Date Initiated by Firm |
August 28, 2017 |
Create Date |
June 12, 2018 |
Recall Status1 |
Terminated 3 on July 08, 2020 |
Recall Number |
Z-0037-2018 |
Recall Event ID |
78093 |
PMA Number |
P880086S283 |
Product Classification |
implantable pacemaker Pulse-generator - Product Code DXY
|
Product |
PM 1260 ASSURITY + These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders. |
Code Information |
Serial Numbers: 4471223 4435620 4471226 4478259 4473923 4465800 4465795 2960123 4471236 4465866 2960119 4480805 4456626 4456605 4456614 4456609 4455665 4470374 2992215 4470251 4478226 2992206 4435532 4436165 4442557 2888746 2992230 4470249 2992219 4481080 4477111 4486603 2992337 4478246 3028382 4470221 2992336 4470256 4470255 4477145 4477154 2992244 4479101 4470215 2817358 2897784 2897779 2897791 2878499 2912102 2912129 2912137 2912163 4427305 2992294 2992270 2992271 2870627 2888748 2912140 2912167 2912110 2912100 4419728 4428451 4495360 4428446 2992205 4479087 4485695 4486629 4486616 4486619 4486415 2992291 2992304 2992208 3028378 2992262 2836144 2992265 2992233 2992331 2992298 2992316 2992225 2992322 2992338 2992221 4480798 2992275 4583284 4507998 4505995 4508010 4540176 4495349 4501470 4510860 4508966 4495359 4535417 4507981 4427067 4508009 4552833 4501456 4505305 4505336 4505325 4505323 4506025 4506017 4507994 4505996 4486592 4568194 4510873 4589129 4548069 4513308 4510893 4450734 4495355 4599671 4538400 4507977 4552843 4505329 4506013 4505297 4505998 4545146 4495352 4505307 4495365 4444189 4420503 2880064 2992334 4486626 4486643 4428621 4486590 2866157 4490599 4510886 4444178 4595435 4435633 4481085 4450718 4450724 4450716 4450730 4478258 4478225 4452570 4444187 4535401 4450688 4508013 4508988 4508971 4508983 4510007 4510031 4510878 4508960 4510011 4522001 4505987 4535413 4535405 4536286 4536287 4538398 2912136 4538415 4510012 4536284 4509990 4510987 4441673 4445638 4445615 4445647 4444191 4444206 4444169 4441652 4441667 2958460 4436172 4436157 4436241 4435240 4435220 4435258 4428465 4428473 4428463 4442581 4427298 4429259 4417850 4419718 4419726 4428439 4428461 4428459 4428619 4429264 2992263 4436159 4502697 4502681 4502703 4502700 4502692 4508961 4513296 4428447 2992279 4436184 2992284 2992223 4417858 2992295 4418150 2992227 2992202 2992285 4462727 2992305 2992247 2992255 4513550 2992333 4445644 4513541 4456618 4455658 4462697 4462677 4462679 4445614 4442609 4473268 4445637 4490597 4445630 4445634 4445646 4442574 4442615 2961574 2961576 2961572 4445622 4480800 4513544 4505301 4505312 4521992 4522002 4490593 4476725 4513324 4479012 4477126 4490598 4490606 4490601 4490613 2959508 4476722 4473911 4475375 4478250 4615403 4506009 4552840 4552824 4508976 4634791 4540167 4621323 4634365 4540184 4538399 4508018 4597323 4589124 4615432 4601896 4638503 4634789 4545147 4513300 2929724 4615408 4507991 4634342 4601959 4552806 4589111 4621334 4552844 4601906 4597324 4552829 4540183 4536290 4615395 4601960 4604147 4604150 4634786 4634773 4604143 4558996 4595443 4552830 4615409 4589125 4615427 4548071 4638486 4552841 4604138 4641844 4599669 4552817 4597317 4632934 4615422 4634334 4641838 4452563 4464622 4621342 4634357 4615412 4574102 4604148 4595438 4634360 4604152 4583289 4548050 4450676 4634355 4601963 4548063 4601894 4583305 4582962 4601903 4582967 4601893 4450844 4621321 4464586 4465789 3028385 4455674 4455668 4462681 4462713 4464620 4464609 4450726 4464604 4450720 4464600 4464338 4450709 4465780 4465821 4465804 4465869 4464373 4464606 4464367 4452595 4452597 4634340 4548068 4452602 4452568 4452559 4452617 4479092 4464591 4462683 4462672 4464619 4464336 4464608 4450706 4450692 4450732 4450690 4452590 4457510 4465791 4455669 4455662 4462731 4464594 4464628 4464376 4456630 4456610 4464362 2915028 4464358 4462714 4464335 4464354 4464590 4464610 4464589 4464342 4464343 4465788 4465778 4464369 4465809 4465808 4464384 4465824 4457519 4452609 4456624 4457502 4464598 4464366 3028381 4456604 4484524 4457508 4457524 2958467 2920003 4457505 4604140 4548064 4595444 4582974 4582977 4582983 4599673 4638511 4634769 4621331 4638498 4601965 4589117 4589128 4545162 4552820 4540168 4574101 4574106 4634331 4601899 4548049 4615431 4638475 4636843 4634793 4558991 4599670 4548059 4638506 4595426 4638494 4548057 4589126 4595427 4595428 4595441 4638497 4638485 4621340 4548062 4480811 4574104 4583281 4621343 4615440 4558999 4558998 4545158 4538417 4583293 4601892 4540174 4480815 4477141 4479093 4479024 4479014 4456601 4452589 2956555 4552813 2958462 4604139 4621324 4510852 4548052 4621344 4545163 4545148 4552839 4582979 2874087 4604145 2915049 4615407 4615405 4589115 4621346 4626457 4634368 4634782 2874100 4597325 4548065 4634752 4638482 4548047 4548051 4548048 4548053 4597322 4601957 4582968 4548067 4583287 4565686 4583288 4615398 4583299 4540173 4540198 4595442 4634754 4634776 4597331 4595431 4548055 4601898 4513331 4552827 4634338 4552825 4604144 4634777 4548066 4634367 4601895 4615425 4552846 4595437 4601900 4597329 4597330 4615415 4582989 4548058 4615420 4601904 4604137 4615417 4559443 4604151 4604146 4638501 4574103 4597318 4601897 4583298 4552821 4599676 4589118 4634364 4634358 4552815 4595430 4552834 4545153 4615430 4552808 4552838 4540190 4536301 4538406 4641847 4536295 4583316 4634351 4615441 4615433 4552831 4548070 4589121 4589119 4582981 4582965 4540196 2952259 4486608 4479086 4484246 3028377 4484513 4487206 4486642 4487214 4487215 4480820 4486599 4483699 4486413 4485704 4486620 4486591 4486411 4485707 4485706 4486596 4483976 4486604 2872076 2912171 2912124 2912158 2912104 2912141 2912105 2912148 2912112 4483979 2872103 4485697 2872090 2846347 2846361 4480825 3025482 4480834 4483957 3025483 4484511 2874092 4483977 4483962 4484512 4484533 4484528 4484523 4483980 4484514 4483954 4485694 4483964 4484535 4483978 4483975 4483692 4483956 4483974 4483970 4477140 4478233 4483700 4486600 2915732 4485702 4486611 4486598 4486410 4486625 4486615 4486621 4486635 4483967 4485698 4484526 4486637 4484534 4484521 4484525 4484517 4484245 4483968 4483969 4485693 4485705 2920032 2927668 2923624 2919965 2919994 2919991 2919975 2920018 2920024 2915700 2919959 2927656 2920034 2920023 2919970 2920028 2920025 2918067 2918056 2918040 2912099 2920033 2929718 2956552 