Date Initiated by Firm | January 02, 2018 |
Date Posted | May 04, 2018 |
Recall Status1 |
Terminated 3 on February 10, 2023 |
Recall Number | Z-1688-2018 |
Recall Event ID |
79505 |
Product Classification |
Radioimmunoassay, dehydroepiandrosterone (free and sulfate) - Product Code JKC
|
Product | ADVIA Centaur DHEA-SO4 |
Code Information |
ADVIA Centaur¿ DHEA-SO4; radioimmunoassay, dehydroepiandrosterone (free and sulfate); SMN 10282227 ADVIA Centaur DHEA-S Kit Lot # 00119053 14041053 80449053 90520053 29750054 30423054 39155054 50533054 63267055 68936055 79273055 86427055 06402057 07109057 14187057 37499057 |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 511 Benedict Ave Tarrytown NY 10591-5005
|
For Additional Information Contact | Ms. Aarti M. Aziz 914-631-8000 |
Manufacturer Reason for Recall | The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere |
FDA Determined Cause 2 | Error in labeling |
Action | Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA |
Distribution | Nationwide and Canada, Mexico, and Thailand |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|