783, 15784, 15785, 15786, 15790, 15791, 15792, 15793, 15794, 15795, 15796, 16505, 16521, 16522, 16523, 16525, 16526, 16527, 16537, 16538, 16539, 16540, 16541, 16542, 16544, 16545, 16546, 16547, 16548, 16549, 16550, 16551, 16552, 16553, 16554, 16555, 16556, 16557, 16558, 16559, 16560, 16561, 16562, 16563, 16564, 16565, 16566, 16567, 16568, 16569, 16570, 16571, 16572, 16573, 16574, 16575, 16576, 16577, 16579, 16580, 16581, 16582, 16583, 16584, 16585, 16586, 16587, 16588, 16589, 16590, 16591, 16592, 16593, 16594, 16595, 16596, 16597, 16598, 16599, 16600, 16601, 16602, 16603, 16604, 16606, 16607, 16608, 16609, 16610, 16611, 16612, 16613, 16614, 16615, 16645, 16670, 16671, 16672, 16673, 16674, 16675, 16676, 16677, 16678, 16679, 16680, 16681, 16682, 16683, 16684, 16723, 16727, 16728, 16729, 16730, 16731, 16756, 16757, 16787, 16788, 16789, 16790, 16791, 16792, 16793, 16794, 16795, 16796, 16797, 16798, 16799, 16800, 16801, 16802, 16803, 16804, 16805, 16806, 16829, 16830, 16831, 16832, 16844, 16845, 16854, 16889, 16890, 16891, 16892, 16893, 16894, 16895, 16896, 16897, 16898, 16899, 16900, 16901, 16902, 16903, 16904, 16905, 16906, 16907, 16908, 16909, 16910, 16911, 16912, 16913, 16914, 16915, 16916, 16917, 16918, 16919, 16920, 16921, 16922, 16923, 16924, 16925, 16926, 16927, 16928, 16929, 16930, 16931, 16932, 16933, 16934, 16935, 16936, 16937, 16938, 16944, 16945, 16946, 16947, 16948, 16949, 16950, 17015, 17016, 17017, 17018, 17019, 17020, 17021, 17022, 17023, 17024, 17025, 17026, 17027, 17028, 17029, 17030, 17031, 17032, 17033, 17034, 17035, 17036, 17037, 17038, 17039, 17040, 17041, 17042, 17043, 17044, 17045, 17046, 17047, 17048, 17049, 17050, 17051, 17052, 17053, 17054, 17063, 17064, 17065, 17066, 17067, 17068, 17069, 17070, 17071, 17072, 17102, 17106, 17107, 17108, 17109, 17110, 17111, 17112, 17113, 17114, 17115, 17116, 17117, 17118, 17119, 17120, 17121, 17122, 17123, 17124, 17125, 17126, 17127, 17128, 17129, 17130, 17131, 17132, 17133, 17134, 17135, 17136, 17137, 17138, 17139, 17140, 17141, 17142, 17143, 17144, 17145, 17146, 17147, 17148, 17149, 17150, 17151, 17152, 17153, 17154, 17155, 17223, 17224, 17225, 17226, 17227, 17228, 17229, 17230, 17231, 17232, 17233, 17234, 17235, 17236, 17237, 17238, 17239, 17240, 17241, 17242, 17243, 17244, 17245, 17246, 17247, 17248, 17249, 17250, 17251, 17252, 17253, 17254, 17255, 17256, 17257, 17258, 17259, 17260, 17261, 17262, 17263, 17264, 17265, 17266, 17267, 17268, 17269, 17270, 17271, 17272, 17273, 17274, 17275, 17276, 17277, 17278, 17279, 17280, 17281, 17282, 17283, 17284, 17285, 17286, 17287, 17288, 17289, 17290, 17291, 17292, 17293, 17294, 17295, 17296, 17297, 17298, 17299, 17300, 17301, 17302, 17303, 17304, 17306, 17307, 17308, 17309, 17310, 17311, 17312, 17313, 17314, 17315, 17316, 17317, 17318, 17319, 17320, 17321, 17322, 17323, 17324, 17325, 17326, 17327, 17328, 17329, 17330, 17331, 17332, 17349, 17362, 17363, 17364, 17365, 17366, 17367, 17368, 17369, 17370, 17371, 17378, 17399, 17400, 17401, 17402, 17403, 17406, 17407, 17408, 17409, 17410, 17411, 17412, 17413, 17414, 17415, 17416, 17417, 17419, 17420, 17421, 17422, 17423, 17424, 17425, 17426, 17427, 17428, 17429, 17430, 17432, 17470, 17471, 17472, 17473, 17474, 17475, 17476, 17477, 17478, 17479, 17481, 17482, 17483, 17521, 17522, 17523, 17533, 17534, 17535, 17536, 17537, 17558, 17559, 17560, 17561, 17562, 17563, 17564, 17565, 17566, 17567, 17596, 17602, 17603, 17604, 17605, 17606, 17607, 17608, 17609, 17610, 17611, 17612, 17613, 17614, 17615, 17616, 17617, 17618, 17619, 17620, 17621, 17622, 17623, 17624, 17625, 17626, 17627, 17628, 17629, 17630, 17631, 17632, 17633, 17634, 17635, 17636, 17637, 17638, 17639, 17640, 17641, 17642, 17643, 17644, 17645, 17646, 17647, 17656, 17701, 17706, 17708, 17709, 17710, 17711, 17712, 17713, 17714, 17715, 17716, 17717, 17718, 17719, 17720, 17721, 17722, 17723, 17724, 17725, 17726, 17727, 17728, 17729, 17730, 17731, 17732, 17734, 17735, 17736, 17737, 17738, 17739, 17740, 17741, 17742, 17743, 17744, 17745, 17746, 17747, 17748, 17749, 17750, 17751, 17752, 17753, 17754, 17755, 17756, 17757, 17758, 17759, 17760, 17761, 17762, 17763, 17764, 17765, 17766, 17767, 17768, 17769, 17770, 17771, 17772, 17773, 17774, 17775, 17776, 17777, 17778, 17779, 17780, 17781, 17782, 17783, 17784, 17785, 17786, 17787, 17788, 17789, 17790, 17791, 17792, 17793, 17794, 17839, 17845, 17846, 17847, 17848, 17849, 17850, 17851, 17852, 17853, 17854, 17855, 17856, 17857, 17858, 17859, 17860, 17861, 17862, 17863, 17864, 17865, 17881, 17882, 17883, 17884, 17885, 17886, 17887, 17888, 17889, 17890, 17891, 17892, 17893, 17894, 17908, 17909, 17910, 17911, 17912, 17913, 17914, 17915, 17916, 17917, 17918, 17919, 18024, 18025, 18026, 18027, 18028, 18029, 18030, 18031, 18032, 18033, 18034, 18035, 18036, 18037, 18038, 18039, 18093, 18094, 18095, 18096, 18097, 18098, 18099, 18100, 18101, 18102, 18103, 18104, 18105, 18106, 18107, 18108, 18109, 18110, 