• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall PROT 1 CON M

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall PROT 1 CON Msee related information
Date Initiated by FirmMay 23, 2017
Create DateNovember 28, 2017
Recall Status1 Terminated 3 on March 30, 2018
Recall NumberZ-0175-2018
Recall Event ID 77381
510(K)NumberK083463 
Product Classification Immunochemical, ceruloplasmin - Product Code CHN
ProductDimension Vista Protein 1 Control M, PROT 1 CON M, Siemens Material Number (SMN) 10445920, Product Code KC716; IVD
Code Information Lot numbers: 6BQM01, 6BQM01A, 6BQM01B, 6BQM01C,6BQM01D, 6BQM01E (exp. date 2017-06-17); 6FQM01, 6FQM01A (exp. date 2017-11-03) ---  Lot # (UDI): 6BQM01 (008427680063336JQL01C20180131),  6BQM01A (008427680063406BQM01A20170617),  6BQM01B (008427680063406BQM01B20170617),  6BQM01C (008427680063406BQM01C20170617),  6BQM01D( 008427680063406BQM01D20170617),  6BQM01E (008427680063406BQM01E20170617),  6FQM01(008427680063406FQM0120171103),  6FQM01A (008427680063406FQM01A20171103).
FEI Number 2517506
Recalling Firm/
Manufacturer
Siemens Healthcare Diagnostics, Inc.
500 Gbc Dr Ms 514
PO BOX 6101
Newark DE 19702-2466
For Additional Information Contact
800-441-9250
Manufacturer Reason
for Recall
Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).
FDA Determined
Cause 2
Nonconforming Material/Component
ActionSiemens issued an Urgent Medical Device Correction (UMDC) letter (PP-17-012.A.US, dated May 23, 2017) to US customers on May 23, 2017 and an Urgent Field Safety Notice (UFSN) to customers Outside the US in May 2017. Customers were instructed in the letter to use re-assigned CER values for the lots indicated of PROT1 CAL and PROT1 CON L, M, and H. In the US, the method of distribution is FEDEX and outside the US, the distribution is determined at the country level. Customers were requested to complete Effectiveness Check questionnaire attached to the UMDC/UFSN and return to Siemens within 30 days.
DistributionWorldwide Distribution - US (nationwide) Internationally to Austria, Australia, Belgium, Canada, France, Germany, Great Britain, Italy, Japan, Netherlands, Portugal, Slovakia, Spain, Switzerland and Republic of Korea South
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = CHN
-
-