| | Class 2 Device Recall Philips Respironics V60 Ventilator |  |
| Date Initiated by Firm | November 02, 2020 |
| Date Posted | January 22, 2021 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0902-2021 |
| Recall Event ID |
86697 |
| 510(K)Number | K082660 |
| Product Classification |
Ventilator, continuous, minimal ventilatory support, facility use - Product Code MNT
|
| Product | Philips Respironics V60 Ventilator Part Number 1053613 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories. |
| Code Information |
Model: V60
Part Number 1053613
V60 Ventilator,Intl Opt: CFLEX,AVAPS,PP
Serial numbers:
100029460
100086827
100106522
100125955
100173096
100173294
100173478
100173526
100173591
100173593
100173595
100174299
100174305
100174308
100190730
100252781
100252782
100252785
100304095
100304097
100304263
100304264
100304446
100304447
100304448
100304451
100304462
201022963
100190253
201024196
201025984
201025987
201025988
100176847
100176856
100252776
100315180
100315377
100315598
100315602
100315607
100315610
100315889
100350701
100361426
100363749
100370895
100375246
100044066
100074200
100074206
100074207
100085241
100085243
100085327
100102145
100102150
100102152
100102155
100114031
100118553
100118676
100118684
100119491
100120774
100138585
100138589
100146847
100146849
100147395
201008175
201008180
201008192
201008218
201008225
201008226
201013645
201013649
201013650
201014018
201014041
201014042
201014047
201031153
201031164
201031195
201031200
201031214
201031215
201031218
100039855
100039944
100040070
100045713
100045728
100046139
100064525
100064561
100064565
100064566
100064717
100071592
100071593
100081759
100082460
100083215
100086784
100086887
100087022
100091384
100091385
100101055
100106443
100106507
100106523
100114030
100125962
100136656
100137203
100180416
100181591
100181614
100189483
100189758
100229116
100230021
100252575
100252788
201004061
201004067
201004363
201004374
201004375
201004687
100087603
100129448
100129779
100129782
100129783
100129784
100129785
100173086
100173089
100173090
100173095
100173097
100173219
100173241
100173246
100173252
100173597
100243515
100243631
100243632
100243640
100243912
201007865
201008232
201011112
100118546
100118694
100260182
100260183
201013158
100189466
201029733
100036150
100036155
100067450
100107006
100160829
100190082
100219300
201026534
100026664
100026678
100029057
100029058
100029066
100029068
100029075
100029077
100029078
100029222
100029240
100037555
100045715
100045723
100059156
100084841
100084845
100084847
100084849
100084850
100085135
100085240
100086454
100114020
100137503
100146859
100181615
100181741
100190804
100229633
201002728
201004021
201004051
201004055
201004056
201004057
201004058
201004060
201004063
201004064
201008173
201029651
201029729
201029762
201029807
100116680
100116681
100116697
100116830
100116831
100116832
100118079
100118080
100118264
100124863
100124968
100147573
201008060
100026672
100026686
100026755
100026760
100029335
100031914
100031916
100037540
100037545
100037546
100037550
100037558
100038202
100038212
100038214
100045670
100045676
100045678
100045681
100045682
100045711
100045714
100045717
100045719
100045721
100045722
100045724
100045727
100045731
100046143
100073255
100085247
100091381
100091383
100104055
100104065
100106520
100114024
100118693
100137022
100137481
100147282
100147563
100174990
100181612
100230020
100237374
100237623
100237625
100237626
100237628
100237629
100237630
100237632
100237635
100237638
100261124
201001547
201001568
201001570
201001576
201002273
201002274
201002276
201002394
201002401
201002576
201002579
201002729
201002740
201002741
201002744
201002746
201002747
201002748
201002749
201002750
201002751
201002754
100038208
100045726
100118695
100125750
201001573
201004062
201004371
100037539
100046136
100046138
100046140
100046142
100046152
100086773
100086886
100091389
100098477
100100310
100103663
100106511
100106517
100118557
100118697
100163749
100164087
100176028
100181587
100181611
100181619
100189468
100189471
100189472
100189480
100189486
100229635
100229636
100229639
100230023
100244201
100251169
100254446
100254791
201002725
201029886
