• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Philips Afiniti Diagnostic Ultrasound Systems

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Philips Afiniti Diagnostic Ultrasound Systemssee related information
Date Initiated by FirmAugust 06, 2021
Create DateSeptember 17, 2021
Recall Status1 Open3, Classified
Recall NumberZ-2484-2021
Recall Event ID 88514
510(K)NumberK182857 
Product Classification System, imaging, pulsed echo, ultrasonic - Product Code IYO
ProductAffiniti Diagnostic Ultrasound Systems: Software Versions 5.0, 5.0.1 and 5.0.2 Models: Affiniti 30 795121 Affiniti 30 795218 Affiniti 50 795208 Affiniti 50 795118 Affiniti 70 795210 Affiniti 70 795119
Code Information Afiniti Diagnostic Ultrasound Systems: Affiniti 30 Model Number: 795121 Serial Numbers/UDI Codes US219E1452 (01)00884838074187(21)US219E1452 US220E0262 (01)00884838074187(21)US220E0262 US320E1098 (01)00884838074187(21)US320E1098 US320E1099 (01)00884838074187(21)US320E1099 US918E0145 (01)00884838074187(21)US918E0145 USN18E0607 (01)00884838074187(21)USN18E0607 USN18E1617 (01)00884838074187(21)USN18E1617 USN19E2219 (01)00884838074187(21)USN19E2219 USN19E2220 (01)00884838074187(21)USN19E2220 USN19E2221 (01)00884838074187(21)USN19E2221 Affiniti 30 Model Number: 795218 Serial Numbers/UDI Codes PU220E1330 N/A PU220E1331 N/A PU220E1332 N/A PU220E1333 N/A PU220E1334 N/A PU220E1335 N/A PU420E0442 N/A PU420E0658 N/A PU420E0659 N/A PU420E0660 (01)00884838097957 PU420E0761 (01)00884838074187 PU520E0703 (01)00884838074187 PU520E0713 N/A PU520E0854 (01)00884838074187 PU520E0855 N/A PU520E0856 N/A PU520E0857 N/A PU520E0858 N/A PU520E0859 N/A PU520E0861 N/A PU520E0965 N/A PU520E0966 N/A PU520E0967 N/A PU520E0968 N/A PU520E1012 N/A PU520E1013 (01)00884838074187 PU520E1014 (01)00884838074187 PU520E1015 N/A PU520E1016 N/A PU520E1017 (01)00884838074187 PU520E2820 (01)00884838097957 PU520E2821 (01)00884838097957 PU520E2831 (01)00884838097957 PU620E0090 (01)00884838097957 PU720E0464 (01)00884838074187 PU720E0468 N/A PU720E0611 (01)00884838097957 PU720E0800 (01)00884838097957 PU720E0801 (01)00884838097957 PU720E0803 N/A PU720E0804 N/A PU820E0248 N/A PU820E0249 N/A PU820E0347 (01)00884838097957 PU820E0349 N/A PU820E0525 N/A PU820E0526 N/A PU820E0535 N/A PU820E0536 N/A PU820E0540 N/A PU820E0541 N/A PU820E0544 (01)00884838074187 PU820E0545 N/A PU820E0654 (01)00884838074187 PUD19E0772 N/A PUD19E0832 N/A PUD19E0833 N/A US117E1004 (01)00884838074187(21)US117E1004 US118E1385 (01)00884838074187(21)US118E1385 US119E0798 (01)00884838074187(21)US119E0798 US119E0912 (01)00884838074187(21)US119E0912 US119E1661 (01)00884838074187(21)US119E1661 US119E1662 (01)00884838074187(21)US119E1662 US120E0309 (01)00884838074187(21)US120E0309 US120E0310 (01)00884838074187(21)US120E0310 US120E0340 (01)00884838074187(21)US120E0340 US120E0341 (01)00884838074187(21)US120E0341 US120E0342 (01)00884838074187(21)US120E0342 US120E0448 (01)00884838074187(21)US120E0448 US120E0449 (01)00884838074187(21)US120E0449 US120E0507 (01)00884838074187(21)US120E0507 US120E0630 (01)00884838074187(21)US120E0630 US120E0651 (01)00884838074187(21)US120E0651 US120E0652 (01)00884838074187(21)US120E0652 US120E0653 (01)00884838074187(21)US120E0653 US120E0654 (01)00884838074187(21)US120E0654 US120E0655 (01)00884838074187(21)US120E0655 US120E0656 (01)00884838074187(21)US120E0656 US120E0657 (01)00884838074187(21)US120E0657 US120E0719 (01)00884838074187(21)US120E0719 US120E0720 (01)00884838074187(21)US120E0720 US120E0721 (01)00884838074187(21)US120E0721 US120E0722 (01)00884838074187(21)US120E0722 US120E0723 (01)00884838074187(21)US120E0723 US120E0724 (01)00884838074187(21)US120E0724 US120E0778 (01)00884838074187(21)US120E0778 US120E0811 (01)00884838074187(21)US120E0811 US120E0812 (01)00884838074187(21)US120E0812 US120E0813 (01)00884838074187(21)US120E0813 US120E0814 (01)00884838074187(21)US120E0814 US120E0815 (01)00884838074187(21)US120E0815 US120E0816 (01)00884838074187(21)US120E0816 US120E0817 (01)00884838074187(21)US120E0817 US120E0818 (01)00884838074187(21)US120E0818 US120E0819 (01)00884838074187(21)US120E0819 US120E0820 (01)00884838074187(21)US120E0820 US120E0842 (01)00884838074187(21)US120E0842 US120E0843 (01)00884838074187(21)US120E0843 US120E0892 (01)00884838074187(21)US120E0892 US120E1001 (01)00884838074187(21)US120E1001 US120E1002 (01)00884838074187(21)US120E1002 US120E1003 (01)00884838074187(21)US120E1003 US120E1044 (01)00884838074187(21)US120E1044 US120E1045 (01)00884838074187(21)US120E1045 US120E1046 (01)00884838074187(21)US120E1046 US120E1047 (01)00884838074187(21)US120E1047 US120E1048 (01)00884838074187(21)US120E1048 US120E1049 (01)00884838074187(21)US120E1049 US120E1050 (01)00884838074187(21)US120E1050 US120E1051 (01)00884838074187(21)US120E1051 US120E1052 (01)00884838074187(21)US120E1052 US120E1053 (01)00884838074187(21)US120E1053 US120E1054 (01)00884838074187(21)US120E1054 US120E1150 (01)00884838074187(21)US120E1150 US120E1151 (01)00884838074187(21)US120E1151 US120E1152 (01)00884838074187(21)US120E1152 US120E1153 (01)00884838074187(21)US120E1153 US120E1189 (01)00884838074187(21)US120E1189 US120E1190 (01)00884838074187(21)US120E1190 US120E1218 (01)00884838074187(21)US120E1218 US120E1298 (01)00884838074187(21)US120E1298 US120E1402 (01)00884838074187(21)US120E1402 US120E1403 (01)00884838074187(21)US120E1403 US120E1404 (01)00884838074187(21)US120E1404 US120E1405 (01)00884838074187(21)US120E1405 US120E1466 (01)00884838074187(21)US120E1466 US120E1467 (01)00884838074187(21)US120E1467 US120E1535 (01)00884838074187(21)US120E1535 US120E1539 (01)00884838074187(21)US120E1539 US120E1575 (01)00884838074187(21)US120E1575 US120E1576 (01)00884838074187(21)US120E1576 US120E1610 (01)00884838074187(21)US120E1610 US120E1614 (01)00884838074187(21)US120E1614 US120E1651 (01)00884838074187(21)US120E1651 US120E1652 (01)00884838074187(21)US120E1652 US120E1725 (01)00884838074187(21)US120E1725 US120E1727 (01)00884838074187(21)US120E1727 US120E1808 (01)00884838074187(21)US120E1808 US120E1954 (01)00884838074187(21)US120E1954 US120E1981 (01)00884838074187(21)US120E1981 US121E0676 (01)00884838074187(21)US121E0676 US218E1315 (01)00884838074187(21)US218E1315 US219E0867 (01)00884838074187(21)US219E0867 US219E0903 (01)00884838074187(21)US219E0903 US219E0915 (01)00884838074187(21)US219E0915 US219E0916 (01)00884838074187(21)US219E0916 US220E0261 (01)00884838074187(21)US220E0261 US220E0279 (01)00884838074187(21)US220E0279 US220E0280 (01)00884838074187(21)US220E0280 US220E0281 (01)00884838074187(21)US220E0281 US220E0357 (01)00884838074187(21)US220E0357 US220E0358 (01)00884838074187(21)US220E0358 US220E0359 (01)00884838074187(21)US220E0359 US220E0456 (01)00884838074187(21)US220E0456 US220E0551 (01)00884838074187(21)US220E0551 US220E0763 (01)00884838074187(21)US220E0763 US220E0913 (01)00884838074187(21)US220E0913 US220E0914 (01)00884838074187(21)US220E0914 US220E1013 (01)00884838074187(21)US220E1013 US220E1014 (01)00884838074187(21)US220E1014 US220E1056 (01)00884838074187(21)US220E1056 US220E1097 (01)00884838074187(21)US220E1097 US220E1098 (01)00884838074187(21)US220E1098 US220E1106 (01)00884838074187(21)US220E1106 US220E1107 (01)00884838074187(21)US220E1107 US220E1217 (01)00884838074187(21)US220E1217 US220E1218 (01)00884838074187(21)US220E1218 US220E1219 (01)00884838074187(21)US220E1219 US220E1303 (01)00884838074187(21)US220E1303 US220E1370 (01)00884838074187(21)US220E1370 US220E1381 (01)00884838074187(21)US220E1381 US220E1382 (01)00884838074187(21)US220E1382 US220E1430 (01)00884838074187(21)US220E1430 US220E1431 (01)00884838074187(21)US220E1431 US220E1432 (01)00884838074187(21)US220E1432 US220E1433 (01)00884838074187(21)US220E1433 US220E1482 (01)00884838074187(21)US220E1482 US220E1487 (01)00884838074187(21)US220E1487 US220E1488 (01)00884838074187(21)US220E1488 US220E1489 (01)00884838074187(21)US220E1489 US220E1539 (01)00884838074187(21)US220E1539 US220E1540 (01)00884838074187(21)US220E1540 US220E1541 (01)00884838074187(21)US220E1541 US220E1542 (01)00884838074187(21)US220E1542 US220E1543 (01)00884838074187(21)US220E1543 US220E1582 (01)00884838074187(21)US220E1582 US220E1608 (01)00884838074187(21)US220E1608 US220E1609 (01)00884838074187(21)US220E1609 US319E0045 (01)00884838074187(21)US319E0045 US319E0480 (01)00884838074187(21)US319E0480 US319E0485 (01)00884838074187(21)US319E0485 US319E0514 (01)00884838074187(21)US319E0514 US319E1034 (01)00884838074187(21)US319E1034 US320E0029 (01)00884838074187(21)US320E0029 US320E0030 (01)00884838074187(21)US320E0030 US320E0031 (01)00884838074187(21)US320E0031 US320E0062 (01)00884838074187(21)US320E0062 US320E0063 (01)00884838074187(21)US320E0063 US320E0064 (01)00884838074187(21)US320E0064 US320E0065 (01)00884838074187(21)US320E0065 US320E0066 (01)00884838074187(21)US320E0066 US320E0067 (01)00884838074187(21)US320E0067 US320E0124 (01)00884838074187(21)US320E0124 US320E0242 (01)00884838074187(21)US320E0242 