| Class 2 Device Recall 91496 Ultraview SL Command Module |  |
Date Initiated by Firm | July 01, 2016 |
Date Posted | August 04, 2016 |
Recall Status1 |
Terminated 3 on May 22, 2017 |
Recall Number | Z-2343-2016 |
Recall Event ID |
74549 |
510(K)Number | K103142 |
Product Classification |
Detector and alarm, arrhythmia - Product Code DSI
|
Product | 91496 Ultraview SL Command Module, Options A, B, C, and I.
Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP.
Option B: ECG/Respiration, Invasive Pressure x2, SpO2, Temperature x2 & adult/neonate NIBP.
Option C: ECG/Respiration, Invasive Pressure x4, SpO2, Cardiac Output, Temperature x2 & adult/neonate NIBP.
Option I: SpO2, Temperature x2, & adult/neonate NIBP. |
Code Information |
Manifold Lot Codes: 128280, 129420, 129940, or 155640. |
Recalling Firm/ Manufacturer |
Spacelabs Healthcare Inc 35301 SE Center St Snoqualmie WA 98065-9216
|
For Additional Information Contact | Technical Support 800-522-7025 Ext. 2 |
Manufacturer Reason for Recall | The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error messages: Inflate Error, HW Error, or No Data. |
FDA Determined Cause 2 | Vendor change control |
Action | Spacelabs expanded the recall and sent an Urgent - Medical Device Correction letter dated August 30, 2016, to additional affected customers via priority mail (return receipt requested).
On September 6, 2016, a customer letter (translated as necessary) was sent via email to all international subsidiaries and distributors of record.
**********************************************************************************************
Spacelabs Healthcare Inc, sent an Urgent - Medical Device Correction letter dated July 1, 2016, to all affected customers (return receipt requested).
On July 8, 2016, a customer letter (translated as necessary) was sent via email to all international subsidiaries and distributors of record.
Spacelabs Healthcare will contact customers to schedule a convenient time for a Spacelabs service technician to update their affected Command Module(s) at no cost.
For additional information or technical assistance, please contact:
Technical Support 1-800-522-7025 and select (2) for Technical Support |
Quantity in Commerce | 3801 modules and 116 PCBAs & valves distributed in the US; 1297 modules and 366 PCBAs & valves distributed outside the US |
Distribution | Worldwide Distribution - USA Distribution and to the countries of :
AFGHANISTAN, ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BRUNEI DARUSSALAM; CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DOMINICAN REPUBLIC;
ECUADOR, EGYPT, ESTONIA, FRANCE, GERMANY, GUATEMALA, HONG KONG, INDIA, IRAN (ISLAMIC REPUBLIC OF), ISRAEL, ITALY, JORDAN, KUWAIT, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, PAKISTAN, PALESTINE, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, Slovakia, SPAIN, SRI LANKA, SWITZERLAND, TAIWAN, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VIET NAM. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = DSI
|
|
|
|