18111, 18112, 18113, 18114, 18115, 18116, 18117, 18123, 18124, 18125, 18126, 18127, 18128, 18129, 18130, 18131, 18132, 18135, 18136, 18137, 18138, 18139, 18140, 18141, 18142, 18143, 18144, 18145, 18149, 18150, 18151, 18152, 18153, 18154, 18155, 18156, 18157, 18158, 18159, 18160, 18161, 18162, 18163, 18164, 18165, 18166, 18167, 18168, 18169, 18170, 18171, 18172, 18177, 18178, 18179, 18180, 18181, 18182, 18183, 18184, 18185, 18186, 18187, 18188, 18189, 18190, 18191, 18192, 18193, 18194, 18195, 18196, 18197, 18198, 18199, 18200, 18201, 18202, 18203, 18204, 18205, 18206, 18207, 18208, 18209, 18210, 18211, 18212, 18213, 18214, 18215, 18216, 18217, 18218, 18219, 18220, 18221, 18222, 18223, 18224, 18225, 18226, 18227, 18228, 18229, 18230, 18231, 18232, 18233, 18234, 18235, 18236, 18237, 18238, 18239, 18240, 18241, 18242, 18243, 18244, 18245, 18246, 18247, 18248, 18249, 18250, 18251, 18252, 18253, 18254, 18255, 18256, 18257, 18258, 18259, 18260, 18261, 18262, 18263, 18264, 18265, 18266, 18267, 18268, 18269, 18270, 18271, 18272, 18273, 18274, 18275, 18276, 18277, 18278, 18279, 18280, 18281, 18282, 18283, 18284, 18285, 18286, 18287, 18288, 18289, 18290, 18291, 18292, 18293, 18294, 18295, 18296, 18297, 18298, 18299, 18300, 18301, 18321, 18322, 18323, 18324, 18325, 18326, 18327, 18435, 18436, 18437, 18438, 18440, 18441, 18442, 18443, 18445, 18446, 18447, 18448, 18449, 18450, 18451, 18452, 18453, 18454, 18455, 18456, 18457, 18458, 18459, 18460, 18461, 18462, 18463, 18464, 18465, 18466, 18467, 18468, 18469, 18470, 18471, 18472, 18473, 18474, 18475, 18476, 18477, 18478, 18480, 18481, 18482, 18483, 18484, 18485, 18486, 18487, 18488, 18489, 18490, 18491, 18492, 18493, 18494, 18495, 18496, 18497, 18498, 18499, 18500, 18501, 18502, 18503, 18504, 18505, 18506, 18507, 18508, 18509, 18510, 18511, 18512, 18513, 18514, 18515, 18516, 18517, 18518, 18519, 18520, 18521, 18522, 18523, 18524, 18525, 18526, 18527, 18528, 18529, 18530, 18531, 18532, 18533, 18534, 18535, 18536, 18537, 18538, 18539, 18540, 18541, 18542, 18543, 18544, 18545, 18546, 18547, 18548, 18549, 18550, 18551, 18552, 18553, 18554, 18555, 18556, 18557, 18558, 18559, 18560, 18561, 18562, 18563, 18564, 18565, 18566, 18567, 18568, 18569, 18570, 18571, 18572, 18573, 18574, 18575, 18576, 18577, 18578, 18579, 18580, 18581, 18582, 18583, 18584, 18585, 18586, 18587, 18588, 18589, 18590, 18591, 18592, 18593, 18594, 18595, 18596, 18597, 18598, 18599, 18600, 18601, 18602, 18603, 18604, 18605, 18606, 18607, 18608, 18609, 18610, 18611, 18612, 18613, 18614, 18615, 18616, 18617, 18618, 18619, 18620, 18621, 18622, 18623, 18624, 18625, 18626, 18627, 18628, 18629, 18630, 18631, 18632, 18633, 18634, 18635, 18636, 18637, 18638, 18639, 18640, 18641, 18642, 18643, 18644, 18645, 18646, 18647, 18648, 18649, 18650, 18651, 18652, 18653, 18654, 18655, 18656, 18657, 18658, 18659, 18660, 18661, 18662, 18663, 18715, 18716, 18717, 18721, 18722, 18723, 18724, 18727, 18748, 18749, 18750, 18751, 18752, 18753, 18754, 18755, 18756, 18757, 18758, 18759, 18760, 18761, 18762, 18817, 18818, 18819, 18820, 18821, 18822, 18823, 18824, 18825, 18826, 18827, 18828, 18829, 18830, 18831, 18832, 18833, 18834, 18835, 18836, 18869, 18872, 18882, 18883, 18890, 18891, 18892, 18904, 18905, 18906, 18907, 18908, 18909, 18915, 18916, 18917, 18918, 18940, 18941, 18946, 18947, 18948, 18949, 18950, 18952, 18953, 18955, 18956, 18957, 19012, 19013, 19014, 19015, 19016, 19023, 19024, 19025, 19026, 19027, 19037, 19038, 19039, 19040, 19041, 19042, 19043, 19045, 19046, 19047, 19053, 19054, 19055, 19056, 19057, 19058, 19059, 19060, 19061, 19067, 19071, 19072, 19073, 19074, 19075, 19076, 19077, 19078, 19079, 19080, 19081, 19082, 19083, 19084, 19089, 19090, 19091, 19092, 19093, 19094, 19095, 19096, 19097, 19098, 19099, 19125, 19135, 19140, 19144, 19145, 19148, 19149, 19151, 19152, 19153, 19154, 19155, 19156, 19157, 19158, 19159, 19160, 19161, 19162, 19163, 19164, 19165, 19166, 19167, 19168, 19169, 19170, 19171, 19172, 19173, 19174, 19175, 19176, 19177, 19178, 19179, 19180, 19181, 19182, 19183, 19184, 19185, 19186, 19187, 19188, 19189, 19190, 19191, 19192, 19193, 19194, 19195, 19196, 19197, 19198, 19199, 19200, 19201, 19202, 19203, 19204, 19205, 19206, 19207, 19208, 19209, 19210, 19211, 19212, 19213, 19214, 19215, 19216, 19217, 19218, 19219, 19220, 19221, 19222, 19223, 19224, 19225, 19226, 19227, 19228, 19229, 19230, 19231, 19232, 19233, 19234, 19235, 19236, 19237, 19238, 19239, 19240, 19241, 19242, 19243, 19244, 19245, 19246, 19247, 19248, 19249, 19250, 19251, 19252, 19253, 19254, 19255, 19256, 19257, 19258, 19259, 19260, 19261, 19262, 19263, 19264, 19265, 19266, 19267, 19268, 19269, 19270, 19271, 19272, 19273, 19274, 19275, 19276, 19277, 19278, 19279, 19280, 19281, 19282, 19283, 19284, 19285, 19286, 19287, 19288, 19289, 19290, 19291, 19292, 19293, 19294, 19295, 19296, 19297, 19298, 19299, 19300, 19301, 19302, 19303, 19304, 19305, 19306, 19307, 19308, 19309, 