• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall MegaPower Electrosurgical Generator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall MegaPower Electrosurgical Generatorsee related information
Date Initiated by FirmMay 15, 2018
Date PostedAugust 13, 2018
Recall Status1 Terminated 3 on January 30, 2024
Recall NumberZ-2792-2018
Recall Event ID 80327
510(K)NumberK050579 
Product Classification Electrosurgical, cutting & coagulation & accessories - Product Code GEI
ProductMegaPower Electrosurgical Generator, Product code 1000, GTIN 10614559103715
Code Information 00001001 00010001 13806001 00053001 00063001 00066001 00070001 00078001 00085001 00091001 00100001 00105001 00110001 00115001 00128001 00132001 00146001 00167001 00181001 00187001 00191001 00200001 00211001 00212001 00223001 00229001 00237001 00240001 00242001 00246001 00248001 00254001 00265001 00278001 00282001 00283001 00308001 00313001 00313001 00314001 00333001 00334001 00379001 00380001 00391001 00398001 00402001 00413001 00416001 00418001 00424001 00432001 00437001 00438001 00458001 00459001 00471001 00479001 00481001 00484001 00524001 00535001 00543001 00551001 00553001 00554001 00574001 00586001 00590001 00619001 00627001 00632001 00638001 00639001 00640001 00658001 00685001 00690001 00693001 00707001 00708001 00710001 00721001 00725001 00728001 00731001 00740001 00779001 00796001 00801001 00804001 00813001 00816001 00829001 00841001 00847001 00848001 00859001 00868001 00871001 00872001 00884001 00890001 00899001 00904001 00905001 00906001 00922001 00932001 00935001 00937001 00943001 00951001 00952001 00953001 00959001 00960001 00978001 00982001 00988001 1003001 1007001 1020001 1023001 1032001 1034001 1043001 1055001 1065001 1069001 1070001 1083001 1088001 1092001 1109001 1110001 1115001 1119001 1120001 1126001 1134001 1135001 1144001 1153001 1158001 1176001 1182001 1190001 1195001 1201001 1213001 1219001 1224001 1227001 1237001 1241001 1279001 1284001 1293001 1294001 1300001 1319001 1320001 1331001 1332001 1337001 1345001 1346001 1351001 1360001 1361001 1390001 1392001 1393001 1422001 1425001 1434001 1440001 1442001 1448001 1451001 1452001 1459001 1462001 1474001 1477001 1484001 1490001 1495001 1497001 1510001 1516001 1520001 1525001 1529001 1536001 1558001 1562001 1566001 1574001 1596001 1598001 1601001 1608001 1612001 1613001 1623001 1625001 1649001 1650001 1656001 1659001 1667001 1674001 1682001 1686001 1727001 1729001 1732001 1752001 1762001 1771001 1773001 1800001 1815001 1823001 1824001 1831001 1833001 1837001 1839001 1844001 1845001 1847001 1880001 1881001 1893001 1900001 1912001 1914001 1915001 1922001 1937001 1959001 1962001 1972001 1973001 1985001 2006001 2010001 2015001 2020001 2024001 2033001 2057001 2058001 2070001 2099001 2101001 2107001 2108001 2120001 2124001 2141001 2148001 2149001 2180001 2199001 2224001 2226001 2252001 2259001 2269001 2277001 2278001 2289001 2292001 2310001 2328001 2371001 2422001 2441001 2446001 2451001 2459001 2461001 2471001 2473001 2480001 2482001 2487001 2490001 2493001 2513001 2519001 2526001 2530001 2531001 2548001 2549001 2574001 2578001 2582001 2593001 2615001 2616001 2621001 2637001 2648001 2655001 2669001 2676001 2679001 2704001 2707001 2708001 2716001 2718001 2725001 2727001 2735001 2746001 2748001 2749001 2751001 2754001 2755001 2766001 2770001 2771001 2772001 2780001 2781001 2782001 2788001 2791001 2795001 2797001 2801001 2803001 2808001 2816001 2819001 2823001 2830001 2833001 2834001 2835001 2848001 2849001 2855001 2863001 2866001 2875001 2878001 2882001 2885001 2886001 2889001 2890001 2901001 2905001 2910001 2913001 2916001 2921001 2924001 2927001 2931001 2933001 2936001 2937001 2943001 2955001 2959001 2966001 2971001 2974001 2977001 2988001 2989001 2990001 2993001 2994001 2997001 3010001 3011001 3012001 3022001 3024001 3027001 3028001 3030001 3035001 3039001 3043001 3051001 3057001 3060001 3066001 3072001 3086001 3089001 3105001 3128001 3129001 3130001 3134001 3139001 3142001 3145001 3155001 3156001 3164001 3167001 3178001 3182001 3200001 3207001 3211001 3214001 3216001 3220001 3235001 3243001 3248001 3249001 3256001 3260001 3275001 3277001 3279001 3284001 3289001 3291001 3296001 3297001 3305001 3309001 3315001 3331001 3332001 3338001 3340001 3341001 3346001 3350001 3353001 3356001 3366001 3367001 3371001 3380001 3382001 3389001 3391001 3397001 3412001 3432001 3435001 3438001 3444001 3446001 3449001 3478001 3480001 3500001 3504001 3520001 3523001 3524001 3527001 3535001 3536001 3549001 3557001 3558001 3570001 3588001 3602001 3607001 3609001 3610001 3623001 3629001 3637001 3638001 3645001 3651001 3652001 3669001 3676001 3685001 3686001 3687001 3694001 3698001 3704001 3708001 3729001 3744001 3763001 3777001 3781001 3787001 3792001 3806001 3814001 3832001 3836001 3841001 3842001 3857001 3861001 3878001 3879001 3884001 3900001 3903001 3904001 3908001 3910001 3923001 3924001 3941001 3945001 3952001 3962001 3963001 3972001 3980001 3983001 3987001 4006001 4030001 4031001 4036001 4039001 4042001 7250001 10002001 10004001 10016001 10017001 10019001 10028001 10058001 10065001 10069001 10090001 10091001 10100001 10102001 10108001 10116001 10120001 10124001 10130001 10131001 10141001 10142001 10147001 10183001 10185001 10196001 10212001 10214001 10216001 10217001 10219001 10234001 10249001 10251001 10262001 10263001 10268001 10272001 10273001 10279001 10286001 10294001 10305001 10309001 10323001 10330001 10340001 10355001 10358001 10363001 10364001 10365001 10382001 10383001 10390001 10393001 10413001 10416001 10422001 10432001 10439001 10447001 10450001 10452001 10473001 10474001 10481001 10494001 10514001 10516001 10522001 10532001 10538001 10541001 10550001 10550001 10551001 10552001 10555001 10575001 10580001 10586001 10587001 10615001 10621001 10622001 10628001 10650001 10665001 10672001 10677001 10684001 10711001 10727001 10741001 10754001 10758001 10773001 10782001 10823001 10851001 10884001 10885001 10886001 10893001 10896001 10931001 10940001 10950001 10960001 10966001 10969001 10970001 10975001 10981001 10989001 10993001 10995001 11005001 11007001 11010001 11013001 11014001 11022001 11023001 11042001 11043001 11047001 11052001 11058001 11061001 11063001 11071001 11072001 11077001 11087001 11092001 11094001 11119001 11121001 11125001 11136001 11141001 11144001 11154001 11155001 11157001 11158001 11174001 11183001 11187001 11192001 11195001 11200001 11207001 11208001 11218001 11223001 11232001 11239001 11240001 11243001 11245001 11247001 11250001 11262001 11268001 11271001 11272001 11290001 11291001 11292001 11301001 11309001 11314001 11317001 11318001 11320001 11335001 11337001 11340001 11341001 11350001 11356001 11363001 11366001 11385001 11416001 11423001 11424001 11439001 11477001 11618001 11628001 11644001 11651001 11652001 11659001 11664001 11668001 11671001 11679001 11690001 11694001 11695001 11713001 11714001 11739001 11740001 11746001 11747001 11761001 11764001 11783001 11803001 11804001 11829001 11830001 11839001 11840001 11893001 11896001 11900001 11905001 11922001 11924001 11926001 11939001 11941001 11944001 11945001 11963001 11974001 11976001 11984001 11998001 12005001 12014001 12030001 12039001 12051001 12053001 12054001 12072001 12082001 12089001 12092001 12095001 12096001 12105001 12115001 12117001 12118001 12121001 12122001 12137001 12156001 12161001 12162001 12163001 12166001 12170001 12171001 12174001 12176001 12185001 12187001 12197001 12198001 12201001 12202001 12264001 12266001 12271001 12293001 12306001 12310001 12311001 12316001 12317001 12320001 12324001 12337001 12343001 12513001 12515001 12518001 12523001 12535001 12536001 12538001 12539001 12540001 12560001 12561001 12566001 12577001 12587001 12593001 12595001 12599001 12601001 12602001 12603001 12604001 12605001 12606001 12607001 12608001 12637001 12638001 12639001 12661001 12705001 12707001 12709001 12710001 12711001 12713001 12719001 12722001 12739001 12741001 12746001 12771001 12777001 12782001 12785001 12793001 12796001 12807001 12811001 12831001 12838001 12841001 12842001 12856001 12866001 12897001 12901001 12918001 12936001 12945001 12946001 12947001 12949001 12950001 12966001 12967001 12973001 12976001 12998001 12999001 13001001 13010001 13011001 13018001 13019001 13023001 13030001 13041001 13050001 13055001 13056001 13057001 13063001 13064001 13086001 13091001 13092001 13097001 13102001 13103001 13104001 13117001 13128001 13143001 13146001 13147001 13148001 13151001 13154001 13162001 13164001 13165001 13166001 13167001 13168001 13171001 13176001 13177001 13180001 13182001 13189001 13202001 13216001 13217001 13218001 13224001 13228001 13229001 13230001 13235001 13238001 13251001 13252001 13272001 13280001 13283001 13288001 13293001 13294001 13295001 13296001 13297001 13298001 13299001 13300001 13301001 13302001 13303001 13304001 13305001 13306001 13307001 13308001 13309001 13310001 13311001 13312001 13313001 13314001 13315001 13316001 13317001 13318001 13319001 13320001 13321001 13322001 13323001 13324001 13325001 13329001 13330001 13331001 13332001 13333001 13334001 13335001 13336001 13339001 13341001 13342001 13343001 13344001 13345001 13346001 13347001 13648001 13655001 13666001 13669001 13670001 13686001 13687001 13705001 13718001 13736001 13739001 13750001 13760001 13765001 13776001 13783001 13785001 13802001 13807001 13817001 13828001 13829001 13834001 13838001 13845001 13849001 13857001 13866001 13875001 13879001 13881001 13882001 13885001 13896001 13898001 13900001 13901001 13902001 13910001 13918001 13921001 13939001 13940001 20000001 20008001 20009001 20012001 20014001 20030001 20043001 20047001 20073001 20092001 20096001 20097001 20112001 20120001 20121001 20123001 20136001 20141001 20164001 20179001 20180001 20191001 20192001 20212001 20215001 20216001 20218001 20237001 20252001 20253001 20259001 20260001 20261001 20262001 20264001 20265001 20266001 20274001 20277001 20300001 20317001 20318001 20335001 20342001 20347001 20361001 20369001 20370001 20382001 20385001 20389001 20392001 20401001 20408001 20409001 20412001 20416001 20421001 20424001 20434001 20436001 20450001 20451001 20456001 20462001 20465001 20466001 20481001 20482001 20483001 20491001 20503001 20505001 20509001 20624001 20794001 20795001 20807001 20830001 20845001 20847001 20854001 20856001 20859001 20861001 20864001 20875001 20878001 20883001 20885001 20892001 20911001 20928001 20944001 20954001 20960001 20961001 20966001 20973001 20983001 20990001 21003001 21009001 21012001 21015001 21046001 21053001 21060001 21134001 21135001 21138001 21140001 21149001 21152001 21159001 21160001 21165001 21180001 21181001 21182001 21192001 21204001 21208001 21209001 21210001 21216001 21219001 21220001 21224001 21229001 21236001 21240001 21255001 21256001 21261001 21267001 21275001 21277001 21285001 21296001 