2952261 2952263 2952255 4456602 2929712 2929720 2929717 2929715 2927669 2929713 2927660 2929721 2929723 2929726 2929725 2927666 2927663 2927662 2927661 2915698 2929714 2920030 2874090 2874089 2874088 2870619 2832948 4480832 2919955 2923627 2918039 2923628 2912156 2919978 2920007 2920011 2920015 2919972 2918065 2919958 2915710 4479036 2923625 2915013 2915742 2915695 2854998 2915040 2915029 2915018 2915017 2915015 2915725 2912165 2915026 2912115 2915011 2915730 2915039 2915044 2915718 2912117 2912133 2912135 2915032 2915706 4464363 4481056 4465790 4464584 4465792 4481076 4465799 4465816 4464359 4477153 4464371 3028383 4464368 4464370 4462701 4464372 4464345 4464337 4462700 4462691 4464380 4464364 4479032 4479022 4478997 4479018 4479043 4479052 4479047 4479031 4480818 4465863 3026689 4470211 4479105 4479056 4471215 4470205 4486605 4479002 4479040 3026691 4476731 3026690 4456622 4456625 4455666 4455655 4456600 4455682 4455678 4455677 4455671 4452598 4456628 4456615 4455660 4455684 4462666 4462725 4462716 4452608 4452569 2872102 4450678 4456617 2959504 4480826 4483693 4455672 4457509 4456631 4455667 4457497 4457491 4456620 4455664 4462726 4455680 4480831 4457517 4456629 4455676 4456632 4455656 4456606 4455657 3028379 4462669 4478239 4475378 4477125 4479025 4479050 4478243 4478248 4476730 4478224 4471261 4478244 4479028 4477122 4477108 4477138 4477144 4477143 4477132 4477151 4476743 4479100 4478222 4480833 4478232 4479051 4480819 4480790 4480794 4481083 4481067 4480816 4480830 4479006 4480796 4479099 4481062 4480813 4479088 4479023 4479027 4479096 4479055 4479053 4471248 4480806 4464605 3025479 2958463 2958468 3026695 2956551 4456607 4464630 4464374 4476742 4464621 4471225 4464615 4456603 4484243 4484520 4484537 4484518 4483698 4483955 2952258 4464378 4470373 4462662 4471199 4471245 4471211 4471242 4470226 4471235 4471230 4471209 2952260 4470203 4471216 4470375 4471214 4471219 4471231 3025480 4486630 4486595 4477112 4479011 4470200 2956559 4452565 4479049 3025481 4479008 4465861 2958469 2959506 2959509 2959497 2959498 2959496 2959495 2959505 2959503 4470257 2956550 2958465 2958464 2958466 2958461 2956553 2956556 2956558 2952257 2952254 2959500 2960120 4479098 4470243 4484509 4486607 4486602 4484248 4484250 4483953 2960117 2961573 2960118 2960109 2960111 2960115 2960105 2960114 2960112 2960106 2960110 4471210 2960113 2952256 2878500 2878496 2878510 4428475 2853271 2853272 2893460 2894444 2894436 2894461 2894458 4428452 2897780 4428441 2897762 4428464 4428438 4428627 4428460 4427053 4428440 4428458 2897792 4435229 2829689 2829481 2820568 2822826 2822828 2820564 4436244 4435260 2846366 4435631 4419700 4435239 4435237 4435207 4428623 4429300 4429287 4429260 4428442 4435538 2912160 2897764 2900976 2902522 2902539 2902520 2902534 2902548 2902544 4427292 4419711 2897778 2912121 2912170 2878471 2903440 2903435 2903439 2903442 2903444 4420025 2897760 2903443 4420041 4420023 4420026 4420526 4420519 2897763 2897759 2897793 2897787 2897767 4419682 4419686 2894439 4419710 2829684 2897772 4436176 2897773 4427070 2897774 2870623 4478236 4488969 2894467 2894460 2893456 2893430 2893425 2893426 2894459 2894457 4450710 2870622 4450713 2893459 4488974 4488984 4488982 4488968 4488985 4488967 4488975 2862244 2865768 4429266 4488965 2894466 2893432 2893433 2893438 2893458 4486414 2897783 4490595 2893446 2893440 2893445 2894473 2894468 2870615 2893451 2893462 2894464 2894465 2894454 4450694 4450685 4450680 4450675 4450851 4450698 2893447 4441633 4488972 4444196 4441655 4441657 4442571 4442569 4442556 4442594 4442602 4442560 4444159 4441676 4444164 4441644 4435621 4436147 2992303 4435533 2846363 4435624 4435528 4436153 4436156 4436183 4441677 2897785 2872075 2872078 2888735 2872080 2872086 2870616 4450852 4452558 4452578 4473916 4444198 4473896 4435628 2870613 2888753 2894435 2894452 2894437 4446206 4442613 4445660 4445656 4442575 4444186 4473881 2992231 4489636 4489642 4490589 2897781 4470212 2897771 2897766 2888731 2888724 2888733 2845619 4484529 4473266 4473917 4473921 4473256 4470208 4473259 4473260 4473261 4473284 4471212 4473241 4473257 4473901 4435238 4435234 4435636 4435622 4435626 4435215 4435216 4435213 4435223 4435554 4470247 2894462 4473235 4487204 4487203 4487207 2894445 2886257 4473887 4473884 4473240 4473269 4473889 4470365 4479097 2894453 2865764 2862258 4479007 4473892 4464592 4465862 4464382 2897775 4465776 2888728 4470372 4470228 4470245 4464340 4465876 4465820 4465815 4465823 4470361 4470364 2897788 2888726 2992280 2829688 2829687 2829685 2829482 2822827 2817365 2817352 2817371 4455661 2992228 2894447 4486628 4479046 4436186 2992326 2992276 2992257 2992256 2992307 2992259 2992216 2992218 4470207 2992232 2897777 4478256 4493139 4490619 4464601 4464593 4493136 4493138 4493152 4490587 2817674 2894455 2829483 2894456 2817355 2870636 2880080 2870634 2870618 2870629 4464597 2894446 2894463 2845604 2817369 4462661 4479004 4479003 4479021 4479020 4479034 4479030 4479026 4464351 2872098 2872101 2894474 4462705 2870621 4462675 4479019 2866152 2870624 4489632 4489631 4489644 4489649 4489635 4489640 2872091 2845629 2880143 4510990 4505980 4583319 4538408 4508970 4509993 4509998 4508969 4508982 4486638 4510868 4508990 4510866 2846358 2845627 2845603 2880135 2865782 2862262 2858076 2862264 2858093 2858084 2831549 2865792 4508989 4507996 2862267 2862247 2862248 4510859 2865763 4510017 2865781 4508963 4508987 4508980 4508978 4510870 4502687 2873282 4508015 4495369 4507974 4508007 4508974 4508967 4510024 4507973 4508022 4508017 4508016 2845611 4507993 2872104 2832936 2880084 2880068 2894441 2894449 2894451 2897789 2897768 2897782 2886261 2886259 2880148 2866155 2880071 2872106 2872083 2872077 4510015 4510030 4508968 4508991 4510027 4510001 4510862 4510876 4510342 2886269 4436175 2880069 2880134 2880144 2817709 2865788 2866160 2865783 2880079 2865785 2870641 2880081 2880087 2846389 2865784 2880138 2880086 2880076 2846377 4489637 2845609 4489648 4489647 4510980 4510992 4510877 4508965 4489650 4486639 4477101 4479013 4487202 2858082 2880063 2817359 2880065 2846362 2811909 2866159 2865778 2865773 2811912 2858079 2855027 2855028 2893424 2893427 2893428 2894470 2893453 2888752 2845617 2845606 2862259 2820561 2811925 4488989 2880073 2853269 2853263 2853262 