19310, 19311, 19312, 19313, 19314, 19315, 19316, 19317, 19318, 19346, 19347, 19348, 19349, 19350, 19351.19352, 19353, 19354, 19355, 19356, 19357, 19358, 19359, 19360, 19371, 19372, 19407, 19408, 19424, 19425, 19426, 19427, 19428, 19429, 19430, 19431, 19432, 19433, 19434, 19435, 19436, 19437, 19438, 19439, 19440, 19441, 19442, 19443, 19444, 19445, 19446, 19447, 19448, 19449, 19450, 19451, 19452, 19453, 19454, 19455, 19456, 19457, 19458, 19459, 19460, 19461, 19462, 19463, 19464, 19465, 19466, 19467, 19468, 19469, 19470, 19471, 19472, 19473, 19474, 19475, 19476, 19477, 19478, 19479, 19480, 19481, 19482, 19483, 19484, 19485, 19486, 19487, 19488, 19489, 19490, 19491, 19492, 19493, 19494, 19495, 19496, 19497, 19498, 19499, 19500, 19501, 19502, 19503, 19504, 19505, 19506, 19507, 19508, 19509, 19510, 19511, 19512, 19513, 19514, 19515, 19516, 19517, 19518, 19520, 19521, 19522, 19523, 19524, 19525, 19526, 19527, 19528, 19529, 19530, 19531, 19532, 19533, 19534, 19535, 19536, 19537, 19538, 19539, 19540, 19541, 19542, 19543, 19544, 19545, 19546, 19547, 19548, 19549, 19550, 19551, 19552, 19553, 19554, 19555, 19556, 19557, 19558, 19559, 19560, 19561, 19562, 19563, 19564, 19565, 19566, 19567, 19568, 19569, 19570, 19571, 19656, 19657, 19658, 19659, 19660, 19661, 19662, 19663, 19664, 19665, 19666, 19667, 19668, 19669, 19670, 19671, 19672, 19673, 19674, 19675, 19676, 19677, 19678, 19679, 19680, 19681, 19682, 19683, 19684, 19685, 19686, 19687, 19688, 19689, 19690, 19691, 19692, 19693, 19694, 19695, 19696, 19697, 19698, 19699, 19700, 19701, 19702, 19703, 19704, 19705, 19706, 19707, 19708, 19709, 19710, 19711, 19712, 19713, 19714, 19715, 19716, 19717, 19718, 19719, 19720, 19721, 19722, 19723, 19724, 19725, 19726, 19727, 19728, 19729, 19730, 19731, 19732, 19733, 19734, 19735, 19736, 19737, 19738, 19739, 19740, 19741, 19742, 19743, 19744, 19745, 19746, 19747, 19748, 19749, 19750, 19751, 19752, 19753, 19754, 19755, 19757, 19758, 19759, 19760, 19761, 19762, 19763, 19764, 19765, 19766, 19767, 19768, 19769, 19770, 19771, 19772, 19773, 19774, 19775, 19776, 19777, 19778, 19779, 19780, 19781, 19782, 19783, 19784, 19785, 19786, 19787, 19788, 19789, 19790, 19791, 19792, 19793, 19794, 19795, 19796, 19797, 19798, 19799, 19800, 19801, 19802, 19803, 19804, 19805, 19806, 19807, 19808, 19809, 19810, 19811, 19812, 19813, 19814, 19815, 19816, 19817, 19818, 19819, 19820, 19821, 19822, 19823, 19824, 19825, 19826, 19827, 19828, 19829, 19830, 19831, 19832, 19833, 19834, 19835, 19836, 19837, 19838, 19839, 19840, 19841, 19842, 19843, 19844, 19845, 19846, 19847, 19848, 19849, 19850, 19851, 19852, 19853, 19854, 19855, 19856, 19857, 19858, 19859, 19860, 19861, 19862, 19863, 19864, 19865, 19866, 19867, 19868, 19869, 19870, 19871, 19872, 19873, 19874, 19875, 19876, 19877, 19878, 19879, 19880, 19881, 19882, 19883, 19884, 19885, 19886, 19887, 19888, 19889, 19890, 19891, 20512, 20513, 20514, 20515, 20516, 20517, 20518, 20519, 20520, 20521, 20522, 20523, 20524, 20525, 20526, 20527, 20528, 20529, 20530, 20531, 20532, 20533, 20534, 20535, 20536, 20537, 20538, 20539, 20540, 20541, 20542, 20543, 20544, 20545, 20546, 20547, 20548, 20549, 20550, 20551, 20552, 20553, 20554, 20555, 20556, 20557, 20558, 20559, 20560, 20561, 20562, 20563, 20564, 20565, 20566, 20567, 20568, 20569, 20570, 20571, 20572, 20573, 20574, 20575, 20576, 20577, 20578, 20579, 20580, 20581, 20582, 20583, 20584, 20585, 20586, 20587, 20588, 20589, 20590, 20591, 20592, 20593, 20594, 20595, 20596, 20597, 20598, 20599, 20600, 20601, 20602, 20603, 20604, 20605, 20606, 20607, 20608, 20609, 20610, 20611, 20612, 20613, 20614, 20615, 20616, 20617, 20618, 20619, 20620, 20621, 20622, 20623, 20624, 20625, 20626, 20627, 20628, 20629, 20630, 20631, 20632, 20633, 20634, 20635, 20636, 20678, 20679, 20680, 20681, 20682, 20683, 20684, 20685, 20686, 20687, 20688, 20689, 20690, 20691, 20692, 20693, 20694, 20695, 20696, 20697, 20698, 20699, 20700, 20701, 20702, 20703, 20704, 20705, 20706, 20707, 20708, 20709, 20730, 20731, 20732, 20733, 20734, 20735, 20736, 20737, 20738, 20739, 20740, 20741, 20742, 20743, 20744, 20745, 20748, 20753, 20754, 20755, 20756, 20757, 20758, 20759, 20760, 20842, 20843, 20844, 20845, 20846, 20847, 20848, 20849, 20850, 20851, 20896, 20899, 20900, 20902, 20903, 20906, 20908, 20909, 20911, 20913, 20914, 20915, 20919, 20920, 20921, 20938, 20939, 20940, 20941, 20942, 20943, 20944, 20945, 20948, 20996, 20997, 20998, 20999, 21000, 21009, 21010, 21011, 21012, 21013, 21014, 21064, 21065, 21066, 21067, 21069, 21070, 21071, 21072, 21073, 21074, 21077, 21078, 21079, 21080, 21081, 21082, 21083, 21084, 21085, 21086, 21087, 21088, 21089, 21090, 21091, 21099, 21130, 21132, 21133, 21134, 21135, 21136, 21137, 21138, 21139, 21142, 21143, 21144, 21145, 21146, 21147, 21173, 21174, 21175, 21176, 21177, 21178, 21179, 21180, 21181, 21182, 21183, 21184, 21185, 21186, 21187, 21188, 21189, 21190, 21192, 21193, 21194, 21195, 21196, 21197, 21198, 