21298001 21302001 21310001 21311001 21318001 21323001 21364001 21377001 21380001 21392001 21400001 21419001 21421001 21432001 21442001 21444001 21447001 21448001 21452001 21453001 21460001 21463001 21464001 21466001 21468001 21492001 21494001 21498001 21501001 21504001 21505001 21518001 21519001 21523001 21525001 21527001 21534001 21535001 21538001 21541001 21547001 21554001 21563001 21575001 21580001 21582001 21594001 21615001 21622001 21675001 21676001 21677001 21684001 21687001 21695001 21700001 21701001 21711001 21714001 21716001 21728001 21739001 21741001 21752001 21757001 21758001 21845001 21847001 21855001 21869001 21878001 21887001 21888001 21892001 21895001 21896001 21899001 21907001 21910001 21947001 21953001 21956001 21967001 21968001 21970001 21971001 21974001 21982001 21984001 21990001 22001001 22005001 22007001 22014001 22027001 22029001 22031001 22049001 22054001 22074001 22076001 22079001 22081001 22088001 22089001 22104001 22122001 22132001 22136001 22138001 22143001 22153001 22154001 22155001 22157001 22168001 22169001 22170001 22181001 22187001 22192001 22201001 22206001 22217001 22218001 22223001 22226001 22231001 22232001 22243001 22244001 22246001 22261001 22277001 22282001 22308001 22314001 22320001 22322001 22325001 22340001 22345001 22346001 22347001 22358001 22366001 22373001 22379001 22385001 22392001 22394001 22395001 22407001 22409001 22410001 22412001 22415001 22420001 22431001 22435001 22447001 22457001 22458001 22465001 22473001 22474001 22476001 22477001 22478001 22493001 22494001 22495001 22501001 22506001 22510001 22523001 22525001 22533001 22549001 22556001 22557001 22566001 22568001 22576001 22581001 22591001 22592001 22611001 22615001 22618001 22626001 22628001 22631001 22632001 22643001 22653001 22654001 22663001 22665001 22681001 22685001 22690001 22701001 22709001 22716001 22722001 22736001 22742001 22744001 22745001 22746001 22759001 22765001 22784001 22824001 22831001 22837001 22840001 22841001 22847001 22848001 22853001 22857001 22863001 22865001 22871001 22892001 22893001 22895001 22898001 22912001 22920001 22928001 22932001 22939001 22942001 22955001 22957001 22960001 22961001 22962001 22975001 22986001 23000001 23001001 23002001 23043001 23047001 23051001 23058001 23059001 23068001 23077001 23080001 23100001 23102001 23110001 23112001 23114001 23115001 23125001 23136001 23137001 23144001 23146001 23155001 23164001 23168001 23169001 23170001 23180001 23184001 23186001 23191001 23203001 23223001 23227001 23228001 23229001 23265001 23267001 23279001 23286001 23294001 23296001 23303001 23316001 23317001 23319001 23325001 23329001 23335001 23340001 23343001 23355001 23357001 23358001 23361001 23371001 23377001 23395001 23402001 23411001 23419001 23427001 23430001 23432001 23434001 23447001 23449001 23451001 23464001 23479001 23481001 23485001 23502001 23510001 23546001 23552001 23555001 23560001 23579001 23587001 23593001 23598001 23611001 23614001 23637001 23661001 23671001 23672001 23681001 23684001 23686001 23692001 23709001 23718001 23723001 23725001 23726001 23775001 23777001 23778001 23788001 23792001 23793001 23802001 23850001 23852001 23865001 23867001 23915001 23923001 23935001 23939001 23946001 23950001 23966001 23982001 23992001 23994001 23998001 24013001 24015001 24024001 24029001 24037001 24040001 24042001 24071001 24094001 24096001 24102001 24117001 24123001 24137001 50726001 50727001 50730001 50733001 50734001 50735001 50745001 50748001 50749001 50750001 50751001 50752001 50753001 50754001 51237001 51237002 51238001 51238002 51239001 51239002 51240002 51241001 51241002 51242002 51243002 51244001 51244002 51245001 51246001 51246002 51344001 51345001 51346001 51347001 51348001 51349001 51350001 51351001 51353001 51354001 51355001 51356001 51357001 51358001 51359001 51360001 51361001 51362001 51363001 51364001 51389001 51390001 51391001 51392001 51393001 51394001 51395001 51396001 51397001 51398001 51399001 51400001 51401001 51402001 51403001 51404001 51405001 51406001 51407001 51408001 51442001 51443001 51444001 51445001 51446001 51447001 51448001 51449001 51450001 51451001 51461001 51462001 51463001 51464001 51465001 51466001 51467001 51468001 51469001 51470001 51471001 51472001 51473001 51474001 51475001 51476001 51477001 51478001 51479001 51480001 51562001 51563001 51564001 51565001 51566001 51567001 51568001 51569001 51570001 51571001 51572001 51573001 51574001 51575001 51576001 51577001 51578001 51579001 51580001 51581001 51602001 51603001 51604001 51605001 51606001 51607001 51608001 51609001 51610001 51611001 51612001 51613001 51614001 51615001 51616001 51617001 51618001 51619001 51620001 51621001 51665001 51666001 51667001 51668001 51669001 51670001 51671001 51672001 51673001 51674001 51675001 51676001 51677001 51678001 51679001 51680001 51681001 51682001 51683001 51684001 51685001 51698001 51699001 51700001 51701001 51702001 51703001 51705001 51706001 51707001 51708001 51709001 51710001 51711001 51740001 51799001 51800001 51801001 51802001 51803001 51804001 51805001 51806001 51807001 51808001 51809001 51810001 51811001 51812001 51813001 51814001 51815001 51816001 51817001 51818001 51819001 51820001 51821001 51822001 51824001 51825001 51826001 51827001 51828001 51829001 51830001 51831001 51832001 51833001 51834001 51835001 51836001 51837001 51838001 51839001 51840001 51841001 51842001 51843001 51844001 51845001 51846001 51847001 51848001 51849001 51850001 51851001 51852001 51853001 51854001 51855001 60014001 60015001 60016001 60017001 60018001 60019001 60020001 60021001 60022001 60023001 60024001 60025001 60026001 60027001 60028001 60160001 60161001 60162001 60165001 60166001 60167001 60168001 60169001 60170001 60171001 60172001 60173001 60174001 60175001 60176001 60177001 60178001 60179001 60264001 60265001 60266001 60267001 60268001 60269001 60270001 60271001 60272001 60274001 60275001 60276001 60277001 60278001 60279001 60280001 60281001 60282001 60283001 60284001 60316001 60317001 60318001 60319001 60320001 60321001 60322001 60323001 60324001 60325001 60326001 60327001 60328001 60329001 60330001 60331001 60332001 60333001 60334001 60335001 60746001 60747001 60748001 60749001 60752001 60753001 60754001 60755001 60756001 60757001 60758001 60759001 60807001 60808001 60809001 60810001 60811001 60812001 60813001 60814001 60815001 60816001 60823001 61117001 61118001 61119001 61120001 61121001 61122001 61123001 61124001 61125001 61126001 61127001 61128001 61129001 61130001 61131001 61132001 61133001 61134001 61135001 61136001 61408001 61409001 61410001 61411001 61412001 61413001 61414001 61415001 61416001 61417001 61418001 61419001 61420001 61421001 61422001 61423001 61424001 61425001 61426001 61427001 61455001 61456001 61457001 61458001 61459001 61460001 61461001 61462001 61463001 61464001 61465001 61466001 61467001 61468001 61469001 61470001 61471001 61472001 61473001 61474001 61686001 61687001 61688001 61689001 61690001 61691001 61692001 61693001 61694001 61695001 61696001 61697001 61698001 61699001 61700001 61701001 61702001 61703001 61704001 61705001 61760001 61761001 61762001 61763001 61764001 61765001 61766001 61767001 61768001 61769001 61770001 61771001 61772001 61773001 61774001 61775001 61776001 61777001 61778001 61779001 61828001 61829001 61830001 61831001 61832001 61833001 61834001 61835001 61836001 61837001 61838001 61839001 61840001 61841001 61842001 61843001 61844001 61845001 61846001 61847001 61883001 61884001 61885001 61886001 61887001 61888001 61889001 61890001 61891001 61892001 61937001 61938001 61939001 61940001 61941001 61942001 61943001 61944001 61945001 61946001 61947001 61948001 61949001 61950001 61951001 61952001 61953001 61954001 61955001 61956001 61957001 62087001 62088001 62089001 62090001 62091001 62092001 62093001 62094001 62095001 62096001 62097001 62098001 62099001 62100001 62101001 62102001 62103001 62104001 62105001 62106001 62107001 62167001 62168001 62169001 62170001 62171001 62172001 62173001 62174001 62175001 62176001 62176001 62176002 62193001 62194001 62195001 62196001 62197001 62198001 62199001 62200001 62201001 62202001 70008001 70009001 70010001 70011001 70012001 70013001 70014001 70015001 70039001 70040001 70041001 70042001 70043001 70044001 70045001 70046001 70047001 70048001 70049001 70189001 70190001 70191001 70192001 70193001 70194001 70195001 70196001 70197001 70198001 70298001 70299001 70300001 70301001 70302001 70303001 70304001 70305001 70306001 70307001 70435001 70436001 70437001 70438001 70439001 70440001 70442001 70443001 70444001 70445001 70446001 70447001 70448001 70449001 70450001 70451001 70452001 70453001 70454001 70455001 70456001 70457001 70927001 70928001 70929001 70930001 70931001 70932001 70933001 70934001 70935001 70936001 71241001 71242001 71243001 71244001 71245001 71246001 71413001 71478001 71479001 71480001 71486001 71487001 71488001 71512001 71513001 71521001 71522001 71523001 71529001 71530001 71531001 71594001 71595001 71596001 71601001 71606001 71627001 71628001 71629001 71634001 71635001 71636001 71639001 71694001 71695001 71696001 71701001 71702001 71707001 71708001 71709001 71746001 71747001 71748001 71752001 71753001 71754001 71759001 71760001 71761001 71768001 71769001 71770001 71771001 71772001 71773001 71774001 71775001 71775001 71776001 71777001 71778001 71779001 71821001 71822001 71823001 71845001 71846001 71847001 71851001 71852001 71853001 71869001 71870001 71871001 71875001 71876001 71877001 71881001 71882001 71883001 71892001 71893001 71894001 71926001 71927001 71928001 71958001 71959001 71960001 71961001 71969001 71970001 71971001 71972001 71973001 71974001 71975001 71976001 71977001 71978001 71979001 71980001 71981001 71982001 71983001 71984001 71985001 71986001 71987001 71988001 71989001 71990001 71991001 71992001 71993001 71994001 71995001 71996001 71997001 71998001 72000001 72001001 72002001 72003001 72004001 72005001 72006001 72007001 72008001 72135001 72136001 72137001 72138001 72139001 72145001 72146001 72147001 72148001 72149001 72199001 72200001 72201001 72201001 72202001 72203001 72203001 72209001 72210001 72211001 72212001 72212001 72213001 72232001 72233001 72234001 72235001 72236001 72248001 72249001 72254001 72255001 72256001 72257001 72331001 72332001 72333001 72334001 72335001 72339001 72340001 72341001 72342001 72343001 72348001 72350001 72351001 72352001 72353001 72354001 72368001 72369001 72370001 72371001 72372001 72383001 72384001 72385001 72386001 72387001 72389001 72390001 72391001 72392001 72393001 72425001 72426001 72427001 72457001 72458001 72459001 72462001 72463001 72464001 72465001 72466001 72467001 72478001 72479001 72480001 72481001 72482001 72494001 72495001 72496001 72497001 72498001 72499001 72500001 72501001 72502001 72503001 72535001 72536001 72537001 72538001 72539001 72540001 72541001 72542001 72543001 