|
Class 2 Device Recall Flexor Radial Access Set |
 |
Date Initiated by Firm |
February 01, 2017 |
Create Date |
March 02, 2017 |
Recall Status1 |
Terminated 3 on September 28, 2018 |
Recall Number |
Z-1390-2017 |
Recall Event ID |
76360 |
510(K)Number |
K132592
|
Product Classification |
Introducer, catheter - Product Code DYB
|
Product |
Flexor Radial Access Set, Individual product is packaged in a Tyvek-film sterilizable outer pouch. The Flexor¿ Radial Hydrophilic Introducer Access Set is intended to introduce diagnostic and interventional devices In radial artery access procedures |
Code Information |
Catalog Identifier: KCFN-4.0(5.0,6.0,7.0)- / -RA-(S)-HC Codes consist of Lot #, RPN #, Manufacture Date and Expiration Date: 4427564X KCFN-4.0-18-13-RA-HC 8/21/2014 8/21/2017 4427564XX KCFN-4.0-18-13-RA-HC 8/21/2014 8/21/2017 4427564XXX KCFN-4.0-18-13-RA-HC 8/21/2014 8/21/2017 4427566X KCFN-4.0-18-13-RA-HC 2014/03 3/1/2017 4427568X KCFN-4.0-18-13-RA-HC 2014/03 3/1/2017 4610803X KCFN-4.0-18-13-RA-HC 2014/03 3/1/2017 4954615 KCFN-4.0-18-13-RA-HC 6/1/2014 6/1/2017 5087583 KCFN-4.0-18-13-RA-HC 8/1/2014 8/1/2017 5087583X KCFN-4.0-18-13-RA-HC 8/1/2014 8/1/2017 5102705 KCFN-4.0-18-13-RA-HC 8/11/2014 8/11/2017 5102705X KCFN-4.0-18-13-RA-HC 8/11/2014 8/11/2017 5260392 KCFN-4.0-18-13-RA-HC 10/6/2014 10/6/2017 5260392X KCFN-4.0-18-13-RA-HC 10/6/2014 10/6/2017 4427454X KCFN-4.0-18-23-RA-HC 8/21/2014 8/21/2017 4427454XX KCFN-4.0-18-23-RA-HC 8/21/2014 8/21/2017 4427456X KCFN-4.0-18-23-RA-HC 2014/03 3/1/2017 4427458X KCFN-4.0-18-23-RA-HC 2014/03 3/1/2017 4667209X KCFN-4.0-18-23-RA-HC 8/5/2014 8/5/2017 4844427 KCFN-4.0-18-23-RA-HC 2014/04 4/1/2017 4900689 KCFN-4.0-18-23-RA-HC 5/1/2014 5/1/2017 5579987 KCFN-4.0-18-23-RA-HC 2/25/2015 2/25/2018 4427654X KCFN-4.0-18-7-RA-HC 8/8/2014 8/8/2017 4427656X KCFN-4.0-18-7-RA-HC 2014/03 3/1/2017 4667251X KCFN-4.0-18-7-RA-HC 8/5/2014 8/5/2017 4767849 KCFN-4.0-18-7-RA-HC 2014/03 3/1/2017 5122477 KCFN-4.0-18-7-RA-HC 8/8/2014 8/8/2017 5167205 KCFN-4.0-18-7-RA-HC 8/29/2014 8/29/2017 5167205X KCFN-4.0-18-7-RA-HC 8/29/2014 8/29/2017 5537258 KCFN-4.0-18-7-RA-HC 2/10/2015 2/10/2018 5580037 KCFN-4.0-18-7-RA-HC 2/10/2015 2/10/2018 5628941 KCFN-4.0-18-7-RA-HC 3/16/2015 3/16/2018 5688179 KCFN-4.0-18-7-RA-HC 4/9/2015 4/9/2018 5797339 KCFN-4.0-18-7-RA-HC 5/1/2015 5/1/2018 5956253 KCFN-4.0-18-7-RA-HC 7/14/2015 7/14/2018 6101759 KCFN-4.0-18-7-RA-HC 9/4/2015 9/4/2018 6183817 KCFN-4.0-18-7-RA-HC 10/1/2015 10/1/2018 6273554 KCFN-4.0-18-7-RA-HC 11/8/2015 11/8/2018 6351813 KCFN-4.0-18-7-RA-HC 12/11/2015 12/11/2018 4427728X KCFN-5.0-18-13-RA-HC 8/21/2014 8/21/2017 4427728XX KCFN-5.0-18-13-RA-HC 8/21/2014 8/21/2017 4427728XXX KCFN-5.0-18-13-RA-HC 8/21/2014 8/21/2017 4427730X KCFN-5.0-18-13-RA-HC 8/21/2014 8/21/2017 4427730XX KCFN-5.0-18-13-RA-HC 8/21/2014 8/21/2017 4427732X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427734X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427736X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427738X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427740X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427744X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427746X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427748X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427750X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427752X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427754X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427756X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427758X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427760X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427762X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427764X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427766X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427768X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427770X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427772X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427774X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4427776X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017 4667336X KCFN-5.0-18-13-RA-HC 8/5/2014 8/5/2017 4846949 KCFN-5.0-18-13-RA-HC 2014/04 4/1/2017 4846951 KCFN-5.0-18-13-RA-HC 2014/04 4/1/2017 4846953 KCFN-5.0-18-13-RA-HC 2014/04 4/1/2017 4933698 KCFN-5.0-18-13-RA-HC 5/1/2014 5/1/2017 4933700 KCFN-5.0-18-13-RA-HC 6/1/2014 6/1/2017 4933702 KCFN-5.0-18-13-RA-HC 6/1/2014 6/1/2017 5034189 KCFN-5.0-18-13-RA-HC 7/14/2014 7/14/2017 5034191 KCFN-5.0-18-13-RA-HC 7/16/2014 7/16/2017 5034193 KCFN-5.0-18-13-RA-HC 7/14/2014 7/14/2017 5034195 KCFN-5.0-18-13-RA-HC 7/21/2014 7/21/2017 5102745 KCFN-5.0-18-13-RA-HC 7/21/2014 7/21/2017 5102745X KCFN-5.0-18-13-RA-HC 7/21/2014 7/21/2017 5112123 KCFN-5.0-18-13-RA-HC 8/8/2014 8/8/2017 5112125 KCFN-5.0-18-13-RA-HC 8/8/2014 8/8/2017 5112127 KCFN-5.0-18-13-RA-HC 8/15/2014 8/15/2017 5167398 KCFN-5.0-18-13-RA-HC 8/29/2014 8/29/2017 5167400 KCFN-5.0-18-13-RA-HC 8/29/2014 8/29/2017 5167402 KCFN-5.0-18-13-RA-HC 8/29/2014 8/29/2017 5167404 KCFN-5.0-18-13-RA-HC 9/5/2014 9/5/2017 5167406 KCFN-5.0-18-13-RA-HC 9/5/2014 9/5/2017 5224131 KCFN-5.0-18-13-RA-HC 9/5/2014 9/5/2017 5224133 KCFN-5.0-18-13-RA-HC 9/15/2014 9/15/2017 5224135 KCFN-5.0-18-13-RA-HC 9/18/2014 9/18/2017 5253913 KCFN-5.0-18-13-RA-HC 10/20/2014 10/20/2017 5253915 KCFN-5.0-18-13-RA-HC 10/14/2014 10/14/2017 5253917 KCFN-5.0-18-13-RA-HC 10/20/2014 10/20/2017 5253919 KCFN-5.0-18-13-RA-HC 10/31/2014 10/31/2017 5425165 KCFN-5.0-18-13-RA-HC 12/9/2014 12/9/2017 5425167 KCFN-5.0-18-13-RA-HC 12/16/2014 12/16/2017 5425169 KCFN-5.0-18-13-RA-HC 12/16/2014 12/16/2017 5493731 KCFN-5.0-18-13-RA-HC 1/14/2015 1/14/2018 5493733 KCFN-5.0-18-13-RA-HC 1/27/2015 1/27/2018 5493735 KCFN-5.0-18-13-RA-HC 1/14/2015 1/14/2018 5580062 KCFN-5.0-18-13-RA-HC 2/10/2015 2/10/2018 5589721 KCFN-5.0-18-13-RA-HC 3/18/2015 3/18/2018 5589723 KCFN-5.0-18-13-RA-HC 4/6/2015 4/6/2018 5589725 KCFN-5.0-18-13-RA-HC 4/9/2015 4/9/2018 5651232 KCFN-5.0-18-13-RA-HC 4/9/2015 4/9/2018 5651234 KCFN-5.0-18-13-RA-HC 4/16/2015 4/16/2018 5651236 KCFN-5.0-18-13-RA-HC 4/16/2015 4/16/2018 5651238 KCFN-5.0-18-13-RA-HC 5/1/2015 5/1/2018 5703845 KCFN-5.0-18-13-RA-HC 5/1/2015 5/1/2018 5703847 KCFN-5.0-18-13-RA-HC 5/1/2015 5/1/2018 5703849 KCFN-5.0-18-13-RA-HC 5/18/2015 5/18/2018 5893889 KCFN-5.0-18-13-RA-HC 7/9/2015 7/9/2018 5893891 KCFN-5.0-18-13-RA-HC 7/8/2015 7/8/2018 5971098 KCFN-5.0-18-13-RA-HC 7/29/2015 7/29/2018 5994633 