201029948
100029459
100036094
100036143
100036153
100036154
100036156
100036161
100036290
100039849
100040072
100040073
100045725
100045732
100046145
100046149
100046900
100050086
100050097
100064567
100064571
100064572
100064574
100064575
100064576
100064669
100064670
100064962
100071513
100071514
100073256
100075378
100075407
100081757
100082431
100082435
100082436
100082462
100087174
100091098
100091099
100091380
100091386
100091387
100091388
100091392
100091396
100100859
100101061
100101068
100101081
100103676
100104050
100104053
100104057
100104068
100104071
100104072
100106524
100106734
100114023
100114026
100114029
100118349
100118687
100118690
100118691
100118692
100125175
100125178
100125721
100125753
100125757
100125758
100125760
100125762
100125852
100125953
100125954
100125956
100125957
100125958
100136805
100136886
100137131
100146858
100147384
100147393
100147566
100147577
100163934
100174175
100174964
100174970
100174974
100174986
100176029
100176846
100180942
100181592
100181600
100189463
100189467
100189476
100189491
100189748
100190008
100190080
100190190
100190739
100190801
100190802
100195440
100195643
100195647
100195651
100195657
100195676
100195867
100195876
100195897
100196405
100196432
100219288
100219294
100219296
100219297
100219298
100219299
100219301
100219305
100219306
100228999
100229101
100230024
100230074
100231171
100231173
100231175
100231176
100231177
100231178
100231202
100231203
100243644
100252774
100252783
100252792
100253495
100292297
100315391
100315604
100350686
100350697
100350757
100354059
100354431
100354434
100354443
100354449
100354457
100361203
100361208
100361587
100362571
100362577
100362711
100362938
100363006
100363023
100363081
100363753
100363784
100367301
100370788
100371646
100373208
100373510
100373515
100373517
100373542
100373544
100373549
100373558
100373559
100373603
100373604
100373608
100373609
100373611
100375002
100375005
100375006
100375011
100375012
100375015
100375016
100375018
100375144
100375146
100375218
100375221
100375235
100375237
100375242
100375247
100375263
201002575
201002682
201002689
201002714
201002727
201004065
201004354
201004376
201004380
201004681
201004684
201022721
201023046
201023103
201025264
201025274
201026475
201026552
201026707
201027791
201029718
201029730
201029731
201029732
201029735
201029738
201029739
201029743
201029747
201029790
201029935
100136654
100137135
100147398
100147562
100163751
100181584
100181590
100026671
100253473
100315036
100315390
100361577
100362560
100363765
100363774
201040666
100071594
100253478
201002753
201004398
100029074
100029457
100029467
100029468
100029473
100037544
100037556
100231168
100091391
100147293
100147388
100146854
100147197
100147286
100147381
100147570
100160825
100160826
100160833
100160836
100181581
100181601
100181606
100252970
100253489
100036151
100071507
100071512
100071515
100071519
100071595
100073206
100073249
100073261
100073262
100084757
100085246
100038199
100038204
100046399
100047067
100055806
100055807
100089641
100089644
100089646
100089649
100089650
100089652
100089653
100089655
100089656
100089657
100094138
100094216
100094394
100094425
100097039
100100832
100100836
100100839
100100840
100100843
100100844
100100845
100100847
100100848
100100849
100107145
100107147
100107148
100107149
100107293
100107294
100107297
100107300
100107302
100107304
100107305
100137839
100138042
100138050
100138053
100138056
100138058
100138286
100138289
100138417
100138581
100138590
100138592
100138593
100138594
100145966
100346938
100346940
100346946
100346948
100346949
100346950
100346953
100346957
100346959
100346962
100346963
100346976
100348093
100348094
100348095
100348096
100348097
100348099
100348101
100348103
100348104
100348105
100348106
100348107
100348108
100348115
100348117
100348134
100348160
100348161
100348162
100348164
100348165
100348166
100348167
100348168
100348169
100348170
100348171
100348172
100359658
100359876
100359951
100360190
100360191
100360211
100360212
100360282
100360342
100360350
100360351
100360354
100360355
100360357
100360359
100360361
100360363
100360380
100360381
100361198
100362046
100362443
100362518
100363078
