| Class 2 Device Recall Lyophilized Chemistry Calibrator Level 1 and Level 2 |  |
Date Initiated by Firm | August 24, 2017 |
Create Date | February 13, 2018 |
Recall Status1 |
Terminated 3 on August 28, 2020 |
Recall Number | Z-0605-2018 |
Recall Event ID |
78699 |
510(K)Number | K043460 |
Product Classification |
Calibrator, multi-analyte mixture - Product Code JIX
|
Product | Lyophilized Chemistry Calibrator Level 1
Product Usage:
Beckman Coulter Chemistry Calibrators are intended for use when calibrating methods run on the Beckman Coulter AU series of chemistry analyzers. Beckman Coulter Chemistry Calibrators are lyophilized, human serum based products formulated for use as a reference material when calibrating Beckman Coulter AU clinical chemistry system assays. The calibrators are used to calibrate the following assays: Albumin (ALB), Bicarbonate (CO2), Direct Bilirubin (DBIL), Total Bilirubin (TBIL), Calcium Arsenazo (CALA), Calcium (CAL oCPC), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Phosphorus (PHOS), Iron, Lactate, Magnesium (MG), Total Protein (TP), Triglyceride (TRIG), Unbound Iron Binding Capacity (UIBC), Urea Nitrogen (BUN), and Uric Acid. Only lots 6101K51 and 6101K61 were subject to recall. |
Code Information |
Lots # 6101K51 and 6101K61 |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 250 S Kraemer Blvd Brea CA 92821-6232
|
For Additional Information Contact | Customer Support 800-854-3633 |
Manufacturer Reason for Recall | Unpaired lots of AU lyophilized chemistry calibrator level 1 and 2 have been distributed. Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. If lot numbers are not correctly paired, calibration can be run using incorrect set points which may generate erroneous results. QC failures could occur if incorrect pairs of calibrators are used. |
FDA Determined Cause 2 | Under Investigation by firm |
Action | On August 28, 2017 a letter was sent to all the customers stating the following:
Attention Beckman Coulter Customer,
Beckman Coulter is initiating a field action for the product listed above. This letter contains important information that needs your immediate attention.
ISSUE:
Unpaired lots of AU Lyophilized Chemistry Calibrator level 1 and 2 have been distributed.
Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination.
Paired calibrator lots are identified by the last two digits of the lot number.
Calibrator Level 1 REF DR0070-1, Lot 6101K51
Calibrator Level 2 REF DR0070-2, Lot 6102K61
Incorrect pair
Calibrator Level 1 REF DR0070-1, Lot 6101K61
Calibrator Level 2 REF DR0070-2, Lot 6102K61
Correct pair
IMPACT:
If lot numbers are not correctly paired, calibration can be run using incorrect set points which may generate erroneous results.
QC failures could occur if incorrect pairs of calibrators are used.
Calibrator Level 1 REF DR0070-1, Lot 6101K51
Calibrator Level 2 REF DR0070-2, Lot 6102K61
Incorrect pair
If the above incorrect pair is run with Level 1 lot 6101K51and Level 2 lot 6102K51 set points, erroneous results could be generated for the following assays:
o Bicarbonate : maximum 11% bias low
o Calcium Arsenazo : maximum 5% bias low
o Calcium oCPC : maximum 5% bias low
o Inorganic Phosphorous : 6.4% bias high
o Magnesium: 3.2% bias low
If the above incorrect pair is run with Level 1 lot 6101K61 and Level 2 lot 6102K61 set points, erroneous results could be generated for the following assays:
o Bicarbonate : maximum 15.8% bias high
o Direct Bilirubin OSR6x181: 8.7% bias low
ACTION: Check the level 1 and level 2 calibrators to make sure you are using
correctly paired lots for calibration.
Ensure correct set points are being used for the lot. Set points are in
the IFU for each lot and can be found in each kit, or on the Beckman
Coulter websites |
Quantity in Commerce | unknown |
Distribution | Worldwide Distribution in the states of AL, AK, AZ, AK, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IO, KS, KY, LA, MA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NY, NM, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VI, WA, WV, WI, and WY
and the countries of Brazil, Canada, Costa Rica, El Salvador, Guam, Paraguay, and Peru |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JIX
|
|
|
|