KCFN-5.0-18-13-RA-HC 8/19/2015 8/19/2018 5994635 KCFN-5.0-18-13-RA-HC 8/4/2015 8/4/2018 6025703 KCFN-5.0-18-13-RA-HC 8/4/2015 8/4/2018 6025705 KCFN-5.0-18-13-RA-HC 8/2/2015 8/2/2018 6025707 KCFN-5.0-18-13-RA-HC 8/2/2015 8/2/2018 6117310 KCFN-5.0-18-13-RA-HC 9/10/2015 9/10/2018 6117312 KCFN-5.0-18-13-RA-HC 9/17/2015 9/17/2018 6176097 KCFN-5.0-18-13-RA-HC 9/27/2015 9/27/2018 6176099 KCFN-5.0-18-13-RA-HC 10/14/2015 10/14/2018 6192470 KCFN-5.0-18-13-RA-HC 10/2/2015 10/2/2018 6192472 KCFN-5.0-18-13-RA-HC 10/2/2015 10/2/2018 6284394 KCFN-5.0-18-13-RA-HC 12/9/2015 12/9/2018 6284396 KCFN-5.0-18-13-RA-HC 12/9/2015 12/9/2018 4427778X KCFN-5.0-18-13-RA-S-HC 2014/03 3/1/2017 4427780X KCFN-5.0-18-13-RA-S-HC 8/21/2014 8/21/2017 4427780XX KCFN-5.0-18-13-RA-S-HC 8/21/2014 8/21/2017 4667340X KCFN-5.0-18-13-RA-S-HC 8/5/2014 8/5/2017 4893616 KCFN-5.0-18-13-RA-S-HC 2014/04 4/1/2017 4897225 KCFN-5.0-18-13-RA-S-HC 2014/04 4/1/2017 4991650 KCFN-5.0-18-13-RA-S-HC 6/1/2014 6/1/2017 5102747 KCFN-5.0-18-13-RA-S-HC 8/1/2014 8/1/2017 5102747X KCFN-5.0-18-13-RA-S-HC 8/1/2014 8/1/2017 5580064 KCFN-5.0-18-13-RA-S-HC 2/10/2015 2/10/2018 4427680X KCFN-5.0-18-23-RA-HC 8/21/2014 8/21/2017 4427684X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4427686X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4427688X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4427690X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4427692X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4427694X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4427696X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4427696X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4427698X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4427700X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4427702X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4605309X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4605311X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4691483X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4691485X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4691487X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017 4893601 KCFN-5.0-18-23-RA-HC 2014/04 4/1/2017 4893603 KCFN-5.0-18-23-RA-HC 2014/04 4/1/2017 4893605 KCFN-5.0-18-23-RA-HC 2014/04 4/1/2017 4900715 KCFN-5.0-18-23-RA-HC 2014/04 4/1/2017 4900717 KCFN-5.0-18-23-RA-HC 2014/04 4/1/2017 4900719 KCFN-5.0-18-23-RA-HC 2014/05 5/1/2017 4905653 KCFN-5.0-18-23-RA-HC 6/1/2014 6/1/2017 4905655 KCFN-5.0-18-23-RA-HC 6/1/2014 6/1/2017 4905657 KCFN-5.0-18-23-RA-HC 6/1/2014 6/1/2017 5094891 KCFN-5.0-18-23-RA-HC 8/8/2014 8/8/2017 5094891 KCFN-5.0-18-23-RA-HC 8/8/2014 8/8/2017 5094893 KCFN-5.0-18-23-RA-HC 8/15/2014 8/15/2017 5102736 KCFN-5.0-18-23-RA-HC 7/21/2014 7/21/2017 5102736X KCFN-5.0-18-23-RA-HC 7/21/2014 7/21/2017 5122485 KCFN-5.0-18-23-RA-HC 8/11/2014 8/11/2017 5122487 KCFN-5.0-18-23-RA-HC 8/15/2014 8/15/2017 5122489 KCFN-5.0-18-23-RA-HC 8/28/2014 8/28/2017 5143028 KCFN-5.0-18-23-RA-HC 8/15/2014 8/15/2017 5143030 KCFN-5.0-18-23-RA-HC 8/28/2014 8/28/2017 5171213 KCFN-5.0-18-23-RA-HC 8/28/2014 8/28/2017 5171215 KCFN-5.0-18-23-RA-HC 8/28/2014 8/28/2017 5171217 KCFN-5.0-18-23-RA-HC 9/12/2014 9/12/2017 5253880 KCFN-5.0-18-23-RA-HC 10/6/2014 10/6/2017 5253882 KCFN-5.0-18-23-RA-HC 10/14/2014 10/14/2017 5253884 KCFN-5.0-18-23-RA-HC 10/3/2014 10/3/2017 5263587 KCFN-5.0-18-23-RA-HC 10/20/2014 10/20/2017 5263589 KCFN-5.0-18-23-RA-HC 10/20/2014 10/20/2017 5298730 KCFN-5.0-18-23-RA-HC 10/20/2014 10/20/2017 5298732 KCFN-5.0-18-23-RA-HC 10/14/2014 10/14/2017 5298732 KCFN-5.0-18-23-RA-HC 10/14/2014 10/14/2017 5298734 KCFN-5.0-18-23-RA-HC 10/31/2014 10/31/2017 5364979 KCFN-5.0-18-23-RA-HC 12/15/2014 12/15/2017 5364981 KCFN-5.0-18-23-RA-HC 12/11/2014 12/11/2017 5364983 KCFN-5.0-18-23-RA-HC 12/9/2014 12/9/2017 5421713 KCFN-5.0-18-23-RA-HC 11/25/2014 11/25/2017 5421715 KCFN-5.0-18-23-RA-HC 12/11/2014 12/11/2017 5421717 KCFN-5.0-18-23-RA-HC 12/9/2014 12/9/2017 5520711 KCFN-5.0-18-23-RA-HC 1/27/2015 1/27/2018 5520711X KCFN-5.0-18-23-RA-HC 1/27/2015 1/27/2018 5520713 KCFN-5.0-18-23-RA-HC 1/27/2015 1/27/2018 5600196 KCFN-5.0-18-23-RA-HC 3/16/2015 3/16/2018 5600198 KCFN-5.0-18-23-RA-HC 3/16/2015 3/16/2018 5600200 KCFN-5.0-18-23-RA-HC 4/8/2015 4/8/2018 5675240 KCFN-5.0-18-23-RA-HC 4/22/2015 4/22/2018 5675242 KCFN-5.0-18-23-RA-HC 4/22/2015 4/22/2018 5675244 KCFN-5.0-18-23-RA-HC 5/18/2015 5/18/2018 5839392 KCFN-5.0-18-23-RA-HC 6/17/2015 6/17/2018 5839394 KCFN-5.0-18-23-RA-HC 6/26/2015 6/26/2018 5842468 KCFN-5.0-18-23-RA-HC 6/29/2015 6/29/2018 5842470 KCFN-5.0-18-23-RA-HC 6/24/2015 6/24/2018 6019212 KCFN-5.0-18-23-RA-HC 7/31/2015 7/31/2018 6019214 KCFN-5.0-18-23-RA-HC 7/31/2015 7/31/2018 6091378 KCFN-5.0-18-23-RA-HC 8/30/2015 8/30/2018 6192461 KCFN-5.0-18-23-RA-HC 10/15/2015 10/15/2018 6192463 KCFN-5.0-18-23-RA-HC 10/15/2015 10/15/2018 6212495 KCFN-5.0-18-23-RA-HC 10/15/2015 10/15/2018 6212497 KCFN-5.0-18-23-RA-HC 10/20/2015 10/20/2018 6237778 KCFN-5.0-18-23-RA-HC 11/8/2015 11/8/2018 6237780 KCFN-5.0-18-23-RA-HC 11/3/2015 11/3/2018 6263210 KCFN-5.0-18-23-RA-HC 11/8/2015 11/8/2018 6263212 KCFN-5.0-18-23-RA-HC 11/25/2015 11/25/2018 6280180 KCFN-5.0-18-23-RA-HC 11/29/2015 11/29/2018 6354966 KCFN-5.0-18-23-RA-HC 12/9/2015 12/9/2018 NS5230210 KCFN-5.0-18-23-RA-HC 10/6/2014 10/6/2017 NS5230210 KCFN-5.0-18-23-RA-HC 10/6/2014 10/6/2017 NS5230210 KCFN-5.0-18-23-RA-HC 10/6/2014 10/6/2017 4427788X KCFN-5.0-18-7-RA-HC 8/21/2014 8/21/2017 4427788XX KCFN-5.0-18-7-RA-HC 8/21/2014 8/21/2017 4667356X KCFN-5.0-18-7-RA-HC 8/5/2014 8/5/2017 4765057 KCFN-5.0-18-7-RA-HC 2014/02 2/1/2017 4838340 KCFN-5.0-18-7-RA-HC 2014/04 4/1/2017 4893627 KCFN-5.0-18-7-RA-HC 2014/04 4/1/2017 5167315 KCFN-5.0-18-7-RA-HC 10/6/2014 10/6/2017 5167315X KCFN-5.0-18-7-RA-HC 10/6/2014 10/6/2017 5357936 KCFN-5.0-18-7-RA-HC 12/9/2014 12/9/2017 5467702 KCFN-5.0-18-7-RA-HC 12/16/2014 12/16/2017 5467702X KCFN-5.0-18-7-RA-HC 12/16/2014 12/16/2017 5493765 KCFN-5.0-18-7-RA-HC 1/13/2015 1/13/2018 5580117 KCFN-5.0-18-7-RA-HC 2/10/2015 2/10/2018 5634995 KCFN-5.0-18-7-RA-HC 3/18/2015 3/18/2018 5666595 KCFN-5.0-18-7-RA-HC 4/15/2015 4/15/2018 5688231 KCFN-5.0-18-7-RA-HC 4/15/2015 4/15/2018 5727259 KCFN-5.0-18-7-RA-HC 5/11/2015 5/11/2018 5736885 KCFN-5.0-18-7-RA-HC 4/15/2015 4/15/2018 5739813 KCFN-5.0-18-7-RA-HC 5/1/2015 5/1/2018 5745146 KCFN-5.0-18-7-RA-HC 5/19/2015 5/19/2018 5748864 KCFN-5.0-18-7-RA-HC 5/12/2015 5/12/2018 5751980 KCFN-5.0-18-7-RA-HC 5/1/2015 5/1/2018 6307647 KCFN-5.0-18-7-RA-HC 12/11/2015 