US320E0243 (01)00884838074187(21)US320E0243 US320E0244 (01)00884838074187(21)US320E0244 US320E0245 (01)00884838074187(21)US320E0245 US320E0246 (01)00884838074187(21)US320E0246 US320E0247 (01)00884838074187(21)US320E0247 US320E0256 (01)00884838074187(21)US320E0256 US320E0260 (01)00884838074187(21)US320E0260 US320E0262 (01)00884838074187(21)US320E0262 US320E0327 (01)00884838074187(21)US320E0327 US320E0331 (01)00884838074187(21)US320E0331 US320E0334 (01)00884838074187(21)US320E0334 US320E0372 (01)00884838074187(21)US320E0372 US320E0373 (01)00884838074187(21)US320E0373 US320E0389 (01)00884838074187(21)US320E0389 US320E0390 (01)00884838074187(21)US320E0390 US320E0391 (01)00884838074187(21)US320E0391 US320E0392 (01)00884838074187(21)US320E0392 US320E0393 (01)00884838074187(21)US320E0393 US320E0394 (01)00884838074187(21)US320E0394 US320E0395 (01)00884838074187(21)US320E0395 US320E0396 (01)00884838074187(21)US320E0396 US320E0398 (01)00884838074187(21)US320E0398 US320E0471 (01)00884838074187(21)US320E0471 US320E0472 (01)00884838074187(21)US320E0472 US320E0651 (01)00884838074187(21)US320E0651 US320E0652 (01)00884838074187(21)US320E0652 US320E0653 (01)00884838074187(21)US320E0653 US320E0654 (01)00884838074187(21)US320E0654 US320E0655 (01)00884838074187(21)US320E0655 US320E0656 (01)00884838074187(21)US320E0656 US320E0657 (01)00884838074187(21)US320E0657 US320E0663 (01)00884838074187(21)US320E0663 US320E0720 (01)00884838074187(21)US320E0720 US320E0721 (01)00884838074187(21)US320E0721 US320E0722 (01)00884838074187(21)US320E0722 US320E0723 (01)00884838074187(21)US320E0723 US320E0789 (01)00884838074187(21)US320E0789 US320E0790 (01)00884838074187(21)US320E0790 US320E0791 (01)00884838074187(21)US320E0791 US320E0820 (01)00884838074187(21)US320E0820 US320E0821 (01)00884838074187(21)US320E0821 US320E0822 (01)00884838074187(21)US320E0822 US320E0933 (01)00884838074187(21)US320E0933 US320E0934 (01)00884838074187(21)US320E0934 US320E1011 (01)00884838074187(21)US320E1011 US320E1016 (01)00884838074187(21)US320E1016 US320E1084 (01)00884838074187(21)US320E1084 US320E1431 (01)00884838074187(21)US320E1431 US320E1432 (01)00884838074187(21)US320E1432 US320E1433 (01)00884838074187(21)US320E1433 US320E1434 (01)00884838074187(21)US320E1434 US320E1435 (01)00884838074187(21)US320E1435 US320E1436 (01)00884838074187(21)US320E1436 US320E1885 (01)00884838074187(21)US320E1885 US320E1887 (01)00884838074187(21)US320E1887 US320E1891 (01)00884838074187(21)US320E1891 US320E1892 (01)00884838074187(21)US320E1892 US320E1893 (01)00884838074187(21)US320E1893 US320E1894 (01)00884838074187(21)US320E1894 US320E1935 (01)00884838074187(21)US320E1935 US417E0707 (01)00884838074187(21)US417E0707 US419E1970 (01)00884838074187(21)US419E1970 US419E1989 (01)00884838074187(21)US419E1989 US420E0086 (01)00884838074187(21)US420E0086 US420E0097 (01)00884838074187(21)US420E0097 US420E0098 (01)00884838074187(21)US420E0098 US420E0099 (01)00884838074187(21)US420E0099 US420E0258 (01)00884838074187(21)US420E0258 US420E0259 (01)00884838074187(21)US420E0259 US420E0503 (01)00884838074187(21)US420E0503 US420E0671 (01)00884838074187(21)US420E0671 US420E0672 (01)00884838074187(21)US420E0672 US420E0673 (01)00884838074187(21)US420E0673 US420E2630 (01)00884838074187(21)US420E2630 US420E2994 (01)00884838074187(21)US420E2994 US520E1450 (01)00884838074187(21)US520E1450 US619E0200 (01)00884838074187(21)US619E0200 US619E0665 (01)00884838074187(21)US619E0665 US619E0666 (01)00884838074187(21)US619E0666 US619E0667 (01)00884838074187(21)US619E0667 US619E1012 (01)00884838074187(21)US619E1012 US619E1792 (01)00884838074187(21)US619E1792 US620E1699 (01)00884838074187(21)US620E1699 US718E0382 (01)00884838074187(21)US718E0382 US719E0599 (01)00884838074187(21)US719E0599 US719E1191 (01)00884838074187(21)US719E1191 US719E1192 (01)00884838074187(21)US719E1192 US719E1389 (01)00884838074187(21)US719E1389 US719E1390 (01)00884838074187(21)US719E1390 US719E1716 (01)00884838074187(21)US719E1716 US719E1817 (01)00884838074187(21)US719E1817 US719E1818 (01)00884838074187(21)US719E1818 US719E1819 (01)00884838074187(21)US719E1819 US719E1864 (01)00884838074187(21)US719E1864 US719E2042 (01)00884838074187(21)US719E2042 US719E2043 (01)00884838074187(21)US719E2043 US719E2044 (01)00884838074187(21)US719E2044 US719E2062 (01)00884838074187(21)US719E2062 US719E2144 (01)00884838074187(21)US719E2144 US719E2148 (01)00884838074187(21)US719E2148 US719E2149 (01)00884838074187(21)US719E2149 US720E1276 (01)00884838074187(21)US720E1276 US720E1392 (01)00884838074187(21)US720E1392 US720E1393 (01)00884838074187(21)US720E1393 US720E1394 (01)00884838074187(21)US720E1394 US720E1395 (01)00884838074187(21)US720E1395 US720E1409 (01)00884838074187(21)US720E1409 US818E0717 (01)00884838074187(21)US818E0717 US818E1715 (01)00884838074187(21)US818E1715 US819E0085 (01)00884838074187(21)US819E0085 US819E0179 (01)00884838074187(21)US819E0179 US819E0180 (01)00884838074187(21)US819E0180 US819E0181 (01)00884838074187(21)US819E0181 US819E0182 (01)00884838074187(21)US819E0182 US819E0369 (01)00884838074187(21)US819E0369 US819E0370 (01)00884838074187(21)US819E0370 US819E0371 (01)00884838074187(21)US819E0371 US819E0384 (01)00884838074187(21)US819E0384 US819E0433 (01)00884838074187(21)US819E0433 US819E0434 (01)00884838074187(21)US819E0434 US819E0435 (01)00884838074187(21)US819E0435 US819E0436 (01)00884838074187(21)US819E0436 US819E0437 (01)00884838074187(21)US819E0437 US819E0438 (01)00884838074187(21)US819E0438 US819E0439 (01)00884838074187(21)US819E0439 US819E0629 (01)00884838074187(21)US819E0629 US819E0630 (01)00884838074187(21)US819E0630 US819E0631 (01)00884838074187(21)US819E0631 US819E0647 (01)00884838074187(21)US819E0647 US819E0648 (01)00884838074187(21)US819E0648 US819E0649 (01)00884838074187(21)US819E0649 US819E0650 (01)00884838074187(21)US819E0650 US819E0708 (01)00884838074187(21)US819E0708 US819E0709 (01)00884838074187(21)US819E0709 US819E0710 (01)00884838074187(21)US819E0710 US819E0711 (01)00884838074187(21)US819E0711 US819E0712 (01)00884838074187(21)US819E0712 US819E0713 (01)00884838074187(21)US819E0713 US819E0714 (01)00884838074187(21)US819E0714 US819E0715 (01)00884838074187(21)US819E0715 US819E0716 (01)00884838074187(21)US819E0716 US819E0815 (01)00884838074187(21)US819E0815 US819E0816 (01)00884838074187(21)US819E0816 US819E0817 (01)00884838074187(21)US819E0817 US819E0818 (01)00884838074187(21)US819E0818 US819E0819 (01)00884838074187(21)US819E0819 US819E0820 (01)00884838074187(21)US819E0820 US819E0827 (01)00884838074187(21)US819E0827 US819E0828 (01)00884838074187(21)US819E0828 US819E0829 (01)00884838074187(21)US819E0829 US819E0830 (01)00884838074187(21)US819E0830 US819E0891 (01)00884838074187(21)US819E0891 US819E0892 (01)00884838074187(21)US819E0892 US819E0893 (01)00884838074187(21)US819E0893 US819E0894 (01)00884838074187(21)US819E0894 US819E0895 (01)00884838074187(21)US819E0895 US819E0896 (01)00884838074187(21)US819E0896 US819E0897 (01)00884838074187(21)US819E0897 US819E0898 (01)00884838074187(21)US819E0898 US819E0899 (01)00884838074187(21)US819E0899 US819E0900 (01)00884838074187(21)US819E0900 US819E0901 (01)00884838074187(21)US819E0901 US819E0902 (01)00884838074187(21)US819E0902 US819E0945 (01)00884838074187(21)US819E0945 US819E0946 (01)00884838074187(21)US819E0946 US819E0947 (01)00884838074187(21)US819E0947 US819E1008 (01)00884838074187(21)US819E1008 US819E1009 (01)00884838074187(21)US819E1009 US819E1010 (01)00884838074187(21)US819E1010 US819E1011 (01)00884838074187(21)US819E1011 US819E1012 (01)00884838074187(21)US819E1012 US819E1013 (01)00884838074187(21)US819E1013 US819E1014 (01)00884838074187(21)US819E1014 US819E1015 (01)00884838074187(21)US819E1015 US819E1016 (01)00884838074187(21)US819E1016 US819E1053 (01)00884838074187(21)US819E1053 US819E1155 (01)00884838074187(21)US819E1155 US819E1212 (01)00884838074187(21)US819E1212 US819E1370 (01)00884838074187(21)US819E1370 US819E1371 (01)00884838074187(21)US819E1371 US819E1468 (01)00884838074187(21)US819E1468 US819E1611 (01)00884838074187(21)US819E1611 US819E1668 (01)00884838074187(21)US819E1668 US819E1669 (01)00884838074187(21)US819E1669 US819E1670 (01)00884838074187(21)US819E1670 US819E1671 (01)00884838074187(21)US819E1671 US819E1672 (01)00884838074187(21)US819E1672 US819E1752 (01)00884838074187(21)US819E1752 US819E1753 (01)00884838074187(21)US819E1753 US819E1754 (01)00884838074187(21)US819E1754 US819E1755 (01)00884838074187(21)US819E1755 US819E1756 (01)00884838074187(21)US819E1756 US819E1757 (01)00884838074187(21)US819E1757 US819E1758 (01)00884838074187(21)US819E1758 US819E1759 (01)00884838074187(21)US819E1759 US819E1775 (01)00884838074187(21)US819E1775 US819E1869 (01)00884838074187(21)US819E1869 US819E1870 (01)00884838074187(21)US819E1870 US819E1871 (01)00884838074187(21)US819E1871 US819E1995 (01)00884838074187(21)US819E1995 US819E2070 (01)00884838074187(21)US819E2070 US819E2071 (01)00884838074187(21)US819E2071 