100363651
100363655
100363742
100363757
100363791
100364520
100364603
100365116
100365175
100365178
100365188
100365192
100365649
100365651
100365652
100366646
100366657
100366658
100366701
100366702
100366705
100366707
100366711
100366712
100366714
100366914
100366917
100366941
100367018
100367020
100367030
100367032
100367270
100367271
100367273
100367274
100369412
100369423
100369429
100369438
100369439
100369442
100369443
100369448
100370288
100370289
100370756
100370804
100370806
100370815
100370837
100370884
100370999
100371100
100371202
201007526
201008056
201008189
201008221
201008224
201008229
201008231
100028670
100029465
100189489
100189659
100026094
100291733
100037543
100037547
100046154
201001571
201002277
100064960
100064988
100071522
100073263
100103585
100103677
100104054
100104059
100104070
201004358
201004369
100120756
100120768
100120856
100120860
100120863
100243639
100237624
100237633
100237634
100237637
100237639
100237640
201034072
201002752
100118345
100125748
100026121
100040071
100118348
100118544
100118558
100118680
100125173
100125177
100125179
100125759
100189759
100190736
100190803
100219302
100219303
100229115
100229631
100229632
100229634
100229637
100229638
100252579
100252778
100252790
100291846
100291849
100291850
100291853
100292298
201004351
201004362
201004366
201004372
201004381
100036096
100036097
100036147
100036157
100036159
100039844
100040074
100040075
100040076
100040078
100040079
100040080
100040081
100040082
100040083
100040084
100040085
100040086
100040087
100040089
100045716
100045718
100045720
100045729
100045733
100046141
100046150
100046151
100064563
100064568
100064570
100064961
100081492
100081691
100083222
100083227
100083228
100083232
100083233
100083234
100083235
100083236
100087013
100091395
100187424
100350146
100350152
100350156
100350157
100350158
100350163
100350667
100350674
100350683
100350688
100350691
100350692
100350699
100350733
100350742
100350743
100350751
100350758
100350932
100351347
100351348
100351351
100351352
100351354
100351357
100351358
100351547
100351562
100351565
100351572
100351575
100351576
100351582
100351584
100351681
100351683
100352744
100353246
100353256
100353308
100353310
100353313
100353317
100353318
100353319
100353586
100353587
100353590
100353591
100353593
100353594
100353595
100353596
100353919
100354440
100354442
100354461
100354468
100354710
100354854
100354887
100354891
100355153
100355165
100355168
100355391
100355398
100355668
100355678
100355800
100355844
100355847
100355848
100355859
100355862
100355866
100355869
100355870
100355871
100355875
100355877
100355881
100355889
100355892
100355910
100355918
100355922
100355924
100355925
100355928
100356019
100356083
100356085
100356086
100356088
100356209
100356246
100356253
100356254
100356275
100356276
100356284
100356289
100356294
100356295
100356299
100356300
100356304
100356765
100356770
100356774
100357766
100359603
100359654
100359768
100359769
100359871
100359877
100359879
100359947
100359948
100359955
100360083
100360086
100360144
100360148
100360204
100360258
100360259
100360280
100360348
100360366
100360382
100360384
100361235
100361399
100361405
100361435
100361441
100361444
100361489
100361495
100361529
100361532
100361541
100361544
100361554
100361572
100361589
100361767
100361770
100361778
100361960
100362000
100362002
100362016
100362029
100362041
100362062
100362383
100362385
100362386
100362394
100362446
100362448
100362519
100362541
100362550
100362552
100362562
100362572
100362715
100362720
100362724
100362730
100362734
100362744
100363009
100363010
100363019
100363031
100363058
100363652
100363654
100363658
100363752
100363762
100363770
100363782
100363789
100363790
100363796
100363803
100363805
100363806
100364253
100364254
100364255
100364257
100364259
100364500
100364501
100364502
100364504
100364505
100364506
100364507
100364508
100364509
100364510
100364513
100364535
100364536
100364612
100364614
100364615
100365104
100365106
100365114
100365118
100365646
100365647
100365736
100366649
100366652
100366654
100366659
100366704
100366920
100366922
100366936
100366940
100366943
100366945
100366948
100367019
100367022
100367024