12/11/2018 4427570X KCFN-6.0-18-13-RA-HC 8/21/2014 8/21/2017 4427570XX KCFN-6.0-18-13-RA-HC 8/20/2014 8/20/2017 4427572X KCFN-6.0-18-13-RA-HC 8/21/2014 8/21/2017 4427572XX KCFN-6.0-18-13-RA-HC 8/21/2014 8/21/2017 4427578X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427580X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427582X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427584X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427586X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427588X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427590X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427594X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427596X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427598X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427600X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427602X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427604X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427606X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427608X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427610X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427612X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427614X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427616X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427618X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427620X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427622X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427624X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427626X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427628X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427630X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427632X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427634X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427636X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427638X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427640X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427642X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427644X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427646X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427648X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427650X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4427652X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017 4667237X KCFN-6.0-18-13-RA-HC 8/5/2014 8/5/2017 4667241X KCFN-6.0-18-13-RA-HC 8/5/2014 8/5/2017 4667247X KCFN-6.0-18-13-RA-HC 8/5/2014 8/5/2017 4667249X KCFN-6.0-18-13-RA-HC 8/5/2014 8/5/2017 4893552 KCFN-6.0-18-13-RA-HC 2014/04 4/1/2017 4893554 KCFN-6.0-18-13-RA-HC 2014/04 4/1/2017 4893556 KCFN-6.0-18-13-RA-HC 2014/04 4/1/2017 4893558 KCFN-6.0-18-13-RA-HC 2014/04 4/1/2017 4893562 KCFN-6.0-18-13-RA-HC 2014/04 4/1/2017 4893562X KCFN-6.0-18-13-RA-HC 4/1/2014 4/1/2017 5005398 KCFN-6.0-18-13-RA-HC 6/1/2014 6/1/2017 5005844 KCFN-6.0-18-13-RA-HC 6/1/2014 6/1/2017 5005846 KCFN-6.0-18-13-RA-HC 7/1/2014 7/1/2017 5005848 KCFN-6.0-18-13-RA-HC 7/1/2014 7/1/2017 5005850 KCFN-6.0-18-13-RA-HC 6/1/2014 6/1/2017 5005852 KCFN-6.0-18-13-RA-HC 7/1/2014 7/1/2017 5034172 KCFN-6.0-18-13-RA-HC 7/24/2014 7/24/2017 5034174 KCFN-6.0-18-13-RA-HC 7/14/2014 7/14/2017 5034176 KCFN-6.0-18-13-RA-HC 7/15/2014 7/15/2017 5034178 KCFN-6.0-18-13-RA-HC 7/14/2014 7/14/2017 5070910 KCFN-6.0-18-13-RA-HC 7/24/2014 7/24/2017 5070912 KCFN-6.0-18-13-RA-HC 7/24/2014 7/24/2017 5070914 KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017 5070916X KCFN-6.0-18-13-RA-HC 8/8/2014 8/8/2017 5070918 KCFN-6.0-18-13-RA-HC 8/1/2014 8/1/2017 5102709 KCFN-6.0-18-13-RA-HC 8/1/2014 8/1/2017 5102709X KCFN-6.0-18-13-RA-HC 8/1/2014 8/1/2017 5102711 KCFN-6.0-18-13-RA-HC 7/24/2014 7/24/2017 5102713 KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017 5102713X KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017 5102715 KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017 5102715X KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017 5112129 KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017 5112131 KCFN-6.0-18-13-RA-HC 8/15/2014 8/15/2017 5112133 KCFN-6.0-18-13-RA-HC 9/15/2014 9/15/2017 5112135 KCFN-6.0-18-13-RA-HC 9/15/2014 9/15/2017 5152643 KCFN-6.0-18-13-RA-HC 9/18/2014 9/18/2017 5152645 KCFN-6.0-18-13-RA-HC 9/18/2014 9/18/2017 5152647 KCFN-6.0-18-13-RA-HC 9/15/2014 9/15/2017 5152649 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017 5152651 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017 5152653 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017 5197821 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017 5197823 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017 5197825 KCFN-6.0-18-13-RA-HC 10/14/2014 10/14/2017 5197827 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017 5197829 KCFN-6.0-18-13-RA-HC 10/14/2014 10/14/2017 5219648 KCFN-6.0-18-13-RA-HC 10/14/2014 10/14/2017 5219650 KCFN-6.0-18-13-RA-HC 10/31/2014 10/31/2017 5219652 KCFN-6.0-18-13-RA-HC 11/6/2014 11/6/2017 5219654 KCFN-6.0-18-13-RA-HC 10/31/2014 10/31/2017 5219656 KCFN-6.0-18-13-RA-HC 11/11/2014 11/11/2017 5219658 KCFN-6.0-18-13-RA-HC 11/6/2014 11/6/2017 5289239 KCFN-6.0-18-13-RA-HC 11/6/2014 11/6/2017 5289241 KCFN-6.0-18-13-RA-HC 11/6/2014 11/6/2017 5289243 KCFN-6.0-18-13-RA-HC 11/10/2014 11/10/2017 5289245 KCFN-6.0-18-13-RA-HC 11/6/2014 11/6/2017 5289247 KCFN-6.0-18-13-RA-HC 11/11/2014 11/11/2017 5289247 KCFN-6.0-18-13-RA-HC 11/11/2014 11/11/2017 5322547 KCFN-6.0-18-13-RA-HC 11/10/2014 11/10/2017 5322549 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017 5322551 KCFN-6.0-18-13-RA-HC 11/11/2014 11/11/2017 5322553 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017 5350870 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017 5350872 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017 5350874 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017 5350876 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017 5350878 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017 5373504 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017 5373506 KCFN-6.0-18-13-RA-HC 12/9/2014 12/9/2017 5373508 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017 5373510 KCFN-6.0-18-13-RA-HC 12/9/2014 12/9/2017 5373512 KCFN-6.0-18-13-RA-HC 12/9/2014 12/9/2017 5432906 KCFN-6.0-18-13-RA-HC 12/9/2014 12/9/2017 5432908 KCFN-6.0-18-13-RA-HC 12/16/2014 12/16/2017 5432910 KCFN-6.0-18-13-RA-HC 1/8/2015 