US820E1471 (01)00884838074187(21)US820E1471 US820E1531 (01)00884838074187(21)US820E1531 US820E1577 (01)00884838074187(21)US820E1577 US820E1578 (01)00884838074187(21)US820E1578 US917E0757 (01)00884838074187(21)US917E0757 US917E1468 (01)00884838074187(21)US917E1468 US917E2018 (01)00884838074187(21)US917E2018 US919E0151 (01)00884838074187(21)US919E0151 US919E0152 (01)00884838074187(21)US919E0152 US919E0153 (01)00884838074187(21)US919E0153 US919E0234 (01)00884838074187(21)US919E0234 US919E0235 (01)00884838074187(21)US919E0235 US919E0236 (01)00884838074187(21)US919E0236 US919E0287 (01)00884838074187(21)US919E0287 US919E0288 (01)00884838074187(21)US919E0288 US919E0289 (01)00884838074187(21)US919E0289 US919E0290 (01)00884838074187(21)US919E0290 US919E0291 (01)00884838074187(21)US919E0291 US919E0292 (01)00884838074187(21)US919E0292 US919E0342 (01)00884838074187(21)US919E0342 US919E0402 (01)00884838074187(21)US919E0402 US919E0430 (01)00884838074187(21)US919E0430 US919E0510 (01)00884838074187(21)US919E0510 US919E0511 (01)00884838074187(21)US919E0511 US919E0659 (01)00884838074187(21)US919E0659 US919E0660 (01)00884838074187(21)US919E0660 US919E0661 (01)00884838074187(21)US919E0661 US919E0738 (01)00884838074187(21)US919E0738 US919E0739 (01)00884838074187(21)US919E0739 US919E0740 (01)00884838074187(21)US919E0740 US919E0741 (01)00884838074187(21)US919E0741 US919E0742 (01)00884838074187(21)US919E0742 US919E0743 (01)00884838074187(21)US919E0743 US919E0744 (01)00884838074187(21)US919E0744 US919E0745 (01)00884838074187(21)US919E0745 US919E0799 (01)00884838074187(21)US919E0799 US919E0800 (01)00884838074187(21)US919E0800 US919E0801 (01)00884838074187(21)US919E0801 US919E0827 (01)00884838074187(21)US919E0827 US919E0899 (01)00884838074187(21)US919E0899 US919E0900 (01)00884838074187(21)US919E0900 US919E0901 (01)00884838074187(21)US919E0901 US919E0902 (01)00884838074187(21)US919E0902 US919E0903 (01)00884838074187(21)US919E0903 US919E0916 (01)00884838074187(21)US919E0916 US919E0917 (01)00884838074187(21)US919E0917 US919E0918 (01)00884838074187(21)US919E0918 US919E0974 (01)00884838074187(21)US919E0974 US919E0975 (01)00884838074187(21)US919E0975 US919E1066 (01)00884838074187(21)US919E1066 US919E1067 (01)00884838074187(21)US919E1067 US919E1158 (01)00884838074187(21)US919E1158 US919E1205 (01)00884838074187(21)US919E1205 US919E1254 (01)00884838074187(21)US919E1254 US919E1271 (01)00884838074187(21)US919E1271 US919E1272 (01)00884838074187(21)US919E1272 US919E1273 (01)00884838074187(21)US919E1273 US919E1274 (01)00884838074187(21)US919E1274 US919E1275 (01)00884838074187(21)US919E1275 US919E1276 (01)00884838074187(21)US919E1276 US919E1277 (01)00884838074187(21)US919E1277 US919E1293 (01)00884838074187(21)US919E1293 US919E1294 (01)00884838074187(21)US919E1294 US919E1295 (01)00884838074187(21)US919E1295 US919E1296 (01)00884838074187(21)US919E1296 US919E1297 (01)00884838074187(21)US919E1297 US919E1301 (01)00884838074187(21)US919E1301 US919E1302 (01)00884838074187(21)US919E1302 US919E1319 (01)00884838074187(21)US919E1319 US919E1439 (01)00884838074187(21)US919E1439 US919E1440 (01)00884838074187(21)US919E1440 US919E1443 (01)00884838074187(21)US919E1443 US919E1444 (01)00884838074187(21)US919E1444 US919E1445 (01)00884838074187(21)US919E1445 US919E1576 (01)00884838074187(21)US919E1576 US919E1650 (01)00884838074187(21)US919E1650 US919E1720 (01)00884838074187(21)US919E1720 US919E1721 (01)00884838074187(21)US919E1721 US919E1749 (01)00884838074187(21)US919E1749 US919E1750 (01)00884838074187(21)US919E1750 US919E1808 (01)00884838074187(21)US919E1808 US919E2009 (01)00884838074187(21)US919E2009 US919E2095 (01)00884838074187(21)US919E2095 US919E2096 (01)00884838074187(21)US919E2096 US919E2097 (01)00884838074187(21)US919E2097 US919E2098 (01)00884838074187(21)US919E2098 US920E1093 (01)00884838074187(21)US920E1093 US920E1094 (01)00884838074187(21)US920E1094 US920E1095 (01)00884838074187(21)US920E1095 US920E1096 (01)00884838074187(21)US920E1096 US920E1097 (01)00884838074187(21)US920E1097 US920E1226 (01)00884838074187(21)US920E1226 US920E1227 (01)00884838074187(21)US920E1227 USD17E0525 (01)00884838074187(21)USD17E0525 USD19E0050 (01)00884838074187(21)USD19E0050 USD19E0051 (01)00884838074187(21)USD19E0051 USD19E0078 (01)00884838074187(21)USD19E0078 USD19E0079 (01)00884838074187(21)USD19E0079 USD19E0080 (01)00884838074187(21)USD19E0080 USD19E0084 (01)00884838074187(21)USD19E0084 USD19E0086 (01)00884838074187(21)USD19E0086 USD19E0087 (01)00884838074187(21)USD19E0087 USD19E0088 (01)00884838074187(21)USD19E0088 USD19E0089 (01)00884838074187(21)USD19E0089 USD19E0097 (01)00884838074187(21)USD19E0097 USD19E0098 (01)00884838074187(21)USD19E0098 USD19E0099 (01)00884838074187(21)USD19E0099 USD19E0100 (01)00884838074187(21)USD19E0100 USD19E0179 (01)00884838074187(21)USD19E0179 USD19E0264 (01)00884838074187(21)USD19E0264 USD19E0307 (01)00884838074187(21)USD19E0307 USD19E0308 (01)00884838074187(21)USD19E0308 USD19E0309 (01)00884838074187(21)USD19E0309 USD19E0310 (01)00884838074187(21)USD19E0310 USD19E0332 (01)00884838074187(21)USD19E0332 USD19E0387 (01)00884838074187(21)USD19E0387 USD19E0388 (01)00884838074187(21)USD19E0388 USD19E0389 (01)00884838074187(21)USD19E0389 USD19E0390 (01)00884838074187(21)USD19E0390 USD19E0391 (01)00884838074187(21)USD19E0391 USD19E0392 (01)00884838074187(21)USD19E0392 USD19E0393 (01)00884838074187(21)USD19E0393 USD19E0490 (01)00884838074187(21)USD19E0490 USD19E0491 (01)00884838074187(21)USD19E0491 USD19E0532 (01)00884838074187(21)USD19E0532 USD19E0533 (01)00884838074187(21)USD19E0533 USD19E0534 (01)00884838074187(21)USD19E0534 USD19E0535 (01)00884838074187(21)USD19E0535 USD19E0536 (01)00884838074187(21)USD19E0536 USD19E0537 (01)00884838074187(21)USD19E0537 USD19E0538 (01)00884838074187(21)USD19E0538 USD19E0539 (01)00884838074187(21)USD19E0539 USD19E0540 (01)00884838074187(21)USD19E0540 USD19E0541 (01)00884838074187(21)USD19E0541 USD19E0542 (01)00884838074187(21)USD19E0542 USD19E0543 (01)00884838074187(21)USD19E0543 USD19E0544 (01)00884838074187(21)USD19E0544 USD19E0545 (01)00884838074187(21)USD19E0545 USD19E0546 (01)00884838074187(21)USD19E0546 USD19E0547 (01)00884838074187(21)USD19E0547 USD19E0548 (01)00884838074187(21)USD19E0548 USD19E0637 (01)00884838074187(21)USD19E0637 USD19E0638 (01)00884838074187(21)USD19E0638 USD19E0639 (01)00884838074187(21)USD19E0639 USD19E0640 (01)00884838074187(21)USD19E0640 USD19E0641 (01)00884838074187(21)USD19E0641 USD19E0642 (01)00884838074187(21)USD19E0642 USD19E0643 (01)00884838074187(21)USD19E0643 USD19E0682 (01)00884838074187(21)USD19E0682 USD19E0683 (01)00884838074187(21)USD19E0683 USD19E0684 (01)00884838074187(21)USD19E0684 USD19E0685 (01)00884838074187(21)USD19E0685 USD19E0686 (01)00884838074187(21)USD19E0686 USD19E0687 (01)00884838074187(21)USD19E0687 USD19E0688 (01)00884838074187(21)USD19E0688 USD19E0689 (01)00884838074187(21)USD19E0689 USD19E0695 (01)00884838074187(21)USD19E0695 USD19E0696 (01)00884838074187(21)USD19E0696 USD19E0697 (01)00884838074187(21)USD19E0697 USD19E0698 (01)00884838074187(21)USD19E0698 USD19E0699 (01)00884838074187(21)USD19E0699 USD19E0700 (01)00884838074187(21)USD19E0700 USD19E0701 (01)00884838074187(21)USD19E0701 USD19E0702 (01)00884838074187(21)USD19E0702 USD19E0703 (01)00884838074187(21)USD19E0703 USD19E0704 (01)00884838074187(21)USD19E0704 USD19E0705 (01)00884838074187(21)USD19E0705 USD19E0706 (01)00884838074187(21)USD19E0706 USD19E0707 (01)00884838074187(21)USD19E0707 USD19E0708 (01)00884838074187(21)USD19E0708 USD19E0709 (01)00884838074187(21)USD19E0709 USD19E0712 (01)00884838074187(21)USD19E0712 USD19E0713 (01)00884838074187(21)USD19E0713 USD19E0822 (01)00884838074187(21)USD19E0822 USD19E0868 (01)00884838074187(21)USD19E0868 USD19E0869 (01)00884838074187(21)USD19E0869 USD19E0871 (01)00884838074187(21)USD19E0871 USD19E0872 (01)00884838074187(21)USD19E0872 USD19E0873 (01)00884838074187(21)USD19E0873 USD19E0874 (01)00884838074187(21)USD19E0874 USD19E0875 (01)00884838074187(21)USD19E0875 USD19E0915 (01)00884838074187(21)USD19E0915 USD19E0916 (01)00884838074187(21)USD19E0916 USD19E0945 (01)00884838074187(21)USD19E0945 USD19E0946 (01)00884838074187(21)USD19E0946 USD19E0947 (01)00884838074187(21)USD19E0947 USD19E0980 (01)00884838074187(21)USD19E0980 USD19E0981 (01)00884838074187(21)USD19E0981 USD19E0982 (01)00884838074187(21)USD19E0982 USD19E0983 (01)00884838074187(21)USD19E0983 USD19E0984 (01)00884838074187(21)USD19E0984 USD19E0985 (01)00884838074187(21)USD19E0985 USD19E1009 (01)00884838074187(21)USD19E1009 USD19E1010 (01)00884838074187(21)USD19E1010 USD19E1110 (01)00884838074187(21)USD19E1110 USD19E1111 (01)00884838074187(21)USD19E1111 USD19E1146 (01)00884838074187(21)USD19E1146 USD19E1147 (01)00884838074187(21)USD19E1147 USD19E1163 (01)00884838074187(21)USD19E1163 USN16E0834 (01)00884838074187(21)USN16E0834 USN16E1033 (01)00884838074187(21)USN16E1033 USN18E0917 (01)00884838074187(21)USN18E0917 