| Class 2 Device Recall Artis Q and Q.Zen | |
Date Initiated by Firm | June 06, 2017 |
Create Date | February 23, 2018 |
Recall Status1 |
Terminated 3 on May 16, 2019 |
Recall Number | Z-0706-2018 |
Recall Event ID |
79097 |
510(K)Number | K123529 |
Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
Product | Artis Q and Q.Zen, Model Numbers:10848280, 10848281, 10848282, 10848353 & 10848355 |
Code Information |
10094135, 10094137, 10094141, 10094142, 10094143, 10280959, 10848280, 10848281, 10848282, 10848353, 10848354, 10848355, 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350, 7728392 |
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc 40 Liberty Blvd Malvern PA 19355-1418
|
For Additional Information Contact | Meredith.adams@siemens.com 610-448-6461 |
Manufacturer Reason for Recall | Potential for liquid entry |
FDA Determined Cause 2 | Device Design |
Action | On June 6, 2017, Siemens Healthineers distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via USPS Certified Mail.
Siemens conducted a corrective field action in June 2015 to protect the Artis table against liquid entry at critical locations. However, even with this corrective action, if there are larger amounts of liquid on the table surface, they can still enter the table. Siemens Medical Solutions USA, Inc. will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment by calling 610-448-6461. |
Distribution | U.S. Nationwide |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = OWB
|
|
|
|