72544001 72545001 72546001 72547001 72548001 72553001 72554001 72555001 72562001 72563001 72564001 72565001 72566001 72567001 72568001 72569001 72570001 72571001 72572001 72573001 72597001 72598001 72599001 72600001 72601001 72602001 72605001 72607001 72608001 72609001 72610001 72611001 72612001 72613001 72614001 72620001 72621001 72622001 72626001 72630001 72631001 72632001 72633001 72634001 72635001 72636001 72638001 80064001 80065001 80103001 80104001 80107001 80108001 80115001 80116001 80117001 80118001 80119001 80143001 80144001 80145001 80146001 80158001 80158001 80159001 80160001 80222001 80223001 80224001 80225001 80226001 80227001 80228001 80229001 80230001 80232001 80233001 80234001 80278001 80279001 80280001 80283001 80284001 80285001 80313001 80314001 80315001 80317001 80318001 80319001 80320001 80334001 80335001 80336001 80337001 80339001 80340001 80341001 80342001 80343001 80356001 80357001 80358001 80359001 80360001 80402001 80403001 80404001 80405001 80406001 80463001 80464001 80465001 80469001 80470001 80471001 80505001 80513001 80519001 80566001 80567001 80570001 80571001 80572001 80573001 80574001 80579001 80580001 80587001 80588001 80589001 80590001 80640001 80641001 80678001 80679001 80682001 80683001 80688001 80699001 80701001 80702001 80754001 80755001 80756001 80763001 80764001 80766001 80818001 80821001 80824001 80825001 80872001 80873001 80874001 80875001 80886001 80887001 80888001 80889001 80917001 80918001 80919001 80920001 80921001 80922001 80924001 80925001 80929001 80930001 80931001 80989001 80990001 80991001 80992001 80993001 80994001 81055001 81056001 81057001 81058001 81067001 81068001 81069001 81078001 81079001 81080001 81111001 81118001 81144001 81145001 81146001 81147001 81148001 81149001 81158001 81173001 81174001 81209001 81210001 81211001 81212001 81213001 81214001 81215001 81216001 81232001 81233001 81234001 81270001 81275001 81276001 81277001 81281001 81282001 81283001 81290001 81331001 81335001 81342001 81343001 81344001 81345001 81346001 81347001 81351001 81352001 81389001 81390001 81391001 81392001 81393001 81394001 81395001 81396001 81397001 81398001 81490001 81491001 81492001 81493001 81495001 81496001 81497001 81498001 81499001 81500001 81501001 81502001 81503001 81513001 81514001 81578001 81579001 81580001 81584001 81585001 81589001 81590001 81591001 81593001 81594001 81595001 81641001 81647001 81648001 81649001 81650001 81651001 81690001 81696001 81697001 81698001 81699001 81700001 81701001 81702001 81703001 81704001 81705001 81755001 81756001 81757001 81758001 81759001 81796001 81800001 81801001 81808001 81809001 81810001 81832001 81833001 81834001 81844001 81845001 81846001 81854001 81891001 81892001 81904001 81905001 81906001 81909001 81917001 81942001 81943001 81948001 81949001 81950001 81958001 81959001 81961001 81962001 81963001 81970001 81971001 81972001 82035001 82073001 82074001 82075001 82076001 82078001 82078001 82079001 82079001 82080001 82088001 82089001 82090001 82091001 82192001 82193001 82200001 82218001 82219001 82220001 82223001 82224001 82231001 82232001 82240001 82241001 82242001 82243001 82277001 82278001 82285001 82286001 82287001 82290001 82291001 82292001 82294001 82320001 82321001 82354001 82355001 82371001 82373001 82408001 82409001 82410001 82411001 82412001 82413001 82453001 82465001 82466001 82512001 82513001 82543001 82544001 82545001 82558001 82559001 82589001 82590001 82619001 82620001 82634001 82637001 82682001 82683001 82692001 82728001 82729001 90119001 90120001 90125001 90126001 90127001 90174001 90175001 90176001 90177001 90178001 90179001 90336001 90342001 90343001 90346001 90347001 90383001 90386001 90387001 90392001 90393001 90449001 90452001 90453001 90455001 90456001 90488001 90489001 90490001 90495001 90496001 90559001 90560001 90566001 90567001 90569001 90570001 90588001 90591001 90594001 90638001 90668001 90670001 90672001 90677001 90678001 90685001 90686001 90688001 90719001 90724001 90725001 90726001 90728001 90734001 90796001 90797001 90798001 90800001 90803001 90804001 90805001 90807001 90811001 90812001 90817001 90820001 90825001 90826001 90831001 90832001 90842001 90847001 90896001 90897001 90910001 90920001 91070001 91083001 91093001 91100001 91118001 91152001 91242001 91245001 91246001 91253001 91282001 91288001 91329001 91331001 91335001 91342001 91349001 91363001 91380001 91382001 91394001 91397001 91414001 91431001 91432001 91445001 91447001 91454001 91456001 91470001 91477001 91488001 91489001 91498001 91503001 91506001 91515001 91522001 91536001 91539001 91547001 91555001 91579001 91588001 91599001 91607001 91617001 91621001 91627001 91649001 91650001 91658001 91659001 91677001 91678001 91705001 91741001 91745001 91751001 91766001 91773001 91784001 91791001 91800001 91811001 91814001 91823001 91829001 91831001 91842001 91847001 91849001 91851001 91855001 91865001 91870001 91872001 91875001 91886001 91893001 91895001 91908001 91913001 91921001 91922001 91935001 91936001 91943001 91944001 91955001 91968001 91970001 91971001 91973001 91986001 91989001 92001001 92004001 92022001 92027001 92028001 92038001 92044001 92054001 92057001 92058001 92059001 92066001 92071001 92158001 92188001 92189001 92249001 92255001 92256001 92259001 92279001 92281001 92294001 92300001 92305001 92308001 92310001 92321001 92322001 92323001 92324001 92325001 92326001 92327001 92333001 92336001 92348001 92353001 92362001 92367001 92378001 92407001 92416001 92417001 92421001 92426001 92432001 92439001 92444001 92445001 92453001 92475001 92476001 92480001 92485001 92495001 92502001 92503001 92504001 92513001 92523001 92533001 92534001 92548001 92566001 92574001 92603001 92619001 92660001 92666001 92675001 92683001 92685001 92696001 92701001 92706001 92707001 92739001 92765001 92770001 92777001 92780001 92787001 92788001 92789001 92800001 92862001 92864001 92865001 92873001 92879001 92889001 92890001 92894001 92901001 92909001 92911001 92935001 92940001 92950001 92954001 92961001 92975001 92982001 92989001 130001001 130010001 130012001 130019001 130024001 130029001 130030001 130065001 130067001 130080001 130084001 130086001 130090001 130094001 130096001 130127001 130130001 130138001 130139001 130140001 130141001 130143001 130159001 130160001 130164001 130168001 130183001 130184001 130191001 130192001 130212001 130222001 130226001 130241001 130252001 130273001 130603001 130604001 130611001 130628001 130629001 130639001 130641001 130642001 130644001 130647001 130667001 130682001 130683001 130688001 130713001 130718001 130775001 130776001 130780001 130806001 130816001 130823001 130833001 130835001 130841001 130845001 130857001 130861001 130867001 130870001 130873001 130879001 130906001 130914001 130922001 130936001 130945001 130948001 130960001 130965001 130966001 130971001 130984001 130987001 130988001 131000001 131007001 131018001 131044001 131047001 131049001 131051001 131055001 131058001 131077001 131079001 131080001 131086001 131087001 131089001 131091001 131092001 131109001 131113001 131151001 131160001 131180001 131185001 131193001 131219001 131226001 131237001 131240001 131242001 131250001 131274001 131280001 131281001 131286001 131294001 131314001 131315001 131330001 131351001 131352001 131354001 131356001 131368001 131377001 131378001 131383001 131399001 131409001 131410001 131413001 131421001 131461001 131661001 131662001 131663001 131668001 131675001 131681001 131682001 131694001 131698001 131739001 131750001 131758001 131759001 131761001 131763001 131769001 131776001 131779001 131790001 131798001 131799001 131802001 131805001 131806001 131820001 131826001 131827001 131837001 131841001 131842001 131858001 131859001 131861001 131882001 131885001 131886001 131906001 131922001 131923001 131938001 131943001 131947001 131954001 131969001 131974001 131984001 131997001 132000001 132003001 132005001 132006001 132007001 132010001 132015001 132029001 132039001 132043001 132055001 132064001 132070001 132071001 132084001 132100001 132102001 132121001 132124001 132129001 132139001 132140001 132144001 132156001 132166001 132168001 132187001 132189001 132196001 132207001 132213001 132220001 132228001 132253001 132257001 132269001 132273001 132284001 132310001 132326001 132327001 132358001 132367001 132368001 132383001 132397001 132398001 132403001 132404001 132408001 132411001 132412001 132413001 132432001 132436001 132457001 132459001 132472001 132474001 132493001 132498001 132500001 132501001 132513001 132521001 132525001 132526001 132551001 132557001 132563001 132569001 132586001 132595001 132614001 132619001 132622001 132624001 132642001 132644001 132665001 132675001 132683001 132687001 132688001 132692001 132710001 132714001 132725001 132733001 132734001 132808001 132810001 132812001 132828001 132831001 132835001 132842001 132843001 132856001 132862001 132864001 132865001 132876001 132878001 132886001 132890001 132897001 132906001 132908001 132920001 132923001 132933001 132939001 132945001 132953001 132984001 132985001 132986001 132989001 132990001 132991001 132993001 133005001 133018001 133029001 133031001 133032001 133043001 133056001 133068001 133069001 133072001 133073001 133085001 133089001 133090001 133091001 133105001 133108001 133109001 133118001 133119001 133152001 133159001 133169001 133173001 133186001 133187001 133188001 133194001 133203001 133205001 133213001 133230001 133236001 133253001 133261001 133263001 133264001 133276001 133278001 133292001 133297001 133300001 133304001 133305001 133310001 133314001 133322001 133324001 133326001 133347001 133349001 133351001 133353001 133361001 133366001 133372001 133386001 133391001 133392001 133394001 133397001 133403001 133415001 133417001 133423001 133435001 133460001 133469001 133480001 133484001 133485001 133486001 133493001 133499001 133527001 133552001 133562001 133570001 133581001 133585001 133586001 133600001 133602001 133605001 133608001 133611001 133622001 133624001 133641001 133656001 133657001 133660001 133665001 133666001 133692001 133696001 133712001 133735001 133736001 133742001 133743001 133761001 133782001 133789001 133794001 133800001 133803001 133804001 133816001 133819001 133834001 133835001 133849001 133850001 133908001 133913001 133927001 133932001 133934001 133949001 133950001 133954001 133981001 133997001 133998001 133999001 134000001 134005001 134013001 134023001 134024001 134047001 134061001 134070001 134073001 134074001 134085001 134087001 134088001 134089001 134117001 134119001 134127001 134137001 134140001 134141001 134142001 134149001 134152001 134153001 134154001 134168001 134170001 134171001 134183001 134184001 134186001 134200001 134208001 134209001 134211001 134212001 134217001 134218001 134219001 134235001 134236001 134239001 134243001 134248001 134249001 134266001 134270001 134273001 134274001 134275001 134292001 