| Class 2 Device Recall MEDTECH ROSA Brain 3.0 | |
Date Initiated by Firm | February 27, 2017 |
Date Posted | March 27, 2017 |
Recall Status1 |
Terminated 3 on August 06, 2019 |
Recall Number | Z-1754-2017 |
Recall Event ID |
76820 |
510(K)Number | K151359 |
Product Classification |
Neurological stereotaxic Instrument - Product Code HAW
|
Product | MEDTECH ROSA Brain 3.0
The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurological procedure in which the use of stereotactic surgery may be appropriate. |
Code Information |
Serial #'s: BR16004, BR16005, BR16006, BR16009, BR16010, BR16011, BR16012, BR16013, BR16014, BR16015, BR16016, BR16017, BR16018, BR16021, BR16023 and BR16025. |
Recalling Firm/ Manufacturer |
Zimmer Biomet, Inc. 1800 W Center St Warsaw IN 46580-2304
|
For Additional Information Contact | Customer Service 800-874-7711 Ext. 9225 |
Manufacturer Reason for Recall | Unapproved change made by the supplier. |
FDA Determined Cause 2 | Component change control |
Action | Zimmer Biomet sent an Urgent Medical Device Recall Correction letter on February 27, 2017,to all affected customers informing them of the following:
Description of the issue:
A robot arm file parameter linked to arm calibration was modified by the robot arm supplier without notification to MEDTECH.
As a result, once the robot arm is on trajectory, an error leading to a device reboot may decrease the devices applicative accuracy, one of the essential performance requirements, to a level above the specification. This may happen after a shutdown of the device, (voluntary or due to an error), without a return of the robot arm position to a Parking position.
In the worst case scenario, this may cause an incorrect positioning of the instruments by the ROSA devices. If this is not detected by the user this could ultimately lead to ineffective treatment, serious injury or even death of the patient.
MEDTECH did not record, in normal conditions of use, any customer reports linked to this issue.
Required actions by users:
Pending the implementation of a permanent correction, MEDTECH is recommending the following actions to all users of the ROSA device:
1- After a shutdown of the device, (voluntary or due to an error), without a return of the robot arm in Parking position, ROSA device must be switched off with the I/O switch button on the device rear panel.
2- Restart the device after 10 seconds, and after full extinction of emergency button light. If the Reset button at the device rear panel is red, restart the ROSA PC by pushing the Reset button.
3- If robot position is NOT Home , load patient folder and main software interface will appear on the screen.
a. Go to the Guidance menu, select trajectories and follow instructions.
b. Click on Start button.
c. Start recovery procedure and follow all instructions displayed on the screen until robot arm position is Home.
d. Close the patient folder
e. Click on Exit and follow all ins |
Quantity in Commerce | 16 |
Distribution | Nationwide Distribution to CA, FL, IL, KY, MA, MN, NC, NY, OH, PA, and District of Columbia. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = HAW
|
|
|
|