1/8/2018 5432912 KCFN-6.0-18-13-RA-HC 1/7/2015 1/7/2018 5449003 KCFN-6.0-18-13-RA-HC 1/12/2015 1/12/2018 5449005 KCFN-6.0-18-13-RA-HC 1/12/2015 1/12/2018 5449007 KCFN-6.0-18-13-RA-HC 1/6/2015 1/6/2018 5449009 KCFN-6.0-18-13-RA-HC 12/31/2014 12/31/2017 5449011 KCFN-6.0-18-13-RA-HC 1/7/2015 1/7/2018 5478604 KCFN-6.0-18-13-RA-HC 1/7/2015 1/7/2018 5478606 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018 5478608 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018 5478610 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018 5478612 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018 5511396 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018 5511396X KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018 5511398 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018 5511400 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018 5511402 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018 5511404 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018 5511406 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018 5533811 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018 5543747 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018 5543751 KCFN-6.0-18-13-RA-HC 2/10/2015 2/10/2018 5543753 KCFN-6.0-18-13-RA-HC 2/10/2015 2/10/2018 5543755 KCFN-6.0-18-13-RA-HC 2/25/2015 2/25/2018 5569070 KCFN-6.0-18-13-RA-HC 2/25/2015 2/25/2018 5569072 KCFN-6.0-18-13-RA-HC 3/19/2015 3/19/2018 5569074 KCFN-6.0-18-13-RA-HC 3/19/2015 3/19/2018 5569074X KCFN-6.0-18-13-RA-HC 3/19/2015 3/19/2018 5569076 KCFN-6.0-18-13-RA-HC 3/19/2015 3/19/2018 5569078 KCFN-6.0-18-13-RA-HC 3/18/2015 3/18/2018 5569080 KCFN-6.0-18-13-RA-HC 4/6/2015 4/6/2018 5580025 KCFN-6.0-18-13-RA-HC 2/12/2015 2/12/2018 5580027 KCFN-6.0-18-13-RA-HC 2/10/2015 2/10/2018 5580029 KCFN-6.0-18-13-RA-HC 2/12/2015 2/12/2018 5580031 KCFN-6.0-18-13-RA-HC 2/10/2015 2/10/2018 5580033 KCFN-6.0-18-13-RA-HC 2/12/2015 2/12/2018 5597115 KCFN-6.0-18-13-RA-HC 3/18/2015 3/18/2018 5597117 KCFN-6.0-18-13-RA-HC 3/18/2015 3/18/2018 5597119 KCFN-6.0-18-13-RA-HC 4/9/2015 4/9/2018 5597121 KCFN-6.0-18-13-RA-HC 3/30/2015 3/30/2018 5597121X KCFN-6.0-18-13-RA-HC 3/30/2015 3/30/2018 5597123 KCFN-6.0-18-13-RA-HC 3/30/2015 3/30/2018 5615321 KCFN-6.0-18-13-RA-HC 3/30/2015 3/30/2018 5632276 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018 5648402 KCFN-6.0-18-13-RA-HC 4/8/2015 4/8/2018 5648404 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018 5648406 KCFN-6.0-18-13-RA-HC 4/16/2015 4/16/2018 5648408 KCFN-6.0-18-13-RA-HC 4/16/2015 4/16/2018 5648410 KCFN-6.0-18-13-RA-HC 4/20/2015 4/20/2018 5648412 KCFN-6.0-18-13-RA-HC 4/16/2015 4/16/2018 5666544 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018 5666546 KCFN-6.0-18-13-RA-HC 4/21/2015 4/21/2018 5666548 KCFN-6.0-18-13-RA-HC 4/21/2015 4/21/2018 5666550 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018 5666552 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018 5684510 KCFN-6.0-18-13-RA-HC 2/28/2015 2/28/2018 5717938 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018 5717940 KCFN-6.0-18-13-RA-HC 5/22/2015 5/22/2018 5717942 KCFN-6.0-18-13-RA-HC 5/12/2015 5/12/2018 5717944 KCFN-6.0-18-13-RA-HC 5/12/2015 5/12/2018 5717946 KCFN-6.0-18-13-RA-HC 5/19/2015 5/19/2018 5758608 KCFN-6.0-18-13-RA-HC 5/26/2015 5/26/2018 5758610 KCFN-6.0-18-13-RA-HC 5/22/2015 5/22/2018 5758612 KCFN-6.0-18-13-RA-HC 5/22/2015 5/22/2018 5758614 KCFN-6.0-18-13-RA-HC 5/22/2015 5/22/2018 5758616 KCFN-6.0-18-13-RA-HC 5/22/2015 5/22/2018 5797329 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018 5797331 KCFN-6.0-18-13-RA-HC 6/17/2015 6/17/2018 5797333 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018 5797335 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018 5797337 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018 5825850 KCFN-6.0-18-13-RA-HC 6/17/2015 6/17/2018 5825852 KCFN-6.0-18-13-RA-HC 6/17/2015 6/17/2018 5825854 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018 5825856 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018 5825858 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018 5845959 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018 5845961 KCFN-6.0-18-13-RA-HC 6/17/2015 6/17/2018 5845963 KCFN-6.0-18-13-RA-HC 7/13/2015 7/13/2018 5845965 KCFN-6.0-18-13-RA-HC 7/13/2015 7/13/2018 5893856 KCFN-6.0-18-13-RA-HC 6/24/2015 6/24/2018 5893858 KCFN-6.0-18-13-RA-HC 7/8/2015 7/8/2018 5893860 KCFN-6.0-18-13-RA-HC 7/8/2015 7/8/2018 5893862 KCFN-6.0-18-13-RA-HC 7/8/2015 7/8/2018 5893864 KCFN-6.0-18-13-RA-HC 7/8/2015 7/8/2018 5909225 KCFN-6.0-18-13-RA-HC 7/13/2015 7/13/2018 5909227 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018 5909229 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018 5909231 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018 5909233 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018 5909235 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018 5952767 KCFN-6.0-18-13-RA-HC 7/16/2015 7/16/2018 5952769 KCFN-6.0-18-13-RA-HC 7/15/2015 7/15/2018 5952771 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018 5952773 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018 5959457 KCFN-6.0-18-13-RA-HC 7/15/2015 7/15/2018 5959459 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018 5959461 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018 5959463 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018 5959465 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018 5984820 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018 5984822 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018 5984824 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018 5984826 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018 5984828 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018 6032995 KCFN-6.0-18-13-RA-HC 8/18/2015 8/18/2018 6032997 KCFN-6.0-18-13-RA-HC 8/2/2015 8/2/2018 6032999 KCFN-6.0-18-13-RA-HC 8/19/2015 8/19/2018 6032999 KCFN-6.0-18-13-RA-HC 8/19/2015 8/19/2018 6033001 KCFN-6.0-18-13-RA-HC 9/16/2015 9/16/2018 6033003 KCFN-6.0-18-13-RA-HC 8/30/2015 8/30/2018 6109716 KCFN-6.0-18-13-RA-HC 9/4/2015 9/4/2018 6109718 KCFN-6.0-18-13-RA-HC 9/16/2015 9/16/2018 6109720 KCFN-6.0-18-13-RA-HC 9/17/2015 9/17/2018 6109722 KCFN-6.0-18-13-RA-HC 9/17/2015 9/17/2018 6109724 KCFN-6.0-18-13-RA-HC 9/10/2015 9/10/2018 6117265 KCFN-6.0-18-13-RA-HC 9/10/2015 9/10/2018 6117267 KCFN-6.0-18-13-RA-HC 9/17/2015 9/17/2018 6117269 KCFN-6.0-18-13-RA-HC 9/17/2015 9/17/2018 6117271 KCFN-6.0-18-13-RA-HC 