  Class 1 Device Recall Sheridan/CF Novaplus see related information
Date Initiated by Firm May 24, 2019
Create Date June 19, 2019
Recall Status1 Open3, Classified
Recall Number Z-1765-2019
Recall Event ID 82964
510(K)Number K822082  
Product Classification Tube, tracheal (w/wo connector) - Product Code BTR
Product Sheridan/CF Novaplus 7.0 mm, Product Code V5-10114

Product Usage:
Tracheal tube/airway management
Code Information Lot Numbers: 73K1600156 73K1600735 73L1600501 73A1700120 73B1700321 73C1700425 73D1700630 73E1700504 73G1700146 73H1700053 73J1700307 73K1700085 73B1800223 73B1800371 73C1800372 73D1800655 73E1800579 73G1800627 73H1800203 73H1800429 73H1800758 73J1800118 73J1800696  
Recalling Firm/
Manufacturer
Teleflex Medical
3015 Carrington Mill Blvd
Morrisville NC 27560-5437
Manufacturer Reason
for Recall
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
FDA Determined
Cause 2
Under Investigation by firm
Action Teleflex sent an Urgent Medical Device Recall letter dated May 24, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use, quarantine the product, and contact Teleflex for return of the product. Distributors were directed to notify their customers and request the customer return the recalled products to them for consolidation and subsequent return to Teleflex.. For questions contact your local sales representative or Customer Service at 1-866-396-2111
Quantity in Commerce 77136 units
Distribution Worldwide - US Nationwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database 510(K)s with Product Code = BTR and Original Applicant = SHERIDAN CATHETER CORP.
-
-