USN18E1175 (01)00884838074187(21)USN18E1175 USN18E1549 (01)00884838074187(21)USN18E1549 USN19E0061 (01)00884838074187(21)USN19E0061 USN19E0062 (01)00884838074187(21)USN19E0062 USN19E0063 (01)00884838074187(21)USN19E0063 USN19E0064 (01)00884838074187(21)USN19E0064 USN19E0065 (01)00884838074187(21)USN19E0065 USN19E0066 (01)00884838074187(21)USN19E0066 USN19E0186 (01)00884838074187(21)USN19E0186 USN19E0187 (01)00884838074187(21)USN19E0187 USN19E0188 (01)00884838074187(21)USN19E0188 USN19E0189 (01)00884838074187(21)USN19E0189 USN19E0190 (01)00884838074187(21)USN19E0190 USN19E0191 (01)00884838074187(21)USN19E0191 USN19E0192 (01)00884838074187(21)USN19E0192 USN19E0193 (01)00884838074187(21)USN19E0193 USN19E0313 (01)00884838074187(21)USN19E0313 USN19E0387 (01)00884838074187(21)USN19E0387 USN19E0467 (01)00884838074187(21)USN19E0467 USN19E0582 (01)00884838074187(21)USN19E0582 USN19E0583 (01)00884838074187(21)USN19E0583 USN19E0584 (01)00884838074187(21)USN19E0584 USN19E0585 (01)00884838074187(21)USN19E0585 USN19E0586 (01)00884838074187(21)USN19E0586 USN19E0587 (01)00884838074187(21)USN19E0587 USN19E0588 (01)00884838074187(21)USN19E0588 USN19E0590 (01)00884838074187(21)USN19E0590 USN19E0692 (01)00884838074187(21)USN19E0692 USN19E0693 (01)00884838074187(21)USN19E0693 USN19E0738 (01)00884838074187(21)USN19E0738 USN19E0739 (01)00884838074187(21)USN19E0739 USN19E0740 (01)00884838074187(21)USN19E0740 USN19E0742 (01)00884838074187(21)USN19E0742 USN19E0743 (01)00884838074187(21)USN19E0743 USN19E0744 (01)00884838074187(21)USN19E0744 USN19E0745 (01)00884838074187(21)USN19E0745 USN19E0746 (01)00884838074187(21)USN19E0746 USN19E0747 (01)00884838074187(21)USN19E0747 USN19E0811 (01)00884838074187(21)USN19E0811 USN19E0812 (01)00884838074187(21)USN19E0812 USN19E0813 (01)00884838074187(21)USN19E0813 USN19E0814 (01)00884838074187(21)USN19E0814 USN19E0815 (01)00884838074187(21)USN19E0815 USN19E0934 (01)00884838074187(21)USN19E0934 USN19E1012 (01)00884838074187(21)USN19E1012 USN19E1013 (01)00884838074187(21)USN19E1013 USN19E1143 (01)00884838074187(21)USN19E1143 USN19E1144 (01)00884838074187(21)USN19E1144 USN19E1145 (01)00884838074187(21)USN19E1145 USN19E1146 (01)00884838074187(21)USN19E1146 USN19E1176 (01)00884838074187(21)USN19E1176 USN19E1177 (01)00884838074187(21)USN19E1177 USN19E1273 (01)00884838074187(21)USN19E1273 USN19E1337 (01)00884838074187(21)USN19E1337 USN19E1338 (01)00884838074187(21)USN19E1338 USN19E1339 (01)00884838074187(21)USN19E1339 USN19E1358 (01)00884838074187(21)USN19E1358 USN19E1450 (01)00884838074187(21)USN19E1450 USN19E1451 (01)00884838074187(21)USN19E1451 USN19E1452 (01)00884838074187(21)USN19E1452 USN19E1453 (01)00884838074187(21)USN19E1453 USN19E1569 (01)00884838074187(21)USN19E1569 USN19E1623 (01)00884838074187(21)USN19E1623 USN19E1624 (01)00884838074187(21)USN19E1624 USN19E1625 (01)00884838074187(21)USN19E1625 USN19E1633 (01)00884838074187(21)USN19E1633 USN19E1634 (01)00884838074187(21)USN19E1634 USN19E1753 (01)00884838074187(21)USN19E1753 USN19E1754 (01)00884838074187(21)USN19E1754 USN19E1814 (01)00884838074187(21)USN19E1814 USN19E1815 (01)00884838074187(21)USN19E1815 USN19E1816 (01)00884838074187(21)USN19E1816 USN19E1915 (01)00884838074187(21)USN19E1915 USN19E1916 (01)00884838074187(21)USN19E1916 USN19E1917 (01)00884838074187(21)USN19E1917 USN19E1918 (01)00884838074187(21)USN19E1918 USN19E1919 (01)00884838074187(21)USN19E1919 USN19E1937 (01)00884838074187(21)USN19E1937 USN19E1938 (01)00884838074187(21)USN19E1938 USN19E1939 (01)00884838074187(21)USN19E1939 USN19E1940 (01)00884838074187(21)USN19E1940 USN19E1941 (01)00884838074187(21)USN19E1941 USN19E1942 (01)00884838074187(21)USN19E1942 USN19E2117 (01)00884838074187(21)USN19E2117 USN19E2137 (01)00884838074187(21)USN19E2137 USN19E2138 (01)00884838074187(21)USN19E2138 USN19E2139 (01)00884838074187(21)USN19E2139 USN19E2203 (01)00884838074187(21)USN19E2203 USN19E2216 (01)00884838074187(21)USN19E2216 USN19E2223 (01)00884838074187(21)USN19E2223 USN20E1214 (01)00884838074187(21)USN20E1214 USN20E1615 (01)00884838074187(21)USN20E1615 USN20E1689 (01)00884838074187(21)USN20E1689 USN20E2096 (01)00884838074187(21)USN20E2096 USN20E2097 (01)00884838074187(21)USN20E2097 USO19E0048 (01)00884838074187(21)USO19E0048 USO19E0049 (01)00884838074187(21)USO19E0049 USO19E0308 (01)00884838074187(21)USO19E0308 USO19E0309 (01)00884838074187(21)USO19E0309 USO19E0310 (01)00884838074187(21)USO19E0310 USO19E0311 (01)00884838074187(21)USO19E0311 USO19E0312 (01)00884838074187(21)USO19E0312 USO19E0313 (01)00884838074187(21)USO19E0313 USO19E0315 (01)00884838074187(21)USO19E0315 USO19E0647 (01)00884838074187(21)USO19E0647 USO19E0648 (01)00884838074187(21)USO19E0648 USO19E0649 (01)00884838074187(21)USO19E0649 USO19E0923 (01)00884838074187(21)USO19E0923 USO19E0924 (01)00884838074187(21)USO19E0924 USO19E0925 (01)00884838074187(21)USO19E0925 USO19E0987 (01)00884838074187(21)USO19E0987 USO19E1032 (01)00884838074187(21)USO19E1032 USO19E1587 (01)00884838074187(21)USO19E1587 USO19E1588 (01)00884838074187(21)USO19E1588 USO19E1589 (01)00884838074187(21)USO19E1589 USO19E1678 (01)00884838074187(21)USO19E1678 USO19E1680 (01)00884838074187(21)USO19E1680 USO19E1681 (01)00884838074187(21)USO19E1681 USO19E1682 (01)00884838074187(21)USO19E1682 USO19E1683 (01)00884838074187(21)USO19E1683 USO19E1714 (01)00884838074187(21)USO19E1714 USO19E1715 (01)00884838074187(21)USO19E1715 USO19E1716 (01)00884838074187(21)USO19E1716 USO19E1717 (01)00884838074187(21)USO19E1717 USO19E1801 (01)00884838074187(21)USO19E1801 USO19E1811 (01)00884838074187(21)USO19E1811 USO19E1812 (01)00884838074187(21)USO19E1812 USO19E1917 (01)00884838074187(21)USO19E1917 USO19E1918 (01)00884838074187(21)USO19E1918 USO19E1919 (01)00884838074187(21)USO19E1919 USO19E1920 (01)00884838074187(21)USO19E1920 USO19E1921 (01)00884838074187(21)USO19E1921 USO19E2003 (01)00884838074187(21)USO19E2003 USO19E2004 (01)00884838074187(21)USO19E2004 USO19E2005 (01)00884838074187(21)USO19E2005 USO19E2006 (01)00884838074187(21)USO19E2006 USO19E2007 (01)00884838074187(21)USO19E2007 USO19E2008 (01)00884838074187(21)USO19E2008 USO19E2009 (01)00884838074187(21)USO19E2009 USO19E2076 (01)00884838074187(21)USO19E2076 USO19E2206 (01)00884838074187(21)USO19E2206 USO19E2207 (01)00884838074187(21)USO19E2207 USO19E2304 (01)00884838074187(21)USO19E2304 USO19E2352 (01)00884838074187(21)USO19E2352 USO19E2353 (01)00884838074187(21)USO19E2353 USO19E2354 (01)00884838074187(21)USO19E2354 USO19E2418 (01)00884838074187(21)USO19E2418 USO19E2419 (01)00884838074187(21)USO19E2419 USO19E2420 (01)00884838074187(21)USO19E2420 USO19E2551 (01)00884838074187(21)USO19E2551 USO19E2552 (01)00884838074187(21)USO19E2552 USO19E2615 (01)00884838074187(21)USO19E2615 USO19E2616 (01)00884838074187(21)USO19E2616 USO19E2624 (01)00884838074187(21)USO19E2624 USO19E2628 (01)00884838074187(21)USO19E2628 USO19E2629 (01)00884838074187(21)USO19E2629 USO19E2630 (01)00884838074187(21)USO19E2630 USO19E2631 (01)00884838074187(21)USO19E2631 USO19E2632 (01)00884838074187(21)USO19E2632 USO19E2770 (01)00884838074187(21)USO19E2770 USO19E2775 (01)00884838074187(21)USO19E2775 USO19E2776 (01)00884838074187(21)USO19E2776 USO19E2777 (01)00884838074187(21)USO19E2777 USO19E2778 (01)00884838074187(21)USO19E2778 USO19E2779 (01)00884838074187(21)USO19E2779 USO19E2782 (01)00884838074187(21)USO19E2782 USO19E2783 (01)00884838074187(21)USO19E2783 USO19E2868 (01)00884838074187(21)USO19E2868 USO19E2869 (01)00884838074187(21)USO19E2869 USO19E2909 (01)00884838074187(21)USO19E2909 USO20E1064 (01)00884838074187(21)USO20E1064 Affiniti 50 Model Number: 795208 Serial Numbers/UDI Codes BZ120D0069 N/A BZ120D0177 N/A BZ120D0179 (01)00884838056244 BZ120D0180 N/A BZ120D0182 N/A BZ120D0293 N/A BZ120D0445 N/A BZ120D0481 N/A BZ217D0676 N/A BZ719D1386 N/A BZ719D1629 N/A BZ720D0995 (01)00884838097759 BZ720D0996 (01)00884838056244 BZ720D1059 (01)00884838097759 BZ819D0890 N/A BZ819D1052 N/A BZ819D1147 N/A BZ819D1205 N/A BZ819D1206 N/A BZ819D1211 N/A BZ916D0335 (01)00884838056244(21)BZ916D0335 BZD18D0191 N/A BZD18D1259 N/A PU220D1306 N/A PU220D1307 N/A PU220D1308 N/A PU220D1310 N/A PU220D1311 N/A PU420D0510 N/A PU420D0511 N/A PU420D0554 (01)00884838056244 PU420D0555 N/A PU420D0556 (01)00884838056244 PU420D0558 N/A PU420D0559 (01)00884838056244 PU420D0560 N/A PU420D0561 (01)00884838056244 PU420D0562 N/A PU420D0563 N/A PU520D0704 (01)00884838056244 PU520D0705 N/A PU520D0706 N/A PU520D0707 (01)00884838056244 PU620D0013 N/A PU620D0014 N/A PU620D0015 N/A PU620D0083 (01)00884838056244 PU720D0358 N/A PU720D0452 N/A PU720D0456 N/A PU720D0808 N/A PU820D0528 N/A PU820D0529 (01)00884838056244 PU820D0645 (01)00884838056244 PU820D0647 N/A PUD19D0827 N/A PUD19D0828 N/A PUD19D0835 N/A PUD19D0836 N/A RU220D0691 N/A RU818D1831 (01)00884838056244(21)RU818D1831 US116D0243 N/A US117D0146 (01)00884838056244(21)US117D0146 US117D0414 (01)00884838056244(21)US117D0414 US118D1125 (01)00884838056244(21)US118D1125 US119D1626 (01)00884838056244(21)US119D1626 US119D1673 (01)00884838056244(21)US119D1673 US120B0918 (01)00884838088658(21)US120B0918 US120D0013 (01)00884838056244(21)US120D0013 US120D0014 (01)00884838056244(21)US120D0014 US120D0066 (01)00884838056244(21)US120D0066 US120D0067 (01)00884838056244(21)US120D0067 US120D0160 (01)00884838056244(21)US120D0160 US120D0300 (01)00884838056244(21)US120D0300 US120D0301 (01)00884838056244(21)US120D0301 US120D0302 (01)00884838056244(21)US120D0302 US120D0303 (01)00884838056244(21)US120D0303 US120D0379 (01)00884838056244(21)US120D0379 US120D0403 (01)00884838056244(21)US120D0403 US120D0405 (01)00884838056244(21)US120D0405 US120D0406 (01)00884838056244(21)US120D0406 US120D0436 (01)00884838056244(21)US120D0436 US120D0437 (01)00884838056244(21)US120D0437 US120D0438 (01)00884838056244(21)US120D0438 US120D0506 (01)00884838056244(21)US120D0506 US120D0512 (01)00884838056244(21)US120D0512 US120D0572 (01)00884838056244(21)US120D0572 US120D0632 (01)00884838056244(21)US120D0632 US120D0642 (01)00884838056244(21)US120D0642 US120D0734 (01)00884838056244(21)US120D0734 US120D0735 (01)00884838056244(21)US120D0735 US120D0748 (01)00884838056244(21)US120D0748 US120D0749 (01)00884838056244(21)US120D0749 US120D0834 (01)00884838056244(21)US120D0834 US120D0886 (01)00884838056244(21)US120D0886 US120D0887 (01)00884838056244(21)US120D0887 US120D0953 (01)00884838056244(21)US120D0953 US120D0954 (01)00884838056244(21)US120D0954 US120D0992 (01)00884838056244(21)US120D0992 US120D0993 (01)00884838056244(21)US120D0993 US120D1062 (01)00884838056244(21)US120D1062 US120D1111 (01)00884838056244(21)US120D1111 US120D1145 (01)00884838056244(21)US120D1145 US120D1146 (01)00884838056244(21)US120D1146 US120D1148 (01)00884838056244(21)US120D1148 US120D1149 (01)00884838056244(21)US120D1149 US120D1207 (01)00884838056244(21)US120D1207 US120D1219 (01)00884838056244(21)US120D1219 US120D1220 (01)00884838056244(21)US120D1220 US120D1221 (01)00884838056244(21)US120D1221 US120D1291 (01)00884838056244(21)US120D1291 US120D1292 (01)00884838056244(21)US120D1292 US120D1293 (01)00884838056244(21)US120D1293 US120D1399 (01)00884838056244(21)US120D1399 US120D1400 (01)00884838056244(21)US120D1400 US120D1459 (01)00884838056244(21)US120D1459 US120D1524 (01)00884838056244(21)US120D1524 US120D1525 (01)00884838056244(21)US120D1525 US120D1526 (01)00884838056244(21)US120D1526 US120D1529 (01)00884838056244(21)US120D1529 US120D1530 (01)00884838056244(21)US120D1530 US120D1531 (01)00884838056244(21)US120D1531 US120D1532 (01)00884838056244(21)US120D1532 US120D1570 (01)00884838056244(21)US120D1570 US120D1571 (01)00884838056244(21)US120D1571 US120D1572 (01)00884838056244(21)US120D1572 US120D1616 (01)00884838056244(21)US120D1616 US120D1617 (01)00884838056244(21)US120D1617 US120D1618 (01)00884838056244(21)US120D1618 US120D1640 (01)00884838056244(21)US120D1640 US120D1641 (01)00884838056244(21)US120D1641 US120D1642 (01)00884838056244(21)US120D1642 US120D1643 (01)00884838056244(21)US120D1643 US120D1644 (01)00884838056244(21)US120D1644 US120D1702 (01)00884838056244(21)US120D1702 US120D1704 (01)00884838056244(21)US120D1704 US120D1705 (01)00884838056244(21)US120D1705 US120D1706 (01)00884838056244(21)US120D1706 US120D1707 (01)00884838056244(21)US120D1707 US120D1708 (01)00884838056244(21)US120D1708 US120D1709 (01)00884838056244(21)US120D1709 US120D1710 (01)00884838056244(21)US120D1710 US120D1711 (01)00884838056244(21)US120D1711 US120D1716 (01)00884838056244(21)US120D1716 US120D1717 (01)00884838056244(21)US120D1717 US120D1718 (01)00884838056244(21)US120D1718 US120D1719 (01)00884838056244(21)US120D1719 US120D1720 (01)00884838056244(21)US120D1720 US120D1721 (01)00884838056244(21)US120D1721 US120D1809 (01)00884838056244(21)US120D1809 US120D1879 (01)00884838056244(21)US120D1879 US120D1974 (01)00884838056244(21)US120D1974 US120D1975 (01)00884838056244(21)US120D1975 US121D0974 (01)00884838056244(21)US121D0974 US215D0269 N/A US215D0324 N/A US216D0179 N/A US216D0656 N/A US216D0901 N/A US217D0351 (01)00884838056244(21)US217D0351 US217D0473 (01)00884838056244(21)US217D0473 US217D0474 (01)00884838056244(21)US217D0474 US217D0475 (01)00884838056244(21)US217D0475 US218D0047 (01)00884838056244(21)US218D0047 US218D0233 (01)00884838056244(21)US218D0233 US218D1309 (01)00884838056244(21)US218D1309 US219D0048 (01)00884838056244(21)US219D0048 US219D0054 (01)00884838056244(21)US219D0054 US220D0113 (01)00884838056244(21)US220D0113 US220D0303 (01)00884838056244(21)US220D0303 US220D0304 (01)00884838056244(21)US220D0304 US220D0305 (01)00884838056244(21)US220D0305 US220D0306 (01)00884838056244(21)US220D0306 US220D0350 (01)00884838056244(21)US220D0350 US220D0351 (01)00884838056244(21)US220D0351 US220D0352 (01)00884838056244(21)US220D0352 US220D0355 (01)00884838056244(21)US220D0355 US220D0356 (01)00884838056244(21)US220D0356 US220D0443 (01)00884838056244(21)US220D0443 US220D0444 (01)00884838056244(21)US220D0444 US220D0468 (01)00884838056244(21)US220D0468 US220D0791 (01)00884838056244(21)US220D0791 US220D0828 (01)00884838056244(21)US220D0828 US220D0829 (01)00884838056244(21)US220D0829 US220D0830 (01)00884838056244(21)US220D0830 US220D0831 (01)00884838056244(21)US220D0831 US220D0832 (01)00884838056244(21)US220D0832 US220D0833 (01)00884838056244(21)US220D0833 US220D0901 (01)00884838056244(21)US220D0901 US220D0956 (01)00884838056244(21)US220D0956 US220D1002 (01)00884838056244(21)US220D1002 US220D1003 (01)00884838056244(21)US220D1003 US220D1004 (01)00884838056244(21)US220D1004 US220D1053 (01)00884838056244(21)US220D1053 US220D1054 (01)00884838056244(21)US220D1054 US220D1055 (01)00884838056244(21)US220D1055 US220D1064 (01)00884838056244(21)US220D1064 US220D1084 (01)00884838056244(21)US220D1084 US220D1085 (01)00884838056244(21)US220D1085 US220D1088 (01)00884838056244(21)US220D1088 US220D1089 (01)00884838056244(21)US220D1089 US220D1090 (01)00884838056244(21)US220D1090 US220D1091 (01)00884838056244(21)US220D1091 US220D1092 (01)00884838056244(21)US220D1092 US220D1093 (01)00884838056244(21)US220D1093 US220D1094 (01)00884838056244(21)US220D1094 US220D1095 (01)00884838056244(21)US220D1095 US220D1096 (01)00884838056244(21)US220D1096 US220D1100 (01)00884838056244(21)US220D1100 US220D1101 (01)00884838056244(21)US220D1101 US220D1102 (01)00884838056244(21)US220D1102 US220D1103 (01)00884838056244(21)US220D1103 US220D1104 (01)00884838056244(21)US220D1104 US220D1105 (01)00884838056244(21)US220D1105 US220D1116 (01)00884838056244(21)US220D1116 US220D1186 (01)00884838056244(21)US220D1186 US220D1206 (01)00884838056244(21)US220D1206 US220D1207 (01)00884838056244(21)US220D1207 US220D1208 (01)00884838056244(21)US220D1208 US220D1209 (01)00884838056244(21)US220D1209 US220D1380 (01)00884838056244(21)US220D1380 US220D1383 (01)00884838056244(21)US220D1383 US220D1384 (01)00884838056244(21)US220D1384 US220D1385 (01)00884838056244(21)US220D1385 US220D1386 (01)00884838056244(21)US220D1386 US220D1387 (01)00884838056244(21)US220D1387 US220D1388 (01)00884838056244(21)US220D1388 US220D1444 (01)00884838056244(21)US220D1444 US220D1544 (01)00884838056244(21)US220D1544 US220D1546 (01)00884838056244(21)US220D1546 US220D1547 (01)00884838056244(21)US220D1547 US220D1548 (01)00884838056244(21)US220D1548 US220D1581 (01)00884838056244(21)US220D1581 US220D1603 (01)00884838056244(21)US220D1603 US220D1604 (01)00884838056244(21)US220D1604 US220D1606 (01)00884838056244(21)US220D1606 US315D1278 N/A US315D1283 N/A US317D0531 (01)00884838056244(21)US317D0531 US317D0767 (01)00884838056244(21)US317D0767 US317D1349 (01)00884838056244(21)US317D1349 US318D2441 (01)00884838056244(21)US318D2441 US319D0032 (01)00884838056244(21)US319D0032 US319D0530 (01)00884838056244(21)US319D0530 US320D0019 (01)00884838056244(21)US320D0019 US320D0020 (01)00884838056244(21)US320D0020 US320D0021 (01)00884838056244(21)US320D0021 US320D0022 (01)00884838056244(21)US320D0022 US320D0023 (01)00884838056244(21)US320D0023 US320D0237 (01)00884838056244(21)US320D0237 US320D0238 (01)00884838056244(21)US320D0238 US320D0254 (01)00884838056244(21)US320D0254 US320D0255 (01)00884838056244(21)US320D0255 US320D0257 (01)00884838056244(21)US320D0257 US320D0259 (01)00884838056244(21)US320D0259 US320D0261 (01)00884838056244(21)US320D0261 US320D0263 (01)00884838056244(21)US320D0263 US320D0264 (01)00884838056244(21)US320D0264 US320D0296 (01)00884838056244(21)US320D0296 US320D0297 (01)00884838056244(21)US320D0297 US320D0328 (01)00884838056244(21)US320D0328 US320D0330 (01)00884838056244(21)US320D0330 US320D0332 (01)00884838056244(21)US320D0332 US320D0335 (01)00884838056244(21)US320D0335 US320D0442 (01)00884838056244(21)US320D0442 US320D0465 (01)00884838056244(21)US320D0465 US320D0613 (01)00884838056244(21)US320D0613 US320D0614 (01)00884838056244(21)US320D0614 US320D0615 (01)00884838056244(21)US320D0615 US320D0616 (01)00884838056244(21)US320D0616 US320D0617 (01)00884838056244(21)US320D0617 