• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Pan Medical US/Zavation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Pan Medical US/Zavationsee related information
Date Initiated by FirmMarch 25, 2021
Create DateJune 01, 2021
Recall Status1 Terminated 3 on March 20, 2023
Recall NumberZ-1733-2021
Recall Event ID 87661
510(K)NumberK141419 
Product Classification Cement, bone, vertebroplasty - Product Code NDN
Product10G BEVELED TROCAR/CANNULA. For use in Orthopedic / spinal procedures.
Code Information Pending
FEI Number 3008583793
Recalling Firm/
Manufacturer
Zavation
220 Lakeland Pkwy
Flowood MS 39232-9552
For Additional Information ContactQuality Department
601-919-1119
Manufacturer Reason
for Recall
Products distributed as sterile may not have been adequately sterilized
FDA Determined
Cause 2
Vendor change control
ActionZavation initially notified their consignees of the problem by email beginning on 03/24/2021 and requested they quarantine the product. The firm sent a subsequent email beginning on 04/13/2021 requesting the return of the product. The recall is being conducted to the user level.
DistributionUS nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = NDN
-
-