Class 2 Device Recall REFLECTION(R) DEAD BLOW MALLET |
 |
Date Initiated by Firm |
November 16, 2016 |
Create Date |
January 11, 2017 |
Recall Status1 |
Terminated 3 on August 19, 2019 |
Recall Number |
Z-1008-2017 |
Recall Event ID |
75751 |
510(K)Number |
K123598
|
Product Classification |
Prosthesis, hip, semi-constrained, metal/polymer, cemented - Product Code JDI
|
Product |
REFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116 |
Code Information |
606861, 606862, 606863, 606864, 701432, 701433, 808421, 808422, 1200840, 50903653, 60708545, 60806634, 61005467, 70704991, 70907344, 71002434, 71106654, 71106655, 80603593, 80603594, 80609278, 80610026, 80807064, 81007463, 81108699, 81202445, 90309401, 90309402, 90309937, 90309941, 90309943, 7136210699, 00701432A, 01BM13815, 01BM13816, 01GM03498, 01GM03503, 01GM03504, 01GM03506, 01JM01382, 02DM06721, 02DM06722, 02EM06595, 02FM01963, 2LM00998, 02LM01099, 02LM01100, 02MM02211, 02MM02211A, 03CM01191, 03EM04694, 03HM11994, 03KM13414, 04AM06979, 04BM09860, 04DM13907, 04DM17920, 04DM17921, 04EM07129, 04EM07132, 04GM03326, 04GM03327, 04GM03328, 04GM03329, 04GM03330, 04KM01861, 04KM13683, 04KM13684, 04MM02700, 04MM02701, 05AM14642, 05AM14643, 05AM14644, 05AM14645, 05AM14646, 05AM14646A, 05EM08660, 05EM08661, 05FM07198, 05FM07199, 05GM00620, 05GM00621,05GM00622, 05GM01203, 05GM01204, 05GM01205, 05GM02670, 05GM02671, 05GM05689, 05GM05690, 05GM05692, 05GM08285, 05GM08286, 05GM08287, 05HM03895, 05HM03896, 05HM07787, 05HM07788, 05HM07789, 05HM07790, 05HM07791, 05HM08138, 05HM08139, 05HM08140, 05HM12179, 05HM12180, 05JM02349, 05JM02350, 05LM11363, 05LM11364, 05LM11365, 05MM07114, 05MM07115, 05MM07116, 06AM12304, 05MM07114, 06DM06240, 06FM08915, 07AM04686, 07EM23206, 05MM07115, 06DM08506, 06GM03246, 07AM04687, 07EM23207, 05MM07116, 06DM08507, 06GM03247, 07AM04688, 07EM23207A, 06AM12304, 06DM08508, 06GM04389, 07AM04689, 07EM23208, 06AM12306, 06DM08510, 06GM04390, 07BM11025, 07FM13912, 06AM12307, 06DM08511, 06HM04336, 07BM11027, 07FM22176, 06BM06458, 06DM08513, 06HM04337, 07BM11028, 07FM22177, 06BM06459, 06DM14258, 06HM13309, 07CM20766, 07FM22177R, 06BM06460, 06EM07997, 06HM13310, 07CM20767, 07FM22179, 06DM03707, 06EM07998, 06HM13311, 07CM20769, 07FM22180, 06DM03709, 06EM07999, 06HM13312, 07CM20770, 07FM22180R, 06DM03710, 06EM08000, 06KM05142, 07CM20772, 07FM22181, 06DM04641, 06EM08001, 06KM05143, 07CM20773, 07FM22181R, 06DM04642, 06EM08913, 06KM05144, 07DM20507, 07GM15214 06DM04643, 06EM08914, 06KM09904, 07DM20511, 07JM08242, 06DM04644, 06EM08915, 06KM09905, 07DM20512, 07JM08244, 06DM04645, 06EM08916, 06LM07300, 07DM20513, 07JM08246, 06DM04646, 06EM08917, 06LM07302, 07DM20517, 07JM08248, 06DM06236, 06EM15667, 06MM06711, 07EM17133, 07JM08253, 06DM06237, 06EM15668, 06MM06712, 07EM17134, 07JM08254, 06DM06238, 06EM18051, 07AM04684, 07EM17135, 07JM08254R, 06DM06239, 06FM05947, 07AM04685, 07EM17139, 07JM08256, 07KM02873, 08GM15339, 09EM01236, 10MM09446, 11MM06903, 07KM02875, 08GM15340, 09EM01237, 10MM09448, 12BM07223, 07KM02876, 08GM15341, 09EM01238, 10MM09450, 12BM10723, 07KM02877, 08GM15342, 09EM01239, 11AM08358, 12BM10724, 07KM17062, 08HC08114, 09EYW0009, 11AM08358A, 12BM17244, 07KM17063, 08HM08096, 09EYW0009A, 11AM08359, 12BM17245, 07KM23733, 08HM08100, 09EYW0009B, 11AM08359A, 12BM18966, 07KM23734, 08HM08101, 09EYW0009C, 11AM08360, 12CM17458, 07KM23735, 08HM08104, 09EYW0009D, 11AM08361, 12DM05835, 07LM10125, 08HM08105, 09EYW0009E, 11AM08364, 12DM05835A, 07LM10126, 08HM08106, 09EYW0010, 11AM10562, 12DM05836, 07LM10127, 08HM08107, 09EYW0010A, 11AM10564, 12DM05836A, 07LM10128, 08HM08108, 09EYW0010B, 