|
Class 1 Device Recall Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir With Balance Biosurface |
 |
Date Initiated by Firm |
February 04, 2021 |
Date Posted |
March 17, 2021 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1191-2021 |
Recall Event ID |
87271 |
510(K)Number |
K183511
|
Product Classification |
Reservoir, blood, cardiopulmonary bypass - Product Code DTN
|
Product |
Medtronic Affinity Pixie BP241 Hollow Fiber Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (OXYGENATOR BBP241 W/CVR NI BB STERILE). The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a single-use device designed to collect and store blood during extracorporeal circulation. Cardiotomy blood is collected, filtered, and defoamed before mixing with the venous blood, which is filtered. The primary blood-contacting surfaces of the CVR are coated with Balance Biosurface to reduce platelet activation and adhesion and preserve platelet function. (Generic Name: Cardiopulmonary bypass oxygenator). |
Code Information |
Description: OXYGENATOR BBP241 W/CVR NI BB STERILE. Model Number: BBP241. GTIN: 20613994617303. Lot Number: 13340434. |
Recalling Firm/ Manufacturer |
Medtronic Perfusion Systems 7611 Northland Dr N Brooklyn Park MN 55428-1088
|
For Additional Information Contact |
Medtronic Customer Service 800-8543570
|
Manufacturer Reason for Recall |
Products being recalled due to potentially elevated level of bacterial endotoxin. Patients exposed to elevated levels of bacterial endotoxins may develop an acute inflammatory response.
|
FDA Determined Cause 2 |
Process control |
Action |
A written Medical Device Correction letter will be mailed via UPS 2-day delivery to consignees who have received affected product according to Medtronic records. Consignees will be asked to complete the enclosed Customer Confirmation Form and email to rs.cfqfca@medtronic.com. Consignees will also be asked to return all unused affected product in your inventory to Medtronic. Contact Medtronic Customer Service at800-854-3570 to initiate a product return. |
Quantity in Commerce |
57 units |
Distribution |
Worldwide distribution: US nationwide, Afghanistan, Argentina, Chile, China, Colombia, Egypt, Germany, Iraq, Italy, Japan, Korea, Kyrgyzstan, Latvia, Nicaragua, Paraguay, Poland, Qatar, Romania, Tanzania, United Arab Emirates, United Kingdom, Uzbekistan, and Viet Nam |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = DTN and Original Applicant = Medtronic Inc
|
|
|
|