Class 1 Device Recall MoniTorr

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall MoniTorr see related information
Date Initiated by Firm April 08, 2019
Date Posted May 20, 2019
Recall Status1 Terminated 3 on January 27, 2021
Recall Number Z-1311-2019
Recall Event ID 82659
510(K)Number K022554  K062599  
Product Classification Shunt, central nervous system and components - Product Code JXG
Product MoniTorr INS1100 INS-1100 CSF Drainage System used with Pole Mount System

Product Usage: The MoniTorr ICP External CSF Drainage and Monitoring Systems provide a simple to use, closed system for the drainage of cerebrospinal fluid (CSF) from ventricles of the brain or the lumbar subarachnoid space to a drainage bag. The system may be used with a pole mounted assembly that allows for simple, quick and accurate alignment with the patient and secure positive or negative pressure level setting. The system also has been designed to provide for ease of patient transport through a compact design and antimicrobial hydrophobic vent feature that resists occlusion.
Indications: The MoniTorr ICP system allows for drainage and monitoring of CSF from the lateral ventricles of the brain and the lumbar subarachnoid space in selected patients to reduce intracranial pressure (ICP), to monitor CSF, to provide temporary drainage of CSF inpatients with infected CSF shunts, and to monitor ICP.
Code Information All lots
Recalling Firm/
Manufacturer
Integra LifeSciences Corp.
311 Enterprise Dr
Plainsboro NJ 08536-3344
For Additional Information Contact Dr. Patricia Kihn
609-275-0500
Manufacturer Reason
for Recall
The firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount stopcock and MoniTorr Panel Mount stopcock.
FDA Determined
Cause 2
Device Design
Action On April 8, 2019, the firm distributed Urgent Medical Device Correction letters to its affected customers. The communication advised customers of the possible breakage of the stopcock. Customers were advised to take the following precautions: When connecting a fluid-filled transducer (FFT) to the Panel Mount/Transducer Mount stopcocks: " Do not apply excessive force while attaching the FFT to the stopcock. " Take care during the priming process to not torque or bend the syringe and FFT. " Ensure luer lock is "finger tight;" do not over-tighten connection at the stopcock junction. When managing an external ventricular or lumbar drain (EVD) with a FFT attached: " Take care when manipulating the stopcock to zero the FFT and to obtain pressure readings, do not apply excessive force to the stopcock junction. " Use caution during patient transportation and during routine ICU care, to avoid direct contact of the FFT with other devices or equipment. " If the FFT must be flushed or replaced, do not apply excessive torque or force to the FFT/ stopcock junction. " If the FFT is connected to the ICU bedside monitor, ensure the interface cable is positioned so that it does not place excess force on the FFT (e.Q. do not allow cable to touch the floor. Customers were asked to determine if they had any affected product on hand, review and understand the precautions, and complete and return a completed acknowledgement form to the firm. Distributors were asked to forward the customer letter to any customers who received the affected product. Should you have any questions regarding these instructions, please contact Customer Service Monday to Friday 8:00 AM - 8:00 PM EST USA: 1-800-654-2873: email: custsvcnj@integralife.com Outside USA: 781-565-1401 ***UPDATE*** On October 22, 2019, the firm sent out an updated notification informing customers that a new stopcock has been designed for the LimiTorr and MoniTorr, and is now available for replacement of th
Quantity in Commerce 42102 units
Distribution Worldwide Distribution - US Nationwide International distribution to Argentina, American Samoa, Australia, Canada, and Guam.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JXG and Original Applicant = INTEGRA LIFESCIENCES CORPORATION
510(K)s with Product Code = JXG and Original Applicant = INTEGRA NEUROSCIENCES
-
-