4484247 4485691 4488973 4485700 4488971 4488966 4488963 4488962 2880152 2888751 4479091 4488976 4488983 4490592 4490591 4479089 4479094 4486644 4479095 4479039 4479029 2858118 4480792 4485696 4666232 4419703 2878506 2893431 2846335 2893461 2862260 2865777 2862245 2865774 2865772 2865767 4419709 4666250 4419699 4666231 4666236 4666241 4666251 4666261 4666233 4666243 4589123 4568196 4568202 4568203 4552845 4666258 2858115 2880151 4476733 4480810 2878502 2878501 4480786 2878514 2858121 2893444 4480812 4419684 4486624 4486614 2903437 2811910 2811916 2858114 4419706 2888747 4486416 4510881 4510993 4509992 2903436 4486640 4486601 4486622 2825188 2820562 2992248 2992327 2992299 2992236 4540200 4489645 2827250 2845625 2845623 2836145 2836155 2836148 2992302 4511624 4552849 4505333 4510000 4508984 4507997 4540187 4538412 2825194 4509995 2825196 2845616 4511619 2845614 2836161 2836162 2846350 4508001 4505320 4540197 4513323 4510871 4510985 3116267 2846375 4442598 4486594 2912169 2912108 2912123 2912149 2912153 2912145 2912114 2912130 2912125 2912122 4429286 4419692 4452572 4457498 2902537 2902540 2927670 2831548 2888737 4428466 2858095 4420048 3028376 4452555 2832939 2992210 4479016 2992281 4510034 4508011 2915038 2915739 4495372 4495371 2832949 4507990 4457493 4510889 4450681 4420508 4420515 4455673 2992320 4506002 4435245 4489641 2897776 2897770 2920002 4420501 4427073 2858128 4429257 2927657 4435545 4435634 4435251 4435244 4435556 4452564 4426784 2874102 2992214 2992289 4450855 2858101 2862266 2865766 2888723 2874097 2992286 2888727 2888729 2884440 2897786 4452567 2918041 2874111 4429274 4427270 4427060 4427065 4427066 4427048 4429290 2878527 4429292 4429277 2858111 2866153 2870631 4428456 4429269 4427295 4427061 2831562 2831551 2831552 4420513 4505335 2919966 2920016 2920008 2992268 2992212 4470235 4470230 4495363 4495357 4495368 4505331 4502708 4508981 4508014 4510857 4475368 4419712 4476727 2846355 2870642 2846365 2845621 4471213 2831554 2831565 4495351 2831543 4417856 2915723 4418141 2878518 2880149 2915033 2915042 2915719 2915738 2915016 4484527 2919983 4486632 4418151 4418140 2927664 4428445 4435264 4435228 4442552
4476721 4442583 2915705 4418154 4471217 4495386 4508002 4508975 4510028 2915037 4418157 4477146 4436180 4445642 4476734 4475379 4477104 4477105 2960107 4479090 4485703 2919967 4475380 2920022 2918057 2918049 2918042 2918064 2920017 2920009 2919997 2920000 2918060 4417854 4417857 4429294 2920014 2915729 2902524 2915699 4427074 4427050 4429293 4429298 4418145 2915743 2915741 2915727 2915733 2915744 2915722 2915702 4429299 2902528 4429283 4428477 4428453 4428448 4428618 4428471 4420033 4420509 4428462 4420510 4429265 4419695 2900977 4420497 4420523 4420518 4420032 4420044 4420494 4420504 4418146 4418156 4418152 4417849 4429279 4429263 4428626 4428622 2880150 4428468 4428620 4418142 4429255 4429270 4429272 4428444 4428443 4428457 4418158 2915740 442965 2888745 2880072 2880083 2880074 2880075 4452581 4452623 4473237 4452576 4493154 2865786 2866161 2865776 2865780 4452588 4452603 4452616 4493147 2900969 4473245 4473253 4473251 4489643 4493157 4493145 4484531 4493153 4493150 4493155 4489639 4493151 4493137 4493142 4452580 4493146 4462729 4452600 4452560 4462676 4462684 4471237 4473243 4450729 4462728 4452545 4462696 4462688 4462724 4462732 4462692 2878516 2878524 4473281 4478242 4473262 4452548 4452624 4452619 4452582 2878519 4452575 2878493 4452614 4477109 4478247 4455683 4462709 4445613 4452592 4452566 2878525 4473890 4484510 4477142 4476748 4473277 4473914 4473899 4473244 4473895 2854981 4473915 4473250 4473255 4473282 4462710 4471240 4473906 2894438 2902535 2902521 2900974 2900979 4429253 2900978 2846346 2897769 2854972 2894448 2893449 2893437 2893435 2880145 2855014 4470204 4420027 4475371 4471208 2880070 2874112 2854983 2878508 2880136 2880078 2874098 2880085 2888756 4476735 4477133 4477127 4476729 4473270 4473279 2878512 4470202 2902543 4471197 4471196 4471238 4471254 4471243 4428455 4471203 4470258 4470222 4470377 4470359 4470219 4471249 2886273 2886271 4470363 4505989 4506007 4506010 4506000 4505994 4506019 4506008 4506012 4505981 4506020 2870628 4505985 4502704 4505306 4502705 4505988 4505992 4505339 4538413 2854979 2854985 2854991 4501455 4506022 4502679 4501463 4501464 4501498 4501485 4502671 4502717 4501459 4502672 4502680 4502701 4506003 4502684 4506024 4502706 4502676 4502699 4502715 4502696 4502711 4502685 4501499 4501462 4505343 4501477 4428472 2846373 2846360 2846336 2846371 2846383 2845615 2845601 2846344 2886260 4615396 2886266 2880139 2886277 4417851 2886270 2886265 4535411 4615413 4615410 4507985 4508000 4506011 2886267 2874099 2872092 4535414 4538397 4615435 4615414 2854968 2878498 2878509 2878521 2874108 2846388 2874101 2846342 2878491 2880146 4538395 4538407 4536305 2880147 2880142 2880137 2880140 4502710 2874103 4471207 2832950 4435529 4462715 4429302 4502719 4501450 4470227 4470234 4435625 2888739 4436168 2888725 2888732 2888722 2858122 4435630 4450719 2858125 2832955 2832954 2884441 4436163 4435231 4435218 4435224 4435225 4436195 4436181 4436185 4436173 4435221 4436162 2832947 4435261 4436155 4435551 4435253 4436167 4436191 4436179 4435263 4435549 4436192 4501452 4475385 2858123 2836141 2880077 2854994 2886274 2886275 2886278 4540178 4540181 2832946 4501493 4476720 2855032 2855020 2855021 2855010 2855015 2854989 2854990 2855009 2855031 2886272 2886262 4540185 4476747 2854975 4462689 4476736 4477106 4477117 4477136 4477150 4477114 4477113 4477129 4477135 4476738 4476737 4476740 4476723 4470236 2846367 2855013 2855022 2854984 2854996 2846359 2854973 2855030 2854967 2858078 2846357 4478260 2858100 2846348 2846380 2817370 2992249 2992266 2992209 2992238 4490614 4490603 2858105 4471193 4420029 4477124 4420527 4490610 4480797 4471204 4480808 4444182 4452551 4444208 4441654 4442589 4436233 2858091 4419702 2846387 2817708 2858108 2992297 2858099 2858112 2854970 2855025 2854992 2855016 2855019 2845608 4436148 4513309 4545154 4558997 4615436 4552826 2846354 2817710 4450845 2817707 4535400 4521995 4513332 2854993 4632928 4510869 4513549 4634337 2992222 4583297 2836152 4506018 4656218 4502695 4502707 2846369 2854965 4452574 4508979 2836142 2836147 4471257 4470362 4442582 4457522 4490586 4436240 2846370 