Date Initiated by Firm |
January 07, 2016 |
Date Posted |
April 14, 2016 |
Recall Status1 |
Terminated 3 on May 03, 2017 |
Recall Number |
Z-1485-2016 |
Recall Event ID |
72992 |
Product Classification |
Orthopedic tray - Product Code OJH
|
Product |
Catalog Number: 900741A ORTHO-IMPLANT SURGICAL PACK Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
|
Code Information |
LOTS: SL15174 (150619406) 30-JUN-16 SL15197 (150719810) 31-JUL-16 SL15201-02 (150719810) 31-JUL-16 SL15231 (150820049) 31-AUG-16 SL15232 (150820049) 31-AUG-16 SL15243 (150820049) 31-AUG-16 SL15265 (150820558) 31-AUG-16 SL15268 (150920711) 30-SEP-16 |
Recalling Firm/ Manufacturer |
Customed, Inc Calle Igualdad Final #7 Fajardo PR 00738
|
Manufacturer Reason for Recall |
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
|
FDA Determined Cause 2 |
Packaging |
Action |
Recall Notification Letter sent to direct accounts via certified mail with return receipt on January 7, 2016 and January 8, 2016. |
Quantity in Commerce |
153 |
Distribution |
Distributed only in Puerto Rico. |