9/16/2015 9/16/2018 6176082 KCFN-6.0-18-13-RA-HC 10/2/2015 10/2/2018 6176084 KCFN-6.0-18-13-RA-HC 10/1/2015 10/1/2018 6176086 KCFN-6.0-18-13-RA-HC 10/2/2015 10/2/2018 6176088 KCFN-6.0-18-13-RA-HC 10/2/2015 10/2/2018 6224299 KCFN-6.0-18-13-RA-HC 10/16/2015 10/16/2018 6224301 KCFN-6.0-18-13-RA-HC 10/29/2015 10/29/2018 6224303 KCFN-6.0-18-13-RA-HC 11/8/2015 11/8/2018 6224305 KCFN-6.0-18-13-RA-HC 11/3/2015 11/3/2018 6240884 KCFN-6.0-18-13-RA-HC 11/3/2015 11/3/2018 6240886 KCFN-6.0-18-13-RA-HC 10/29/2015 10/29/2018 6240888 KCFN-6.0-18-13-RA-HC 10/29/2015 10/29/2018 6240890 KCFN-6.0-18-13-RA-HC 11/8/2015 11/8/2018 6251597 KCFN-6.0-18-13-RA-HC 11/25/2015 11/25/2018 6251599 KCFN-6.0-18-13-RA-HC 11/29/2015 11/29/2018 6251601 KCFN-6.0-18-13-RA-HC 11/25/2015 11/25/2018 6251601X KCFN-6.0-18-13-RA-HC 11/25/2015 11/25/2018 6251603 KCFN-6.0-18-13-RA-HC 11/25/2015 11/25/2018 6293425 KCFN-6.0-18-13-RA-HC 12/9/2015 12/9/2018 6293427 KCFN-6.0-18-13-RA-HC 12/9/2015 12/9/2018 4427704X KCFN-6.0-18-13-RA-S-HC 8/21/2014 8/21/2017 4427704XX KCFN-6.0-18-13-RA-S-HC 8/21/2014 8/21/2017 4427706X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017 4427708X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017 4427710X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017 4427712X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017 4427714X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017 4427716X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017 4427718X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017 4427720X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017 4427722X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017 4427724X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017 4667286X KCFN-6.0-18-13-RA-S-HC 8/5/2014 8/5/2017 4667292X KCFN-6.0-18-13-RA-S-HC 8/5/2014 8/5/2017 4897210 KCFN-6.0-18-13-RA-S-HC 2014/04 4/1/2017 4897212 KCFN-6.0-18-13-RA-S-HC 2014/04 4/1/2017 4900721 KCFN-6.0-18-13-RA-S-HC 2014/04 4/1/2017 4900723 KCFN-6.0-18-13-RA-S-HC 4/1/2014 4/1/2017 4900725 KCFN-6.0-18-13-RA-S-HC 2014/04 4/1/2017 4900727 KCFN-6.0-18-13-RA-S-HC 2014/04 4/1/2017 4905659 KCFN-6.0-18-13-RA-S-HC 7/1/2014 7/1/2017 4905661 KCFN-6.0-18-13-RA-S-HC 7/1/2014 7/1/2017 5005399 KCFN-6.0-18-13-RA-S-HC 6/1/2014 6/1/2017 5102738 KCFN-6.0-18-13-RA-S-HC 8/1/2014 8/1/2017 5102740 KCFN-6.0-18-13-RA-S-HC 8/11/2014 8/11/2017 5102740X KCFN-6.0-18-13-RA-S-HC 8/11/2014 8/11/2017 5520550 KCFN-6.0-18-13-RA-S-HC 1/14/2015 1/14/2018 5520552 KCFN-6.0-18-13-RA-S-HC 1/14/2015 1/14/2018 5520554 KCFN-6.0-18-13-RA-S-HC 1/27/2015 1/27/2018 5580057 KCFN-6.0-18-13-RA-S-HC 2/25/2015 2/25/2018 4427538X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4427540X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4427542X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4427546X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4427548X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4556115X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4556117X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4556119X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4556121X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4556123X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4556123X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4600056XX KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4600058X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4600060X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4600062X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4600064X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4600066X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4635509X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4635511X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4635513X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4635515X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4635517X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4676509X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4718572 KCFN-6.0-18-23-RA-HC 2014/01 1/1/2017 4718572X KCFN-6.0-18-23-RA-HC 2014/01 1/1/2017 4718574 KCFN-6.0-18-23-RA-HC 2014/02 2/1/2017 4718576 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4718578 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4718580 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4770554 KCFN-6.0-18-23-RA-HC 2014/02 2/1/2017 4770556 KCFN-6.0-18-23-RA-HC 2014/02 2/1/2017 4770558 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4770560 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4770562 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4770564 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4770566 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017 4844431 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4844431X KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4844433 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4844435 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4844437 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4844439 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4844441 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4844443 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4879430 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4879432 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4879434 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4879436 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4879438 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4879440 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017 4917588 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017 4917590 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017 4917592 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017 4917594 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017 4917596 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017 4917598 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017 4917600 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017 4917602 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017 4977711 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 4977713 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 4977715 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 4977717 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 4977719 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 4981344 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 