US320D0618 (01)00884838056244(21)US320D0618 US320D0619 (01)00884838056244(21)US320D0619 US320D0648 (01)00884838056244(21)US320D0648 US320D0718 (01)00884838056244(21)US320D0718 US320D0719 (01)00884838056244(21)US320D0719 US320D0724 (01)00884838056244(21)US320D0724 US320D0770 (01)00884838056244(21)US320D0770 US320D0771 (01)00884838056244(21)US320D0771 US320D0772 (01)00884838056244(21)US320D0772 US320D0773 (01)00884838056244(21)US320D0773 US320D0774 (01)00884838056244(21)US320D0774 US320D0776 (01)00884838056244(21)US320D0776 US320D0824 (01)00884838056244(21)US320D0824 US320D0825 (01)00884838056244(21)US320D0825 US320D0866 (01)00884838056244(21)US320D0866 US320D0868 (01)00884838056244(21)US320D0868 US320D0869 (01)00884838056244(21)US320D0869 US320D0870 (01)00884838056244(21)US320D0870 US320D0871 (01)00884838056244(21)US320D0871 US320D0872 (01)00884838056244(21)US320D0872 US320D0873 (01)00884838056244(21)US320D0873 US320D0876 (01)00884838056244(21)US320D0876 US320D0972 (01)00884838056244(21)US320D0972 US320D0973 (01)00884838056244(21)US320D0973 US320D1012 (01)00884838056244(21)US320D1012 US320D1083 (01)00884838056244(21)US320D1083 US320D1177 (01)00884838056244(21)US320D1177 US320D1178 (01)00884838056244(21)US320D1178 US320D1180 (01)00884838056244(21)US320D1180 US320D1181 (01)00884838056244(21)US320D1181 US320D1182 (01)00884838056244(21)US320D1182 US320D1183 (01)00884838056244(21)US320D1183 US320D1184 (01)00884838056244(21)US320D1184 US320D1185 (01)00884838056244(21)US320D1185 US320D1186 (01)00884838056244(21)US320D1186 US320D1191 (01)00884838056244(21)US320D1191 US320D1192 (01)00884838056244(21)US320D1192 US320D1193 (01)00884838056244(21)US320D1193 US320D1194 (01)00884838056244(21)US320D1194 US320D1195 (01)00884838056244(21)US320D1195 US320D1196 (01)00884838056244(21)US320D1196 US320D1197 (01)00884838056244(21)US320D1197 US320D1439 (01)00884838056244(21)US320D1439 US320D1440 (01)00884838056244(21)US320D1440 US320D1441 (01)00884838056244(21)US320D1441 US320D1443 (01)00884838056244(21)US320D1443 US320D1444 (01)00884838056244(21)US320D1444 US320D1445 (01)00884838056244(21)US320D1445 US320D1555 (01)00884838056244(21)US320D1555 US320D1558 (01)00884838056244(21)US320D1558 US320D1748 (01)00884838056244(21)US320D1748 US320D1879 (01)00884838056244(21)US320D1879 US320D1880 (01)00884838056244(21)US320D1880 US320D1881 (01)00884838056244(21)US320D1881 US320D1882 (01)00884838056244(21)US320D1882 US320D1883 (01)00884838056244(21)US320D1883 US320D1886 (01)00884838056244(21)US320D1886 US320D1888 (01)00884838056244(21)US320D1888 US320D1979 (01)00884838056244(21)US320D1979 US415D0023 N/A US415D0487 N/A US415D0837 N/A US416D0183 N/A US416D0967 N/A US416D0968 N/A US417D0561 (01)00884838056244(21)US417D0561 US417D0562 (01)00884838056244(21)US417D0562 US417D0702 (01)00884838056244(21)US417D0702 US417D0946 (01)00884838056244(21)US417D0946 US418D0319 (01)00884838056244(21)US418D0319 US418D0725 (01)00884838056244(21)US418D0725 US418D1482 (01)00884838056244(21)US418D1482 US418D1483 (01)00884838056244(21)US418D1483 US419D0229 (01)00884838056244(21)US419D0229 US419D1087 (01)00884838056244(21)US419D1087 US419D2033 (01)00884838056244(21)US419D2033 US420D0101 (01)00884838056244(21)US420D0101 US420D0102 (01)00884838056244(21)US420D0102 US420D0161 (01)00884838056244(21)US420D0161 US420D0162 (01)00884838056244(21)US420D0162 US420D0163 (01)00884838056244(21)US420D0163 US420D0164 (01)00884838056244(21)US420D0164 US420D0165 (01)00884838056244(21)US420D0165 US420D0447 (01)00884838056244(21)US420D0447 US420D0448 (01)00884838056244(21)US420D0448 US420D0449 (01)00884838056244(21)US420D0449 US420D0499 (01)00884838056244(21)US420D0499 US420D0540 (01)00884838056244(21)US420D0540 US420D0665 (01)00884838056244(21)US420D0665 US516D0111 N/A US516D0508 N/A US516D0518 N/A US516D0961 N/A US516D1033 N/A US516D1153 N/A US517D1159 (01)00884838056244(21)US517D1159 US518D1565 (01)00884838056244(21)US518D1565 US519D1550 (01)00884838056244(21)US519D1550 US520D0343 (01)00884838056244(21)US520D0343 US520D1704 (01)00884838056244(21)US520D1704 US615D0473 N/A US615D0894 N/A US615D1123 N/A US615D1201 N/A US615D1202 N/A US615D1203 N/A US615D1204 N/A US615D1244 N/A US615D1245 N/A US615D1246 N/A US615D1347 N/A US615D1348 N/A US615D1349 N/A US615D1350 N/A US615D1351 N/A US616D0893 N/A US617D0412 (01)00884838056244(21)US617D0412 US618D0686 (01)00884838056244(21)US618D0686 US618D0720 (01)00884838056244(21)US618D0720 US619D0655 (01)00884838056244(21)US619D0655 US619D0656 (01)00884838056244(21)US619D0656 US619D0657 (01)00884838056244(21)US619D0657 US619D0668 (01)00884838056244(21)US619D0668 US619D0684 (01)00884838056244(21)US619D0684 US619D0685 (01)00884838056244(21)US619D0685 US619D0686 (01)00884838056244(21)US619D0686 #N/A (01)00884838056244(21)US619D1430 US619D1793 (01)00884838056244(21)US619D1793 US619D1855 (01)00884838056244(21)US619D1855 US620D0648 (01)00884838056244(21)US620D0648 US620D1473 (01)00884838056244(21)US620D1473 US715D0185 N/A US715D0186 N/A US715D0797 N/A US716D0560 N/A US717D1022 (01)00884838056244(21)US717D1022 US719D1181 (01)00884838056244(21)US719D1181 US719D1182 (01)00884838056244(21)US719D1182 US719D1183 (01)00884838056244(21)US719D1183 US719D1184 (01)00884838056244(21)US719D1184 US719D1395 (01)00884838056244(21)US719D1395 US719D1433 (01)00884838056244(21)US719D1433 US719D1434 (01)00884838056244(21)US719D1434 US719D1435 (01)00884838056244(21)US719D1435 US719D1470 (01)00884838056244(21)US719D1470 US719D1471 (01)00884838056244(21)US719D1471 US719D1472 (01)00884838056244(21)US719D1472 US719D1473 (01)00884838056244(21)US719D1473 US719D1487 (01)00884838056244(21)US719D1487 US719D1488 (01)00884838056244(21)US719D1488 US719D1489 (01)00884838056244(21)US719D1489 US719D1490 (01)00884838056244(21)US719D1490 US719D1491 (01)00884838056244(21)US719D1491 US719D1816 (01)00884838056244(21)US719D1816 US719D1840 (01)00884838056244(21)US719D1840 US719D1841 (01)00884838056244(21)US719D1841 US719D1842 (01)00884838056244(21)US719D1842 US719D1843 (01)00884838056244(21)US719D1843 US719D1844 (01)00884838056244(21)US719D1844 US719D1845 (01)00884838056244(21)US719D1845 US719D1846 (01)00884838056244(21)US719D1846 US719D1860 (01)00884838056244(21)US719D1860 US719D1939 (01)00884838056244(21)US719D1939 US719D1966 (01)00884838056244(21)US719D1966 US719D2027 (01)00884838056244(21)US719D2027 US719D2028 (01)00884838056244(21)US719D2028 US719D2029 (01)00884838056244(21)US719D2029 US719D2030 (01)00884838056244(21)US719D2030 US719D2031 (01)00884838056244(21)US719D2031 US719D2032 (01)00884838056244(21)US719D2032 US719D2033 (01)00884838056244(21)US719D2033 US719D2063 (01)00884838056244(21)US719D2063 US719D2124 (01)00884838056244(21)US719D2124 US719D2125 (01)00884838056244(21)US719D2125 US719D2126 (01)00884838056244(21)US719D2126 US719D2127 (01)00884838056244(21)US719D2127 US719D2145 (01)00884838056244(21)US719D2145 US719D2146 (01)00884838056244(21)US719D2146 US719D2147 (01)00884838056244(21)US719D2147 US720D0810 (01)00884838056244(21)US720D0810 US720D1061 (01)00884838056244(21)US720D1061 US720D1062 (01)00884838056244(21)US720D1062 US814D0473 N/A US815D0494 N/A US817D0294 (01)00884838056244(21)US817D0294 US819D0066 (01)00884838056244(21)US819D0066 US819D0154 (01)00884838056244(21)US819D0154 US819D0162 (01)00884838056244(21)US819D0162 US819D0163 (01)00884838056244(21)US819D0163 US819D0164 (01)00884838056244(21)US819D0164 US819D0288 (01)00884838056244(21)US819D0288 US819D0360 (01)00884838056244(21)US819D0360 US819D0362 (01)00884838056244(21)US819D0362 US819D0363 (01)00884838056244(21)US819D0363 US819D0365 (01)00884838056244(21)US819D0365 US819D0385 (01)00884838056244(21)US819D0385 US819D0389 (01)00884838056244(21)US819D0389 US819D0428 (01)00884838056244(21)US819D0428 US819D0429 (01)00884838056244(21)US819D0429 US819D0489 (01)00884838056244(21)US819D0489 US819D0490 (01)00884838056244(21)US819D0490 US819D0491 (01)00884838056244(21)US819D0491 US819D0492 (01)00884838056244(21)US819D0492 US819D0493 (01)00884838056244(21)US819D0493 US819D0494 (01)00884838056244(21)US819D0494 US819D0495 (01)00884838056244(21)US819D0495 US819D0636 (01)00884838056244(21)US819D0636 US819D0637 (01)00884838056244(21)US819D0637 US819D0638 (01)00884838056244(21)US819D0638 US819D0639 (01)00884838056244(21)US819D0639 US819D0640 (01)00884838056244(21)US819D0640 US819D0641 (01)00884838056244(21)US819D0641 US819D0642 (01)00884838056244(21)US819D0642 US819D0701 (01)00884838056244(21)US819D0701 US819D0773 (01)00884838056244(21)US819D0773 US819D0774 (01)00884838056244(21)US819D0774 US819D0775 (01)00884838056244(21)US819D0775 US819D0776 (01)00884838056244(21)US819D0776 US819D0777 (01)00884838056244(21)US819D0777 US819D0778 (01)00884838056244(21)US819D0778 US819D0779 (01)00884838056244(21)US819D0779 US819D0780 (01)00884838056244(21)US819D0780 US819D0781 (01)00884838056244(21)US819D0781 US819D0782 (01)00884838056244(21)US819D0782 US819D0783 (01)00884838056244(21)US819D0783 US819D0784 (01)00884838056244(21)US819D0784 US819D0785 (01)00884838056244(21)US819D0785 US819D0786 (01)00884838056244(21)US819D0786 US819D0821 (01)00884838056244(21)US819D0821 US819D0883 (01)00884838056244(21)US819D0883 US819D0884 (01)00884838056244(21)US819D0884 US819D0885 (01)00884838056244(21)US819D0885 US819D0929 (01)00884838056244(21)US819D0929 US819D0997 (01)00884838056244(21)US819D0997 US819D0998 (01)00884838056244(21)US819D0998 US819D1040 (01)00884838056244(21)US819D1040 US819D1041 (01)00884838056244(21)US819D1041 US819D1101 (01)00884838056244(21)US819D1101 US819D1158 (01)00884838056244(21)US819D1158 US819D1159 (01)00884838056244(21)US819D1159 US819D1160 (01)00884838056244(21)US819D1160 US819D1161 (01)00884838056244(21)US819D1161 US819D1162 (01)00884838056244(21)US819D1162 US819D1163 (01)00884838056244(21)US819D1163 US819D1197 (01)00884838056244(21)US819D1197 US819D1198 (01)00884838056244(21)US819D1198 US819D1329 (01)00884838056244(21)US819D1329 US819D1330 (01)00884838056244(21)US819D1330 US819D1332 (01)00884838056244(21)US819D1332 US819D1334 (01)00884838056244(21)US819D1334 US819D1335 (01)00884838056244(21)US819D1335 US819D1361 (01)00884838056244(21)US819D1361 US819D1362 (01)00884838056244(21)US819D1362 US819D1363 (01)00884838056244(21)US819D1363 US819D1448 (01)00884838056244(21)US819D1448 US819D1449 (01)00884838056244(21)US819D1449 US819D1450 (01)00884838056244(21)US819D1450 US819D1451 (01)00884838056244(21)US819D1451 US819D1509 (01)00884838056244(21)US819D1509 US819D1510 (01)00884838056244(21)US819D1510 US819D1511 (01)00884838056244(21)US819D1511 US819D1513 (01)00884838056244(21)US819D1513 US819D1597 (01)00884838056244(21)US819D1597 US819D1677 (01)00884838056244(21)US819D1677 US819D1678 (01)00884838056244(21)US819D1678 US819D1743 (01)00884838056244(21)US819D1743 US819D1744 (01)00884838056244(21)US819D1744 US819D1745 (01)00884838056244(21)US819D1745 US819D1774 (01)00884838056244(21)US819D1774 US819D1818 (01)00884838056244(21)US819D1818 US819D1819 (01)00884838056244(21)US819D1819 US819D1820 (01)00884838056244(21)US819D1820 US819D1821 (01)00884838056244(21)US819D1821 US819D1822 (01)00884838056244(21)US819D1822 US819D1823 (01)00884838056244(21)US819D1823 US819D1857 (01)00884838056244(21)US819D1857 US819D1858 (01)00884838056244(21)US819D1858 US819D1859 (01)00884838056244(21)US819D1859 US819D1860 (01)00884838056244(21)US819D1860 US819D1861 (01)00884838056244(21)US819D1861 US819D1862 (01)00884838056244(21)US819D1862 US819D1916 (01)00884838056244(21)US819D1916 US819D1917 (01)00884838056244(21)US819D1917 US819D1918 (01)00884838056244(21)US819D1918 US819D1921 (01)00884838056244(21)US819D1921 US819D1922 (01)00884838056244(21)US819D1922 US819D1985 (01)00884838056244(21)US819D1985 US819D1986 (01)00884838056244(21)US819D1986 US819D1987 (01)00884838056244(21)US819D1987 US819D2058 (01)00884838056244(21)US819D2058 US819D2059 (01)00884838056244(21)US819D2059 US819D2060 (01)00884838056244(21)US819D2060 US914D0256 N/A US915D0113 N/A US915D0235 N/A US915D0749 N/A US915D0761 N/A US915D0884 N/A US915D1004 N/A US915D1159 N/A US916D0036 (01)00884838056244(21)US916D0036 US916D0683 (01)00884838056244(21)US916D0683 US916D0863 (01)00884838056244(21)US916D0863 US916D0864 (01)00884838056244(21)US916D0864 US917D0040 (01)00884838056244(21)US917D0040 US917D0041 (01)00884838056244(21)US917D0041 US917D0760 (01)00884838056244(21)US917D0760 US917D0854 (01)00884838056244(21)US917D0854 US917D1469 (01)00884838056244(21)US917D1469 US918D0173 (01)00884838056244(21)US918D0173 US918D0374 (01)00884838056244(21)US918D0374 US918D0524 (01)00884838056244(21)US918D0524 US919D0071 (01)00884838056244(21)US919D0071 US919D0072 (01)00884838056244(21)US919D0072 US919D0145 (01)00884838056244(21)US919D0145 US919D0146 (01)00884838056244(21)US919D0146 US919D0147 (01)00884838056244(21)US919D0147 US919D0188 (01)00884838056244(21)US919D0188 US919D0226 (01)00884838056244(21)US919D0226 US919D0227 (01)00884838056244(21)US919D0227 US919D0228 (01)00884838056244(21)US919D0228 US919D0274 (01)00884838056244(21)US919D0274 US919D0276 (01)00884838056244(21)US919D0276 US919D0318 (01)00884838056244(21)US919D0318 US919D0337 (01)00884838056244(21)US919D0337 US919D0338 (01)00884838056244(21)US919D0338 US919D0404 (01)00884838056244(21)US919D0404 US919D0422 (01)00884838056244(21)US919D0422 US919D0497 (01)00884838056244(21)US919D0497 US919D0498 (01)00884838056244(21)US919D0498 US919D0499 (01)00884838056244(21)US919D0499 US919D0500 (01)00884838056244(21)US919D0500 US919D0541 (01)00884838056244(21)US919D0541 US919D0542 (01)00884838056244(21)US919D0542 US919D0543 (01)00884838056244(21)US919D0543 US919D0544 (01)00884838056244(21)US919D0544 US919D0570 (01)00884838056244(21)US919D0570 US919D0571 (01)00884838056244(21)US919D0571 US919D0652 (01)00884838056244(21)US919D0652 US919D0731 (01)00884838056244(21)US919D0731 US919D0732 (01)00884838056244(21)US919D0732 US919D0733 (01)00884838056244(21)US919D0733 US919D0734 (01)00884838056244(21)US919D0734 US919D0735 (01)00884838056244(21)US919D0735 US919D0736 (01)00884838056244(21)US919D0736 US919D0737 (01)00884838056244(21)US919D0737 US919D0785 (01)00884838056244(21)US919D0785 US919D0786 (01)00884838056244(21)US919D0786 US919D0787 (01)00884838056244(21)US919D0787 US919D0788 (01)00884838056244(21)US919D0788 US919D0789 (01)00884838056244(21)US919D0789 US919D0866 (01)00884838056244(21)US919D0866 US919D0885 (01)00884838056244(21)US919D0885 US919D0886 (01)00884838056244(21)US919D0886 US919D0887 (01)00884838056244(21)US919D0887 US919D0888 (01)00884838056244(21)US919D0888 US919D0889 (01)00884838056244(21)US919D0889 US919D0890 (01)00884838056244(21)US919D0890 US919D0891 (01)00884838056244(21)US919D0891 US919D0977 (01)00884838056244(21)US919D0977 US919D1056 (01)00884838056244(21)US919D1056 US919D1068 (01)00884838056244(21)US919D1068 US919D1069 (01)00884838056244(21)US919D1069 US919D1090 (01)00884838056244(21)US919D1090 US919D1091 (01)00884838056244(21)US919D1091 US919D1092 (01)00884838056244(21)US919D1092 US919D1093 (01)00884838056244(21)US919D1093 US919D1094 (01)00884838056244(21)US919D1094 US919D1095 (01)00884838056244(21)US919D1095 US919D1159 (01)00884838056244(21)US919D1159 US919D1160 (01)00884838056244(21)US919D1160 US919D1204 (01)00884838056244(21)US919D1204 US919D1246 (01)00884838056244(21)US919D1246 US919D1247 (01)00884838056244(21)US919D1247 US919D1248 (01)00884838056244(21)US919D1248 US919D1249 (01)00884838056244(21)US919D1249 US919D1269 (01)00884838056244(21)US919D1269 US919D1270 (01)00884838056244(21)US919D1270 US919D1278 (01)00884838056244(21)US919D1278 US919D1313 (01)00884838056244(21)US919D1313 US919D1314 (01)00884838056244(21)US919D1314 US919D1315 (01)00884838056244(21)US919D1315 US919D1316 (01)00884838056244(21)US919D1316 US919D1317 (01)00884838056244(21)US919D1317 US919D1446 (01)00884838056244(21)US919D1446 US919D1448 (01)00884838056244(21)US919D1448 US919D1449 (01)00884838056244(21)US919D1449 US919D1450 (01)00884838056244(21)US919D1450 US919D1451 (01)00884838056244(21)US919D1451 US919D1452 (01)00884838056244(21)US919D1452 US919D1453 (01)00884838056244(21)US919D1453 US919D1454 (01)00884838056244(21)US919D1454 US919D1455 (01)00884838056244(21)US919D1455 US919D1456 (01)00884838056244(21)US919D1456 US919D1457 (01)00884838056244(21)US919D1457 US919D1458 (01)00884838056244(21)US919D1458 US919D1465 (01)00884838056244(21)US919D1465 US919D1616 (01)00884838056244(21)US919D1616 US919D1619 (01)00884838056244(21)US919D1619 US919D1620 (01)00884838056244(21)US919D1620 US919D1621 (01)00884838056244(21)US919D1621 US919D1716 (01)00884838056244(21)US919D1716 US919D1809 (01)00884838056244(21)US919D1809 US919D1810 (01)00884838056244(21)US919D1810 US919D1811 (01)00884838056244(21)US919D1811 US919D1866 (01)00884838056244(21)US919D1866 US919D1867 (01)00884838056244(21)US919D1867 US919D1868 (01)00884838056244(21)US919D1868 US919D1948 (01)00884838056244(21)US919D1948 US919D1949 (01)00884838056244(21)US919D1949 US919D1950 (01)00884838056244(21)US919D1950 US919D1955 (01)00884838056244(21)US919D1955 US919D1956 (01)00884838056244(21)US919D1956 US919D1957 (01)00884838056244(21)US919D1957 US919D2082 (01)00884838056244(21)US919D2082 USD14D1061 N/A USD15D0186 N/A USD15D0187 N/A USD15D0188 N/A USD15D0673 N/A USD15D0674 N/A USD15D0675 N/A USD15D0676 N/A USD15D0677 N/A USD16D0414 (01)00884838056244(21)USD16D0414 USD16D1038 (01)00884838056244(21)USD16D1038 USD16D1041 (01)00884838056244(21)USD16D1041 USD17D0881 (01)00884838056244(21)USD17D0881 USD18D0246 (01)00884838056244(21)USD18D0246 USD18D0247 (01)00884838056244(21)USD18D0247 USD19D0047 (01)00884838056244(21)USD19D0047 USD19D0048 (01)00884838056244(21)USD19D0048 USD19D0049 (01)00884838056244(21)USD19D0049 USD19D0081 (01)00884838056244(21)USD19D0081 USD19D0082 (01)00884838056244(21)USD19D0082 USD19D0101 (01)00884838056244(21)USD19D0101 USD19D0124 (01)00884838056244(21)USD19D0124 USD19D0170 (01)00884838056244(21)USD19D0170 