Date Initiated by Firm | April 06, 2020 |
Date Posted | May 06, 2020 |
Recall Status1 |
Terminated 3 on September 16, 2022 |
Recall Number | Z-1922-2020 |
Recall Event ID |
85513 |
510(K)Number | K060937 |
Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
Product | Philips Brilliance iCT SP Computed Tomography X-Ray System; Model # 728311 |
Code Information |
Serial # 200044 Additional unit as off 5/22/20: 200026 |
Recalling Firm/ Manufacturer |
Philips North America, LLC 3000 Minuteman Rd Andover MA 01810-1032
|
For Additional Information Contact | Ms. Holly Wright Lee 440-869-4807 |
Manufacturer Reason for Recall | The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems. |
FDA Determined Cause 2 | Process control |
Action | Philips sent a Medical Device Correction-Urgent Field Safety Notice dated April 6, 2020 to all affected customers. The firm instructed customers to review the information with all members of the staff who need to be aware of the contents of the notice and to maintain a copy with the equipment in the Instructions for Use. Customers were directed to complete and return the form stating that they received, read, and understood the contents of the letter and have taken appropriate actions.
URGENT - Field Safety Notice Medical Device Correction letter dated 5/20/20 was sent to customers informing them of two additional affected systems.
There is no action required by the customer. Testing and engineering analysis demonstrate that the under-torqued bolts do not pose any immediate safety risk as the mechanical connection between the cooling unit and the rotor continues to meet applicable safety standards. The affected systems may continue to be used according to the IFU.
Philips is distributing this FSN to all affected customers/users and will correct the issue via Field Change Order FCO 72800725.
Urgent Field Safety Notice Medical Device Correction dated 5/20/20 was distributed to notify additional systems not previously included in the original notification.
There is no action required by the customer. Testing and engineering analysis demonstrate that the under-torqued bolts do not pose any immediate safety risk as the mechanical connection between the cooling unit and the rotor continues to meet applicable safety standards. The affected systems may continue to be used according to the IFU.
Philips is distributing this FSN to all affected customers/users and will correct the issue via Field Change Order FCO 72800725.
If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the
Customer Care Solutions Center (1-800-722-9377, follow the prompts). |
Quantity in Commerce | 2 |
Distribution | Worldwide distribution including one unit distributed to Iowa. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JAK
|