Class 2 Device Recall MicroVention Terumo Microplex 18, Cosmos 18 |
|
Date Initiated by Firm |
November 22, 2019 |
Create Date |
February 27, 2020 |
Recall Status1 |
Terminated 3 on January 26, 2022 |
Recall Number |
Z-1399-2020 |
Recall Event ID |
84727 |
510(K)Number |
K153594 K132952 K050954 K090891 K131948
|
Product Classification |
Device, neurovascular embolization - Product Code HCG
|
Product |
MicroVention Terumo Microplex 18, Platinum Embolization Coil, Cosmos 18, Sterile, Rx, REF numbers 180619CS-V, 180723CS-V, 180827CS-V, 180931CS-V, 181036CS-V, 181139CS-V, 181243CS-V, 181347CS-V, 181451CS-V, 181652CS-V, 181859CS-V, 182065CS-V, 182263CS-V, and 182468CS-V;
Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The firm name on the label is MicroVention, Inc., Tustin, CA. |
Code Information |
Catalog/lot numbers: 180619CS-V - 1903065W1, 1904015P1, 1904225W1, 1904295W1, 1905085P1, 1905085W1, 1905175W1, and 1905205W1; 180723CS-V - 1903085W1, 1903185W1, 1904015P1, 1904175P1, 1904295W1, 1905085W1, and 1905175W1; 180827CS-V - 1903205W1, 1904015P1, 1905155W1, and 1905205W1; 180931CS-V - 1904225R1, 1903205W1, 1905175W1, and 1905225W1; 181036CS-V - 1904035R1, 1903085W1, 1903185W1, 1904225W1, 1904265W1, 1905035W1, 1905065W1, 1905175W1, and 1905205W1; 181139CS-V - 1903015L1, 1903015W1, 1903085P1, 1903085W1, 1903185W1, 1904015P1, 1904245P1, and 1904245W1; 181243CS-V - 1903015L1, 1903015W1, 1903065W1, 1903185W1, 1904015P1, 1904175P1, 1904225W1, 1904265W1, 1904295W1, 1905155W1, 1905175W1, and 1905225W1; 181347CS-V - 1903085V1, 1903135R1, 1904035R1, 1903015W1, 1903085W1, 1903205P1, 1903205W1, and 1904225W1; 181451CS-V - 1903085W1, 1903225W1, 1904015P1, 1904175P1, 1904225W1, 1904295W1, 1905105W1, and 1905135W1; 181652CS-V - 1905135R1, 1903015W1, 1903045W1, 1903085W1, 1903205W1, 1903225W1, 1904105W1, 1905105W1, and 1905225W1; 181859CS-V - 1903155P1, 1903155W1, 1903185H1, and 1905105W1; 182065CS-V - 1903045L1, 1904085Q1, 1903155W1, 1903185H1, and 1903205H1; 182263CS-V - 1904035R1; and 182468CS-V - 1903015L1, 1903135R1, 1904055R1, 1905205R1, 1904295H1, 1905025G1, 1905025H1, 1905105H1, 1905105W1, 1905135H1, and 1905205H1. |
Recalling Firm/ Manufacturer |
Microvention, Inc. 35 Enterprise Aliso Viejo CA 92656-2601
|
For Additional Information Contact |
Dr. Irina Kulinets 714-247-8159
|
Manufacturer Reason for Recall |
The devices may be missing the implant coil.
|
FDA Determined Cause 2 |
Process design |
Action |
The recalling firm issued letters beginning 11/22/2019 dated 11/26/2019 via email using different flags according to the region the foreign consignee was located, i.e. 'URGENT: FIELD SAFETY CORRECTION ACTION" (for EMEA), "URGENT: MEDICAL DEVICE RECALL" (for Japan), or "FIELD SAFETY ADVISORY NOTICE" (for Rest of World). |
Quantity in Commerce |
403 units |
Distribution |
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = HCG and Original Applicant = MICROVENTION, INC.
|
|
|
|