134337001 134338001 134354001 134355001 134357001 134361001 134365001 134378001 134393001 134394001 134399001 134401001 134404001 134409001 134410001 134425001 134426001 134427001 134446001 134449001 134452001 134458001 134470001 134471001 134496001 134505001 134508001 134510001 134528001 134531001 134533001 134541001 134542001 134548001 134549001 134552001 134553001 134580001 140004001 140015001 140021001 140052001 140054001 140060001 140062001 140075001 140077001 140078001 140079001 140085001 140098001 140099001 140100001 140108001 140123001 140124001 140130001 140132001 140154001 140160001 140161001 140162001 140184001 140209001 140210001 140211001 140214001 140245001 140296001 140305001 140315001 140318001 140341001 140346001 140347001 140349001 140350001 140369001 140373001 140374001 140376001 140412001 140419001 140427001 140430001 140439001 140468001 140479001 140481001 140509001 140512001 140522001 140524001 140526001 140539001 140567001 140570001 140580001 140585001 140615001 140619001 140620001 140624001 140633001 140634001 140636001 140647001 140649001 140666001 140686001 140688001 140717001 140732001 140733001 140738001 140740001 140743001 140756001 140758001 140760001 140761001 140782001 140785001 140787001 140789001 140808001 140811001 140812001 140827001 140828001 140835001 140840001 140845001 140853001 140929001 140934001 140948001 140956001 140961001 140969001 140971001 140974001 140976001 140984001 140985001 140989001 140990001 140991001 140995001 140997001 141012001 141015001 141019001 141020001 141021001 141047001 141050001 141055001 141061001 141068001 141069001 141078001 141089001 141090001 141095001 141096001 141097001 141098001 141099001 141100001 141108001 141117001 141119001 141130001 141134001 141135001 141156001 141162001 141163001 141164001 141168001 141185001 141190001 141215001 141229001 141241001 141246001 141260001 141261001 141309001 141311001 141314001 141334001 141335001 141338001 141339001 141362001 141366001 141368001 141372001 141380001 141387001 141401001 141403001 141412001 141416001 141435001 141455001 141483001 141484001 141497001 141530001 141534001 141536001 141537001 141540001 141541001 141544001 141552001 141557001 141558001 141564001 141565001 141573001 141574001 141579001 141585001 141596001 141599001 141604001 141608001 141613001 141615001 141616001 141617001 141619001 141620001 141621001 141632001 141640001 141641001 141648001 141682001 141687001 141688001 141690001 141698001 141703001 141707001 141709001 141720001 141729001 141736001 141737001 141740001 141741001 141742001 141760001 141763001 141772001 141785001 141823001 141824001 141825001 141845001 141846001 141848001 141849001 141850001 141853001 141878001 141879001 141881001 141887001 141900001 141902001 141904001 141905001 141910001 141913001 141915001 141916001 141917001 141925001 141930001 141931001 141933001 141952001 141954001 141956001 141957001 141963001 141971001 141976001 141978001 141990001 141993001 142000001 142001001 142002001 142007001 142009001 142010001 142011001 142018001 142033001 142035001 142037001 142041001 142061001 142065001 142069001 142071001 142096001 142097001 142099001 142112001 142113001 142114001 142130001 142137001 142139001 142144001 142162001 142164001 142168001 142184001 142188001 142189001 142205001 142209001 142214001 142220001 142230001 142232001 142235001 142265001 142267001 142271001 142279001 142280001 142282001 142283001 142285001 142286001 142287001 142288001 142291001 142307001 142308001 142309001 142322001 142327001 142341001 142345001 142347001 142354001 142356001 142363001 142366001 142367001 142371001 142372001 142381001 142383001 142384001 142389001 142400001 142413001 142415001 142417001 142427001 142428001 142429001 142434001 142436001 142447001 142458001 142460001 142461001 142485001 142492001 142494001 142498001 142506001 142516001 142524001 142584001 142844001 142846001 142854001 142855001 142980001 143019001 143020001 143022001 143025001 143040001 143042001 143052001 143053001 143058001 143059001 143063001 143064001 143069001 143070001 143072001 143073001 143074001 143108001 143109001 143125001 143131001 143135001 143137001 143140001 143153001 143154001 143156001 143157001 143165001 143182001 143185001 143197001 143206001 143209001 143214001 143215001 143228001 143232001 143234001 143243001 143246001 143253001 143259001 143261001 143264001 143265001 143274001 143282001 143283001 143285001 143286001 143297001 143304001 143305001 143306001 143314001 143316001 143324001 143329001 143339001 143340001 143350001 143351001 143362001 143365001 143366001 143367001 143377001 143379001 143381001 143391001 143398001 143399001 143405001 143407001 143416001 143418001 143423001 143434001 143435001 143446001 143447001 143461001 143463001 143465001 143499001 143502001 143504001 143524001 143526001 143527001 143559001 143570001 143572001 143573001 143613001 143614001 143615001 143616001 143621001 143625001 143634001 143644001 143648001 143649001 143651001 143652001 143670001 143697001 143698001 143699001 143708001 143717001 143719001 143721001 143722001 143725001 143727001 143733001 143735001 143736001 143737001 143738001 143739001 143740001 143741001 143742001 143745001 143747001 143760001 143762001 143768001 143771001 143779001 143790001 143794001 143805001 143808001 143809001 143812001 143845001 143846001 143847001 143864001 143884001 143885001 143889001 143892001 143898001 143911001 143914001 143924001 143972001 143976001 143980001 143983001 143992001 143994001 143997001 144012001 144014001 144015001 144019001 144036001 144040001 144045001 144047001 144048001 144066001 144068001 144070001 144077001 144083001 144084001 144085001 144089001 144090001 144097001 144099001 144101001 144103001 144104001 144105001 144107001 144112001 144154001 144165001 144169001 144183001 144184001 144235001 144239001 144355001 144368001 144378001 144382001 144383001 144385001 144404001 144414001 144421001 144440001 144444001 144445001 144462001 144466001 144470001 144488001 144495001 144521001 144526001 144532001 144533001 144535001 144536001 144549001 144550001 144551001 144566001 144574001 144575001 144576001 144583001 144595001 144599001 144601001 144602001 144612001 144613001 144624001 144636001 144638001 144646001 144660001 144661001 144680001 144683001 144688001 144689001 144690001 144691001 144692001 144707001 144711001 144721001 144722001 144732001 144734001 144744001 144747001 144750001 144755001 144756001 144757001 144758001 144779001 144780001 144782001 144784001 144790001 144811001 144812001 144818001 144819001 144821001 144835001 144836001 144838001 144841001 144845001 144850001 144854001 144855001 144856001 144864001 144865001 144867001 144880001 144889001 144890001 144911001 144914001 144915001 144921001 144925001 144926001 144928001 144929001 144930001 144931001 144932001 144933001 150014001 150016001 150023001 150029001 150044001 150045001 150054001 150070001 150073001 150086001 150098001 150107001 150115001 150127001 150139001 150142001 150143001 150147001 150149001 150151001 150171001 150175001 150177001 150189001 150196001 150208001 150211001 150221001 150235001 150247001 150254001 150259001 150263001 150297001 150300001 150303001 150307001 150317001 150328001 150333001 150336001 150345001 150347001 150350001 150367001 150370001 150373001 150381001 150408001 150414001 150434001 150437001 150445001 150464001 150466001 150473001 150478001 150483001 150485001 150486001 150488001 150508001 150518001 150519001 150520001 150531001 150542001 150543001 150544001 150545001 150556001 150572001 150579001 150584001 150592001 150596001 150597001 150609001 150610001 150613001 150618001 150630001 150640001 150643001 150655001 150657001 150659001 150661001 150666001 150668001 150672001 150680001 150688001 150694001 150713001 150714001 150715001 150716001 150729001 150740001 150743001 150744001 150756001 150762001 150766001 150784001 150793001 150803001 150808001 150814001 150815001 150816001 150817001 150818001 150830001 150835001 150837001 150838001 150839001 150840001 150854001 150858001 150863001 150866001 150868001 150873001 150874001 150875001 150876001 150884001 150885001 150891001 150894001 150895001 150898001 150911001 150922001 150924001 150926001 150938001 150939001 150940001 150941001 150942001 150943001 150970001 150972001 150973001 150974001 150975001 150976001 150977001 150986001 150987001 150988001 150989001 150991001 151010001 151012001 151031001 151033001 151040001 151044001 151049001 151064001 151066001 151067001 151068001 151070001 151076001 151082001 151083001 151090001 151114001 151119001 151120001 151128001 151133001 151138001 151139001 151140001 151141001 151142001 151143001 151144001 151153001 151154001 151166001 151167001 151172001 151177001 151195001 151200001 151204001 151206001 151210001 151217001 151219001 151222001 151223001 151224001 151232001 151233001 151234001 151236001 151239001 151241001 151242001 151243001 151245001 151246001 151247001 151248001 151249001 151250001 151251001 151263001 151264001 151268001 151270001 151275001 151281001 151291001 151295001 151298001 151303001 151310001 151320001 151321001 151326001 151338001 151339001 151345001 151355001 151356001 151357001 151358001 151359001 151360001 151370001 151371001 151372001 151373001 151382001 151384001 151386001 151402001 151403001 151404001 151405001 151406001 151413001 151421001 151431001 151436001 151437001 151444001 151454001 151455001 151459001 151460001 151462001 151466001 151467001 151471001 151474001 151524001 151525001 151548001 151549001 151555001 151559001 151562001 151569001 151573001 151574001 151578001 151579001 151581001 151582001 151583001 151585001 151593001 151594001 151595001 151599001 151611001 151612001 151616001 151620001 151635001 151642001 151643001 151712001 151727001 151735001 151738001 151739001 151743001 151744001 151745001 151747001 151760001 151761001 151763001 151770001 151771001 151781001 151784001 151786001 151790001 151791001 151792001 151793001 151797001 151798001 151803001 151804001 151809001 151810001 151811001 151814001 151816001 151824001 151827001 151839001 151844001 151845001 151846001 151847001 151848001 151863001 151864001 151865001 151866001 151867001 151868001 151869001 151896001 151898001 151899001 151900001 151901001 151902001 151903001 151912001 151913001 151914001 151915001 151916001 151917001 151932001 151933001 151934001 151935001 151936001 151937001 151953001 151955001 151961001 151962001 151964001 151965001 151966001 151967001 151968001 151969001 151997001 151998001 151999001 152000001 152001001 152002001 152016001 152017001 152018001 152019001 152020001 152021001 152030001 152031001 152032001 152033001 152034001 152035001 152057001 152058001 152059001 152060001 152061001 152093001 152094001 152095001 152096001 