| | Class 2 Device Recall Uric Acid (URCA) |  |
| Date Initiated by Firm | March 17, 2016 |
| Create Date | May 11, 2016 |
| Recall Status1 |
Terminated 3 on September 25, 2017 |
| Recall Number | Z-1666-2016 |
| Recall Event ID |
73667 |
| 510(K)Number | K061655 |
| Product Classification |
Acid, uric, uricase (colorimetric) - Product Code KNK
|
| Product | Dimension Vista Assays: Uric Acid (URCA) |
| Code Information |
Device Listing: D011436 - Uric Acid (URCA), Catalog # K1077 |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 500 Gbc Dr Ms 514 PO BOX 6101 Newark DE 19702-2466
|
| For Additional Information Contact | Mindy Losapio 914-524-2312 |
Manufacturer Reason for Recall | Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific |
FDA Determined Cause 2 | Device Design |
| Action | Siemens mailed an Urgent Medical Device Correction letter to customers making them aware of the problem. Special instructions were also given for the customer to review the letter with their Medical Director and ensure Venipuncture occurs before drug administration of NAC or Metamizole as indicated in the letter. Baseline assay values before administration of NAC or Metamizole therapy would not be affected. Customers were asked to complete and return the Field Correction Effectiveness Check Form attached with the letter within 30 days. Furthermore, if customers received any complaints of illness or adverse events associated with the specified lots, they were to immediately contact their local Siemens Customer Care Center or local Siemens technical support representative. |
| Quantity in Commerce | 137422 units total |
| Distribution | Nationwide: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, St. Thomas, VI, Guam, Pacific Mariana Islands |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KNK
|
|
|
|