21199, 21200, 21201, 21202, 21203, 21204, 21205, 21206, 21207, 21208, 21209, 21210, 21211, 21212, 21213, 21214, 21215, 21216, 21217, 21218, 21219, 21220, 21221, 21222, 21223, 21224, 21225, 21226, 21227, 21228, 21229, 21230, 21231, 21232, 21233, 21234, 21235, 21236, 21237, 21238, 21239, 21240, 21241, 21242, 21243, 21244, 21245, 21246, 21247, 21248, 21279, 21280, 21305, 21306, 21307, 21308, 21309, 21311, 21312, 21317, 21318, 21319, 21320, 21321, 21322, 21323, 21324, 21325, 21326, 21327, 21328, 21329, 21330, 21331, 21332, 21333, 21334, 21335, 21336, 21337, 21338, 21339, 21340, 21341, 21342, 21343, 21344, 21349, 21350, 21351, 21352, 21353, 21355, 21357, 21358, 21359, 21360, 21361, 21362, 21365, 21366, 21369, 21370, 21375, 21376, 21377, 21378, 21379, 21380, 21381, 21382, 21383, 21384, 21385, 21386, 21387, 21388, 21389, 21390, 21391, 21392, 21393, 21394, 21395, 21396, 21397, 21398, 21399, 21400, 21401, 21402, 21403, 21404, 21405, 21406, 21419, 21420, 21421, 21422, 21423, 21424, 21425, 21426, 21427, 21428, 21429, 21430, 21431, 21432, 21433, 21434, 21435, 21437, 21439, 21440, 21441, 21442, 21443, 21444, 21445, 21446, 21447, 21448, 21449, 21450, 21451, 21452, 21453, 21454, 21455, 21456, 21457, 21458, 21459, 21460, 21461, 21462, 21463, 21464, 21465, 21466, 21467, 21468, 21469, 21470, 21471, 21472, 21473, 21474, 21475, 21476, 21477, 21478, 21479, 21480, 21481, 21482, 21483, 21484, 21485, 21486, 21487, 21488, 21489, 21490, 21491, 21492, 21493, 21494, 21495, 21496, 21497, 21498, 21499, 21500, 21501, 21502, 21503, 21504, 21505, 21506, 21507, 21508, 21509, 21510, 21511, 21512, 21513, 21514, 21515, 21516, 21517, 21518, 21519, 21520, 21521, 21522, 21523, 21524, 21525, 21526, 21527, 21528, 21529, 21530, 21531, 21532, 21533, 21534, 21535, 21536, 21537, 21538, 21539, 21540, 21541, 21542, 21543, 21544, 21545, 21546, 21547, 21548, 21549, 21550, 21551, 21552, 21553, 21554, 21555, 21556, 21557, 21558, 21559, 21560, 21561, 21562, 21563, 21564, 21565, 21566, 21567, 21568, 21569, 21570, 21571, 21572, 21573, 21574, 21575, 21576, 21577, 21578, 21579, 21580, 21581, 21582, 21583, 21584, 21585, 21586, 21587, 21588, 21589, 21590, 21591, 21592, 21593, 21594, 21595, 21596, 21597, 21598, 21599, 21600, 21601, 21602, 21603, 21604, 21605, 21606, 21607, 21608, 21609, 21610, 21611, 21612, 21613, 21615, 21616, 21617, 21618, 21619, 21620, 21621, 21622, 21623, 21624, 21625, 21626, 21627, 21628, 21629, 21630, 21631, 21632, 21633, 21652, 21663, 21697, 21698, 21699, 21700, 21701, 21702, 21703, 21704, 21705, 21706, 21707, 21708, 21709, 21710, 21711, 21712, 21713, 21714, 21715, 21716, 21717, 21718, 21719, 21720, 21721, 21722, 21723, 21724, 21725, 21726, 21727, 21728, 21729, 21730, 21731, 21732, 21733, 21734, 21735, 21736, 21737, 21748, 21749, 21750, 21751, 21752, 21753, 21754, 21755, 21756, 21757, 21758, 21759, 21760, 21761, 21762, 21763, 21764, 21765, 21766, 21767, 21768, 21769, 21770, 21771, 21772, 21773, 21774, 21775, 21776, 21777, 21778, 21779, 21780, 21781, 21782, 21853, 21854, 21855, 21856, 21857, 21858, 21859, 21860, 21861, 21862, 21863, 21864, 21865, 21866, 21867, 21869, 21916, 21917, 21918, 21919, 21920, 21921, 21922, 21923, 21924, 21925, 21926, 21927, 21928, 21929, 21930, 21931, 21932, 21933, 21934, 21935, 21936, 21937, 21938, 21939, 21940, 21941, 21942, 21943, 21944, 21945, 21946, 21947, 21948, 21949, 21950, 21960, 21961, 21962, 21963, 21964, 21965, 21966, 21967, 21968, 21969, 21970, 21981, 21982, 21983, 21984, 21985, 21986, 21988, 22014, 22015, 22016, 22017, 22018, 22019, 22020, 22021, 22022, 22023, 22024, 22025, 22026, 22027, 22028, 22039, 22040, 22041, 22056, 22060, 22065, 22066, 22067, 22068, 22069, 22078, 22079, 22080, 22096, 22097, 22098, 22124, 22125, 22145, 22146, 22147, 22148, 22149, 22150, 22151, 22152, 22182, 22185, 22186, 22190, 22191, 22222, 22229, 22230, 22231, 22232, 22239, 22240, 22241, 22242, 22243, 22244, 22245, 22246, 22247, 22248, 22249, 22250, 22251, 22252, 22253, 22360, 22364, 22365, 22366, 22367, 22368, 22369, 22370, 22371, 22392, 22462, 22463, 22466, 22473, 22474, 22491, 22523, 22526, 22546, 22550, 22551, 22553, 22554, 22555, 22556, 22557, 22558, 22559, 22560, 22561, 22562, 22563, 22564, 22566, 22567, 22597, 22601, 22602, 22604, 22605, 22606, 22614, 22615, 22616, 22617, 22618, 22619, 22620, 22621, 22622, 22623, 22624, 22625, 22626, 22627, 22666, 22670, 22681, 22682, 22692, 22693, 22694, 22695, 22696, 22697, 22698, 22699, 22709, 22710, 22711, 22713, 22714, 22715, 22716, 22717, 22718, 22719, 22720, 22721, 22722, 22723, 22724, 22725, 22726, 22727, 22731, 22732, 22733, 22734, 22735, 22736, 22737, 22738, 22739, 22740, 22741, 22742, 22743, 22744, 22745, 22746, 22747, 22748, 22749, 22750, 22751, 22752, 22753, 22754, 22755, 22756, 22757, 22758, 22759, 22760, 22761, 22762, 22763, 22764, 22765, 22766, 22767, 22768, 22769, 22770, 22771, 22772, 22773, 22774, 22775, 22776, 22777, 22778, 22779, 22780, 22781, 22782, 22783, 22784, 22785, 22786, 22787, 