| | Class 3 Device Recall BC Phoenix AP |  |
| Date Initiated by Firm | June 01, 2017 |
| Create Date | December 06, 2017 |
| Recall Status1 |
Terminated 3 on May 09, 2018 |
| Recall Number | Z-0198-2018 |
| Recall Event ID |
78102 |
| Product Classification |
Nephelometer, for clinical use - Product Code JQX
|
| Product | BD PhoenixTM AP, Catalog Number 448010
The BD Phoenix" AP instrument is designed for use with the BD Phoenix" system is intended to standardize ID broth tube inoculum, add AST indicator to the AST broth tube and transfer an aliquot of ID broth to AST broth tubes, as required for preparing samples for use on the BD Phoenix" system, which performs identification and susceptibility testing |
| Code Information |
Serial Numbers: AP0447, AP0448, AP0449, AP0450, AP0451, AP0458 & AP0463. |
| FEI Number |
1119779
|
Recalling Firm/ Manufacturer |
Becton Dickinson & Co. 7 Loveton Cir Sparks MD 21152-9212
|
| For Additional Information Contact | Gail Griffiths 410-316-4258 |
Manufacturer Reason for Recall | Potential unexpected movement of robot arm |
FDA Determined Cause 2 | Process control |
| Action | BD Diagnostic Systems telephoned all customers on June 1, 2017 and sent an Urgent Product Recall letter dated May 31, 2017, and Response Forms to their customers via email. BD recently confirmed a manufacturing issue with the motor that drives movement of the robot arm. The issue may result in the inability of this motor to withstand the lateral forces encountered during normal operations. Should this occur, the instrument will indicate an E20 error-Pipettor step loss during operation and stop functioning. While there is no loss of patient sample, the samples would need to be manually processed until the instrument can be repaired. To limit any disruption in the workload, BD Service will contact customers to schedule the instrument replacement.
Customers are advised to complete and return the response form via email to BDRC2@aol.com. Customers may continue to use the instrument until replacement is performed. Customers with questions may contact BD (Customer/Technical) Support at 1-800-638-8663.
For questions regarding this recall call 410-316-4258. |
| Quantity in Commerce | 7 units |
| Distribution | Worldwide Distribution - US including CA, NC & TN
Internationally: Canada |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|