  Class 2 Device Recall Zimmer see related information
Date Initiated by Firm January 11, 2016
Date Posted March 12, 2016
Recall Status1 Terminated 3 on April 13, 2017
Recall Number Z-1109-2016
Recall Event ID 72900
510(K)Number K960278  
Product Classification Prosthesis, hip, semi-constrained, metal/polymer, cemented - Product Code JDI
Product Product 11 consists of all product under product code JDI
, and same usage:
Item no:
902602500 6 DEGREE COCR FEM HEAD 26
902602600 6 DEGREE COCR FEM HEAD 26
902602700 6 DEGREE COCR FEM HEAD 26
902602800 6 DEGREE COCR FEM HEAD 28
902602900 6 DEGREE COCR FEM HEAD 28
902603000 6 DEGREE COCR FEM HEAD 28
902603300 6 DEGREE COCR FEM HEAD 32
902603200 6 DEGREE COCR FEM HEAD 32

Product Usage:
For use in total or hemi hip arthroplasty































































































Code Information lot no:61603870 60973105N 60892289N 60900891N 62279763 62093449 62185961 62279764 62384218 62082438 62268537 62384219 
Recalling Firm/
Manufacturer
Zimmer Manufacturing B.V.
Turpeaux Industrial Park Rd 1 km 123.4
Mercedita PR 00715
For Additional Information Contact Consumer Relations Call Center
800-447-5633
Manufacturer Reason
for Recall
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
FDA Determined
Cause 2
Packaging
Action Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
Quantity in Commerce 366
Distribution US Nationwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JDI and Original Applicant = ZIMMER, INC.
-
-