22788, 22789, 22790, 22791, 22792, 22793, 22794, 22795, 22796, 22797, 22798, 22799, 22800, 22801, 22802, 22803, 22804, 22805, 22806, 22807, 22808, 22809, 22810, 22811, 22812, 22813, 22844, 22845, 22849, 22850, 22851, 22854, 22855, 22861, 22862, 22863, 22864, 22866, 22867, 22868, 22869, 22871, 22872, 22873, 22877, 22878, 22879, 22880, 22881, 22882, 22883, 22884, 22885, 22886, 22887, 22888, 22889, 22890, 22891, 22892, 22893, 22894, 22895, 22896, 22897, 22903, 22904, 22905, 22906, 22907, 22908, 22909, 22910, 22911, 22912, 22913, 22914, 22915, 22916, 22934, 22935, 22936, 22937, 22939, 22940, 22941, 22942, 22943, 22944, 22945, 22946, 22947, 22948, 22949, 22950, 22951, 22952, 22953, 22954, 22955, 22956, 22957, 22958, 22959, 22960, 22961, 22962, 22963, 22964, 22965, 22966, 22967, 22968, 22969, 22970, 22971, 22972, 22973, 22974, 22975, 22976, 22977, 22978, 22979, 22981, 22982, 22983, 22984, 22985, 22986, 22987, 22988, 22989, 22990, 22991, 22992, 22993, 22994, 22995, 22996, 22997, 22998, 22999, 23000, 23001, 23002, 23003, 23004, 23005, 23006, 23007, 23008, 23009, 23177, 23178, 23179, 23180, 23181, 23182, 23183, 23184, 23185, 23186, 23199, 23200, 23201, 23202, 23203, 23204, 23205, 23206, 23207, 23208, 23209, 23210, 23211, 23212, 23213, 23214, 23215, 23216, 23217, 23218, 23219, 23220, 23221, 23222, 23223, 23224, 23225, 23226, 23227, 23228, 23229, 23230, 23231, 23232, 23233, 23234, 23235, 23236, 23237, 23238, 23239, 23240, 23241, 23242, 23243, 23244, 23245, 23246, 23247, 23248, 23249, 23250, 23251, 23252, 23253, 23254, 23284, 23285, 23286, 23288, 23290, 23292, 23299, 23300, 23301, 23304, 23306, 23319, 23338, 23339, 23346, 23347, 23348, 23349, 23350, 23351, 23352, 23353, 23354, 23355, 23356, 23357, 23358, 23359, 23360, 23388, 23389, 23390, 23391, 23392, 23400, 23410, 23413, 23414, 23415, 23416, 23417, 23418, 23419, 23420, 23421, 23422, 23423, 23424, 23425, 23426, 23427, 23428, 23429, 23430, 23431, 23432, 23433, 23434, 23435, 23436, 23437, 23438, 23439, 23440, 23441, 23442, 23443, 23444, 23445, 23446, 23447, 23448, 23466, 23470, 23472, 23474, 23475, 23476, 23477, 23482, 23484, 23601, 23662, 23683, 23684, 23685, 23686, 23687, 23688, 23689, 23690, 23691, 23692, 23693, 23694, 23695, 23696, 23698, 23699, 23700, 23702, 23703, 23704, 23706, 23707, 23709, 23732, 23733, 23734, 23735, 23736, 23737, 23738, 23739, 23740, 23741, 23742, 23743, 23744, 23745, 23746, 23747, 23748, 23749, 23750, 23751, 23752, 23753, 23754, 23755, 23756, 23757, 23758, 23759, 23760, 23761, 23762, 23763, 23764, 23765, 23766, 23767, 23768, 23769, 23770, 23771, 23772, 23773, 23774, 23775, 23776, 23777, 23778, 23779, 23780, 23781, 23782, 23783, 23784, 23785, 23786, 23787, 23788, 23789, 23790, 23802, 23830, 23831, 23832, 23833, 23834, 23835, 23836, 23837, 23838, 23839, 23840, 23841, 23842, 23843, 23844, 23845, 23846, 23847, 23848, 23849, 23850, 23851, 23852, 23853, 23854, 23855, 23856, 23857, 23858, 23859, 23872, 23873, 23880, 23881, 23882, 23884, 23885, 23886, 23887, 23888, 23890, 23891, 23893, 23895, 23896, 23897, 23905, 23906, 23908, 23909, 23913, 23914, 23921, 23925, 23946, 23947, 23948, 23949, 23950, 23951, 23952, 23953, 23954, 23955, 23956, 23957, 23958, 23959, 23960, 23961, 23962, 23963, 23964, 23965, 23966, 23971, 23975, 23982, 23984, 23985, 23989, 23990, 23995, 23996, 23997, 24001, 24003, 24004, 24005, 24006, 24007, 24008, 24009, 24010, 24011, 24012, 24013, 24014, 24015, 24016, 24017, 24018, 24019, 24020, 24021, 24022, 24023, 24024, 24025, 24026, 24027, 24028, 24029, 24030, 24031, 24032, 24033, 24034, 24035, 24036, 24037, 24038, 24039, 24040, 24112, 24113, 24114, 24115, 24116, 24117, 24118, 24119, 24120, 24121, 24122, 24123, 24124, 24125, 24126, 24127, 24128, 24129, 24130, 24131, 24132, 24133, 24134, 24135, 24136, 24160, 24161, 24162, 24186, 24187, 24188, 24191, 24192, 24193, 24194, 24195, 24198, 24200, 24201, 24202, 24205, 24207, 24208, 24209, 24213, 24214, 24217, 24218, 24219, 24285, 24286, 24287, 24289, 24290, 24293, 24294, 24295, 24296, 24299, 24321, 24326, 24329, 24330, 24332, 24335, 24338, 24339, 24341, 24343, 24344, 24347, 24348, 24349, 24350, 24352, 24353, 24356, 24357, 24358, 24359, 24362, 24364, 24365, 24366, 24367, 24368, 24371, 24374, 24375, 24377, 24380, 24383, 24384, 24386, 24388, 24389, 24392, 24393, 24395, 24397, 24398, 24401, 24402, 24403, 24404, 24406, 24407, 24410, 24411, 24413, 24415, 24416, 24418, 24497, 24503, 24506, 24509, 24515, 24518, 24563, 24671, 24673, 24674, 24675, 24677, 24680, 24681, 24682, 24683, 24685, 24695, 24698, 24699, 24701, 24703, 24704, 24707, 24708, 24710, 24713, 24716, 24717, 24719, 24722, 24725, 24726, 24728, 24729, 24731, 24734, 24735, 24737, 24739, 24740, 24743, 24746, 24748, 24749, 24752, 24755, 24757, 24758, 24760, 24761, 24764, 24766, 24767, 24770, 24772, 24773, 24774, 24775, 24776, 24779, 24780, 24781, 24782, 24783, 24784, 24785, 24788, 24791, 24792, 24793, 24794, 24796, 24797, 24799, 24800, 24801, 24835, 24836, 24839, 24842, 