| | Class 2 Device Recall VersaVit |  |
| Date Initiated by Firm | August 11, 2016 |
| Date Posted | October 18, 2016 |
| Recall Status1 |
Terminated 3 on April 27, 2017 |
| Recall Number | Z-0121-2017 |
| Recall Event ID |
74918 |
| 510(K)Number | K121675 |
| Product Classification |
Instrument, vitreous aspiration and cutting, ac-powered - Product Code HQE
|
| Product | 25 gauge VersaVit 2.0 Procedure Pack with Photon Wide Field Illuminator, Product/Part Number 70025SP, packaged in HIPS rigid tray with a Tyvek lid sterile barrier system. |
| Code Information |
Lot number M543990 (1 each or box of 6 each), exp 10/2018 Lot number M545780 (1 each or box of 6 each), exp 10/2018 Lot number M569020 (1 each or box of 6 each), exp 01/2019 Lot number M578140 (1 each or box of 6 each), exp 02/2019 |
| FEI Number |
1000119053
|
Recalling Firm/ Manufacturer |
Synergetics Inc
3845 Corporate Centre Dr
O Fallon
MO
63368-8678
|
| For Additional Information Contact | Steve Puljak 314-939-5100 |
Manufacturer Reason for Recall | Vitrectomy cutters have been found not to be operating within the correct pressure pulse per specification. |
FDA Determined Cause 2 | Other |
| Action | A customer notification letter was sent out on 8/11/2016 to the hospital/surgical center level. Customers were asked to fill out and return a recall acknowledgment form and send back any affected devices. |
| Quantity in Commerce | 255 units |
| Distribution | Devices were distributed to KY, VA, TN, OH, IN, CA, NJ, NY, TX, SC, AL, MS, MI, and UT. Foreign distribution to the UK, France, Mexico, and India. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|