2846381 2845618 2846337 2846351 4475384 2855029 2855017 2858120 2854971 2846376 4490585 2811924 4444199 2845626 4490608 4450704 4490584 3096799 3096810 3101791 3071063 3071054 3071051 3071050 3096796 3071041 2866158 4508973 2858085 2858124 2858102 4436235 3071053 3086947 2836150 3096804 2865790 3099854 3100111 3099153 3097816 3071042 3097830 3101810 3099150 3097812 3097808 3097826 3071040 3071064 3071046 3071044 3071052 4501473 3102330 2811927 3071056 3101806 3101803 3071045 3071059 3071058 3071065 3071061 3071048 3071047 3071057 3071062 3071066 2811906 4511625 4535416 2822831 2822824 2853268 2915024 2853270 2854980 4419705 2853266 2811923 4538405 4508008 4538414 4510022 2811917 2811902 2811913 2811914 2811900 2811911 2873273 2872073 2854982 2854963 2854997 2858103 4538411 3101789 2846349 2872089 2846382 2845600 2873269 2873276 4501484 4495377 4508024 4508021 4495350 4508006 4510882 4507976 4510851 4507979 4513330 4507986 4513548 4513543 4513320 4509999 4505326 4505302 4505341 4510864 2870635 4507971 3096809 3079373 3079372 3079734 2956554 3148604 3097827 2845605 3099158 2817374 3097831 3097834 3100108 3099849 3099852 3099851 3099170 3071060 3096802 3099146 3097814 3099169 3099151 3099141 3097832 3100092 3099163 2915708 2912138 2912144 2912120 2912164 2912166 2912157 2912106 2836160 2912177 2912152 2912139 2897790 2915714 2919974 2918054 2846379 2920005 2919989 2918063 2918053 2915724 2918051 2915704 2915005 2915006 2915014 2915003 2915703 2912174 2918052 4475388 2912168 2811919 2992224 2992235 2992203 2858117 4475372 2874093 4475366 4465803 4465868 2865789 4475389 4473880 2992273 2874104 2912173 2888743 2878494 2878513 2878529 2811922 2878522 2912175 2825186 2865770 2873277 2873272 2873275 2873274 2878526 4442549 4436229 4436151 4436238 2811908 4420045 4427307 4435226 4435537 4435635 4435547 4436182 4435235 4435623 2919960 4445643 3153673 3153672 3153668 3071049 4442573 4442586 4445629 4442563 4444160 4444188 4444170 4444218 4444201 4435546 4446193 4420506 4436236 4420021 4420024 4419721 4419722 4418148 4420496 4420525 4420495 4419720 4420489 4418139 4419690 4418153 4420490 4427310 4465782 4435262 4435617 4429261 4427063 4420499 4427291 4436177 4427311 4427075 4427071 4427051 4420521 4420035 4427294 2858097 4456634 2836149 2822825 2854987 4475397 4475373 4476741 2870640 2872079 2855035 4490594 4490590 2873281 4479005 4481082 4478253 2872097 4478237 2870620 2870625 4450703 4480803 4478227 4480788 4481077 4450856 2858110 2858092 2855008 4450731 3071043 4488988 4476732 2870639 2862253 2836164 2854995 2862256 2853260 2870630 2992243 2845622 2870617 2858086 2853273 2855023 4486618 4450728 4486407 4486631 4486627 4486412 2811905 2855018 2855037 4486641 4486633 4486606 4485699 2858096 2853259 2855024 4457503 2817706 2870632 2865779 2858087 2862249 2870637 4462733 2854964 2855006 4493149 4456616 4456611 2858075 4450841 4462671 2872074 2817354 4471198 4471200 4471194 4471253 2870638 2858077 4462670 2872105 2872096 4470369 4470371 4470242 4470366 4470201 2870626 2820569 4452605
2873279 4484519 2866154 4452587 2872082 4452571 2858116 4450854 2855011 4450853 4450849 4465875 4479017 2874110 4477137 4479041 4478261 4473236 4473278 4452607 4441637 4450842 4452557 4452615 2874091 2873280 4452606 2832941 2872072 3071055 2992311 3026694 3026692 3026693 3026698 3026697 4897827 4897791 4897792 4897793 4897817 4897828 4897830 4897836 4897849 4897842 4897856 4897786 4897782 4897840 4897848 4897778 4897816 4897795 4897814 4897818 4897780 4897789 4897807 4897802 4897825 4897851 4897794 4897784 4897812 4902280 4897838 4897783 4897788 4897846 4897841 4897801 4897799 4902278 4902287 4902293 4902286 4902283 4897806 4902271 4897833 4902268 4902269 4902267 4902264 4902261 4902277 4897796 4897790 4897779 4897798 4897821 4897823 4902262 4897847 4897819 4897839 4897854 4897850 4897826 4897781 4897815 4897811 4897853 4897832 4897845 4897803 3157023 3139865 3157043 4897835 4897824 4897855 4897797 4897822 4897852 4897808 4897787 4897805 4897820 4897831 4897829 4897844 4490617 2817351 4490618 4479033 4478999 4479000 2858126 4457499 4455675 4455663 4490620 2858083 4490612 2811915 4481081 4481078 4481079 2858098 2836156 2855003 2836153 2854969 2817349 4480824 4457520 4456613 2846378 4490588 2854988 2855007 2874106 4427296 2858074 2858104 2832945 4452612 4481073 2832953 2858127 2836146 2817362 2836143 2853267 4452621 4452549 2855001 4452622 2866156 4452573 4452610 4452594 2836158 2874113 2811932 2836163 2836151 4479035 4487212 4452593 2811918 4495364 4505986 4510872 4513555 4495374 4502702 4502689 4502678 4502683 2811903 4621339 4457507 2811898 2832935 2832956 2846345 2811907 2811931 2817360 2832944 2858113 2845624 4510879 4479042 2862271 2817361 2854978 4441630 2811901 2811921 2811926 2836159 4535415 4495383 4511623 4495367 4509991 4501486 4508962 4510002 4508959 4502670 4495387 4495353 4495385 2836165 4513539 4457494 2874094 2920010 4436187 2811920 2811929 2811904 2811899 2832933 2853264 2855004 4427046 2865775 4427293 2888749 2893439 2912161 2912147 2912176 2912101 2915707 2915736 2915720 2915012 2862261 4420031 2915043 2915709 2918047 2919998 2920013 2920020 2919986 2919954 2919973 4436231 4420028 2922005 4420502 4420498 4420512 4420034 4420038 4419704 4420505 4420511 4417852 4420528 2918062 4446210 2919963 4442591 4444172 4444165 4441624 4444173 4444217 4444184 4444162 4441674 4446202 4441669 4445619 4446208 4446194 4446196 4445649 4445657 4445618 4471192 2858081 4457514 4446199 4436164 2922006 4419701 4420522 4420520 4420507 4420039 4426781 4427263 4427285 4441642 4435530 4436150 4436243 4436228 4442588 4441650 4441635 4441623 4441643 4441646 4441658 4441653 4427049 4489633 4470220 4471201 4473888 4473891 4473912 2854977 2853265 4471227 4479044 2858089 4481084 2992246 2992315 2992272 2992201 2992306 2992308 2992229 4420043 2831558 2855000 4480821 2854976 4464361 4436161 2915027 4464365 4464347 4464352 4464353 4456608 4464350 4473886 4479001 2858094 2854966 2853261 2858107 2811930 2836157 4470368 4470254 4471218 4471247 4471250 4479015 2902530 2893457 2888736 2900971 2900972 2900982 2900970 2902526 2902527 2902529 2893436 2902542 2902523 2902538 2878490 2878483 2878484 2912142 2858109 2992204 2912150 2912162 2915047 2902536 2865769 2862269 2862273 2862268 