  Class 2 Device Recall SIGNA Vibrant see related information
Date Initiated by Firm December 23, 2019
Date Posted January 28, 2020
Recall Status1 Terminated 3 on December 11, 2020
Recall Number Z-0877-2020
Recall Event ID 84655
510(K)Number K121676  K052293  
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
Product SIGNA Vibrant Nuclear Magnetic Resonance Imaging System, 1.5T Signa HDx, 1.5T Signa HDxt, Signa HDi - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Code Information Mfg. Lot or Serial # System ID 22D-4-0868 901BMHCRITMR HDGOC1900010SC 209521SGMR HDGOC1900015SC 4577711MR11 002776WHF 954916MR1 022D-4-1175 808243MR 857 LONGWOOD30MR2 00000290815MR0 508580BMR3 002801WHF 734246MR 002930WHF 845279MR 00000269040MR2 412749MR1 022D-4-1125 801RIVMR2 HDGOC1900003SC 206386SMR H05114 4815915MR11 00000277443MR8 304526SMR1 To be provided GON4796599 To be provided 424881MR01 To be provided 083027829345912 To be provided 083027862100408 
Recalling Firm/
Manufacturer
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information Contact
800-437-1171
Manufacturer Reason
for Recall
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
FDA Determined
Cause 2
Software design
Action This issue will be resolved by sending an Urgent Medical Device Correction letter to Medical Facility staff (Director of Clinical/Radiology, Risk Manager/Hospital Administrator, Director of Biomedical Engineering) with affected systems. This letter will inform the customer of the issue and instruct the user to ensure the displayed system date is correct. The long-term solution is to correct affected systems with updated software.
Quantity in Commerce 62 (30 US, 32 OUS) total devices
Distribution Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.