| Date Initiated by Firm | January 07, 2016 |
| Date Posted | April 14, 2016 |
| Recall Status1 |
Terminated 3 on May 03, 2017 |
| Recall Number | Z-1410-2016 |
| Recall Event ID |
72992 |
| Product Classification |
General surgery tray (kit) - Product Code LRO
|
| Product | Catalog Number: 9001091 PERY-GYN SURGICAL PACK
Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. |
| Code Information |
LOTS: SL15202 (150719622) 31-JUL-16 SL15211 (150720064) 31-JUL-16 SL15212 (150720065) 31-JUL-16 SL15245 (150820677) 31-AUG-16 SL15245 (150820676) 31-AUG-16 |
Recalling Firm/ Manufacturer |
Customed, Inc Calle Igualdad Final #7 Fajardo PR 00738
|
Manufacturer Reason for Recall | Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile. |
FDA Determined Cause 2 | Packaging |
| Action | Recall Notification Letter sent to direct accounts via certified mail with return receipt on January 7, 2016 and January 8, 2016. |
| Quantity in Commerce | 80 |
| Distribution | Distributed only in Puerto Rico. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|