2902545 2893452 2862272 2862250 2862252 2862255 2862265 2865765 2874105 2888742 2862246 2888744 2865771 2888738 2858106 2873271 2888754 2888755 2888741 4565687 4536306 4510892 4429280 4535418 4535409 4535410 4535407 4535412 4538403 4535402 4535419 4536298 4538396 4538418 4538404 4538401 4540165 4641824 4536289 4565688 4558992 4601958 4601961 4656229 4522009 4597327 4513542 4565695 4548060 4565699 4552814 4656237 4583295 4583282 4583292 4595434 4513305 4589114 4634763 4565698 4473876 4540180 4540179 4540169 4540162 4540182 4540192 4540189 4538409 4545159 4548056 4540175 4442618 4465786 4583308 4583311 4656243 4552842 4552809 4552811 4545156 4510885 4510021 4508964 4510339 4510349 4510005 4508993 4510016 4510350 4510359 4510038 4510357 4510037 4510029 4510986 4510853 4510875 4510865 4508012 4510982 4513317 4510340 4510351 4505979 4506001 4505993 4505983 4505999 4506027 4507980 4510020 4510352 4510354 4510345 4510362 4510035 4510344 4510013 4509994 4510004 4510019 4510355 4505309 4510356 4513315 4511618 4507992 4595445 4595432 4536299 4513540 4513556 4505311 4513321 4513554 4508003 4510883 4522005 4521998 4522010 4522008 4522013 4522011 4522014 4522853 4522006 4510891 4513546 4513328 4513312 4513313 4513301 4513302 4513306 4513329 4508004 4510032 4522850 4513303 4513314 4583312 4513318 4513333 4513307 4513552 4513297 4506016 4513545 4450723 4450725 4450717 4450687 2817353 2855036 4445607 2817357 4450707 4488961 4450682 4450684 4450693 4441675 4452561 2827251 4462703 4450733 4488978 4478255 4478223 4478230 4450691 4450859 4450843 4450695 4427286 4450683 2820563 4488970 4488964 4488990 4488981 4488977 4488980 4464625 4450715 4419689 4464595 4487205 2878504 2817368 2817373 2817356 2820570 4452554 4462665 4452585 4465781 4462730 4471241 4473905 4473883 4506014 4452604 2817367 4452562 2893455 4464583 4464585 2831561 2831563 2832943 2832937 4487213 2832913 4464612 4452586 4452611 4452613 4452579 4452550 4452547 4452546 2832938 4464346 4435555 4559444 4632935 4559440 4510867 4565693 2912159 4656222 4418149 4540201 4464618 4462686 2915731 2831544 2831547 4476739 4478257 2919977 4510026 4508025 4522851 4506021 4522852 4540199 4634758 4510358 4559442 4540171 2918055 4536292 4513304 4508992 4507978 4502716 4621326 4597335 4495382 4464603 4442608 4450689 4450696 4450840 4450702 4450848 4450714 4465818 4450686 4450712 4464349 4462704 4465798 4465779 2820566 2822830 4589113 4450858 4457516 2918066 4476724 4473907 2817372 4457513 4457492 4441648 4450722 4457512 2902546 2820567 4441647 4465873 4465864 4465807 4450679 4450699 4445612 4477116 2874107 2874109 4456633 4452591 2873270 2873278 2870614 2886276 2858119 2858080 2878507 2880153 2872085 2831541 3028384 2874096 2886279 2874085 2886263 4427052 4429288 4429297 4429295 2884442 2884444 4471260 4502674 2846384 4510363 4505340 4505294 4574105 4502667 4502677 4502686 4502673 4589116 4502675 4510353 4470199 4477110 2846339 4477147 4470224 4502668 4470237 4470370 2918068 4471251 4471220 4470229 2825185 4604141 4470376 2832934 4589120 4568198 4595439 4513319 4565692 4521990 4559441 4471221 4473878 4505324 4427279 4589112 4632933 4521999 4545145 4510003 4501492 4522000 4471224 4568193 4501495 4495384 4495354 4495356 4495378 4495375 4495381 4495361 4522007 4495362 4615393 4510360 4589130 4510036 4597319 4565689 4552807 4513326 4510887 4574107 4508020 4568209 4535403 4507975 4552823 4565694 4495376 4545161 4505984 4495379 4505321 4505338 4505342 4505313 4505314 4505319 4505303 4501444 4502712 4506015 4506004 4506005 4505982 4505997 4505990 4506028 4505298 4501467 4568199 4495358 4501468 4501479 4501441 4501460 4501465 4505300 4501480 4495380 4501472 4501490 4502666 4502669 4501475 4502688 4502709 4501442 4505316 4540172 4565685 4509997 4583313 4510854 4495370 4510033 4574108 4505318 4502720 4505330 4505295 4505296 4495366 4545155 4568208 4595433 4505308 4501453 4506026 4442565 4501458 4495348 4501466 4501483 4501449 4568201 4501454 4568207 4501447 4502693 4502690 4501476 4502694 4501471 4508972 4501478 4582971 4513325 4513298 4545149 4552837 4582987 4582975 4582963 4540177 4540195 4626455 4583306 4583307 4583294 4583283 4583318 4582973 4470206 4582970 4597334 4501489 4540194 4505317 4505322 4505315 4501482 4597326 4513299 4597333 4510849 4506006 4568200 4583315 4513553 4634761 4552836 4634750 4597332 4436178 4428625 4435541 4435542 4436149 4436169 4436158 4435209 4436194 4435618 2825191 2825195 2825192 2825183 2825184 2827253 4436232 4435250 2915728 4427265 4429281 4429275 4429291 4429284 4435212 4435247 4445609 4435548 4435266 4435254 4435259 4435256 4435632 4435629 2923620 2915697 2915711 2918046 2918044 2919999 2920001 2832951 2920019 2915712 2922003 2922004 2922002 2923621 2923619 2923626 2920012 2893442 4429267 2878530 2878528 2878505 2880066 2888734 2915696 2893441 2915010 2902525 2912127 2912128 2912111 2915035 2915046 2845620 2893463 2919981 4429276 2920027 2919990 2919980 2920036 2919982 2920035 2919961 2920029 2919985 2920021 2920031 2919996 4473264 4420030 2919957 2919962 2915025 2915008 2915713 2915734 2915701 2915721 2920037 2919953 4419679 2919964 2919956 2919968 2920004 2919969 2919979 2918059 4427300 4426783 4427306 4427047 4427278 4427266 4427271 4420046 4427077 4419723 4427281 4427062 4428476 4428454 4427056 4429273 4427272 4419696 4420019 4419687 4419708 4419717 4419725 4419698 4419685 4419714 4420036 4419681 4419680 4418143 4419694 4419697 4419715 4420040 4419713 4444180 2923623 4444209 4444210 4444192 4444193 4444167 4441651 4442604 4442584 4444211 4444202 4444205 4444216 4444166 4444181 4444183 4442596 4441622 4441664 4442603 4441634 4441641 4442553 4444214 4441626 4442572 4441671 4442597 4441668 4441640 4442617 4441659 4435627 4470367 4445627 4445640 4445633 4445662 4445648 2992283 4442611 4478240 4442568 4471228 4471232 4481068 2992264 4505332 4470210 4465825 4445628 4442600 4442570 4442614 4445620 4445610 4445621 4445617 4445639 4445631 4445624 4445645 4445650 4445608 4445659 4445663 4442593 4445636 4427273 4417855 4426782 4427058 4427059 4427284 4426779 4428469 4427267 4427301 4427302 4427274 4427269 4427280 4427277 4441656 4427282 2846352 4420022 4420020 4419727 4419707 4419693 4419683 4427308 2855012 4427309 2915716 2886264 4427275 4427287 4427069 4427057 4428624 