152097001 152098001 152109001 152110001 152111001 152112001 152114001 152129001 152130001 152131001 152132001 152133001 152134001 152143001 152144001 152145001 152162001 152163001 152165001 152167001 152178001 152180001 152181001 152182001 152183001 152184001 152185001 152206001 152207001 152208001 152209001 152210001 152211001 152212001 152247001 152250001 152253001 152254001 152256001 152257001 152258001 152259001 152260001 152261001 152266001 152267001 152268001 152269001 152270001 152271001 152272001 152273001 152309001 152310001 152322001 152324001 152325001 152326001 152338001 152339001 152340001 152341001 152342001 152343001 152344001 152362001 152373001 152374001 152375001 152376001 152377001 152378001 152389001 152390001 152391001 152392001 152393001 152394001 152395001 152435001 152436001 152437001 152438001 152439001 152440001 152457001 152458001 152459001 152460001 152476001 152477001 152478001 152479001 152480001 152481001 152482001 152514001 152515001 152516001 152517001 152518001 152519001 152520001 152532001 152533001 152534001 152535001 152536001 152537001 152538001 152563001 152564001 152565001 152593001 152594001 152602001 152603001 152604001 152605001 152619001 152621001 152622001 152623001 152647001 152648001 152649001 152650001 152651001 152652001 152653001 152687001 152696001 152697001 152698001 152713001 152714001 152720001 152721001 152723001 152726001 152754001 152757001 152758001 152759001 152762001 152769001 152787001 152789001 152794001 152795001 152796001 152802001 152803001 152804001 152805001 152806001 152807001 152824001 152841001 152842001 152843001 152857001 152859001 152860001 152861001 152862001 152863001 152864001 152888001 152889001 152890001 152891001 152892001 152893001 152894001 152895001 152957001 152958001 152961001 152963001 152966001 152967001 152968001 152971001 152973001 152995001 152997001 152998001 153001001 153002001 153003001 153004001 153012001 153013001 153014001 153015001 153016001 153017001 153018001 153024001 153039001 153040001 153041001 153042001 153043001 153063001 153073001 153082001 153085001 153086001 153087001 153088001 153089001 153090001 153091001 153092001 153117001 153146001 153147001 153149001 153150001 153151001 153152001 153153001 153174001 153198001 153200001 153210001 153214001 153222001 153223001 153236001 153237001 153245001 153245001 153247001 153251001 153253001 153255001 153256001 153272001 153274001 153275001 153276001 153281001 153293001 153308001 153312001 153313001 153319001 153321001 153322001 153363001 153364001 153368001 153369001 153370001 153381001 153382001 153383001 153385001 153387001 153394001 153408001 153427001 153430001 153432001 153433001 153434001 153451001 153455001 153500001 153505001 153506001 153519001 153523001 153525001 153527001 153531001 153543001 153544001 153549001 153557001 153558001 153594001 153596001 153597001 153599001 153600001 153601001 153602001 153618001 153619001 153621001 153624001 153625001 153626001 153639001 153653001 153654001 153655001 153658001 153659001 153662001 153677001 153679001 153681001 153682001 153683001 153684001 153691001 153692001 153693001 153694001 153707001 153713001 153715001 153716001 153718001 153720001 153749001 153755001 153759001 153760001 153765001 153786001 153790001 153793001 153796001 153798001 153799001 153815001 153825001 153829001 153831001 153839001 153840001 153841001 153842001 153846001 153847001 153848001 153893001 153895001 153896001 153898001 153900001 153901001 153916001 153922001 153923001 153924001 153925001 153949001 153950001 153964001 153967001 153968001 153970001 153971001 153973001 153974001 153976001 153979001 153986001 153995001 153996001 153998001 154006001 154023001 154024001 154026001 154033001 154034001 154035001 154036001 154049001 154050001 154054001 154055001 154059001 154069001 154071001 154072001 154074001 154076001 154093001 154101001 154102001 154109001 154113001 154114001 154125001 154127001 154128001 154138001 154139001 154140001 154172001 154176001 154177001 154178001 154196001 154197001 154199001 154202001 154220001 154221001 154223001 154233001 154251001 154253001 154254001 154262001 154263001 154264001 154265001 154266001 154267001 154303001 154304001 154307001 154310001 154317001 154328001 154330001 154334001 154337001 154372001 154381001 154382001 154388001 154389001 154390001 154398001 154411001 154413001 154419001 154420001 154432001 154435001 154444001 154445001 154447001 154449001 154450001 154465001 154466001 154467001 154470001 154471001 154472001 154483001 154486001 154487001 154488001 154489001 154526001 154531001 154532001 154535001 154537001 154538001 154547001 154561001 154563001 154567001 154571001 154573001 154582001 154584001 154593001 154594001 154595001 154596001 154597001 154598001 154599001 154605001 154606001 154615001 154630001 154631001 154632001 154634001 154637001 154651001 154656001 154657001 154669001 154670001 154681001 155380001 155383001 155387001 155396001 155397001 155401001 155426001 155427001 155430001 155432001 155434001 155435001 155455001 155457001 155467001 155473001 155483001 155485001 155494001 155502001 155507001 155508001 155586001 155587001 155588001 155589001 155591001 155595001 155596001 155605001 155606001 155607001 155613001 155615001 155616001 155618001 155636001 155637001 155638001 155641001 155643001 155654001 155655001 155664001 155668001 155671001 155674001 155675001 155679001 155680001 155682001 155693001 155705001 155706001 155707001 155714001 160017001 160018001 160020001 160021001 160031001 160036001 160047001 160049001 160055001 160056001 160058001 160059001 160065001 160068001 160076001 160101001 160107001 160111001 160123001 160124001 160125001 160127001 160128001 160129001 160159001 160160001 160163001 160165001 160166001 160168001 160173001 160174001 160184001 160185001 160187001 160281001 160283001 160285001 160290001 160294001 160297001 160298001 160305001 160306001 160309001 160314001 160319001 160323001 160326001 160327001 160328001 160333001 160336001 160337001 160338001 160339001 160349001 160366001 160371001 160372001 160380001 160383001 160387001 160393001 160399001 160401001 160402001 160406001 160407001 160410001 160417001 160426001 160436001 160439001 160447001 160462001 160463001 160471001 160472001 160480001 160481001 160482001 160501001 160510001 160516001 160519001 160523001 160526001 160529001 160545001 160546001 160556001 160563001 160576001 160579001 160582001 160584001 160586001 160601001 160602001 160603001 160613001 160620001 160621001 160623001 160625001 160626001 160627001 160628001 160639001 160640001 160641001 160673001 160678001 160683001 160687001 160689001 160702001 160705001 160706001 160707001 160708001 160709001 160710001 160721001 160729001 160738001 160741001 160746001 160750001 160751001 160753001 160766001 160767001 160772001 160781001 160788001 160789001 160791001 160799001 160801001 160804001 160812001 160813001 160814001 160815001 160836001 160842001 160843001 160845001 160854001 160863001 160864001 160866001 160874001 160875001 160876001 160880001 160887001 160888001 160891001 160896001 160897001 160900001 160905001 160926001 160927001 160928001 160931001 160932001 160935001 160968001 160970001 160973001 160974001 160975001 160977001 160978001 160994001 161009001 161012001 161013001 161014001 161019001 161022001 161027001 161033001 161039001 161041001 161042001 161051001 161067001 161069001 161070001 161071001 161083001 161094001 161096001 161113001 161114001 161124001 161140001 161141001 161143001 161145001 161146001 161157001 161168001 161169001 161170001 161177001 161178001 161179001 161194001 161195001 161196001 161237001 161239001 161266001 161270001 161279001 161288001 161296001 161297001 161298001 161301001 161314001 161325001 161328001 161332001 161344001 161347001 161350001 161366001 161522001 161525001 161535001 161539001 161540001 161543001 161545001 161548001 161553001 161562001 161563001 161564001 161568001 161576001 161578001 161579001 161581001 161588001 161589001 161602001 161603001 161606001 161607001 161613001 161622001 161627001 161641001 161642001 161648001 161664001 161665001 161666001 161670001 161678001 161688001 161689001 161695001 161696001 161697001 161699001 161712001 161715001 161717001 161720001 161735001 161752001 161757001 161761001 161764001 161783001 161790001 161791001 161794001 161798001 161807001 161812001 161813001 161814001 161815001 161828001 161831001 161840001 161841001 161845001 161853001 161854001 161857001 161861001 161863001 161864001 161865001 161866001 161867001 161868001 161872001 161873001 161874001 161877001 161887001 161888001 161894001 161900001 161908001 161909001 161910001 161913001 161914001 161915001 161928001 161934001 161935001 161944001 161948001 161961001 161966001 161969001 161973001 161993001 161994001 161996001 161998001 162009001 162011001 162019001 162021001 162024001 162026001 162030001 162031001 162032001 162034001 162053001 162066001 162067001 162071001 162076001 162077001 162082001 162093001 162101001 162102001 162103001 162104001 162114001 162116001 162123001 162124001 162125001 162130001 162149001 162151001 162152001 162155001 162165001 162176001 162185001 162186001 162187001 162192001 162203001 162204001 162205001 162206001 162209001 162212001 162226001 162228001 162239001 162244001 162247001 162264001 162267001 162268001 162270001 162343001 162345001 162347001 162348001 162349001 162361001 162365001 162366001 162372001 162373001 162379001 162380001 162388001 162389001 162391001 162392001 162394001 162400001 162408001 162409001 162417001 162418001 162421001 162422001 162437001 162438001 162442001 162455001 162456001 162462001 162476001 162477001 162478001 162479001 162481001 162490001 162500001 162501001 162502001 162504001 162507001 162514001 162713001 162714001 162716001 162794001 162811001 162820001 162838001 162844001 162892001 162910001 162914001 162922001 162929001 162939001 162951001 163185001 163186001 163198001 163200001 163209001 163223001 163227001 163230001 163235001 163246001 163247001 163250001 163258001 163259001 163265001 163266001 163269001 163276001 163286001 163291001 163292001 163296001 163298001 163305001 163306001 163314001 163315001 163318001 163323001 163324001 163336001 163338001 163348001 163355001 163356001 163359001 163360001 163368001 163379001 163380001 163382001 163387001 163395001 163399001 163406001 163414001 163416001 163421001 163422001 163424001 163426001 163430001 163431001 163432001 163452001 163453001 163454001 163468001 163470001 163473001 163474001 163475001 163509001 163514001 163516001 163517001 163519001 163520001 163565001 163579001 163581001 163594001 163597001 163600001 163611001 163612001 163615001 163626001 163632001 163634001 163643001 163648001 163654001 163656001 163674001 163680001 