 Class 2 Device Recall Biomet ARTHROSIMPLICITY KIT EU Item Number: 244055see related information
Date Initiated by FirmSeptember 04, 2020
Create DateOctober 08, 2020
Recall Status1 Terminated 3 on August 31, 2021
Recall NumberZ-0084-2021
Recall Event ID 86424
Product Classification unknown device name - Product Code N/A
ProductBiomet ARTHROSIMPLICITY KIT EU Item Number: 24-4055
Code Information Lot Numbers: 894660A 894660B 894670B 894700B
FEI Number 1825034
Recalling Firm/
Manufacturer
Biomet, Inc.
56 E Bell Dr
Warsaw IN 46582-6989
For Additional Information Contact411 Technical Services
574-371-3071
Manufacturer Reason
for Recall
Product not properly being aligned with the adequate gamma sterilization group and result in the product not being properly sterilized
FDA Determined
Cause 2
Process change control
ActionZimmer Biomet notified on 9/2/20 Distributors with direct sales and/or fielded inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory, as well as hospitals that have consumed products in the past 12 months will be notified via FedEx. Surgeons that have consumed products with in the past 12 months notified via FedEx. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1  Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1  Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you ha
Quantity in Commerce4 units OUS
DistributionNationwide Foreign: Canada, FINLAND, NETHERLANDS, UNITED KINGDOM

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-