Class 3 Device Recall BC Phoenix AP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 3 Device Recall BC Phoenix AP see related information
Date Initiated by Firm June 01, 2017
Create Date December 06, 2017
Recall Status1 Terminated 3 on May 09, 2018
Recall Number Z-0198-2018
Recall Event ID 78102
Product Classification Nephelometer, for clinical use - Product Code JQX
Product BD PhoenixTM AP, Catalog Number 448010
The BD Phoenix" AP instrument is designed for use with the BD Phoenix" system is intended to standardize ID broth tube inoculum, add AST indicator to the AST broth tube and transfer an aliquot of ID broth to AST broth tubes, as required for preparing samples for use on the BD Phoenix" system, which performs identification and susceptibility testing
Code Information Serial Numbers: AP0447, AP0448, AP0449, AP0450, AP0451, AP0458 & AP0463.
Recalling Firm/
Manufacturer
Becton Dickinson & Co.
7 Loveton Cir
Sparks MD 21152-9212
For Additional Information Contact Gail Griffiths
410-316-4258
Manufacturer Reason
for Recall
Potential unexpected movement of robot arm
FDA Determined
Cause 2
Process control
Action BD Diagnostic Systems telephoned all customers on June 1, 2017 and sent an Urgent Product Recall letter dated May 31, 2017, and Response Forms to their customers via email. BD recently confirmed a manufacturing issue with the motor that drives movement of the robot arm. The issue may result in the inability of this motor to withstand the lateral forces encountered during normal operations. Should this occur, the instrument will indicate an E20 error-Pipettor step loss during operation and stop functioning. While there is no loss of patient sample, the samples would need to be manually processed until the instrument can be repaired. To limit any disruption in the workload, BD Service will contact customers to schedule the instrument replacement. Customers are advised to complete and return the response form via email to BDRC2@aol.com. Customers may continue to use the instrument until replacement is performed. Customers with questions may contact BD (Customer/Technical) Support at 1-800-638-8663. For questions regarding this recall call 410-316-4258.
Quantity in Commerce 7 units
Distribution Worldwide Distribution - US including CA, NC & TN Internationally: Canada
Total Product Life Cycle TPLC Device Report