11AM10566, 12DM10416, 07LM10129, 08HM08109, 07LM13415, 08HM08110, 07LM13416, 08HM08111, 07LM13417, 08HM08113, 07LM13418, 08HM08114, 07LM13419, 08HM08115, 07LM13420, 08HM08116, 07MM13910, 08MM15605, 07MM13911, 08MM15605A, 07MM13911R, 08MM15606, 07MM13912 , 08MM15608, 07MM13912R, 08MM15609, 07MM13914, 08MM15611, 07MM13915, 09AM19274, 07MM13917, 09AM19276, 07MM13917R, 09BM09445, 08AM04897, 09BM09447, 08EM13465, 09BM09449, 08EM19975, 09BM09450, 08FM09824, 09BM15217, 08FM09826, 09BM15219, 08FM09827, 09BM15220, 08FM09829, 09BMI5217, 08FM11005, 09BMI5219, 08FM11006, 09BMI5220, 08FM16340, 09BMO9445, 08FM16341, 09BMO9450, 08FM16342, 09CM18713, 08FM16343, 09CM18715, 08FM16344, 09CM18717, 08FM16345, 09CMI8715, 08GM15336, 09DM13396, 08GM15337, 09DM13399, 08GM15338, 09DM13400, 12KM16514, 12KM16515, 12KM17287, 12LM02029, 12LM02109, 12LM07522, 13BM15733, 13CM12241, 13CM12244, 13DM05887, 13DM05888, 13DM09094, 13DM11253, 13DM19796, 13DM20982, 13DM20982A, 13EM09454, 13EM09455, 13FM02852, 13FM18101, 13FM18102, 13FM18102A, 13GM09614, 13GM09615, 13GM10866, 13GM20305, 13HM01391,13HM04334, 13HM20062, 13JM09524, 08HM08109, 08HM08110, 08HM08111, 08HM08113, 08HM08114, 08HM08115, 08HM08116, 08MM15605, 08MM15605A, 08MM15606, 08MM15608, 08MM15609, 08MM15611, 09AM19274, 09AM19276, 09BM09445, 09BM09447, 09BM09449, 09BM09450, 09BM15217, 09BM15219, 09BM15220, 09BMI5217, 09BMI5219, 09BMI5220, 09BMO9445, 09BMO9450, 09CM18713, 09CM18715, 09CM18717, 09CMI8715, 09DM13396, 09DM13399, 09DM13400, 14BM01667, 14BM01668, 14BM01669, 14BM01670, 14BM01671, 14BM01672, 14BM01673, 14BM04157, 14BM04158, 14BM04159, 14BM04160, 14BM04160R, 14CM16664, 14CM17194, 14DM00236, 14DM08760, 14DM08761, 14DM13479, 14EM05451, 14EM06451, 14EM11507,14EM18730, 14EM22863, 14EM22864, 14FM01204, 14FM01205, 14FM17260, 14GM01262, 14GM03572, 14GM10619, 09EYW0010C, 09EYW0011, 09EYW0011A, 09EYW0011B, 09EYW0011C, 09EYW0011D, 09EYW0011E, 09EYW0011F, 09EYW0012, 09EYW0012A, 09EYW0012B, 09EYW0012C, 09EYW0012D, 09EYW0013, 09EYW0013A, 09EYW0013B, 09EYW0013C, 09MM09122, 10GM03167, 10GM05180, 10HM00478, 10HM00480, 10HM01652, 10HM08505, 10HM08505R, 10HM11149, 10HM11149R, 10HM16824, 10HM16824R, 10HM16824S, 10LM01590, 10MM09443, 10MM09444, 10MM09445, 11AM10566A, 11AM11219, 11AM11221, 11CM04420, 11CM06074, 11CM15011, 11DM02748, 11EM14498, 11EM16916, 11FM04754, 11FM04759, 11FM04761, 11FM04767, 11GM05674, 11GM05675, 11GM05676, 11GM10213, 11HM09823, 11HM11556, 11HM15618, 11HM15619, 11HM15620, 11HM15620A, 11JM06646, 11KM03092, 11KM03094, 11KM03095, 11KM03096, 11KM03097, 11MM06746, 11MM06746A, 11MM06894, 11MM06898, 11MM06901, 12DM10417, 12DM10420, 12DM10420A, 12EM00715, 12EM00717, 12EM00717A, 12EM04424, 12EM04425, 12EM07109, 12EM07110, 12EM07111, 12GM06808, 12GM11585, 12HM06211, 12HM06213, 12HM06213A, 12HM06215, 12HM06216, 12HM10408, 12HM11856, 12HM11857, 12HM11857R, 12JM00568, 12JM00569, 12JM00569A, 12JM00570, 12JM00571, 12JM00572, 12JM00573, 12JM04979, 12JM04981, 12JM04983, 12KM16512, 12KM16513, 14MM09041, 14MM13547, 15AM02615, 15AM08843, 15CM15790, 15DM07667, 15DM08587, 15DM10613, 15DM12768, 15DM13536, 15DM13537, 15DM20374, 15EM03979, 15FM16705, 15FM16706, 15JM03486, 15JM03487, 15JM03488, 15JM05522, 15JM05523, 15JM05524, 15JM05525, 15JM05526, 15JM14855, 15JM14999, 15JM15003, 15JM15004, 15JM15005, 15JM15006, 15JM15008, 16BM15809, 16BM15810, 16BM15811, 16BM20892, 16CM00513, 16CM01482, 16CM01483, 16CM08603, 16EM01010, 16EM01012, 16EM13209, 16EM13210, 16EM13211, 16EM18044, 16EM18046, 16FM01913, 16FM03541, 16FM11256, 16FM11257, 16FM11258, 16FM15697, 16FM15698, 16FM15699, 16FM15700, 16GM14186, 16GM14187, 16GM18830, 61005467A, 70907344A, 