4420047 4436196 4435208 4435257 4452577 4436190 4436154 4436239 4428467 4436171 4435619 4435552 3079369 4441649 4441662 4441678 2927659 4436237 4435540 4427289 4428474 4429289 4429268 4429282 4435550 4435217 4435536 4435222 4435531 4435249 4435265 4435255 4435232 4435534 4441670 4435543 4473273 4471233 4473875 4473898 4473238 4475391 4475363 4473893 4470213 4465819 4475361 4464616 4464599 4462674 4464602 4464614 4475398 2878517 4490600 4471256 4471244 4471259 4471234 4471258 4473267 4473904 4464357 2992261 4465806 4465805 2992335 4477134 4465865 4475367 2992237 3079371 4471239 4450708 2825189 4476746 4477102 4464631 4465794 4475390 4477103 4478251 4465814 4465813 4465810 4465822 2831559 3028380 4470225 4473280 4465874 4475376 4470232 4475396 4435219 4470253 4462719 2915000 2992325 4473913 4473263 4477123 4473900 4464627 4473894 4452601 4471252 4470209 4473254 4462718 4473275 4473239 4473258 3026696 4473265 4473908 4462685 4462702 4457504 2992199 2992300 4470252 4473909 4473248 4473902 2915023 4473249 4470246 4473922 4473283 4470231 4457506 4452618 2831550 4464632 4464613 4464375 4464611 4473924 4442592 2846356 2846368 4456623 4455681 4457501 4457518 4476728 4457500 4435230 4478241 4471229 4464587 2845613 2846353 2831556 4457496 4462698 4475370 4457490 4457515 4457521 4452584 4457523 2845628 4464383 4475399 2817364 2831560 2831564 2846338 4456612 4429258 4457511 4457495 2878511 2855005 2855002 2854974 2855033 2878492 2878503 2894442 2880067 2836154 2888730 2893450 2893429 2893448 2831545 2817703 2880141 2846374 2822832 2825197 2827252 2827254 2817363 2831555 2846364 2845607 2846385 2846386 2845612 2846343 2846340 2845610 2845602 2894443 2832942 2915715 2912143 2912132 2915002 2915030 2915031 2915041 2894440 2915036 2912119 2915717 4475364 4456621 2915737 2915022 2915735 2915045 2900973 2902531 2902532 2902547 2902533 2900983 2900975 2912134 2900981 2912155 2912109 2912116 2912118 2912146 2912154 2912126 2893434 2900980 4475383 4452596 4473920 4470248 4477152 4473274 4475394 4470218 4477131 4471255 4475369 4477118 4477121 4477107 4477148 4478228 4478231 4476749 4473242 4490581 4478249 4477120 4475381 2893443 4442587 4470233 4473918 4473879 4473247 4473910 4473903 4473276 4473272 4473246 4464341 2992287 4484522 2992313 2992314 2992317 4470250 4473885 4473882 2992290 4485708 3025486 2992239 3025484 2992274 4473919 2992226 4462667 2992293 4471668 4478245 4488986 2992240 4488987 4490615 2992267 2992251 2992319 2992288 2992292 2992258 2992260 2992269 2992312 2992242 3096405 4435544 4455670 2915009 3101802 3096801 3102326 3102318 3102323 4481065 4481075 4481072 4490604 4462695 2878523 4480804 4481060 4462660 4462711 4462708 4462663 4464381 4487209 4462717 4465817 4480829 4450697 4489634 4487211 4490605 4465811 4465777 4464629 4450701 4465867 4481070 4450705 4450850 4450847 4450727 4450677 4450711 4477130 4462687 2992253 4490583 4476726 4479045 4493135 4489638 4490616 4490602 4490596 4493156 2992282 4473897 2992323 4475393 2992328 2992234 2992213 2992321 2992330 2992277 2992207 2992211 4470378 4502691 2992200 4480835 4470239 4470214 4478998 4475386 4470223 4470238 4478238 4475382 4450846 4479009 4475387 4478234 4470360 4465870 4465787 4465871 2872100 4476745 4471205 4475365 2992310 4475392 4475400 4478252 4510888 4521994 4505991 4507989 4507999 4507982 4507984 4507988 4507972 4508023 4507995 4502714 4510855 4501443 4510348 4510014 4510025 4510983 4510850 4510008 4511627 4510856 4521991 4501446 4508986 4501474 4510343 4510006 4510009 4510010 4513547 4510347 4501481 4568197 4493159 4501457 4502698 4501488 4522003 4501469 4501451 4501448 4502682 4501445 4501461 4495373 4436234 4501491 2992245 4501496 4486636 4521993 4481066 4480814 4480801 4480789 2886280 4493141 4493144 4493143 4493140 4568204 4486634 4565691 4485709 4486617 4485692 4486612 4490609 4490582 4490611 4444163 4445626 2872099 4493158 4538410 4522004 4521996 4535404 4535406 4536307 4536304 4536288 4536285 4536291 4536294 4480823 4536296 4476744 4540163 4536302 4552816 4552822 4552812 4536303 4558994 4565690 4501440 4540164 4536297 4436188 4441665 4436152 4435553 4435211 4436174 4436227 4436189 4436160 4435527 4436245 4435227 4436170 4435233 4441632 4441621 4441660 4441663 4442566 4442607 4442590 4442559 4442564 4442558 4436246 4444200 4429254 4442562 4442595 4442601 4444203 4442577 4441627 4441619 4441645 4441666 4436242 4442576 4441628 2862251 2878520 4420037 4420514 4426780 4427068 4428449 4429262 4435539 4435248 4441661 4446192 4442580 4445658 4445655 4445652 4445632 4446205 4446198 4445635 4445654 4445665 4445606 4446191 4445641 4446209 4445623 4441672 4442561 4444185 4446204 4445611 4442605 2992278 2992254 4446207 4445664 4441629 4441625 4441620 4442606 4441639 4442578 4442585 4442610 4442550 4444197 4445616 4442567 4441638 4445653 4442612 4446203 4446195 4446201 4445661 4436193 4435535 4435236 4445651 4442551 2915007 4420516 2912172 2912098 2912107 2912113 2912151 2915004 2915001 2915021 2915034 2902549 2915020 2903438 2918043 2918058 2918061 2918050 2918048 2919988 2927667 4419688 4419691 4442579 2915019 2872094 2992296 2992220 2992217 4464377 2992324 4462694 2862257 2865791 2862254 2872081 2894472 2870633 4420500 2872088 2872087 2865787 2874086 2874095 2874084 2886281 2886258 2902541 2903441 2872095 4441636 4419724 4435246 4442554 4441631 4444161 4444213 4444215 4444194 4444171 4444212 4429285 4442555 4429256 4444190 4444195 4444168 4444175 4444176 4444204 4444207 4444177 4444174 4444179 4442599 4427045 4420524 4420492 4420493 2929716 4420042 4420491 4420517 4419719 4427283 4427054 4513322 4427055 2992252 4427072 4427264 4427304 4427290 4427276 4427076 4427044 4427297 4427303 4427299 4427288 4436230 4513311 4510979 4565697 4508019 4510361 4510890 4509989 4480787 2872093 4568195 4502713 4510341 4510984 4480795 4510861 2884443 4510991 4511622 4511626 4511621 4513316 4513335 4634330 4513310 4513327 4513334 4513557 4513558 4513551 4510981 4565696 2886268 3149125 3143261 3147601 3148274 3147599 3153667 3139851 3157047 3149121 3157036 3157015 3141738 3146350 3146612 3147598 3139834 3139869 3157044 3157022 3156432 3149120 3157048 3139845 3157028 3139867 3139881 3156428 3147597 3156429 3141708 3124389 3154504 3154502 3157030 3154503 3157031 3157054 3141728 3153670 3148620 