24843, 24845, 24848, 24851, 24854, 24856, 24900, 24901, 24902, 24905, 24908, 24909, 24910, 24911, 24914, 24915, 24916, 24917, 24920, 24923, 24926, 24928, 24929, 24931, 24932, 24935, 24938, 24942, 24944, 24945, 24947, 24950, 24953, 24954, 24955, 24956, 24959, 24960, 24962, 24965, 24990, 25037, 25040, 25041, 25043, 25046, 25049, 25052, 25055, 25058, 25059, 25061, 25063, 25064, 25067, 25068, 25198, 25199, 25200, 25201, 25202, 25203, 25204, 25205, 25206, 25207, 25208, 25209, 25210, 25211, 25212, 25213, 25214, 25215, 25216, 25217, 25218, 25219, 25220, 25221, 25222, 25223, 25225, 25226, 25227, 25228, 25230, 25231, 25232, 25233, 25234, 25235, 25236, 25237, 25238, 25239, 25240, 25241, 25242, 25243, 25244, 25245, 25246, 25247, 25249, 25250, 25251, 25252, 25253, 25254, 25255, 25256, 25257, 25258, 25259, 25260, 25261, 25262, 25263, 25264, 25265, 25266, 25267, 25268, 25269, 25270, 25271, 25272, 25273, 25274, 25275, 25276, 25277, 25278, 25279, 25280, 25281, 25282, 25283, 25284, 25285, 25286, 25287, 25288, 25289, 25290, 25291, 25341, 25342, 25343, 25344, 25345, 25346, 25347, 25348, 25349, 25350, 25351, 25352, 25353, 25354, 25355, 25357, 25358, 25359, 25360, 25361, 25362, 25363, 25364, 25365, 25366, 25367, 25368, 25369, 25370, 25371, 25372, 25373, 25374, 25375, 25376, 25377, 25378, 25379, 25380, 25381, 25382, 25383, 25384, 25385, 25386, 25387, 25388, 25389, 25390, 25391, 25408, 25409, 25410, 25411, 25412, 25413, 25414, 25415, 25416, 25417, 25446, 25447, 25448, 25449, 25450, 25451, 25481, 25482, 25493, 25494, 25495, 25496, 25497, 25498, 25499, 25500, 25501, 25502, 25503, 25504, 25505, 25506, 25507, 25508, 25509, 25510, 25511, 25512, 25513, 25514, 25515, 25516, 25517, 25519, 25520, 25521, 25522, 25523, 25524, 25525, 25526, 25527, 25528, 25529, 25530, 25531, 25532, 25533, 25534, 25535, 25536, 25537, 25538, 25539, 25540, 25541, 25542, 25543, 25544, 25545, 25546, 25547, 25548, 25549, 25550, 25551, 25552, 25558, 25559, 25577, 25578, 25587, 25588, 25589, 25590, 25591, 25592, 25593, 25594, 25595, 25596, 25597, 25598, 25599, 25600, 25601, 25602, 25603, 25604, 25605, 25606, 25607, 25608, 25609, 25610, 25611, 25612, 25613, 25614, 25615, 25616, 25617, 25618, 25619, 25620, 25621, 25622, 25623, 25624, 25625, 25626, 25646, 25647, 25648, 25649, 25650, 25651, 25652, 25653, 25654, 25655, 25656, 25657, 25658, 25659, 25660, 25664, 25665, 25666, 25667, 25668, 25669, 25670, 25671, 25672, 25673, 25695, 25728, 25729, 25730, 25731, 25732, 25733, 25734, 25735, 25736, 25737, 25738, 25739, 25740, 25741, 25742, 25743, 25744, 25745, 25746, 25747, 25748, 25749, 25750, 25751, 25752, 25753, 25754, 25755, 25756, 25757, 25758, 25759, 25760, 25761, 25762, 25763, 25764, 25765, 25766, 25767, 25768, 25769, 25770, 25771, 25773, 25774, 25775, 25776, 25777, 25795, 25796, 25797, 25826, 25874, 25875, 25876, 25877, 25883, 25887, 25888, 25889, 25890, 25891, 25916, 25917, 25946, 25947, 25948, 25949, 25950, 25951, 25952, 25953, 25954, 25955, 25959, 25973, 25974, 25975, 25976, 25977, 25978, 25979, 25981, 25982, 25983, 25984, 25985, 25986, 25987, 25988, 25989, 25990, 25991, 25992, 25993, 25994, 25995, 25996, 25997, 25998, 25999, 26000, 26001, 26002, 26003, 26004, 26005, 26006, 26007, 26008, 26009, 26010, 26011, 26012, 26013, 26014, 26015, 26016, 26017, 26018, 26019, 26020, 26021, 26022, 26023, 26025, 26026, 26027, 26028, 26029, 26030, 26031, 26032, 26033, 26034, 26035, 26036, 26037, 26038, 26040, 26041, 26042, 26043, 26044, 26045, 26046, 26047, 26048, 26049, 26050, 26051, 26052, 26053, 26054, 26055, 26056, 26057, 26058, 26059, 26060, 26066, 26069, 26070, 26071, 26072, 26073, 26074, 26075, 26076, 26077, 26080, 26081, 26082, 26083, 26084, 26086, 26087, 26088, 26089, 26090, 26116, 26118, 26119, 26120, 26155, 26156, 26157, 26160, 26162, 26163, 26164, 26166, 26172, 26173, 26177, 26180, 26182, 26185, 26186, 26187, 26189, 26190, 26191, 26192, 26193, 26195, 26196, 26198, 26199, 26205, 26206, 26207, 26209, 26210, 26211, 26212, 26213, 26214, 26215, 26217, 26218, 26221, 26223, 26225, 26226, 26227, 26229, 26231, 26233, 26234, 26235, 26236, 26237, 26238, 26239, 26240, 26241, 26242, 26243, 26245, 26246, 26248, 26249, 26250, 26251, 26252, 26253, 26254, 26255, 26256, 26257, 26258, 26259, 26260, 26261, 26262, 26263, 26264, 26265, 26266, 26267, 26268, 26269, 26270, 26272, 26273, 26274, 26275, 26276, 26277, 26278, 26279, 26280, 26281, 26282, 26283, 26284, 26285, 26286, 26287, 26288, 26289, 26290, 26291, 26292, 26293, 26294, 26296, 26297, 26298, 26299, 26300, 26301, 26302, 26303, 26304, 26305, 26306, 26307, 26308, 26309, 26310, 26311, 26312, 26313, 26314, 26315, 26317, 26318, 26340, 26342, 26344, 26345, 26346, 26387, 26390, 26391, 26392, 26396, 26397, 26398, 26399, 26400, 26402, 26403, 26404, 26405, 26406, 26407, 26408, 26409, 26435, 26436, 26437, 26438, 26439, 26481, 26482, 26483, 26484, 26485, 26486, 26487, 26488, 26489, 26490, 26521, 26522, 26523, 26524, 26525, 26526, 26547, 26548, 26549, 26550, 26551, 26552, 