71106654A, 13JM09525, 14GM12600, 15JM15788, 71106655A, 13JM09526, 14GM14974, 15JM15789, 80603594A, 13JM10159, 14GM18996, 15JM15790, 80609278A, 13KM08048, 14HM04314, 15JM15791, 13KM08052, 14HM18762, 15LM09208, 13KM10776, 14JM06466, 15LM10375, 13KM22390, 14JM09489, 15LM10376, 13KM22391, 14JM10770, 15LM11432, 13LM07967, 14LM01406, 15LM11433, 13LM20117, 14LM01406R, 15LM19635, 13MM07087, 14LM09219, 15MM11292, 13MM07510, 14LM09227, 15MM11295, 13MM16218, 14LM09235, 15MM11296, 14AM09464, 14LM09245, 16AM00592, 14BM00550, 14LM09247, 16AM01881, 14BM00551, 14LM10680, 16BM15807, 14BM01666, 14MM03602, 16BM15808 |
Recalling Firm/ Manufacturer |
Smith & Nephew, Inc. 1450 E Brooks Rd Memphis TN 38116-1804
|
For Additional Information Contact |
David M. Snyder 978-749-1440
|
Manufacturer Reason for Recall |
The firm received complaints of cracks in the weld on the head of the mallet. In the reported cases, some of the lead beads and particles escaped from the mallet into the surgical wound.
|
FDA Determined Cause 2 |
Process control |
Action |
Smith and Nephew sent an Urgent - Product Recall 1 st Notification dated November 16, 2016, to all affected customers.
Actions for Hospital Representatives:
1. Please inspect your inventory and locate any unused devices from the above listed product and batch numbers, and quarantine them immediately.
2. Complete the last two columns in the Inventory Return Certification Form on the following page, indicating the quantities
that need to be returned. If you do not have product to return, please place an x in the column No Product to Return. Include your contact information in the spaces provided at the bottom of the form.
3. Please contact Smith & Nephews Field Actions Department via email at FieldActions@smith-nephew.com or fax +1- 901-566-7975 to return the completed Inventory Return Certification Form and obtain a return authorization (RA) number if you have product to return.
4. Return any affected product to the address listed on the Inventory Return Certification Form. Please indicate the RA number on your return shipment.
Actions for Smith & Nephew Affiliates/Distributors
1.Send this letter to your customers and ask them to return the recalled product.
2. Please inspect your inventory and locate any unused devices from the above listed product and batch numbers, cease distribution and use of those products, and quarantine them immediately.
3. Complete the last two columns in the Inventory Return Certification Form above, indicating the quantities that need to be returned. If you do not have product to return, please place an x in the column No Product to Return. Include your contact information in the spaces provided at the bottom of the form.
4. Please contact Smith & Nephews Field Actions Department via email at FieldActions@smith-nephew.com or fax +1- 901-566-7975 to return the completed Inventory Return Certification Form and obtain a return authorization (RA) number if you have product to return.
5. Return any affected product |
Quantity in Commerce |
4444 units |
Distribution |
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the countries of : Argentina, Austria, Australia, Belgium, Bahrain, Brazil, Canada, China, Colombia, Costa Rica, Dubai, Denmark, Germany, Finland, France, Great Britain, Greece, Hong Kong, Ireland, Israel, India, Italy, Japan, Mexico, Malaysia, Netherlands, Norway, Poland, Portugal, Puerto Rico, South Africa, Sweden, Singapore, Switzerland, Thailand, Turkey and Venezuela. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JDI and Original Applicant = SMITH & NEPHEW, INC.
|
|
|
|