100367027
100367031
100367142
100368045
100368047
100368059
100368074
100368552
100368720
100368726
100370786
100370789
100370791
100370792
100370795
100370802
100370805
100370808
100370809
100370811
100370812
100370813
100370822
100370828
100370830
100370834
100370840
100370841
100370842
100370892
100370893
100370896
100370960
100370985
100370986
100370990
100370992
100370994
100370998
100371002
100371105
100371123
100371127
100371141
100371143
100371172
100371211
100371212
100371409
100371411
100371415
100371419
100371613
100371640
100371642
100371648
100371652
100371657
100371658
100371659
100371680
100371687
100371689
100371692
100371695
100371699
100371701
100371704
100372281
100372283
100372286
100372287
100372289
100372291
100372292
100372293
100372896
100372898
100373204
100373209
100373211
100373224
100373228
100373232
100373234
100373235
100373433
100373449
100373459
100373462
100373463
100373464
100373465
100373507
100373512
100373513
100373516
100373518
100373606
100373610
100373680
100373682
100373683
100373684
100373941
100373942
100373943
100373945
100373946
100375003
100375017
100375150
100375151
100375229
100375231
100375245
100375264
100375277
100375408
100377558
100377885
100377892
100377894
100377896
100377906
100377909
100377912
100377914
100377915
100377916
100377917
100377918
100377922
100377923
100377924
100377925
100377926
100377928
100377930
100377932
100377936
100377939
100378009
100378017
100378061
100378073
100378074
100378076
100378079
100378082
100378201
100378205
100378228
100378229
100378232
100378234
100378235
100378285
100378286
100378287
100378289
100378299
100378846
100378861
100378929
100380818
100381372
100381375
100381376
100381734
100381828
100381833
100381836
100381837
100381838
100381846
100381849
100381851
100381855
100381858
100381859
100381860
100381863
100381865
100381869
100381900
100381902
100381907
100381913
100381915
100381920
100381927
100381929
100382180
100382183
100382185
100382188
100382192
100382196
100383931
100385164
100385316
100385318
100385319
100385320
100385321
100385324
100385325
100385326
100385329
100385339
100385347
100385351
100385353
100385360
100385361
100385363
100387273
100388723
100388735
100389237
100389382
100389386
100389392
100389398
100389408
100389409
100389432
100389442
100389444
100389445
100389446
100389447
100389458
100389459
100389465
100389466
100389473
100389481
100389485
100389486
100389489
100389491
100389496
100389499
100389500
100389501
100389502
100389506
100389507
100389510
100389511
100389513
100389514
100389517
100389518
100389540
100389541
100389542
100389543
100389545
100389546
100389547
100389548
100389992
100390092
100390129
100390161
100390182
100390414
100390484
100390492
100390505
100391041
100391043
100391050
100391051
100391052
100391053
100391056
100391057
100391059
100391071
100391072
100391073
100391074
100391075
100391076
100391077
100391078
100391079
100391081
100391083
100391085
100391086
100391089
100391091
100391092
100391097
100391098
100391099
100391100
100391102
100391103
100391104
100391105
100391109
100391110
100391111
100391116
100391119
100391200
100391201
100391202
100391204
100391205
100391206
100391207
100391208
100391209
100391210
100391211
100391213
100391215
100391216
100391217
100391219
100391363
100391364
100391365
100391366
100391367
100391368
100391370
100391371
100391372
100391375
100391376
100391464
100391472
100391474
100391478
100391480
100391483
100391489
100391490
100391492
100391494
100391496
100391497
100391498
100391499
100391501
100391503
100391505
100391507
100391510
100391511
100391513
100391516
100391517
100391518
100391519
100391521
100391522
100391523
100391525
100391526
100391527
100391528
100391531
100391532
100391533
100391534
100391536
100391538
100391539
100391540
100391541
100391543
100391544
100391547
100391548
100391549
100391552
100391839
100391861
100391862
100391867
100391868
100391869
100391870
100391871
100391874
100391877
100391883
100391884
100391889
100391894
100391900
100391905
100391908
100391909
100391910
100391911
100391912
100391913
100391914
100391916
100391917
100391960
100391962
100391975
100391976
100391980
100391981
