| Class 2 Device Recall Elekta Unity | |
Date Initiated by Firm | March 12, 2019 |
Create Date | April 08, 2019 |
Recall Status1 |
Terminated 3 on November 03, 2023 |
Recall Number | Z-1109-2019 |
Recall Event ID |
82384 |
510(K)Number | K182076 |
Product Classification |
Accelerator, linear, medical - Product Code IYE
|
Product | Elekta Unity, Image-Guided Radiation Therapy System
Product Usage:
Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems. |
Code Information |
Unity Systems: 13553-006/600003, 13371-002/600021, 10817-011/600010, 10562-011/600016, 10719-T03001/600013, 30004070-001/600023, 10420-AVL-U/600008, 11014-UMCU-U/600007, 10157-006/600014, 11611-40/600011, 30001461-J001/600002, 11489-21/600009, |
Recalling Firm/ Manufacturer |
Elekta Limited Linac House Fleming Way Crawley United Kingdom
|
Manufacturer Reason for Recall | Users need to be aware when using these protocols for daily on line plan adaptation that:
1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration.
2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering.
Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan. |
FDA Determined Cause 2 | Software design |
Action | The firm issued a field safety notice on 03/12/2019 by email. The notice explained the issue and cautioned the consignee to select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan. |
Quantity in Commerce | 12 units |
Distribution | Worldwide Distribution - US Nationwide in the states of TX, NY, WI, and Denmark, Germany, Italy, Netherlands, Sweden, United Kingdom |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = IYE
|
|
|
|