  Class 2 Device Recall ORA System with VerifEye see related information
Date Initiated by Firm June 30, 2017
Create Date August 21, 2017
Recall Status1 Terminated 3 on August 27, 2019
Recall Number Z-3050-2017
Recall Event ID 77823
Product Classification intraocular lens surgery system - Product Code NCF
Product ORA System with VerifEye, Catalog Number 8065998300 For use during intraocular lens surgery
Code Information Serial Numbers: C0002,C2006,C2020,C2026,C2054,C3009R,C3010R,C3024,C3036,C3066,C3067,C3085R,C3090,C3093,C3108,C3136,C3153,C4001,C4003,C4005,C4006,C4007,C4009,C4010,C4011,C4012,C4013,C4014,C4015,C4016,C4017,C4019,C4020,C4021,C4022,C4023,C4024,C4025,C4026,C4027,C4028,C4029,C4030,C4031,C4035,C4036,C4037,C4038,C4039,C4040,C4043,C4045,C4046R,C4047R,C4049R,C4050R,C4052R,C4053R,C4054R,C4055R,C4056R,C4057R,C4058R,C4059R,C4060R,C4061R,C4062R,C4065R,C4066R,C4067R,C4068R,C4069R,C4070R,C4071R,C4072R,C4073R,C4074R,C4075R,C4076R,C4077R,C4078R,C4079R,C4080R,C4083R,C4085R,C4086R,C4087RC4088R,C4089R,C4090R,C4091R,C4092R,C4093R,C4094R,C4095R,C4096R,C4097R,C4098R,C4099R,C4100R,C4101R,C4102R,C4103R,C4104R,C4105R,C4106R,C4107R,C4108R,C4109R,C4110R,C4111R,C4112R,C4113R,C4115R,C4117R,C4118R,C4119R,C4120R,C4121R,C4122R,C4123R,C4124R,C4126R,C4127R,C4128R,C4129R,C4130R,C4132R,C4133R,C4134R,C4135R,C4136R,C4138R,C4140,C4142,C4143,C4144,C4145,C4146,C4147,C4148,C4149,C4150,C4151,C4152R,C4153,C4154,C4155,C4156,C4157R,C4158,C4159,C4160R,C4161R,C4162R,C4163R,C4164R,C4165R,C4166R,C4167R,C4168R,C4169R,C4170R,C4173R,C4174R,C4175,C4177R,C4178R,C4179R,C4180R,C4181R,C4182R,C4183R,C4184R,C4185R,C4186R,C4187R,C4188R,C4189R,C4190R,C4191R,C4192R,C4193R,C4195R,C4196R,C4197R,C4198,C4199R,C4200R,C4201,C4202R,C4203R,C4204R,C4205R,C4206R,C4207R,C4208R,C4209R,C4210R,C4211R,C4212R,C4213R,C4214R,C4215R,C4216R,C4217R,C4218R,C4219R,C4220R,C4221R,C4222R,C4223R,C4224R,C4225R,C4226R,C4227R,C4228R,C4229R,C4230R,C4231R,C4232R,C4233R,C4234R,C4235R,C4238R,C4239R,C4240R,C4241R,C4242R,C4243R,C4244R,C4245R,C4246R,C4247R,C4248R,C4249R,C4251R,C4252R,C4253R,C4254R,C4255R,C4256R,C4257R,C4258R,C4259R,C4260R,C4261R,C4262R,C4264R,C4266R,C4267R,C4268R,C4269R,C4271R,C4272R,C4273R,C4274R,C4275R,C4276R,C4278R,C4279R,C4280R,C4283R,C4284R,C4285R,C4286R,C4287R,C4288R,C4289R,C4290R,C4291R,C4292R,C4293R,C4294R,C4295R,C4296R,C4297R,C4298R,C4299R,C4300R,C4301R,C4302R,C4303R,C4304R,C4305R,C4306R,C4307R,C4308R,C4309R,C4310R,C4311R,C4312R,C4315R,C4316R,C4317R,C4318R,C4319R,C4320R,C4321R,C4323R,C4324R,C4326R,C4327R,C4328R,C4331R,C4332R,C4334R,C4337R,C4339R,C4340R,C4341R,C4342R,C4343R,C4344R,C4345R,C4346R,C4347R,C4348R,C4349R,C4352R,C4354R,C4355R,C4356R,C4357R,C4358R,C4359R,C4360R,C4361R,C4363R,C4364R,C4365R,C4366R,C4367R,C4368R,C4371R,C4372R,C4373R,C4374R,C4376R,C4377R,C4378R,C4379R,C4380R,C4381R,C4382R,C4383R,C4384R,C4385R,C4387R,C4388R,C4389R,C4394,C4396,C4397,C4399,C4406R,C4407R,C4408R,C4415,C4418,C4420,C4421,C4422,C4423,C4424,C4425,C4426
Recalling Firm/
Manufacturer
Alcon Research, Ltd.
6201 South Fwy
Fort Worth TX 76134-2099
For Additional Information Contact Wes Warnock
817-615-2501
Manufacturer Reason
for Recall
Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.
FDA Determined
Cause 2
Software design
Action The Market Action was initiated with Initial telephone contact to all affected customers starting June 30, 2017. These customers were informed of the issue, and advised not to use their ORA Carts for calculations with the lens model affected on their specific system. A confirmatory letter (Attachment 4) was also sent via overnight mail to these customers July 12, 2017. As of July 12, 2017, Alcon has identified 8 ORA Carts that are affected by the software coding error. Alcon has reset the IOL databases on all 8 of the identified affected ORA Carts.
Quantity in Commerce 367 units
Distribution Worldwide distribution. US Nationwide and Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, France, India, Japan, Mexico, Netherlands, Panama, Portugal, Romania, Spain, Thailand, UAE, and United Kingdom
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-