  Class 2 Device Recall MDEleva, UDEleva see related information
Date Initiated by Firm March 26, 2018
Date Posted May 03, 2018
Recall Status1 Terminated 3 on June 12, 2020
Recall Number Z-1609-2018
Recall Event ID 79783
510(K)Number K050151  
Product Classification System, x-ray, stationary - Product Code KPR
Product MultiDiagnostic-Eleva with Flat Detector 708035

As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.
Code Information Pending
Recalling Firm/
Manufacturer
Philips Medical Systems Nederlands
Veenpluis 4-6
Best Netherlands
Manufacturer Reason
for Recall
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
FDA Determined
Cause 2
Nonconforming Material/Component
Action The firm, BU IGT Systems, sent an "URGENT- Field Safety Notice" letter dated 3/23/2018 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to stop using the system and call your Philips representative if the following situations occurs: -the table height or tilt movement is not working. - you notice a cracking or snapping sound during system operation. - you experience a blockade in the table height or tilt movement during system movements. -the system is having an unexpected collision or the system had an unexpected collision in the past 2 months. - you have used a CPR stand that obstructed the system, or - you notice any (other) unusual system behavior beyond the normal use of the system. The customers should also avoid collisions with the table and base stand of the system by ensuring that no obstacles are placed around the system. In addition, ensure that all staff with access to the affected systems are informed of the content of this Field Safety Notice. The firm stated: -All possibly affected products in the field will be corrected by means of a field change order free of charge. -A firm representative will replace affected material on the upper and lower tilt actuator in the affected systems. -You will be contacted by our local Philips representative to schedule this corrective action. -This action will start effective April 2018. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722- 9377, 8am to 8pm (EST) Monday thru Friday.
Quantity in Commerce 1
Distribution Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = KPR and Original Applicant = PHILIPS MEDICAL SYSTEMS NORTH AMERICA, INC.
-
-