4981346 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 4981348 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 4981350 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 4981352 KCFN-6.0-18-23-RA-HC 7/1/2014 7/1/2017 4981354 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 4981356 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 5008370 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 5008372 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 5008374 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 5008376 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 5008378 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 5008380 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 5008382 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 5008384 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 5008386 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017 5042551 KCFN-6.0-18-23-RA-HC 7/14/2014 7/14/2017 5042553 KCFN-6.0-18-23-RA-HC 7/15/2014 7/15/2017 5042555 KCFN-6.0-18-23-RA-HC 7/15/2014 7/15/2017 5042557 KCFN-6.0-18-23-RA-HC 7/14/2014 7/14/2017 5042559 KCFN-6.0-18-23-RA-HC 7/15/2014 7/15/2017 5042561 KCFN-6.0-18-23-RA-HC 7/21/2014 7/21/2017 5070896 KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017 5070898 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017 5070900 KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017 5070902 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017 5070904 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017 5070906 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017 5076722 KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017 5076724 KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017 5076726 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017 5076728 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017 5076730 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017 5102703 KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017 5102703X KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017 5112137 KCFN-6.0-18-23-RA-HC 8/15/2014 8/15/2017 5112139 KCFN-6.0-18-23-RA-HC 8/8/2014 8/8/2017 5112141 KCFN-6.0-18-23-RA-HC 8/11/2014 8/11/2017 5112143 KCFN-6.0-18-23-RA-HC 8/8/2014 8/8/2017 5112145 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017 5112147 KCFN-6.0-18-23-RA-HC 9/8/2014 9/8/2017 5126714 KCFN-6.0-18-23-RA-HC 8/15/2014 8/15/2017 5126716 KCFN-6.0-18-23-RA-HC 8/15/2014 8/15/2017 5126718 KCFN-6.0-18-23-RA-HC 8/11/2014 8/11/2017 5126720 KCFN-6.0-18-23-RA-HC 8/28/2014 8/28/2017 5126722 KCFN-6.0-18-23-RA-HC 8/15/2014 8/15/2017 5126724 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017 5126726 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017 5126728 KCFN-6.0-18-23-RA-HC 9/9/2014 9/9/2017 5152608 KCFN-6.0-18-23-RA-HC 8/15/2014 8/15/2017 5152610 KCFN-6.0-18-23-RA-HC 9/18/2014 9/18/2017 5152612 KCFN-6.0-18-23-RA-HC 9/15/2014 9/15/2017 5152614 KCFN-6.0-18-23-RA-HC 9/15/2014 9/15/2017 5152616 KCFN-6.0-18-23-RA-HC 9/15/2014 9/15/2017 5152618 KCFN-6.0-18-23-RA-HC 9/8/2014 9/8/2017 5152620 KCFN-6.0-18-23-RA-HC 9/15/2014 9/15/2017 5152622 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017 5152624 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017 5152626 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017 5178902 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017 5178904 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017 5178906 KCFN-6.0-18-23-RA-HC 9/15/2014 9/15/2017 5178908 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017 5178910 KCFN-6.0-18-23-RA-HC 10/3/2014 10/3/2017 5178912 KCFN-6.0-18-23-RA-HC 10/3/2014 10/3/2017 5210537 KCFN-6.0-18-23-RA-HC 10/31/2014 10/31/2017 5210539 KCFN-6.0-18-23-RA-HC 10/31/2014 10/31/2017 5210541 KCFN-6.0-18-23-RA-HC 10/31/2014 10/31/2017 5210543 KCFN-6.0-18-23-RA-HC 10/31/2014 10/31/2017 5210545 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017 5210547 KCFN-6.0-18-23-RA-HC 10/31/2014 10/31/2017 5210549 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017 5244230 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017 5244232 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017 5244234 KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017 5244236 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017 5244238 KCFN-6.0-18-23-RA-HC 11/10/2014 11/10/2017 5244240 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017 5263531 KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017 5263531X KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017 5263533 KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017 5263535 KCFN-6.0-18-23-RA-HC 11/10/2014 11/10/2017 5263535 KCFN-6.0-18-23-RA-HC 11/10/2014 11/10/2017 5263537 KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017 5263539 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017 5263541 KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017 5325914 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017 5325916 KCFN-6.0-18-23-RA-HC 11/25/2014 11/25/2017 5325918 KCFN-6.0-18-23-RA-HC 11/25/2014 11/25/2017 5325920 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017 5325922 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017 5347661 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017 5347663 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017 5347665 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017 5347667 KCFN-6.0-18-23-RA-HC 11/25/2014 11/25/2017 5347669 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017 5347671 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017 5376572 KCFN-6.0-18-23-RA-HC 11/25/2014 11/25/2017 5376574 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017 5376576 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017 5376578 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017 5376580 KCFN-6.0-18-23-RA-HC 12/8/2014 12/8/2017 5376582 KCFN-6.0-18-23-RA-HC 12/9/2014 12/9/2017 5376584 KCFN-6.0-18-23-RA-HC 12/9/2014 12/9/2017 5394985 KCFN-6.0-18-23-RA-HC 12/9/2014 12/9/2017 5394987 KCFN-6.0-18-23-RA-HC 12/8/2014 12/8/2017 5394989 KCFN-6.0-18-23-RA-HC 12/8/2014 12/8/2017 5394991 KCFN-6.0-18-23-RA-HC 12/11/2014 12/11/2017 5394993 KCFN-6.0-18-23-RA-HC 12/9/2014 12/9/2017 5394995 KCFN-6.0-18-23-RA-HC 12/8/2014 12/8/2017 5394997 KCFN-6.0-18-23-RA-HC 12/11/2014 12/11/2017 5491309 KCFN-6.0-18-23-RA-HC 1/12/2015 1/12/2018 5491311 KCFN-6.0-18-23-RA-HC 1/12/2015 1/12/2018 5491313 