| Class 2 Device Recall BrightView X |  |
Date Initiated by Firm | May 25, 2016 |
Create Date | September 03, 2016 |
Recall Status1 |
Terminated 3 on July 02, 2018 |
Recall Number | Z-2735-2016 |
Recall Event ID |
74894 |
510(K)Number | K062298 |
Product Classification |
System, tomography, computed, emission - Product Code KPS
|
Product | BrightView X, model #882478; gamma camera designed for single or dual detector nuclear imaging |
Code Information |
11000003, 11000008, 11000030, 11000004, 11000005, 11000006, 11000007, 11000009, 11000010, 11000011, 11000012, 11000013, 11000014, 11000016, 11000017, 11000018, 11000019, 11000020, 11000022, 11000023, 11000024, 11000025, 11000026, 11000027, 11000028, 11000029, 11000031, 11000032, 11000033, 11000034, 11000035, 11000036, 11000037, 11000038, 11000039, 11000040, 11000041, 11000042, 11000043, 11000044, 11000045, 11000046, 11000047, 11000048, 11000049, 11000050, 11000051, 11000052, 11000054, 11000055, 11000056, 11000057, 11000058, 11000059, 11000060, 11000061, 11000062, 11000063, 11000064, 11000065, 11000066, 11000067, 11000068, 11000069, 11000070, 11000071, 11000072, 11000073, 11000074, 11000075, 11000076, 11000077, 11000078, 11000079, 11000080, 11000081, 11000083, 11000084, 11000085, 11000086, 11000087, 11000088, 11000089, 11000090, 11000091, 11000092, 11000093, 11000094, 11000095, 11000096, 11000097, 11000098, 11000099, 11000100, 11000101, 11000102, 11000103, 11000104, 11000105, 11000106, 11000107, 11000108, 11000109, 11000110, 11000112, 11000113, 11000114, 11000115, 11000116, 11000117, 11000118, 11340001, 11340002, 11340003, 11340004, 11340005, 11000015 |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Rd Cleveland OH 44143-2131
|
For Additional Information Contact | Holly Wright Lee 440-483-7600 |
Manufacturer Reason for Recall | The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script. |
FDA Determined Cause 2 | Software design |
Action | The firm is informing the consignee of this issue through Customer Information Letter. |
Quantity in Commerce | 117 units |
Distribution | Worldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republic of Korea, Republic of Lebanon, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|