163685001 163696001 163699001 163701001 163705001 163707001 163717001 163718001 163720001 163731001 163754001 163757001 163759001 163779001 163781001 163784001 163785001 163806001 163810001 163811001 163813001 163823001 163824001 163830001 163832001 163838001 163840001 163853001 163853001 163857001 163858001 163860001 163866001 163868001 163874001 163879001 163880001 163881001 163883001 163892001 163899001 163900001 163903001 163906001 163914001 163919001 163922001 163924001 163934001 163943001 163956001 163958001 163969001 163973001 163980001 163982001 163986001 164002001 164004001 164005001 164011001 164018001 164019001 164025001 164032001 164033001 164038001 164039001 164045001 164058001 164059001 164067001 164074001 164081001 164085001 164089001 164090001 164093001 164096001 164106001 164107001 164118001 164126001 164129001 164131001 164132001 164133001 164151001 164161001 164168001 164171001 164174001 164182001 164183001 164185001 164194001 164208001 164212001 164215001 164218001 164219001 164236001 164242001 164246001 164247001 164254001 164266001 164267001 164269001 164275001 164288001 164292001 164295001 164297001 164305001 164306001 164312001 164318001 164320001 164323001 164333001 164334001 164335001 164337001 164338001 164340001 164347001 164355001 164360001 164362001 164364001 164370001 164372001 164377001 164381001 164385001 164388001 164425001 164437001 164441001 164447001 164458001 164465001 164466001 164468001 164482001 164484001 164491001 164495001 164521001 164642001 164643001 164651001 164662001 164663001 164666001 164667001 164668001 164684001 164692001 164695001 164704001 164706001 164718001 164734001 164743001 164764001 164770001 164800001 164809001 164810001 164825001 164828001 164831001 164852001 164856001 164857001 164862001 164867001 164875001 164876001 164879001 164889001 164893001 165055001 165056001 165057001 165058001 165059001 165060001 165061001 165062001 165063001 165064001 170001001 170002001 170003001 170004001 170005001 170006001 170007001 170008001 170009001 170010001 170011001 170012001 170013001 170014001 170015001 170016001 170017001 170018001 170019001 170020001 170021001 170022001 170023001 170024001 170025001 170026001 170027001 170034001 170040001 170042001 170060001 170064001 170082001 170083001 170088001 170093001 170111001 170112001 170116001 170118001 170121001 170137001 170140001 170150001 170151001 170161001 170163001 170164001 170165001 170166001 170227001 170232001 170233001 170239001 170275001 170282001 170283001 170294001 170305001 170306001 170307001 170309001 170310001 170320001 170401001 170406001 170407001 170439001 170440001 170444001 170445001 170448001 170464001 170466001 170468001 170472001 170473001 170474001 170478001 170497001 170499001 170502001 170503001 170508001 170541001 170543001 170545001 170565001 170570001 170580001 170581001 170587001 170588001 170604001 170605001 170607001 170615001 170616001 170617001 170632001 170636001 170637001 170639001 170648001 170653001 170662001 170673001 170674001 170690001 170691001 170706001 170713001 170724001 170725001 170726001 170727001 170728001 170730001 170740001 170749001 170750001 170757001 170761001 170762001 170764001 170783001 170787001 170788001 170790001 170796001 170800001 170806001 170808001 170814001 170816001 170818001 170819001 170827001 170838001 170845001 170846001 170847001 170848001 170857001 170864001 170865001 170868001 170872001 170880001 170882001 170889001 170890001 170895001 170897001 170898001 170912001 170916001 170921001 170927001 170928001 170938001 170942001 170944001 170945001 170946001 170948001 170960001 170961001 170964001 170965001 170966001 170976001 170979001 170981001 170982001 170984001 170991001 170996001 170999001 171001001 171011001 171014001 171015001 171017001 171022001 171029001 171033001 171034001 171035001 171040001 171045001 171047001 171049001 171050001 171051001 171052001 171054001 171056001 171057001 171063001 171067001 171070001 171079001 171081001 171087001 171088001 171090001 171091001 171092001 171099001 171110001 171116001 171118001 171123001 171124001 171137001 171140001 171144001 171147001 171148001 171154001 171162001 171170001 171171001 171188001 171189001 171205001 171206001 171213001 171214001 171215001 171216001 171218001 171219001 171226001 171234001 171235001 171239001 171243001 171248001 171249001 171250001 171259001 171268001 171270001 171271001 171272001 171278001 171285001 171300001 171301001 171304001 171315001 171324001 171330001 171331001 171333001 171334001 171348001 171350001 171352001 171353001 171356001 171359001 171360001 171366001 171371001 171373001 171381001 171384001 171391001 171399001 171404001 171416001 171417001 171418001 171419001 171426001 171427001 171444001 171445001 171446001 171447001 171451001 171463001 171469001 171473001 171474001 171481001 171493001 171498001 171500001 171503001 171505001 171508001 171517001 171527001 171542001 171561001 171562001 171569001 171572001 171573001 171574001 171578001 171587001 171590001 171591001 171610001 171635001 171638001 171639001 171644001 171647001 171655001 171657001 171669001 171676001 171682001 171683001 171685001 171690001 171705001 171706001 171712001 171714001 171716001 171721001 171735001 171736001 171738001 171740001 171746001 171753001 171759001 171761001 171770001 171774001 171785001 171788001 171790001 171807001 171808001 171821001 171822001 171824001 171826001 171832001 171834001 171842001 171849001 171850001 171874001 171882001 171887001 171905001 171907001 171910001 171912001 171913001 171924001 171925001 171926001 171930001 171931001 171952001 171959001 171960001 171961001 171975001 171981001 171982001 171985001 171986001 171987001 172016001 172018001 172023001 172026001 172028001 172032001 172033001 172034001 172035001 172036001 172038001 172039001 172040001 172041001 172048001 172056001 172058001 172061001 172064001 172074001 172078001 172089001 172095001 172099001 172101001 172102001 172110001 172111001 172113001 172118001 172119001 172138001 172142001 172143001 172146001 172147001 172149001 172150001 172151001 172152001 172161001 172178001 172179001 172180001 172182001 172184001 172194001 172197001 172201001 172208001 172211001 172216001 172227001 172232001 172233001 172234001 172239001 172246001 172257001 172259001 172260001 172264001 172267001 172268001 172269001 172270001 172271001 172273001 172301001 172302001 172305001 172310001 172314001 172330001 172333001 172334001 172339001 172343001 172344001 172345001 172346001 172347001 172348001 172349001 172360001 172361001 172362001 172362001 172363001 172364001 172365001 172366001 172395001 172396001 172397001 172398001 172399001 172400001 172424001 172425001 172426001 172427001 172428001 172430001 172431001 172432001 172450001 172453001 172454001 172456001 172459001 172475001 172497001 172498001 172499001 172501001 172502001 172505001 172520001 172521001 172522001 172524001 172525001 172526001 172527001 172548001 172549001 172550001 172551001 172552001 172557001 172558001 172559001 172560001 172561001 172562001 172563001 172609001 172613001 172615001 172619001 172623001 172624001 172625001 172639001 172640001 172641001 172642001 172643001 172644001 172646001 172647001 172648001 172649001 172674001 172675001 172676001 172677001 172678001 172679001 172680001 172714001 172715001 172716001 172717001 172718001 172728001 172729001 172730001 172731001 172732001 172733001 172734001 172749001 172750001 172751001 172752001 172756001 172757001 172758001 172759001 172760001 172761001 172786001 172787001 172789001 172791001 172792001 172793001 172794001 172795001 172796001 172797001 172798001 172875001 172878001 172879001 172883001 172884001 172885001 172887001 172888001 172889001 172890001 172891001 172898001 172899001 172903001 172904001 172905001 172907001 172909001 172910001 172911001 172912001 172913001 172914001 172915001 172938001 172941001 172942001 172943001 172944001 172945001 172946001 172947001 172948001 172950001 172966001 172967001 172968001 172969001 172970001 172971001 172972001 172973001 172974001 172975001 173001001 173002001 173012001 173015001 173022001 173042001 173043001 173044001 173045001 173046001 173057001 173058001 173059001 173060001 173061001 173062001 173063001 173064001 173065001 173066001 173120001 173122001 173123001 173134001 173135001 173136001 173137001 173138001 173139001 173140001 173141001 173162001 173177001 173178001 173182001 173184001 173185001 173186001 173187001 173188001 173189001 173190001 173191001 173192001 173193001 173226001 173232001 173233001 173234001 173235001 173236001 173237001 173246001 173248001 173250001 173254001 173256001 173257001 173271001 173272001 173283001 173284001 173288001 173294001 173297001 173298001 173301001 173302001 173303001 173304001 173305001 173306001 173307001 173315001 173316001 173317001 173318001 173319001 173320001 173321001 173322001 173323001 173324001 173325001 173326001 173379001 173381001 173383001 173403001 173404001 173405001 173409001 173411001 173414001 173417001 173418001 173427001 173429001 173430001 173431001 173432001 173433001 173434001 173442001 173443001 173444001 173445001 173446001 173447001 173448001 173449001 173450001 173463001 173486001 173487001 173488001 173489001 173490001 173491001 173492001 173493001 173494001 173495001 173496001 173497001 173498001 173499001 173511001 173513001 173543001 173544001 173546001 173551001 173553001 173554001 173574001 173576001 173577001 173580001 173581001 173582001 173583001 173598001 173599001 173600001 173601001 173602001 173603001 173604001 173605001 173606001 173607001 173608001 173609001 173610001 173626001 173629001 173630001 173631001 173632001 173633001 173634001 173651001 173652001 173653001 173654001 173655001 173656001 173657001 173658001 173659001 173660001 173662001 173663001 173664001 173665001 173666001 173696001 173697001 173700001 173701001 173702001 173703001 173704001 173705001 173706001 173720001 173721001 173722001 173723001 173724001 173725001 173760001 173763001 173772001 173782001 173786001 173788001 173789001 173790001 173794001 173809001 173811001 173813001 173821001 173822001 173825001 173843001 173844001 173849001 173851001 173852001 173863001 173864001 173868001 173869001 173870001 173871001 173872001 173880001 173884001 173885001 173886001 173887001 173888001 173889001 173890001 173891001 173892001 173893001 173894001 173917001 173918001 173919001 173920001 173921001 173922001 173923001 173924001 173925001 173926001 173945001 173947001 173951001 173957001 173958001 173959001 173960001 173961001 173962001 173963001 173965001 173966001 173967001 173990001 173991001 173992001 173993001 173994001 173995001 173996001 173997001 173998001 173999001 174000001 174001001 174019001 174020001 174021001 174022001 174029001 174031001 