KCFN-6.0-18-23-RA-HC 1/12/2015 1/12/2018 5491315 KCFN-6.0-18-23-RA-HC 1/14/2015 1/14/2018 5491317 KCFN-6.0-18-23-RA-HC 1/12/2015 1/12/2018 5491319 KCFN-6.0-18-23-RA-HC 1/14/2015 1/14/2018 5514280 KCFN-6.0-18-23-RA-HC 1/14/2015 1/14/2018 5514282 KCFN-6.0-18-23-RA-HC 1/27/2015 1/27/2018 5514284 KCFN-6.0-18-23-RA-HC 1/14/2015 1/14/2018 5514286 KCFN-6.0-18-23-RA-HC 1/12/2015 1/12/2018 5514288 KCFN-6.0-18-23-RA-HC 1/27/2015 1/27/2018 5514290 KCFN-6.0-18-23-RA-HC 1/23/2015 1/23/2018 5514292 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018 5547576 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018 5547578 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018 5547580 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018 5547582 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018 5547584 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018 5573174 KCFN-6.0-18-23-RA-HC 2/25/2015 2/25/2018 5573176 KCFN-6.0-18-23-RA-HC 2/25/2015 2/25/2018 5573178 KCFN-6.0-18-23-RA-HC 2/25/2015 2/25/2018 5573180 KCFN-6.0-18-23-RA-HC 3/16/2015 3/16/2018 5573182 KCFN-6.0-18-23-RA-HC 3/16/2015 3/16/2018 5573182X KCFN-6.0-18-23-RA-HC 3/16/2015 3/16/2018 5573184 KCFN-6.0-18-23-RA-HC 3/16/2015 3/16/2018 5619200 KCFN-6.0-18-23-RA-HC 3/16/2015 3/16/2018 5619202 KCFN-6.0-18-23-RA-HC 3/30/2015 3/30/2018 5619204 KCFN-6.0-18-23-RA-HC 4/8/2015 4/8/2018 5619206 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018 5619206 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018 5619208 KCFN-6.0-18-23-RA-HC 3/30/2015 3/30/2018 5637359 KCFN-6.0-18-23-RA-HC 4/9/2015 4/9/2018 5637361 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018 5637363 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018 5637365 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018 5637367 KCFN-6.0-18-23-RA-HC 4/16/2015 4/16/2018 5637369 KCFN-6.0-18-23-RA-HC 4/22/2015 4/22/2018 5637371 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018 5703791 KCFN-6.0-18-23-RA-HC 5/19/2015 5/19/2018 5703793 KCFN-6.0-18-23-RA-HC 5/18/2015 5/18/2018 5703795 KCFN-6.0-18-23-RA-HC 5/19/2015 5/19/2018 5703797 KCFN-6.0-18-23-RA-HC 5/18/2015 5/18/2018 5703799 KCFN-6.0-18-23-RA-HC 5/18/2015 5/18/2018 5703801 KCFN-6.0-18-23-RA-HC 5/18/2015 5/18/2018 5703803 KCFN-6.0-18-23-RA-HC 5/19/2015 5/19/2018 5834809 KCFN-6.0-18-23-RA-HC 6/24/2015 6/24/2018 5834811 KCFN-6.0-18-23-RA-HC 6/26/2015 6/26/2018 5834813 KCFN-6.0-18-23-RA-HC 6/26/2015 6/26/2018 5834815 KCFN-6.0-18-23-RA-HC 6/26/2015 6/26/2018 5861895 KCFN-6.0-18-23-RA-HC 6/24/2015 6/24/2018 5861897 KCFN-6.0-18-23-RA-HC 6/26/2015 6/26/2018 5861899 KCFN-6.0-18-23-RA-HC 6/24/2015 6/24/2018 5861901 KCFN-6.0-18-23-RA-HC 7/8/2015 7/8/2018 5861903 KCFN-6.0-18-23-RA-HC 7/8/2015 7/8/2018 5861905 KCFN-6.0-18-23-RA-HC 7/8/2015 7/8/2018 5899836 KCFN-6.0-18-23-RA-HC 7/9/2015 7/9/2018 5899838 KCFN-6.0-18-23-RA-HC 7/9/2015 7/9/2018 5899840 KCFN-6.0-18-23-RA-HC 7/8/2015 7/8/2018 5899842 KCFN-6.0-18-23-RA-HC 7/13/2015 7/13/2018 5899844 KCFN-6.0-18-23-RA-HC 7/13/2015 7/13/2018 5916332 KCFN-6.0-18-23-RA-HC 7/14/2015 7/14/2018 5916334 KCFN-6.0-18-23-RA-HC 7/14/2015 7/14/2018 5916336 KCFN-6.0-18-23-RA-HC 7/13/2015 7/13/2018 5916338 KCFN-6.0-18-23-RA-HC 7/13/2015 7/13/2018 5988945 KCFN-6.0-18-23-RA-HC 7/30/2015 7/30/2018 5988947 KCFN-6.0-18-23-RA-HC 7/29/2015 7/29/2018 5988949 KCFN-6.0-18-23-RA-HC 7/31/2015 7/31/2018 5988951 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018 5988953 KCFN-6.0-18-23-RA-HC 7/31/2015 7/31/2018 6004265 KCFN-6.0-18-23-RA-HC 7/29/2015 7/29/2018 6004267 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018 6004269 KCFN-6.0-18-23-RA-HC 7/31/2015 7/31/2018 6004271 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018 6004273 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018 6004275 KCFN-6.0-18-23-RA-HC 8/2/2015 8/2/2018 6011946 KCFN-6.0-18-23-RA-HC 7/31/2015 7/31/2018 6011948 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018 6011950 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018 6011952 KCFN-6.0-18-23-RA-HC 8/2/2015 8/2/2018 6011954 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018 6011956 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018 6011958 KCFN-6.0-18-23-RA-HC 8/18/2015 8/18/2018 6055589 KCFN-6.0-18-23-RA-HC 8/24/2015 8/24/2018 6055591 KCFN-6.0-18-23-RA-HC 8/17/2015 8/17/2018 6055593 KCFN-6.0-18-23-RA-HC 8/18/2015 8/18/2018 6055595 KCFN-6.0-18-23-RA-HC 8/18/2015 8/18/2018 6055597 KCFN-6.0-18-23-RA-HC 8/17/2015 8/17/2018 6055599 KCFN-6.0-18-23-RA-HC 8/30/2015 8/30/2018 6055601 KCFN-6.0-18-23-RA-HC 9/4/2015 9/4/2018 6080971 KCFN-6.0-18-23-RA-HC 9/4/2015 9/4/2018 6080973 KCFN-6.0-18-23-RA-HC 9/4/2015 9/4/2018 6080975 KCFN-6.0-18-23-RA-HC 9/4/2015 9/4/2018 6080977 KCFN-6.0-18-23-RA-HC 9/9/2015 9/9/2018 6080979 KCFN-6.0-18-23-RA-HC 9/9/2015 9/9/2018 6150471 KCFN-6.0-18-23-RA-HC 9/16/2015 9/16/2018 6150473 KCFN-6.0-18-23-RA-HC 9/25/2015 9/25/2018 6150475 KCFN-6.0-18-23-RA-HC 9/9/2015 9/9/2018 6150477 KCFN-6.0-18-23-RA-HC 9/18/2015 9/18/2018 6154447 KCFN-6.0-18-23-RA-HC 9/18/2015 9/18/2018 6154449 KCFN-6.0-18-23-RA-HC 9/18/2015 9/18/2018 6154451 KCFN-6.0-18-23-RA-HC 9/27/2015 9/27/2018 6154451X KCFN-6.0-18-23-RA-HC 9/27/2015 9/27/2018 6154453 KCFN-6.0-18-23-RA-HC 10/2/2015 10/2/2018 6183805 KCFN-6.0-18-23-RA-HC 9/27/2015 9/27/2018 6183807 KCFN-6.0-18-23-RA-HC 10/2/2015 10/2/2018 6183809 KCFN-6.0-18-23-RA-HC 10/4/2015 10/4/2018 6183811 KCFN-6.0-18-23-RA-HC 10/1/2015 10/1/2018 6217250 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018 6217252 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018 6217254 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018 6217256 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018 6217258 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018 6217260 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018 6217262 KCFN-6.0-18-23-RA-HC 11/3/2015 11/3/2018 6266738 KCFN-6.0-18-23-RA-HC 11/8/2015 11/8/2018 6266740 KCFN-6.0-18-23-RA-HC 11/29/2015 11/29/2018 6266742 KCFN-6.0-18-23-RA-HC 11/25/2015 11/25/2018 6266744 KCFN-6.0-18-23-RA-HC 11/29/2015 11/29/2018 6266746 KCFN-6.0-18-23-RA-HC 12/11/2015 12/11/2018 6266746 KCFN-6.0-18-23-RA-HC 12/11/2015 12/11/2018 6266748 KCFN-6.0-18-23-RA-HC 11/25/2015 11/25/2018 6266750 KCFN-6.0-18-23-RA-HC 11/25/2015 11/25/2018 6280152 KCFN-6.0-18-23-RA-HC 11/25/2015 11/25/2018 6280154 KCFN-6.0-18-23-RA-HC 12/9/2015 12/9/2018 6280156 KCFN-6.0-18-23-RA-HC 12/8/2015 12/8/2018 6280156 KCFN-6.0-18-23-RA-HC 12/8/2015 12/8/2018 6280158 KCFN-6.0-18-23-RA-HC 