26553, 26554, 26555, 26556, 26559, 26560, 26561, 26562, 26563, 26564, 26565, 26566, 26567, 26568, 26569, 26570, 26571, 26572, 26573, 26574, 26575, 26576, 26577, 26578, 26579, 26580, 26581, 26582, 26583, 26584, 26585, 26586, 26587, 26588, 26589, 26590, 26591, 26592, 26593, 26594, 26596, 26597, 26598, 26599, 26600, 26601, 26602, 26603, 26604, 26605, 26606, 26607, 26608, 26609, 26610, 26611, 26612, 26613, 26614, 26615, 26616, 26617, 26618, 26619, 26620, 26621, 26622, 26623, 26624, 26625, 26626, 26627, 26628, 26629, 26630, 26631, 26632, 26633, 26634, 26635, 26636, 26637, 26638, 26639, 26640, 26641, 26642, 26643, 26644, 26645, 26646, 26647, 26648, 26649, 26650, 26651, 26652, 26653, 26654, 26655, 26656, 26657, 26658, 26675, 26676, 26758, 26759, 26760, 26761, 26762, 26763, 26764, 26765, 26766, 26767, 26768, 26769, 26770, 26771, 26772, 26781, 26782, 26783, 26792, 26849, 26856, 26857, 26858, 26859, 26860, 26861, 26862, 26863, 26864, 26865, 26898, 26938, 26939, 26940, 26941, 26942, 26978, 26979, 26980, 26981, 26982, 26983, 26984, 26985, 26986, 26987, 26988, 26989, 26990, 26991, 26992, 26993, 26994, 26995, 26996, 26997, 26998, 26999, 27000, 27001, 27002, 27003, 27004, 27005, 27006, 27007, 27008, 27009, 27010, 27011, 27012, 27013, 27014, 27015, 27016, 27017, 27018, 27019, 27020, 27021, 27022, 27023, 27024, 27025, 27026, 27027, 27040, 27041, 27042, 27043, 27044, 27045, 27046, 27047, 27048, 27049, 27051, 27064, 27065, 27066, 27067, 27068, 27069, 27070, 27071, 27072, 27073, 27074, 27075, 27076, 27077, 27078, 27079, 27080, 27081, 27082, 27083, 27084, 27085, 27086, 27087, 27088, 27089, 27091, 27092, 27093, 27094, 27095, 27096, 27097, 27098, 27099, 27100, 27101, 27102, 27103, 27104, 27107, 27108, 27109, 27132, 27133, 27134, 27135, 27136, 27137, 27148, 27149, 27150, 27151, 27152, 27153, 27154, 27155, 27156, 27158, 27159, 27160, 27161, 27162, 27163, 27164, 27165, 27166, 27167, 27168, 27169, 27170, 27171, 27172, 27173, 27174, 27175, 27176, 27177, 27178, 27179, 27180, 27181, 27182, 27183, 27184, 27185, 27186, 27187, 27188, 27189, 27190, 27191, 27192, 27193, 27194, 27195, 27196, 27197, 27216, 27217, 27218, 27221, 27222, 27223, 27224, 27225, 27226, 27227, 27261, 27262, 27263, 27264, 27265, 27266, 27267, 27268, 27269, 27270, 27271, 27272, 27273, 27274, 27275, 27316, 27317, 27318, 27319, 27320, 27321, 27322, 27323, 27324, 27325, 27326, 27327, 27328, 27329, 27330, 27331, 27332, 27333, 27334, 27335, 27336, 27337, 27338, 27339, 27340, 27341, 27342, 27343, 27344, 27345, 27346, 27347, 27348, 27349, 27350, 27351, 27352, 27353, 27354, 27355, 27356, 27357, 27358, 27359, 27360, 27361, 27362, 27363, 27364, 27365, 27366, 27367, 27368, 27369, 27370, 27371, 27372, 27373, 27376, 27377, 27380, 27381, 27384, 27385, 27386, 27387, 27388, 27389, 27390, 27391, 27392, 27393, 27394, 27395, 27396, 27397, 27398, 27399, 27400, 27401, 27402, 27403, 27404, 27405, 27406, 27407, 27408, 27409, 27410, 27411, 27412, 27413, 27414, 27415, 27466, 27467, 27474, 27475, 27505, 27506, 27507, 27508, 27509, 27510, 27511, 27512, 27542, 27543, 27544, 27545, 27546, 27576, 27577, 27578, 27579, 27601, 27602, 27603, 27604, 27605, 27606, 27612, 27613, 27614, 27615, 27650, 27651, 27672, 27673, 27674, 27675, 27676, 27677, 27678, 27679, 27680, 27681, 27682, 27683, 27684, 27685, 27686, 27770, 27771, 27772, 27773, 27774, 27775, 27776, 27777, 27778, 27779, 27780, 27781, 27782, 27783, 27784, 27785, 27786, 27787, 27788, 27789, 27790, 27791, 27792, 27793, 27794, 27795, 27796, 27797, 27798, 27799, 27800, 27801, 27802, 27803, 27804, 27805, 27806, 27807, 27808, 27809, 27810, 27811, 27812, 27813, 27814, 27815, 27816, 27817, 27818, 27819, 27820, 27821, 27822, 27823, 27824, 27825, 27826, 27827, 27828, 27829, 27830, 27831, 27832, 27833, 27834, 27836, 27837, 27838, 27839, 27840, 27841, 27842, 27843, 27844, 27845, 27846, 27847, 27848, 27849, 27850, 27851, 27852, 27853, 27854, 27855, 27856, 27857, 27858, 27859, 27860, 27861, 27862, 27863, 27864, 27865, 27866, 27867, 27868, 27871, 27872, 27873, 27874, 27875, 27876, 27877, 27878, 27879, 27880, 27881, 27882, 27883, 27885, 27886, 27887, 27888, 27889, 27890, 27891, 27892, 27893, 27894, 27895, 27896, 27897, 27898, 27899, 27900, 27901, 27902, 27903, 27904, 27905, 27906, 27907, 27908, 27909, 27910, 27911, 27912, 27913, 27914, 27915, 27916, 27917, 27918, 27919, 27920, 27930, 27931, 27964, 27965, 27966, 27967, 27968, 27969, 27970, 27971, 27972, 27973, 27974, 27975, 27976, 27977, 27978, 27979, 27980, 27981, 27982, 27983, 27984, 27985, 27986, 27987, 27988, 27989, 27990, 27991, 27992, 27993, 27994, 27995, 27996, 27997, 27998, 27999, 28000, 28001, 28002, 28003, 28004, 28005, 28006, 28007, 28008, 28009, 28010, 28011, 28012, 28013, 28014, 28015, 28016, 28017, 28018, 28019, 28020, 28021, 28022, 28023, 28024, 28025, 28026, 28027, 28028, 28052, 28053, 28054, 28055, 28056, 28057, 28058, 28059, 28060, 28061, 28062, 28063, 28064, 28065, 28066, 28067, 28068, 28069, 28070, 28071, 28072, 