174032001 174033001 174034001 174035001 174036001 174037001 174038001 174076001 174078001 174079001 174080001 174081001 174082001 174083001 174084001 174085001 174086001 174087001 174098001 174099001 174100001 174101001 174103001 174104001 174105001 174106001 174107001 174109001 174135001 174136001 174137001 174138001 174139001 174140001 174141001 174142001 174144001 174145001 174217001 174219001 174220001 174352001 174401001 174402001 174407001 174411001 174412001 174416001 174420001 174421001 174422001 174429001 174457001 174501001 174509001 174511001 174520001 174527001 174530001 174534001 174561001 174564001 174579001 174580001 174581001 174591001 174592001 174601001 174611001 174614001 174621001 174622001 174643001 174656001 174665001 174678001 174679001 174697001 174702001 174709001 174710001 174716001 174721001 174803001 174836001 174845001 174846001 174848001 174868001 174869001 174870001 174871001 174874001 174875001 174876001 174877001 174892001 174894001 174898001 174901001 174902001 174918001 174921001 174925001 174937001 174940001 174941001 174953001 174976001 174979001 174980001 174982001 175002001 175003001 175007001 175010001 175013001 175021001 175025001 175026001 175027001 175029001 175040001 175042001 175043001 175045001 175067001 175068001 175070001 175071001 175074001 175075001 175078001 175082001 175093001 175096001 175097001 175098001 175110001 175117001 175118001 175120001 175122001 175123001 175124001 175125001 175126001 175127001 175128001 175152001 175153001 175157001 175160001 175165001 175166001 175167001 175168001 175188001 175193001 175199001 175202001 175204001 175205001 175206001 175207001 175208001 175226001 175233001 175234001 175235001 175236001 175237001 175244001 175245001 175260001 175261001 175262001 175263001 175264001 175271001 175272001 175274001 175282001 175283001 175284001 175289001 175290001 175291001 175293001 175294001 175300001 175318001 175319001 175321001 175322001 175324001 175325001 175326001 175334001 175336001 175337001 175338001 175339001 175340001 175341001 175342001 175343001 175348001 175350001 175351001 175367001 175380001 175383001 175384001 175385001 175386001 175387001 175388001 175389001 175406001 175407001 175408001 175409001 175410001 175412001 175414001 175415001 175416001 175417001 175452001 175453001 175454001 175462001 175463001 175464001 175471001 175475001 175476001 175484001 175485001 175500001 175501001 175506001 175512001 175513001 175514001 175515001 175516001 175530001 175531001 175532001 175543001 175552001 175553001 175574001 175575001 175576001 175577001 175578001 175599001 175605001 175619001 175621001 175636001 175638001 175646001 175651001 175652001 175658001 175665001 175666001 175672001 175674001 175682001 175686001 175687001 175690001 175691001 175692001 175700001 175705001 175706001 175711001 175718001 175719001 175720001 175729001 175731001 175732001 175733001 175749001 175750001 175760001 175763001 175764001 175765001 175766001 175767001 175768001 175790001 175791001 175792001 175804001 175805001 175811001 175812001 175824001 175825001 175834001 175835001 175836001 175837001 175838001 175840001 175841001 175846001 175848001 175849001 175851001 175852001 175853001 175854001 175857001 175858001 175862001 175863001 175866001 175867001 175868001 175871001 175873001 175874001 175875001 175877001 175888001 175889001 175890001 175891001 175892001 175893001 175894001 175895001 175909001 175911001 175918001 175920001 175924001 175936001 175937001 175938001 175939001 175940001 175946001 175947001 175948001 175949001 175955001 175956001 175957001 175958001 175971001 175972001 175973001 175974001 175975001 175976001 175988001 175990001 175992001 175993001 175994001 175997001 175998001 175999001 176000001 176001001 176002001 176010001 176011001 176012001 176013001 176014001 176015001 176016001 176017001 176018001 176019001 176020001 176021001 176046001 176047001 176048001 176049001 176050001 176051001 176052001 176053001 176054001 176055001 176056001 176057001 176077001 176078001 176079001 176080001 176081001 176082001 176083001 176084001 176094001 176095001 176096001 176097001 176098001 176100001 176102001 176103001 176107001 176119001 176120001 176121001 176131001 176132001 176133001 176134001 176135001 176136001 176137001 176150001 176152001 176153001 176154001 176163001 176168001 176169001 176172001 176173001 176174001 176175001 176176001 176177001 176178001 176180001 176186001 176187001 176189001 176195001 176196001 176198001 176201001 176202001 176205001 176207001 176215001 176216001 176218001 176223001 176228001 176232001 176233001 176234001 176235001 176236001 176239001 176240001 176244001 176245001 176246001 176253001 176255001 176264001 176267001 176268001 176270001 176275001 176281001 176282001 176283001 176284001 176285001 176290001 176292001 176293001 176301001 176303001 176304001 176305001 176307001 176308001 176313001 176315001 176316001 176319001 176322001 176324001 176325001 176332001 176333001 176362001 176379001 176396001 176397001 176398001 176409001 176427001 176429001 176432001 176434001 176438001 176439001 176448001 176449001 176451001 176460001 176461001 176462001 176464001 176476001 176486001 176487001 176489001 176491001 176495001 176501001 176535001 176536001 176537001 176540001 176541001 176543001 176546001 176547001 176548001 176549001 176551001 176552001 176553001 176560001 176562001 176563001 176568001 176569001 176570001 176571001 176572001 176576001 176582001 176586001 176588001 176589001 176590001 176591001 176592001 176593001 176594001 180003001 180006001 180017001 180021001 180030001 180032001 180041001 180049001 180052001 180054001 180055001 180056001 180057001 180064001 180065001 180076001 180084001 180089001 180093001 180094001 180095001 180097001 180099001 180108001 180109001 180116001 180118001 180119001 180120001 180126001 180127001 180128001 180133001 180134001 180135001 180142001 180143001 180144001 180145001 180156001 180157001 180166001 180167001 180168001 180169001 180171001 180175001 180176001 180178001 180194001 180195001 180197001 180198001 180205001 180206001 180230001 180231001 180240001 180242001 180248001 180249001 180251001 180252001 180253001 180254001 180255001 180287001 180290001 180291001 180293001 180294001 180295001 180298001 180300001 180307001 180308001 180309001 180317001 180320001 180323001 180332001 180334001 180335001 180338001 180344001 180345001 180346001 180347001 180348001 180354001 180355001 180356001 180357001 180361001 180367001 180375001 180376001 180377001 180378001 180380001 180381001 180382001 180384001 180388001 180392001 180394001 180395001 180397001 180405001 180407001 180411001 180421001 180422001 180423001 180423001 180426001 180428001 180430001 180434001 180436001 180444001 180445001 180446001 180447001 180449001 180470001 180471001 180472001 180473001 180481001 180483001 180486001 180496001 180498001 180499001 180503001 180505001 180506001 180507001 180517001 180518001 180520001 180521001 180522001 180526001 180527001 180531001 180532001 180544001 180545001 180546001 180548001 180550001 180558001 180559001 180561001 180568001 180569001 180570001 180573001 180575001 180576001 180577001 180578001 180579001 180588001 180595001 180596001 180597001 180598001 180611001 180615001 180616001 180617001 180618001 180631001 180632001 180633001 180640001 180641001 180644001 180647001 180659001 180666001 180674001 180675001 180678001 180679001 180680001 180682001 180684001 180685001 180686001 180688001 180689001 180690001 180692001 180693001 180698001 180708001 180710001 180711001 180715001 180716001 180717001 180725001 180727001 180730001 180732001 180736001 180737001 180740001 180751001 180752001 180753001 180754001 180755001 180764001 180769001 180770001 180771001 180772001 180775001 180777001 180779001 180789001 180792001 180801001 180808001 180809001 180811001 180812001 180813001 180814001 180816001 180817001 180818001 180821001 180833001 180835001 180836001 180844001 180848001 180850001 180854001 180858001 180859001 180860001 180862001 180863001 180865001 180875001 180877001 180878001 180879001 180889001 180890001 180891001 180892001 180894001 180895001 180906001 180907001 180914001 180916001 180917001 180926001 180927001 180928001 180931001 180932001 180941001 180942001 180943001 180946001 180947001 180949001 180950001 180951001 180967001 180969001 180978001 180983001 180984001 180985001 180986001 180990001 180991001 180992001 180993001 180999001 181000001 181001001 181002001 181003001 181015001 181016001 181020001 181034001 181035001 181036001 181037001 181038001 181039001 181047001 181053001 181055001 181056001 181058001 181060001 181061001 181067001 181068001 181076001 181077001 181078001 181082001 181083001 181084001 181086001 181087001 181093001 181095001 181096001 181099001 181115001 181116001 181123001 181125001 181126001 181127001 181128001 181131001 181136001 181143001 181144001 181145001 181146001 181149001 181150001 181151001 181152001 181159001 181169001 181171001 181172001 181178001 181179001 181182001 181183001 181186001 181188001 181189001 181190001 181191001 181193001 181194001 181208001 181215001 181216001 181229001 181230001 181231001 181233001 181234001 181238001 181246001 181247001 181250001 181251001 181258001 181260001 181261001 181262001 181263001 181264001 181265001 181267001 181270001 181271001 181273001 181274001 181275001 181277001 181278001 181279001 181280001 181282001 181283001 181296001 181298001 181300001 181302001 181304001 181310001 181311001 181312001 181313001 181314001 181316001 181317001 181330001 181332001 181333001 181334001 181335001 181336001 181338001 181340001 181341001 181342001 181343001 181344001 181345001 181346001 181347001 181358001 181359001 181361001 181363001 181367001 181368001 181370001 181371001 181373001 181381001 181387001 181388001 181391001 181393001 181407001 181409001 181410001 181415001 181416001 181417001 181418001 181420001 181421001 181422001 181423001 181424001 181425001 181426001 181427001 181428001 181429001 181436001 181448001 181454001 181455001 181457001 181458001 181462001 181464001 181465001 181466001 181467001 181468001 181469001 181470001 181471001 181472001 181473001 181474001 181476001 181483001 181484001 181487001 181491001 181492001 181499001 181502001 181503001 181504001 181505001 181506001 181507001 181508001 181509001 181522001 181524001 181525001 181529001 181530001 181531001 181532001 181543001 181544001 181545001 181546001 181547001 181548001 181556001 181557001 181562001 181563001 181568001 181583001 181584001 181585001 181586001 181588001 181593001 181615001 181616001 181620001 181621001 181635001 181636001 181637001 181638001 181650001 181651001 181653001 181663001 181665001 181667001 181668001 181669001 181670001 181671001 181676001 181678001 