• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall POWERPORT(R) ClearVUE(R) Slim with Smooth Septum, 6F ChronoFlex(TM) Silk, Polyurethane Catheter with

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall POWERPORT(R) ClearVUE(R) Slim with Smooth Septum, 6F ChronoFlex(TM) Silk, Polyurethane Catheter withsee related information
Date Initiated by FirmFebruary 08, 2018
Create DateMay 09, 2018
Recall Status1 Completed
Recall NumberZ-1769-2018
Recall Event ID 79841
510(K)NumberK122899 
Product Classification Port & catheter, implanted, subcutaneous, intravascular - Product Code LJT
ProductPOWERPORT ClearVUE Slim with Smooth Septum, 6F ChronoFlex Silk, Polyurethane Catheter with Open Suture Holes, REF 5676301
Code Information Lot Number Unique Device Identifier (UDI) REAN0066 (01)00801741110795(17)170528(10)REAN0066 REAP0629 (01)00801741110795(17)170728(10)REAP0629 REAY1483 (01)00801741110795(17)180531(10)REAY1483 REAY2494 (01)00801741110795(17)180331(10)REAY2494 REBS0163 (01)00801741110795(17)181031(10)REBS0163 REBT0364 (01)00801741110795(17)181130(10)REBT0364 REBT1664 (01)00801741110795(17)181231(10)REBT1664 REBT2149 (01)00801741110795(17)181231(10)REBT2149 REBW0874 (01)00801741110795(17)190331(10)REBW0874 REBY1231 (01)00801741110795(17)190430(10)REBY1231 
Recalling Firm/
Manufacturer
Bard Peripheral Vascular Inc
1625 W 3rd St Bldg 1
Tempe AZ 85281-2438
Manufacturer Reason
for Recall
The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.
FDA Determined
Cause 2
Error in labeling
ActionThe firm initiated the recall by letter on 02/08/2018. The letter explained the problem and the products involved and requested the following actions: "Please check all inventory locations within your institution for affected product code / lot number combinations listed in the field correction notice. If you have further distributed any of the product code / lot numbers, please immediately contact that location, advise them of this field correction notice, and forward these instructions. For any identified product on your shelves, please replace the products Patient ID Card with the corrected Patient ID Card enclosed with this field correction notice. Enclosed with this field correction notice is a Patient Communication Packet that contains one patient letter, one corrected Patient ID Card, and one POWERPORT Patient Guide. Please contact all patients that have received any product listed in this field correction notice and provide them with the Patient Communication Packet."
Quantity in Commerce835 units
DistributionNationwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.