28073, 28074, 28076, 28077, 28078, 28079, 28080, 28081, 28082, 28083, 28084, 28085, 28086, 28087, 28088, 28089, 28090, 28091, 28092, 28093, 28094, 28095, 28096, 28144, 28145, 28146, 28147, 28158, 28159, 28160, 28161, 28179, 28180, 28181, 28182, 28183, 28184, 28185, 28212, 28213, 28214, 28215, 28216, 28217, 28218, 28220, 28221, 28222, 28223, 28225, 28226, 28227, 28249, 28250, 28251, 28280, 28281, 28282, 28299, 28300, 28301, 28302, 28303, 28304, 28305, 28306, 28307, 28308, 28309, 28310, 28311, 28312, 28313, 28328, 28329, 28330, 28331, 28332, 28333, 28334, 28335, 28336, 28337, 28338, 28339, 28340, 28341, 28342, 28343, 28344, 28352, 28353, 28354, 28355, 28356, 28357, 28358, 28359, 28360, 28380, 28381, 28382, 28391, 28392, 28393, 28394, 28395, 28396, 28397, 28398, 28399, 28400, 28401, 28404, 28405, 28423, 28424, 28439, 28440, 28441, 28442, 28443, 28444, 28445, 28446, 28447, 28448, 28449, 28450, 28451, 28452, 28453, 28454, 28455, 28456, 28457, 28458, 28459, 28460, 28461, 28462, 28463, 28476, 28477, 28478, 28479, 28480, 28502, 28503, 28504, 28505, 28506, 28508, 28509, 28510, 28511, 28512, 28565, 28566, 28636, 28637, 28638, 28639, 28640, 28641, 28642, 28643, 28644, 28645, 28646, 28647, 28648, 28649, 28650, 28651, 28652, 28653, 28654, 28655, 28656, 28657, 28658, 28662, 28664, 28665, 28666, 28667, 28668, 28669, 28670, 28671, 28672, 28673, 28674, 28675, 28676, 28679, 28680, 28685, 28686, 28687, 28688, 28689, 28690, 28692, 28693, 28694, 28695, 28704, 28705, 28706, 28707, 28708, 28709, 28710, 28711, 28712, 28713, 28714, 28715, 28716, 28721, 28722, 28723, 28724, 28725, 28726, 28727, 28728, 28729, 28730, 28731, 28732, 28733, 29112, 29113, 29114, 29115, 29116, 29117, 29118, 29119, 29120, 29121, 29122, 29123, 29124, 29125, 29126, 29127, 29128, 29129, 29130, 29202, 50001, 50002, 50003, 50004, 50005, 50006, 50007, 50008, 50009, 50010, 50011, 50012, 50013, 50014, 50015, 50016, 50017, 50018, 50019, 50020, 50021, 50022, 50023, 50024, 50025, 50026, 50027, 50028, 50029, 50030, 50031, 50032, 50033, 50034, 50035, 50036, 50037, 50038, 50039, 50040, 50041, 50042, 50043, 50044, 50045, 50046, 50047, 50048, 50049, 50050, 50051, 50052, 50053, 50054, 50055, 50056, 50057, 50058, 50059, 50060, 50061, 50062, 50063, 50064, 50065, 50066, 50067, 50068, 50069, 50070, 50071, 50072, 50073, 50074, 50075, 50076, 50077, 50078, 50079, 50080, 50081, 50082, 50083, 50084, 50085, 50086, 50087, 50088, 50089, 50090, 50091, 50092, 50093, 50094, 50095, 50096, 50097, 50098, 50101, 50161, 50162, 50163, 50189, 50190, 50191, 50195, 50196, 50197, 50198, 50199, 50200, 50201, 50202, 50203, 50204, and 50205. |
Recalling Firm/ Manufacturer |
Natus Medical Incorporated 5900 1st Ave S Seattle WA 98108-3248
|
For Additional Information Contact |
Natus Technical Service 888-496-2887
|
Manufacturer Reason for Recall |
The firm is adding the warning statement to the Instruction for Use for the neoBLUE LED Phototherapy System and provide a laminated Quick Guide to affix to each device.
|
FDA Determined Cause 2 |
Labeling False and Misleading |
Action |
The URGENT: FIELD SAFETY NOTICE PRODUCT LABELING CHANGE
neoBLUE LED Phototherapy System, dated April 2016, was went to US customers between April 11, 2016 and April 27, 2016.
Natus distribution partners outside the US will receive the Letter to Distributors and they are requested to contact their downstream customers to distribute the information. The estimated date of notifying international consignees is May 31, 2016.
Enclosed with this letter, customers would receive:
1) Quick Guides (2 each) which provide operating steps, information about measuring intensity, and warnings (PN 051693)
2) User manual addendum
3) Customer reply form
Natus requests that this information be distributed to all associated parties within your organization that use or service this product. Please note: this communication is in addition to any prior labeling changes that you may have received from Natus.
Customers should submit the Reply Form via fax to 847-949-6031 or email to
mailto:Natus_Quality_Programs@natus.com.
If there are any questions regarding this letter and addendum, please call Natus Technical Service at 888-496-2887. |
Quantity in Commerce |
13541 units total (7840 units in the US and 5701 units outside the US) |
Distribution |
Worldwide Distributed - US Nationwide in the US including American Samoa, Guam Puerto Rico and in the following countries:
Argentina, Australia, Austria, Bangladesh, Belgium, Bermuda, Bolivia, Brunei Darussalam, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, , Kenya, , Kuwait, Latvia, Lithuania, Malaysia, Moldova Republic of, Morocco, Namibia, Netherlands, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia (Slovak Republic), Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, Virgin Islands (British), and Yemen. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = LBI and Original Applicant = NATUS MEDICAL, INC.
|
|
|
|