181688001 181689001 181690001 181704001 181710001 181712001 181719001 181720001 181726001 181739001 181743001 181745001 181746001 181747001 181748001 181749001 181752001 181753001 181754001 181760001 181762001 181763001 181764001 181765001 181772001 181773001 181780001 181781001 181785001 181786001 181787001 181788001 181789001 181790001 181796001 181798001 181799001 181800001 181801001 181803001 181806001 181807001 181809001 181820001 181823001 181825001 181826001 181832001 181833001 181834001 181835001 181837001 181845001 181848001 181849001 181851001 181867001 181868001 181869001 181870001 181881001 181885001 181886001 181887001 181888001 181889001 181892001 181897001 181898001 181899001 181901001 181912001 181914001 181915001 181916001 181917001 181928001 181929001 181930001 181931001 181937001 181938001 181940001 181941001 181942001 181943001 181952001 181957001 181958001 181961001 181962001 181963001 181972001 181973001 181974001 181975001 181976001 181982001 181987001 181991001 181993001 181998001 181999001 182001001 182005001 182007001 182008001 182009001 182016001 182019001 182037001 182038001 182041001 182042001 182043001 182048001 182049001 182050001 182054001 182063001 182064001 182067001 182071001 182072001 182080001 182081001 182082001 182084001 182092001 182099001 182100001 182110001 182111001 182112001 182113001 182115001 182116001 182117001 182125001 182126001 182127001 182130001 182131001 182136001 182149001 182150001 182152001 182156001 182157001 182158001 182159001 182160001 182161001 182167001 182171001 182174001 182177001 182180001 182181001 182182001 182184001 182186001 182187001 182188001 182189001 182190001 182193001 182206001 182208001 182211001 182214001 182215001 182216001 182220001 182221001 182222001 182223001 182224001 182225001 182243001 182244001 182248001 182252001 182253001 182254001 182257001 182258001 182268001 182277001 182278001 182279001 182280001 182281001 182282001 182283001 182284001 182285001 182286001 182287001 182288001 182290001 182295001 182300001 182303001 182304001 182305001 182306001 182307001 182308001 182309001 182311001 182312001 182319001 182332001 182333001 182338001 182339001 182340001 182351001 182361001 182367001 182368001 182370001 182371001 182372001 182377001 182378001 182379001 182380001 182381001 182382001 182404001 182408001 182409001 182410001 182411001 182412001 182414001 182415001 182420001 182422001 182428001 182429001 182430001 182431001 182439001 182440001 182441001 182442001 182443001 182444001 182445001 182446001 182457001 182458001 182464001 182466001 182467001 182473001 182496001 182503001 182516001 182517001 182527001 182528001 182547001 218880001 176588001  174030001 174143001 176265001 180485001 180942001 180962001 181037001 181095001 181096001 181099001 181115001 181123001 181125001 181126001 181127001 181131001 181143001 181149001 181150001 181151001 181159001 181179001 181189001 181216001 181230001 181233001 181234001 181247001 181261001 181262001 181265001 181267001 181270001 181271001 181273001 181274001 181275001 181277001 181278001 181279001 181280001 181282001 181283001 181296001 181298001 181300001 181302001 181304001 181310001 181311001 181312001 181313001 181314001 181316001 181317001 181330001 181332001 181333001 181334001 181335001 181336001 181338001 181340001 181341001 181342001 181344001 181345001 181346001 181347001 181358001 181359001 181367001 181368001 181370001 181371001 181373001 181408001 181409001 181424001 181426001 181427001 181428001 181436001 181455001 181469001 181487001 181492001 181499001 181502001 181503001 181505001 181506001 181507001 181508001 181509001 181522001 181524001 181525001 181529001 181530001 181532001 181544001 181568001 181583001 181588001 181615001 181635001 181636001 181637001 181638001 181650001 181651001 181653001 181663001 181665001 181667001 181668001 181669001 181670001 181671001 181676001 181678001 181688001 181689001 181690001 181704001 181710001 181712001 181719001 181720001 181726001 181739001 181745001 181746001 181747001 181748001 181753001 181754001 181760001 181762001 181763001 181764001 181772001 181773001 181780001 181781001 181801001 181803001 181807001 181809001 181820001 181825001 181826001 181835001 181837001 181845001 181848001 181851001 181867001 181868001 181869001 181870001 181881001 181885001 181886001 181887001 181888001 181889001 181892001 181897001 181898001 181899001 181901001 181912001 181914001 181915001 181916001 181917001 181928001 181929001 181930001 181937001 181938001 181940001 181941001 181942001 181943001 181952001 181958001 181961001 181962001 181963001 181972001 181973001 181974001 181975001 181976001 181982001 181987001 181991001 181992001 181993001 181998001 181999001 182001001 182005001 182007001 182008001 182019001 182028001 182037001 182038001 182041001 182043001 182048001 182049001 182050001 182054001 182063001 182064001 182067001 182071001 182072001 182080001 182081001 182082001 182084001 182092001 182099001 182100001 182111001 182112001 182113001 182115001 182116001 182117001 182125001 182126001 182127001 182130001 182131001 182136001 182149001 182150001 182152001 182156001 182157001 182158001 182159001 182160001 182161001 182167001 182171001 182174001 182177001 182180001 182181001 182182001 182184001 182186001 182187001 182188001 182189001 182190001 182193001 182206001 182208001 182211001 182214001 182215001 182216001 182220001 182221001 182222001 182223001 182224001 182225001 182243001 182244001 182248001 182252001 182253001 182254001 182257001 182258001 182268001 182277001 182278001 182279001 182280001 182281001 182282001 182283001 182284001 182285001 182286001 182287001 182288001 182290001 182295001 182300001 182303001 182304001 182305001 182306001 182307001 182308001 182309001 182311001 182312001 182319001 182332001 182333001 182338001 182339001 182340001 182351001 182361001 182367001 182368001 182370001 182371001 182372001 182377001 182378001 182379001 182380001 182381001 182382001 182389001 182404001 182408001 182409001 182410001 182411001 182412001 182414001 182415001 182420001 182422001 182428001 182429001 182430001 182431001 182439001 182440001 182441001 182443001 182444001 182445001 182446001 182457001 182458001 182464001 182466001 182467001 182468001 182471001 182473001 182474001 182476001 182548001 182557001 182562001 182564001 182567001 182586001 182594001 182595001 182597001 182602001 182614001 182622001 182623001 182624001 182626001 182627001 182629001 182637001 182647001 182653001 182654001 182662001 182663001 182666001 182668001 182669001 182683001 182688001 182692001 182693001 182694001 182697001 182709001 182715001 182716001 182719001 182728001 182729001 182730001 182731001 182733001 182734001 182735001 182737001 182740001 182747001 182748001 182749001 182750001 182753001 182754001 182755001 182756001 182757001 182758001 182759001 182760001 182769001 182784001 182787001 182788001 182792001 182815001 182816001 182817001 182824001 182828001 182830001 182833001 182834001 182845001 182847001 182848001 182849001 182852001 182853001 182864001 182866001 182867001 182870001 182889001 182912001 182975001 183019001 183071001 183078001 183081001 183083001 183093001 183094001 183095001 183101001 183102001 183104001 183105001 183137001 183138001 183140001 183192001 183218001 183224001 183238001 183240001 183241001 183242001 183243001 183244001 183254001 183255001 183260001 183263001 183267001 183268001 183273001 183275001 183276001 183277001 183278001 183288001 183298001 183304001 183306001 183309001 183313001 183335001 183354001 183372001 183376001 183377001 183382001 183384001 183398001 183399001 183402001 183403001 183437001 183555001 183556001 183557001 183561001 183570001 183571001 183572001 183573001 183574001 183582001 183585001 183587001 183610001 183625001 183626001 183644001 183673001 183699001 183706001 183713001 183721001 183722001 183728001 183730001 183733001 183734001 183740001 183753001 183784001 183785001 183786001 183794001 183798001 183818001 183844001 183845001 183847001 183848001 183851001 183861001 183872001 183873001 183895001 183896001 183909001 183917001 183919001 183920001 183921001 183975001 183976001 183991001 184013001 184014001 184015001 184022001 184023001 184024001 184025001 184028001 184029001 184030001 184033001 184036001 184147001 184160001 184163001 184165001 184187001 184209001 184211001 184213001 184214001 184215001 184220001 184222001 184226001 184227001 184230001 184231001 184233001 184234001 184235001 184264001 184265001 184268001 184274001 184276001 184278001 184279001 184300001 184301001 184309001 184310001 184311001 184312001 184313001 184314001 184321001 184322001 184323001 184328001 184329001 184330001 184331001 184332001 184333001 184336001 184342001 184352001 184353001 184355001 184356001 184357001 184359001 184361001 184362001 184373001 184374001 184378001 184387001 184388001 184390001 184399001 184402001 184412001 184413001 184416001 184417001 184418001 184419001 184422001 184432001 184433001 184455001 184456001 184513001 184514001 184522001 184523001 184532001 184534001 184535001 184536001 184537001 184538001 184539001 184557001 184566001 184603001 184604001 184605001 184606001 184607001 184624001 184627001 184628001 184629001 184630001 184631001 184632001 184639001 184640001 184641001 184643001 184675001 184685001 184686001 184688001 184689001 184690001 184695001 184696001 184698001 184783001 184786001 184790001 184792001 184799001 184800001 184813001 184827001 184828001 184832001 184834001 184843001 184855001 184856001 184857001 184858001 184859001 184880001 184883001 184886001 184901001 184903001 184904001 184930001 184931001 184932001 184944001 184970001 184981001 184990001 185012001 185014001 185052001 185054001 185057001 185078001  
Recalling Firm/
Manufacturer
Megadyne Medical Products, Inc.
11506 S State St
Draper UT 84020-9453
For Additional Information Contact
801-576-9669
Manufacturer Reason
for Recall
If two devices are plugged into the same channel/port in the generator, both devices activate. If one device is left laying on the patient, the result could be burning of the patient.
FDA Determined
Cause 2
Use error
ActionThe firm sent letters in June 2018, informing their customers of the issue. The firm will send labels that can be applied to the product to remind them not to plug 2 active devices into the same channel at a time. Updated 08/13/19 Starting on 06/29/18, updated Medical Device Correction Notifications and labels were mailed via second day service. Customers were asked to complete and return Business Reply Forms.
Quantity in Commerce8380 units
DistributionWorldwide distribution. US Nationwide, Mexico, Europe, Middle East, Latin America, Africa, and Asia Pacific Markets
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = GEI
-
-