Date Initiated by Firm | July 05, 2017 |
Create Date | August 16, 2017 |
Recall Status1 |
Terminated 3 on October 05, 2020 |
Recall Number | Z-3041-2017 |
Recall Event ID |
77711 |
510(K)Number | K102936 |
Product Classification |
System, peritoneal, automatic delivery - Product Code FKX
|
Product | HomeChoice Pro Automated PD system Model numbers: 5C8310, 5C8310R
Product Usage:
The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis. |
Code Information |
all codes |
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
For Additional Information Contact | Center for One Baxter 800-422-9837 |
Manufacturer Reason for Recall | Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient. |
FDA Determined Cause 2 | Use error |
Action | The firm notified their consignees by letter on 07/05/2017 of the potential for damaging the disposable set if tools are used to open the packaging and directing them to only open the packaging by hand by letter. |
Quantity in Commerce | 23732 units |
Distribution | Worldwide Distribution |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = FKX
|