• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall LinearTM 34

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall LinearTM 34see related information
Date Initiated by FirmApril 24, 2019
Create DateMay 26, 2020
Recall Status1 Terminated 3 on January 18, 2022
Recall NumberZ-2121-2020
Recall Event ID 84983
PMA NumberP030017 
Product Classification Stimulator, spinal-cord, totally implanted for pain relief - Product Code LGW
ProductLinearTM 3-4 50 cm 8 Contact Lead Kit, part number M365SC2352500. Leads function as a component of Boston Scientific's Spinal Cord Stimulation (SCS) systems by delivering electrical stimulation to the nerve structures in the dorsal aspect of the spinal cord, resulting in an inhibition of pain sensation. Package Contents Percutaneous Permanent Lead Kits (1) Percutaneous Lead with pre-loaded Curved Stylet (1) Stylet Ring with a Curved and a Straight Stylet (4) Suture Sleeves (1) Insertion Needle with Stylet (1) Lead Blank (1) Steering Cap (1) OR Cable Assembly (2) Lead Position Labels left and right (non-sterile) (1) Device Registration Form/Temporary Patient Identification Card (1) Manual
Code Information Part number M365SC2352500; serial numbers 1039736,  1039779, and 1042340.
FEI Number 3006630150
Recalling Firm/
Manufacturer
Boston Scientific Neuromodulation Corporation
25155 Rye Canyon Loop
Valencia CA 91355-5004
For Additional Information ContactNicole Pshon
661-949-4000
Manufacturer Reason
for Recall
The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.
FDA Determined
Cause 2
Labeling Change Control
ActionCustomers to the user level are being notified of the recall via letter sent by DHL courier beginning on April 24, 2019. A third-party recalling company is not being used. The Urgent Medical Device Product Removal Letter instructs customers to immediately cease further distribution or use of any remaining product affected by this removal. Customers who identify any affected products within their inventory are instructed to segregate them immediately and return the product(s) to the recalling firm in accordance with enclosed instructions. Customers will receive replacements for all removed products returned to the manufacturer. Customers who are distributors are requested to notify any of their customers who have received the affected product(s). Recipients of the recall notice are requested to pass the notice to any healthcare professional from their organization who needs to be aware and to any organization where the potentially affected devices have been transferred, if appropriate. In addition, it is requested that customers provide the recalling firm with details of any affected devices that have been transferred to other organizations. Customers are requested to email or fax the Reply Verification Tracking Form to the BSC Field Action Center to BCSFieldActionCenter@bsci.com or by fax to 1-866-213-1806. Return of the RVTF is requested even if customers do not have any affected product to return.
Quantity in Commerce3 units
DistributionDistributed OUS only. to 14 countries.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
PMA DatabasePMAs with Product Code = LGW
-
-