3157032 3146614 3131160 3157016 3157050 3157026 3157042 3146336 3157020 3154509 3154507 3157017 3148289 3154499 3148277 3149126 3157038 4583309 3139827 3104691 3148617 3157037 3157052 3157033 3141740 3143259 3157041 3143492 3148616 3157018 3157025 3157040 3154510 3154506 3154500 3143267 3157045 3150097 3154498 3116261 3089718 3150096 3148272 3146339 3148276 3143260 3148601 3149119 3149127 3157051 3149111 3139828 3139873 3157034 3157024 3136853 4583317 4621318 4666259 4583302 4634771 4668077 4638505 4621317 4668076 4668067 4638491 4638481 4621319 4638478 4641821 4621327 4636846 4659170 4548054 4583291 4636850 4638508 4589122 4583301 4615397 4582976 4638510 4621325 4661234 4634344 4668071 4666242 4641841 4638499 4615424 4615411 4668064 4634796 4621341 4661239 4632930 4638509 4641848 4621336 4668081 4621348 4615429 4615404 4661237 4552848 4583303 4582985 4583285 4615401 4601962 4540170 4599667 4641818 4615437 4501487 4634762 4501497 4595440 4641827 4638479 4536293 4634349 4506023 4641819 4510989 4621337 4656221 4582990 4582986 4621338 4638516 4641851 4641832 4634756 4636841 4552819 4634764 4636844 4668080 4558993 4599674 4599672 4634775 4583286 4661232 4656233 4621332 4621345 4661225 4638514 4659167 4641825 4615439 4668075 4615416 4615423 4545152 4583304 4601905 4638484 4659161 4668087 4656236 4621335 4615428 4595429 3105869 4634770 4634794 4636852 3141726 4615399 3141741 4661230 4632921 4632923 4632922 4632931 4632938 4634755 4659174 4540186 4634333 4634759 4661240 4626456 4450700 4545160 4636847 3089902 3099152 3100101 3146615 3086958 3120305 4634779 4552818 4634766 4661226 4545144 4634781 4634354 4659175 4621320 4668074 4638487 4641842 4659168 4656231 4656228 4552828 4659165 4632925 4668069 4634361 4668082 4668084 4510023 4604149 4666253 4634366 4666257 4636853 4634350 4638474 4659169 4522012 4666239 4634788 4659173 4666834 4601901 4501494 4641850 4659171 4634792 4597316 4666252 4636845 4540188 4666835 4666838 4638489 4621322 4661229 4656235 4641839 4634784 4656219 4511620 4615394 4668072 4634778 4538416 4656224 4656226 4638488 4661231 4634790 4632927 4634363 4634359 4634346 4666839 4636848 4666240 4666246 4583300 4661228 4595436 4641831 4641828 4668086 4558995 4545150 4582964 4666256 4552832 4666247 4656241 4597320 4601902 4668078 4510018 4582972 4559000 4540193 4636851 4634336 4582961 4634757 4638507 4666833 4638504 3120785 4462723 3139864 3146352 3146613 3136836 3105809 3157035 4465785 3154508 3147616 3090073 3130804 4455679 3118367 3090067 3102333 4638483 4465793 4465872 3086955 3086954 3146617 3124407 3146616 4465783 3139856 4456627 3089900 4465801 3131173 2960116 4621330 4656234 4656215 4638515 4668083 4632936 2919976 4634780 4666249 4638512 4632939 4632929 4668070 4641846 4510858 4634348 3104678 4464607 3099157 3101800 3104688 3104699 3120779 4634753 3086968 4666255 4464360 4464355 2956557 4636849 4601964 3120299 3128369 3105862 4465784 3131159 3101799 3131158 3101804 3089408 3116274 3141709 3086978 4464626 3111178 3120295 3096398 3097824 3105858 3116271 3096411 3101787 3139843 3146351 4481071 3102319 3099856 3090069 3116247 3141730 3111180 3141739 3141694 3141743 3118371 3139857 3104692 3116264 3139836 3139833 3125615 3139872 3105868 3131161 3147600 4462721 4462690 3115077 4464356 4464588 3089402 3115086 4634787 3099156 3089416 3126482 3100090 3105863 3102313 3148271 3124391 3115084 3141736 3096419 3086966 3148280 3113558 3125624 4632940 4638500 4638502 4656240 4656216 4656220 4638496 4634343 4582982 4508005 4601966 4634774 4582969 4634765 4659172 4656238 4661235 4552835 4599668 4641817 4641837 4641820 4599675 4666238 4510880 4656230 4641826 4666832 4666237 4621329 4659164 4621328 3141712 4641843 4552847 3118377 3096800 3104689 4634798 3104700 4582984 4634332 4615392 4641822 4604142 4668065 4615421 3089405 4548061 4582988 4638517 4668068 4615406 4666260 4641835 4641830 4638493 4638495 4582966 4510863 4582978 4597321 4634751 4666245 4634795 4638477 4659163 4636842 4615426 4666837 4634353 4666244 4641823 4661233 4634356 4668079 4659166 4582980 4583314 4589127 4668073 4661227 4632937 4597328 4668066 4661241 4661238 4615434 4641840 4632924 4632932 4634767 4615438 4641833 4634329 4634772 4634335 4538402 4656225 4638480 4536308 4545151 4666248 4634749 4634797 4634347 4666836 4641834 4615419 4634760 4638476 4659162 4638490 4666254 4661236 4668088 4583296 4634341 4641829 4641845 4634345 4666235 4634362 4641836 4638513 4545157 4615402 4634339 4656242 4641849 3139829 3126484 3141735 3086944 3120310 3104693 3116255 3139854 3139838 3118388 3118372 3128452 3141742 3141686 3141731 3139862 3116262 3148287 3148283 3089411 3101790 3131163 3128445 3120306 3089719 3097797 3119353 3119359 3096807 3097828 3120778 3097802 3141687 3125619 3124412 3104686 3150102 3143266 3141688 3115074 3102335 3131165 3128366 3089908 3141733 4638492 4656227 3131178 3124403 3148602 3122112 3139852 3115087 3147609 3130796 3101797 3090051 3090047 3124410 3146349 3131174 3116275 3116272 3099162 3102320 3096412 3089710 3102329 3113562 3146609 3111171 3101795 3096400 3118380 3090057 3089918 3089905 3101792 3089714 3090090 3126481 4668085 4634768 4656244 3147604 4615400 4656232 3102336 4634352 3089400 4632926 3105866 3096402 3097833 3090080 3111179 4634785 3111170 3116269 3115080 3120777 3122117 3136844 3097821 3157021 3116270 3115072 3147620 3141737 3090072 3090048 3141725 3113564 3090082 3143262 4483972 3101811 3136850 3089399 3118379 3090049 3096414 3101793 3157046 3141732 3118385 3118382 3139842 3116263 3118363 3090044 3118390 4483958 2960121 2960108 2959507 2959502 2952262 3116260 3099161 2929722 2929719 3143493 2919992 3154501 3128450 3148613 3104690 2912103 3116248 2920006 3086961 3143269 3148606 3141705 3148275 3105870 3089903 3086965 3089917 3089909 3128444 3118364 3115078 3141706 3150100 3118369 3118375 3090050 3090071 3100089 3089723 3148290 3153671 3089404 3143490 3099142 4464623 4479103 3102327 3141707 3141744 3116253 3116251 3097801 4484532 3097800 3090083 3086981 3086957 3105859 3086983 3090068 3090052 3104696 3146342 3100109 3086956 3148605 3086945 3136845 3086971 3116258 3124409 3131170 3146353 3115069 3113560 3099159 3086980 3124406 4486593 4486610 4480791 3136842 3131179 3100098 3153669 3089414 3136851 3116266 3139860 4477128 3090077 3090079 