12/8/2015 12/8/2018 6280160 KCFN-6.0-18-23-RA-HC 12/9/2015 12/9/2018 NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017 NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017 NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017 NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017 NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017 NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017 NS5230212 KCFN-6.0-18-23-RA-HC 11/13/2014 11/13/2017 NS5230212 KCFN-6.0-18-23-RA-HC 11/13/2014 11/13/2017 NS5230212 KCFN-6.0-18-23-RA-HC 11/13/2014 11/13/2017 NS5230212 KCFN-6.0-18-23-RA-HC 11/13/2014 11/13/2017 NS6061690 KCFN-6.0-18-23-RA-HC 8/28/2015 8/28/2018 4427660X KCFN-6.0-18-23-RA-S-HC 8/21/2014 8/21/2017 4427662X KCFN-6.0-18-23-RA-S-HC 8/21/2014 8/21/2017 4427662XX KCFN-6.0-18-23-RA-S-HC 8/21/2014 8/21/2017 4427664X KCFN-6.0-18-23-RA-S-HC 2014/03 3/1/2017 4667267X KCFN-6.0-18-23-RA-S-HC 8/5/2014 8/5/2017 4793616 KCFN-6.0-18-23-RA-S-HC 2014/04 4/1/2017 4893580 KCFN-6.0-18-23-RA-S-HC 2014/04 4/1/2017 4905625 KCFN-6.0-18-23-RA-S-HC 6/1/2014 6/1/2017 6025691 KCFN-6.0-18-23-RA-S-HC 8/4/2015 8/4/2018 6025693 KCFN-6.0-18-23-RA-S-HC 7/31/2015 7/31/2018 6041168 KCFN-6.0-18-23-RA-S-HC 8/18/2015 8/18/2018 4427658X KCFN-6.0-18-7-RA-HC 8/21/2014 8/21/2017 4427658XX KCFN-6.0-18-7-RA-HC 8/21/2014 8/21/2017 4893578 KCFN-6.0-18-7-RA-HC 2014/04 4/1/2017 5102732 KCFN-6.0-18-7-RA-HC 7/25/2014 7/25/2017 5102732X KCFN-6.0-18-7-RA-HC 7/25/2014 7/25/2017 5147478 KCFN-6.0-18-7-RA-HC 9/15/2014 9/15/2017 5585602 KCFN-6.0-18-7-RA-HC 2/10/2015 2/10/2018 5700318 KCFN-6.0-18-7-RA-HC 5/1/2015 5/1/2018 5842464 KCFN-6.0-18-7-RA-HC 6/19/2015 6/19/2018 NS6035989 KCFN-6.0-18-7-RA-HC 8/18/2015 8/18/2018 4427452X KCFN-7.0-18-13-RA-HC 8/21/2014 8/21/2017 4427452XX KCFN-7.0-18-13-RA-HC 8/21/2014 8/21/2017 4667207X KCFN-7.0-18-13-RA-HC 8/5/2014 8/5/2017 4893510 KCFN-7.0-18-13-RA-HC 2014/05 5/1/2017 4893512 KCFN-7.0-18-13-RA-HC 2014/04 4/1/2017 5143024 KCFN-7.0-18-13-RA-HC 8/15/2014 8/15/2017 5529606 KCFN-7.0-18-13-RA-HC 1/27/2015 1/27/2018 5566469 KCFN-7.0-18-13-RA-HC 2/25/2015 2/25/2018 5579985 KCFN-7.0-18-13-RA-HC 2/10/2015 2/10/2018 5634985 KCFN-7.0-18-13-RA-HC 5/1/2015 5/1/2018 5644741 KCFN-7.0-18-13-RA-HC 5/1/2015 5/1/2018 5675202 KCFN-7.0-18-13-RA-HC 5/1/2015 5/1/2018 5855292 KCFN-7.0-18-13-RA-HC 7/14/2015 7/14/2018 6205756 KCFN-7.0-18-13-RA-HC 10/2/2015 10/2/2018 4427782X KCFN-7.0-18-23-RA-HC 8/21/2014 8/21/2017 4427784X KCFN-7.0-18-23-RA-HC 2014/03 3/1/2017 4427786X KCFN-7.0-18-23-RA-HC 8/6/2014 8/6/2017 4556159X KCFN-7.0-18-23-RA-HC 2014/03 3/1/2017 4667354X KCFN-7.0-18-23-RA-HC 8/5/2014 8/5/2017 4814134 KCFN-7.0-18-23-RA-HC 2014/03 3/1/2017 4905684 KCFN-7.0-18-23-RA-HC 5/1/2014 5/1/2017 5126734 KCFN-7.0-18-23-RA-HC 8/1/2014 8/1/2017 5147561 KCFN-7.0-18-23-RA-HC 8/28/2014 8/28/2017 5156356 KCFN-7.0-18-23-RA-HC 8/28/2014 8/28/2017 5306553 KCFN-7.0-18-23-RA-HC 10/14/2014 10/14/2017 5316191 KCFN-7.0-18-23-RA-HC 11/10/2014 11/10/2017 5354346 KCFN-7.0-18-23-RA-HC 11/6/2014 11/6/2017 5493761 KCFN-7.0-18-23-RA-HC 1/13/2015 1/13/2018 5562614 KCFN-7.0-18-23-RA-HC 2/10/2015 2/10/2018 5585643 KCFN-7.0-18-23-RA-HC 3/16/2015 3/16/2018 5641608 KCFN-7.0-18-23-RA-HC 3/16/2015 3/16/2018 5641610 KCFN-7.0-18-23-RA-HC 3/16/2015 3/16/2018 5644799 KCFN-7.0-18-23-RA-HC 5/18/2015 5/18/2018 5682869 KCFN-7.0-18-23-RA-HC 5/18/2015 5/18/2018 5684511 KCFN-7.0-18-23-RA-HC 2/28/2015 2/28/2018 5842502 KCFN-7.0-18-23-RA-HC 7/14/2015 7/14/2018 6012026 KCFN-7.0-18-23-RA-HC 8/4/2015 8/4/2018 6188966 KCFN-7.0-18-23-RA-HC 10/1/2015 10/1/2018 6284401 KCFN-7.0-18-23-RA-HC 11/29/2015 11/29/2018 |
Recalling Firm/ Manufacturer |
Cook Inc. 750 N Daniels Way Bloomington IN 47404-9120
|
For Additional Information Contact |
Cook Medical Customer Relations Departme 812-339-2235
|
Manufacturer Reason for Recall |
Devices using a specific hemostatic valve design, referred to as a blue valve or polyisoprene valve, have an increase in reports of blood loss associated with the devices which could result in delay in procedure and blood loss from patients.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Cook Inc. is initiating a voluntary recall of multiple products following the identification of an increase in reports of blood loss associated with devices using a specific hemostatic valve design referred to as the blue valve or polyisoprene valve. Potential adverse events that may occur if these devices were used in the arterial system (Flexor Radial Access Sets) include delay in procedure and blood loss. If devices were used in the central venous system, adverse events that may occur include delay in procedure, blood loss, or air embolism. The firm is distributing notices via mail courier on 02/01/2017. Customers are being instructed to share the recall notice with appropriate personnel, including down to the user level, within the customer's organization or to any organization where the potentially affected devices have been transferred.
1. Examine inventory immediately to determine if customers have affected product and quarantine affected product(s).
2. Return affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit.
NOTE: Returned un-affected product will not be credited.
3. Even if customers do not have affected product on hand, customers must still complete the Acknowledgement and Receipt Form and return via fax 812.339.7316 or email to Fieldactionsna@cookmedical.com.
4. Report adverse events to COOK Medical Customer Relations at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time), or by email to: customerrelationsna@cookmedical.com.
Adverse events or quality problems experienced with the use of this product may also be reported to the FDA:
Online at: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail)
Call the FDA at: 1-800-FDA-1088
Effectiveness checks of this recall will be done by response form mailed. All recalled products will be destroyed at the regional distribution warehouses. A cer |
Quantity in Commerce |
41,975 |
Distribution |
Worldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and Internationally to Canada, Armenia, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Spain, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Kenya, Kuwait, Lithuania, Mauritius, Netherlands, Norway, Oman, Poland, Reunion, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom
VA/DOD: Yes |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = DYB and Original Applicant = COOK, INC.
|
|
|
|