100392130
100392134
100392135
100392136
100392137
100392138
100392240
100392241
100392242
100392243
100392244
100392245
100392248
100392250
100392252
100392253
100392254
100392257
100392258
100392268
100392276
100392287
100392292
100392293
100392295
100392296
100392299
100392613
100392624
100392634
100392635
100392636
100392637
100392638
100392640
100392691
100392693
100392694
100392696
100392698
100392699
100393577
100393579
100393641
100393642
100393643
100393649
100393650
100393651
100393653
100393656
100393657
100393658
100393700
100393701
100393702
100393703
100393704
100393705
100393707
100393708
100393709
100393710
100393711
100393715
100393861
100393863
100393868
100393869
100393870
100393873
100393874
100393875
100393877
100393878
100393879
100393880
100393881
100393882
100393883
100393884
100393886
100393888
100393889
100393891
100393894
100393895
100393897
100393898
100393899
100394361
100394363
100394364
100394365
100394366
100394367
100394368
100394370
100394371
100394372
100394373
100394374
100394376
100394377
100394379
100394381
100394384
100394385
100394386
100394387
100394388
100394392
100394394
100394395
100394397
100394399
100394454
100394455
100394456
100394458
100394459
100394461
100394463
100394464
100394465
100394466
100394467
100394469
100394472
100394473
100394474
100394475
100394478
100394482
100394483
100394485
100394489
100394490
100394491
100394492
100394494
100394497
100394505
100394506
100394507
100394510
100394512
100394514
100394515
100394517
100394523
100394525
100394531
100394547
100394548
100394550
100394552
100394668
100394669
100394670
100394672
100394673
100394674
100394675
100394676
100394679
100394948
100394949
100394950
100394951
100394954
100394964
100394965
100394968
100394978
100394980
100394981
100394983
100394984
100394985
100395000
100395001
100395002
100395003
100395005
100395006
100395068
100395069
100395074
100395075
100395320
100395321
100395323
100395329
100395330
100395331
100395342
100395343
100395344
100395345
100395346
100395347
100395351
100395353
100395355
100395358
100395416
100395418
100395560
100395567
100395570
100395571
100395572
201002683
100037537
100037538
100037554
100060417
100060477
100060478
100060479
100060489
100060580
100060581
100060582
100060583
100060584
100060586
100060587
100060588
100060589
100060590
100060592
100060595
100060596
100060598
100060601
100060602
100060603
100061650
100061651
100061775
100061780
100061781
100061782
100061783
100061784
100061785
100061786
100061787
100061788
100061789
100061790
100061791
100061793
100061795
100061799
100029469
100125959
100125965
100180970
100181586
100181605
100189473
100026086
100266744
100045730
100049073
100049074
100049192
100057214
100057353
100064559
100064838
100065594
100065597
100067440
100067531
100067535
100073203
100084823
100084831
100091401
100174176
100174178
100174961
201015523
100083223
100085474
100098474
100100132
100100173
100114021
100114022
100114040
100137026
201002217
201002386
201004059
100394470
100394508
100051959
100051964
100051967
100051968
100094219
100094345
100095647
100095648
100095649
100097381
100181618
201002380
100072480
100072481
100072482
100072484
100073253
100115401
100115402
100115404
100115405
100115406
100115409
100115412
100115461
100115464
100115798
100115802
100115803
100115806
100115809
100115810
100115811
100118343
100118550
100119490
100120751
100120759
100120854
100120855
100120857
100120861
100120865
100120868
100122342
100122343
100122346
100122582
100122585
100132201
100132205
100132208
100136647
100136649
100136653
100136809
100147107
100147205
100147283
100147285
100147383
100147386
100147396
100147397
100147564
100154610
100154611
100159226
100159228
100159236
100159238
100159560
100159561
100159563
100159566
100159568
100159920
100164088
100174307
100189485
100190806
100190818
100190821
100228824
100230070
100230076
100230167
100231170
100256987
100257092
100257809
100257811
100257818
100278754
100278785
100278794
100278799
100278809
100350165
100350169
100350171
100350174
100350179
100350180
100350187
100350700
100350745
100354033
100354426
100354430
100354433
100354436
100354441
100354446
100354447
100354450