3116259 3100094 3115085 4465802 3115079 3118378 3141713 4484530 3089912 3104683 3099148 3102332 3096417 3146354 3139875 3089914 3097817 3090061 3096422 3124404 3118381 3089910 3090066 3149123 3115091 3102328 3143489 4477139 3099160 3090053 3154494 3101798 3089709 3104685 3101808 3139878 3097806 3147606 3147608 3131177 3115089 3096808 3100097 3102316 3101809 3086959 3122113 3147610 3113566 3105864 3115081 3143273 3131169 3118370 3132253 3097805 3086963 3147607 3147617 3115075 3116257 3147612 3113565 3115071 3097796 3118384 3143268 3100107 3099848 3086969 3118386 3139840 3131157 3099145 3097822 3097825 3097818 3099167 3136846 3146347 3097813 3122115 3113555 3113556 3100110 3136847 3089711 3147602 3113567 3118387 3089403 3100099 3118391 3118374 3130802 3097819 3150098 3157027 3111174 3116268 3141696 3086976 3102321 3100093 3102315 3089904 3089901 3090089 4484536 3147611 3148286 3125616 3115070 3096404 3118376 3099165 3100095 3148611 3096408 3090070 3118365 3089707 3122110 3143263 3115083 3118373 3025487 3148608 3120301 3099166 3124392 3101796 3139841 3125623 3146346 3090062 3115076 3136852 3119356 3097799 3089398 3118366 4485701 3087757 3119357 3141704 3146341 3139849 3139879 3124411 3120308 3139877 3146610 3119360 3139846 3100096 3086977 3104695 3116252 3139871 3099853 3111182 3136848 3105860 3099144 3099154 3115082 3096407 3120783 3139858 3128449 3141729 3141699 3141689 4483696 3096413 3157049 3089713 3139850 3148612 3119355 3148285 3130797 3105810 3148610 3148619 3120311 3086979 3086970 3125622 3086975 3090088 3143270 3146340 3148614 3147614 3141734 4484244 4471246 3154497 3086948 4480807 3149122 3113559 3149117 4483959 3079374 3139839 3089721 3086946 3096401 3111173 3139874 3105808 3128451 3120782 3089899 3097807 3141685 3116250 4483966 3141700 3139861 3086962 3139844 3104682 3096427 3101805 3025485 3100087 3148284 3149128 4462707 3097811 3104679 3136840 3136837 3105865 3086953 3090042 3090043 3120303 3097809 3099155 3149113 3101801 3148288 3111177 4456619 3104680 3148618 3139847 3089708 3089715 3115088 3089406 3149114 3126486 3104697 3099850 3089705 3146343 3089409 3141701 3086960 3089410 3089913 3089717 3086967 3146345 3089401 3090054 3097804 3115073 3105853 3104687 3096797 3131164 3090085 3087758 3111176 3099149 3099147 3086973 3087755 3086982 4483960 3100102 3086951 3105852 3090055 3118383 3102334 3099143 4480817 3102322 3116276 3154512 3105806 3139837 3139853 3097823 4486609 3125620 3105854 3139848 3139830 3124408 3149124 3097820 3105851 3139832 3119362 3096421 3141698 3096410 3089915 3119354 3105807 4484538 3154495 4481074 4477149 4483982 3146348 4480827 4483961 4481059 4484242 3146618 4486408 4486623 4484241 3096805 4483697 4478235 3089720 3099847 3089706 2912131 4487210 4480822 3090045 3099164 4478254 4478229 4479054 4481069 4481058 4481064 3090064 3104698 3089716 3141697 3105856 3104694 3146337 3090058 4464339 3090081 3104702 4464344 3089916 3131176 3150099 3139831 4464596 3116256 4486417 3130803 3146344 3089415 2960122 3096426 3141703 3089911 3089412 3096409 3090091 3089407 3090059 4481061 4483971 4480799 3089704 3101807 3089906 4487208 3086949 4481057 4470244 3100105 3131168 3105855 4480802 4483694 3141692 3141702 3096406 3130801 3118362 3131166 3090076 3105857 3089417 3105861 3111183 3141711 3100100 3130798 3130800 3130799 3139868 4471222 3097803 3099168 3149109 3139826 3147618 3146611 3147619 3131172 3120302 3125617 3120786 3119358 3120780 3148603 3118368 3126483 3150101 3124390 3141695 3141714 3115090 3119361 3147621 3143272 3148609 3143271 3100103 3149118 3139866 3120313 3147605 3143265 3122111 3147615 3143491 3096416 3102314 3116273 3139876 3128448 3089413 3122114 3105867 3131156 3126485 3120298 3120307 3141693 3139859 3141690 3148279 3148273 3148278 3139870 3139880 3087759 3147603 3149110 3149115 3131175 3124388 3128367 3128368 3147613 3102325 3154496 3136843 3090074 3090063 3096425 3090084 3102317 3101794 3086974 3096803 3125625 3128453 3139863 3090078 3096798 3096806 3097829 3086950 3089722 3102324 3090046 3096424 3096418 3090086 3090065 3111175 3097798 3090087 3104681 3124405 3148607 3096403 3086964 3087756 3120784 3089712 3139825 3111172 3125621 3118389 3141691 3148615 3090060 3097810 3120297 3090075 3096423 3096420 3096415 3101788 3157029 3079370 3136839 3100106 3096399 3154511 3157053 3146335 3146338 3148281 3149116 3111181 3122119 3143264 3141710 3097835 3113563 3154505 3136838 3128447 3139855 3139835 3113561 3125614 3131167 3113557 3136849 3100104 3131162 3131171 3116254 3102312 3148282 3136841 3141727 3116265 3122116 3104684 3104701 3116249 3122118 3102331 4479010 4540166 4470216 4473271 2829480 2888740 4462668 4462706 4462699 4462734 4462682 4462680 4568206 4480828 2880082 4535408 4508977 4508985 4505299 4505304 4505337 4495388 4615418 4480809 2894469 2832940 4479102 4429271 4479104 4538419 2854999 2992332 2992309 2992301 4493148 4486409 2992329 2919995 4475377 4475362 4897800 4902295 3086952 4902276 4902272 4902273 4897834 4902274 4902288 4902292 4902290 4902275 4902270 4902294 4902289 4902297 4897857 4902266 4902285 4902281 4902282 4902284 4902279 4902265 4902263 3157039 3157019 4902296 4897843 4902291 4897810 4897809 4897804 4897785 |
Recalling Firm/ Manufacturer |
St Jude Medical Inc. 15900 Valley View Ct Sylmar CA 91342-3577
|
For Additional Information Contact |
Candace Steele Flippin 651-756-6293
|
Manufacturer Reason for Recall |
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
|
FDA Determined Cause 2 |
Unknown/Undetermined by firm |
Action |
Abbott sent an Important Cybersecurity Advisory dated August 28, 2017, to all affected customers to notify customers of the availability of the programmer software update and associated pacemaker firmware update. The notification material is in the form of a Physician Letter and Hospital Letter and will be delivered to physicians and hospitals with inventory in the US by overnight service. Customers with questions were instructed to contact their Abbott representative or the customers technical support hotline at 1-800-722-3774. |
Quantity in Commerce |
6051 units |
Distribution |
Worldwide Distribution |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
PMA Database |
PMAs with Product Code = DXY and Original Applicant = Abbott Medical
|