 Class 1 Device Recall AC3 Series IABP Systemsee related information
Date Initiated by FirmMay 20, 2020
Date PostedJune 26, 2020
Recall Status1 Terminated 3 on January 07, 2023
Recall NumberZ-2417-2020
Recall Event ID 85684
510(K)NumberK162820 
Product Classification System, balloon, intra-aortic and control - Product Code DSP
ProductArrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/AJLA, Product Code IAP-0701
Code Information Serial Number Ranges:  AC3 Optimus IABP NA/AJLA IAP-0701 170831F through 170840F  171011F through 171020F  171121F through 171125F  180106F through 180110F  180126F through 180130F  180426F through 180430F  180606F through 180610F  180731F through 180740F  180826F through 180830F  180901F through 180910F  180912F through 180921F  181001F through 181010F  190136F through 190140F  190336F through 190340F  190646F through 190650F  190731F through 190736F  190931F through 190936F  191019F through 191024F  191055F through 191060F  200107F through 200112F  200308F through 200311F 
FEI Number 3015859709
Recalling Firm/
Manufacturer
Arrow International Inc
2400 Bernville Rd
Reading PA 19605-9607
For Additional Information Contact
610-378-0131
Manufacturer Reason
for Recall
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
FDA Determined
Cause 2
Component design/selection
ActionUrgent Medical Device Correction notification letters dated 5/20/20 were sent to customers. ACTIONS TO BE TAKEN BY FACILITIES (Immediate): Please immediately check your inventory of Arrow AutoCAT2 and Arrow AC3 Optimus IABPs, whether stored or in use, and determine if you have an IABP with a model number listed above. Please see Appendix 1 for product codes and serial numbers. If an IABP in the scope of this Medical Device Correction displays a System Error 3 or High Baseline alarm, now or at any point in the future until the Long-Term Corrective Action detailed below occurs, immediately quarantine the device and contact Teleflex at 1-855-419-8507 or recalls@teleflex.com to receive support for inspection and servicing of the impacted device. We recommend against using an IABP for ground or air transport between medical facilities pending implementation of the Customer Action (Long Term). For IABP use during ground or air transport prior to implementation of the Customer Action (Long Term), it is recommended that the risks and benefits of using the IABP be assessed by the medical team treating the patient and that alternative circulatory support devices be considered. Customer Action (Subsequent): Teleflex will contact each impacted facility to schedule inspection and servicing of all IABP units that contain this internal component. This inspection and servicing will be performed onsite at facilities, for all IABP units. Facilities should adhere to the following instructions pending implementation of the Teleflex Long-Term Corrective Action described later in this letter: 1. Ensure that a backup IABP is available as instructed within the Operator Manual. If no such replacement IABP is immediately available, it is recommended that the risks and benefits of using the IABP be assessed by the medical team treating the patient and that alternative circulatory support devices be considered. 2. IABP units should be closely monitored during delivery of IAB
Quantity in Commerce2123 total
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DSP
-
-