100354451
100354455
100354456
100354460
100354474
100355919
100359648
100359872
100360358
100363028
100363029
100363055
100363783
100363795
100364619
100366700
100367141
100367300
100367306
100367308
100368377
100368522
100370807
100370814
100370889
100371176
100371433
100371708
100373552
100373553
100373554
100373605
100373944
100375013
100375140
100375143
100375228
100375230
100375239
100375244
100375276
100375287
100375288
100375297
100375380
100377880
100377887
100378217
100378222
100378226
100378230
100378231
100378233
100378282
100378284
100378288
100378292
100378304
100381374
100381834
100381910
100382187
100387153
100387243
100387249
100391915
100392695
201008038
201008185
201008230
201009337
201009339
201010599
201014978
201015059
201015061
201023059
201028002
201029728
201030728
201030884
201030940
201030942
201030947
201030952
201030961
201031063
201031068
201031074
201031161
201031193
201031196
201031201
201031204
201031205
201031206
201031212
201031213
201031217
201031444
201031446
201031447
201031448
201031449
201031450
201031452
201031454
201031455
201031458
201031520
201031527
201031528
201031531
201031532
201031535
201031538
Amended 07/29/2021: US Products that Faceplates will be replaced
Model: V60
Part Number 1053613
V60 Ventilator,Intl Opt: CFLEX,AVAPS,PP
US Serial Numbers:
100051959;
100051964;
100051967;
100051968;
100094219;
100094345;
100095647;
100095648;
100095649;
100097381;
100181618;
100228824;
|
Recalling Firm/ Manufacturer |
Respironics California, LLC 2271 Cosmos Ct Carlsbad CA 92011-1517
|
| For Additional Information Contact | 800-722-9377 |
Manufacturer Reason for Recall | Potential issues with ventilator navigation ring interface and liquid ingress:
1) The navigation ring may become inoperative or operate intermittently;
2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments. |
FDA Determined Cause 2 | Device Design |
| Action | On 11/02/2020, the firm sent an "URGENT-MEDICAL DEVICE CORRECTION" Notification Letter via priority mail informing customer that the V60 Ventilators may experience two potential issues with the navigation ring interface (see table below for additional information).
1. The navigation ring may become inoperative or begin operating intermittently.
2. The navigation ring may not behave as expected, making it difficult for the user to change ventilator settings via the touchscreen.
and if clinicians accept an incorrect setting and leaves the patient on the ventilator without responding to the resulting alarms, the patient could sustain serious injuries.
Customers are instructed:
1. When cleaning V60 ventilators, do not allow cleaning agents to pool on the
ventilator surfaces. This is especially important on the device interface.
2. Check setting changes before and after accepting the changes.
3. If the navigation ring becomes inoperative or operates intermittently during
clinical use, use the touchscreen for all settings changes and navigation,
and contact Philips to schedule service following clinical use.
4. If the navigation ring interferes with your ability to adjust therapy settings via
touchscreen, remove the ventilator from service and contact Philips for
repair.
and to please review the following sections of the user manual:
1. Chapter 10: Care and maintenance. To prevent possible damage to the
ventilator when cleaning, use only those cleaning and disinfecting agents
listed, and do not drip or spray any liquids directly onto any ventilator
surface.
2. Chapter 1: Warnings, cautions, and notes. During clinical use of the
ventilator, monitor patients for changes in respiratory status with each
ventilator setting adjustment. Use an external oxygen monitor to verify
oxygen concentration, and have an alternative means of ventilation available
whenever a ventilator is in use.
Customers are also instructed to complete, sign and return the Acknowledgement and Receipt |
| Quantity in Commerce | 2,137 systems (Amended on 07/29/2021 - 12 U.S. systems for faceplate replacement) |
| Distribution | Worldwide distribution - US Nationwide including Guam and Puerto Rico; Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = MNT
|
|
|
|