• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Intravascular Administration Set

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Intravascular Administration Set see related information
Date Initiated by Firm October 04, 2019
Create Date May 06, 2020
Recall Status1 Open3, Classified
Recall Number Z-1913-2020
Recall Event ID 85053
510(K)Number K964435  
Product Classification Stopcock, i.V. Set - Product Code FMG
Product REF: B90240-50, 96" (244 cm) Appx 11.2 ml, 10 Drop Admin Set w/Coiled Tubing, Pre-Slit Port, Clave, Rotating Luer, Sterile R, UDI: (01)00887709077058
Code Information Lot #: 4168837, 4169606, 4168835
Recalling Firm/
Manufacturer
ICU Medical, Inc.
951 Calle Amanecer
San Clemente CA 92673-6212
For Additional Information Contact Customer Service
1866-829-9025 Ext. 8
Manufacturer Reason
for Recall
There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.
FDA Determined
Cause 2
Mixed-up of materials/components
Action On October 4, 2019, a "Urgent: Medical Device Recall" Letter was sent to all affected customers via US Mail. In addition to providing customers information on the recalled products, the letter asked customers to do the following: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) Return affected product using the return label provided with this letter. Contact Stericycle at 1-866-367-4703 (Monday - Friday, 8:00 am-5:00 pm ET) if you have not received a return label or require additional labels for returning the affected product. The return labels are for single use only. Please do not reproduce. Please visit http://expertezlabel.com to request additional labels for returning affected product. To ensure proper and timely credit, follow the instructions on the return label for returning product. Upon receipt of the completed response form and return of the affected product, ICU Medical will credit you for any product returned. You will only receive credit for product that you return. NOTE: Credits for product purchased through distributor will be credited by the distributor. 4) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product / Table 1 sections of this notification and ask them to contact Stericycle at 1-866-367-4703 (M-F, 8am-5pm ET) to obtain a response form. 5) Replacement product is available. Please contact customer service using the information provided below. - Global Complaint Management - 1-844-654-7780 or ProductComplaintsPP@icumed.com, To report adverse events or product complaints - C
Quantity in Commerce 14,000 units
Distribution US:AK,AL,AR,AZ,CA,CO,CT,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,SD,TN,TX,UT,VA,WA,WI,WY OUS: Canada *** "REF" numbers AH7117 and B30211 were not marketed or distributed within the US***
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database 510(K)s with Product Code = FMG and Original Applicant = ICU MEDICAL, INC.
-
-