USD19D0180 (01)00884838056244(21)USD19D0180 USD19D0181 (01)00884838056244(21)USD19D0181 USD19D0255 (01)00884838056244(21)USD19D0255 USD19D0256 (01)00884838056244(21)USD19D0256 USD19D0257 (01)00884838056244(21)USD19D0257 USD19D0331 (01)00884838056244(21)USD19D0331 USD19D0333 (01)00884838056244(21)USD19D0333 USD19D0334 (01)00884838056244(21)USD19D0334 USD19D0394 (01)00884838056244(21)USD19D0394 USD19D0397 (01)00884838056244(21)USD19D0397 USD19D0398 (01)00884838056244(21)USD19D0398 USD19D0399 (01)00884838056244(21)USD19D0399 USD19D0400 (01)00884838056244(21)USD19D0400 USD19D0401 (01)00884838056244(21)USD19D0401 USD19D0402 (01)00884838056244(21)USD19D0402 USD19D0481 (01)00884838056244(21)USD19D0481 USD19D0482 (01)00884838056244(21)USD19D0482 USD19D0483 (01)00884838056244(21)USD19D0483 USD19D0484 (01)00884838056244(21)USD19D0484 USD19D0485 (01)00884838056244(21)USD19D0485 USD19D0486 (01)00884838056244(21)USD19D0486 USD19D0510 (01)00884838056244(21)USD19D0510 USD19D0552 (01)00884838056244(21)USD19D0552 USD19D0572 (01)00884838056244(21)USD19D0572 USD19D0626 (01)00884838056244(21)USD19D0626 USD19D0627 (01)00884838056244(21)USD19D0627 USD19D0628 (01)00884838056244(21)USD19D0628 USD19D0629 (01)00884838056244(21)USD19D0629 USD19D0630 (01)00884838056244(21)USD19D0630 USD19D0631 (01)00884838056244(21)USD19D0631 USD19D0632 (01)00884838056244(21)USD19D0632 USD19D0634 (01)00884838056244(21)USD19D0634 USD19D0635 (01)00884838056244(21)USD19D0635 USD19D0690 (01)00884838056244(21)USD19D0690 USD19D0691 (01)00884838056244(21)USD19D0691 USD19D0724 (01)00884838056244(21)USD19D0724 USD19D0823 (01)00884838056244(21)USD19D0823 USD19D0863 (01)00884838056244(21)USD19D0863 USD19D0864 (01)00884838056244(21)USD19D0864 USD19D0865 (01)00884838056244(21)USD19D0865 USD19D0866 (01)00884838056244(21)USD19D0866 USD19D0867 (01)00884838056244(21)USD19D0867 USD19D0917 (01)00884838056244(21)USD19D0917 USD19D0971 (01)00884838056244(21)USD19D0971 USD19D0973 (01)00884838056244(21)USD19D0973 USD19D0998 (01)00884838056244(21)USD19D0998 USD19D0999 (01)00884838056244(21)USD19D0999 USD19D1000 (01)00884838056244(21)USD19D1000 USD19D1001 (01)00884838056244(21)USD19D1001 USD19D1002 (01)00884838056244(21)USD19D1002 USD19D1003 (01)00884838056244(21)USD19D1003 USD19D1004 (01)00884838056244(21)USD19D1004 USD19D1075 (01)00884838056244(21)USD19D1075 USD19D1080 (01)00884838056244(21)USD19D1080 USD19D1081 (01)00884838056244(21)USD19D1081 USD19D1099 (01)00884838056244(21)USD19D1099 USD19D1100 (01)00884838056244(21)USD19D1100 USD19D1101 (01)00884838056244(21)USD19D1101 USD19D1102 (01)00884838056244(21)USD19D1102 USD19D1141 (01)00884838056244(21)USD19D1141 USD19D1142 (01)00884838056244(21)USD19D1142 USD19D1143 (01)00884838056244(21)USD19D1143 USD19D1219 (01)00884838056244(21)USD19D1219 USD19D1220 (01)00884838056244(21)USD19D1220 USD19D1221 (01)00884838056244(21)USD19D1221 USD19D1240 (01)00884838056244(21)USD19D1240 USD20D0118 (01)00884838056244(21)USD20D0118 USD20D0218 (01)00884838056244(21)USD20D0218 USD20D0426 (01)00884838056244(21)USD20D0426 USN15D0092 N/A USN15D0099 N/A USN15D0283 N/A USN15D0971 N/A USN15D0972 N/A USN15D0973 N/A USN18D1482 (01)00884838056244(21)USN18D1482 USN18D1745 (01)00884838056244(21)USN18D1745 USN18D2003 (01)00884838056244(21)USN18D2003 USN19D0041 (01)00884838056244(21)USN19D0041 USN19D0067 (01)00884838056244(21)USN19D0067 USN19D0068 (01)00884838056244(21)USN19D0068 USN19D0069 (01)00884838056244(21)USN19D0069 USN19D0070 (01)00884838056244(21)USN19D0070 USN19D0173 (01)00884838056244(21)USN19D0173 USN19D0174 (01)00884838056244(21)USN19D0174 USN19D0175 (01)00884838056244(21)USN19D0175 USN19D0176 (01)00884838056244(21)USN19D0176 USN19D0240 (01)00884838056244(21)USN19D0240 USN19D0241 (01)00884838056244(21)USN19D0241 USN19D0242 (01)00884838056244(21)USN19D0242 USN19D0243 (01)00884838056244(21)USN19D0243 USN19D0244 (01)00884838056244(21)USN19D0244 USN19D0245 (01)00884838056244(21)USN19D0245 USN19D0297 (01)00884838056244(21)USN19D0297 USN19D0298 (01)00884838056244(21)USN19D0298 USN19D0299 (01)00884838056244(21)USN19D0299 USN19D0300 (01)00884838056244(21)USN19D0300 USN19D0301 (01)00884838056244(21)USN19D0301 USN19D0468 (01)00884838056244(21)USN19D0468 USN19D0510 (01)00884838056244(21)USN19D0510 USN19D0511 (01)00884838056244(21)USN19D0511 USN19D0512 (01)00884838056244(21)USN19D0512 USN19D0513 (01)00884838056244(21)USN19D0513 USN19D0514 (01)00884838056244(21)USN19D0514 USN19D0515 (01)00884838056244(21)USN19D0515 USN19D0516 (01)00884838056244(21)USN19D0516 USN19D0517 (01)00884838056244(21)USN19D0517 USN19D0518 (01)00884838056244(21)USN19D0518 USN19D0577 (01)00884838056244(21)USN19D0577 USN19D0578 (01)00884838056244(21)USN19D0578 USN19D0579 (01)00884838056244(21)USN19D0579 USN19D0591 (01)00884838056244(21)USN19D0591 USN19D0592 (01)00884838056244(21)USN19D0592 USN19D0593 (01)00884838056244(21)USN19D0593 USN19D0594 (01)00884838056244(21)USN19D0594 USN19D0595 (01)00884838056244(21)USN19D0595 USN19D0596 (01)00884838056244(21)USN19D0596 USN19D0677 (01)00884838056244(21)USN19D0677 USN19D0678 (01)00884838056244(21)USN19D0678 USN19D0679 (01)00884838056244(21)USN19D0679 USN19D0680 (01)00884838056244(21)USN19D0680 USN19D0681 (01)00884838056244(21)USN19D0681 USN19D0682 (01)00884838056244(21)USN19D0682 USN19D0683 (01)00884838056244(21)USN19D0683 USN19D0691 (01)00884838056244(21)USN19D0691 USN19D0729 (01)00884838056244(21)USN19D0729 USN19D0775 (01)00884838056244(21)USN19D0775 USN19D0801 (01)00884838056244(21)USN19D0801 USN19D0802 (01)00884838056244(21)USN19D0802 USN19D0803 (01)00884838056244(21)USN19D0803 USN19D0804 (01)00884838056244(21)USN19D0804 USN19D0816 (01)00884838056244(21)USN19D0816 USN19D0817 (01)00884838056244(21)USN19D0817 USN19D0818 (01)00884838056244(21)USN19D0818 USN19D0860 (01)00884838056244(21)USN19D0860 USN19D0863 (01)00884838056244(21)USN19D0863 USN19D0864 (01)00884838056244(21)USN19D0864 USN19D0889 (01)00884838056244(21)USN19D0889 USN19D0890 (01)00884838056244(21)USN19D0890 USN19D0891 (01)00884838056244(21)USN19D0891 USN19D0892 (01)00884838056244(21)USN19D0892 USN19D0893 (01)00884838056244(21)USN19D0893 USN19D0894 (01)00884838056244(21)USN19D0894 USN19D0895 (01)00884838056244(21)USN19D0895 USN19D0896 (01)00884838056244(21)USN19D0896 USN19D1119 (01)00884838056244(21)USN19D1119 USN19D1120 (01)00884838056244(21)USN19D1120 USN19D1121 (01)00884838056244(21)USN19D1121 USN19D1122 (01)00884838056244(21)USN19D1122 USN19D1123 (01)00884838056244(21)USN19D1123 USN19D1174 (01)00884838056244(21)USN19D1174 USN19D1175 (01)00884838056244(21)USN19D1175 USN19D1254 (01)00884838056244(21)USN19D1254 USN19D1255 (01)00884838056244(21)USN19D1255 USN19D1256 (01)00884838056244(21)USN19D1256 USN19D1257 (01)00884838056244(21)USN19D1257 USN19D1258 (01)00884838056244(21)USN19D1258 USN19D1292 (01)00884838056244(21)USN19D1292 USN19D1293 (01)00884838056244(21)USN19D1293 USN19D1294 (01)00884838056244(21)USN19D1294 USN19D1296 (01)00884838056244(21)USN19D1296 USN19D1297 (01)00884838056244(21)USN19D1297 USN19D1298 (01)00884838056244(21)USN19D1298 USN19D1299 (01)00884838056244(21)USN19D1299 USN19D1300 (01)00884838056244(21)USN19D1300 USN19D1301 (01)00884838056244(21)USN19D1301 USN19D1302 (01)00884838056244(21)USN19D1302 USN19D1393 (01)00884838056244(21)USN19D1393 USN19D1434 (01)00884838056244(21)USN19D1434 USN19D1435 (01)00884838056244(21)USN19D1435 USN19D1436 (01)00884838056244(21)USN19D1436 USN19D1437 (01)00884838056244(21)USN19D1437 USN19D1438 (01)00884838056244(21)USN19D1438 USN19D1439 (01)00884838056244(21)USN19D1439 USN19D1440 (01)00884838056244(21)USN19D1440 USN19D1441 (01)00884838056244(21)USN19D1441 USN19D1442 (01)00884838056244(21)USN19D1442 USN19D1518 (01)00884838056244(21)USN19D1518 USN19D1615 (01)00884838056244(21)USN19D1615 USN19D1616 (01)00884838056244(21)USN19D1616 USN19D1626 (01)00884838056244(21)USN19D1626 USN19D1627 (01)00884838056244(21)USN19D1627 USN19D1628 (01)00884838056244(21)USN19D1628 USN19D1629 (01)00884838056244(21)USN19D1629 USN19D1630 (01)00884838056244(21)USN19D1630 USN19D1745 (01)00884838056244(21)USN19D1745 USN19D1795 (01)00884838056244(21)USN19D1795 USN19D1893 (01)00884838056244(21)USN19D1893 USN19D1912 (01)00884838056244(21)USN19D1912 USN19D1913 (01)00884838056244(21)USN19D1913 USN19D1914 (01)00884838056244(21)USN19D1914 USN19D1943 (01)00884838056244(21)USN19D1943 USN19D1944 (01)00884838056244(21)USN19D1944 USN19D1965 (01)00884838056244(21)USN19D1965 USN19D2102 (01)00884838056244(21)USN19D2102 USN19D2103 (01)00884838056244(21)USN19D2103 USN19D2104 (01)00884838056244(21)USN19D2104 USN19D2106 (01)00884838056244(21)USN19D2106 USN19D2107 (01)00884838056244(21)USN19D2107 USN19D2108 (01)00884838056244(21)USN19D2108 USN19D2109 (01)00884838056244(21)USN19D2109 USN19D2110 (01)00884838056244(21)USN19D2110 USN19D2111 (01)00884838056244(21)USN19D2111 USN19D2112 (01)00884838056244(21)USN19D2112 USN19D2126 (01)00884838056244(21)USN19D2126 USN19D2127 (01)00884838056244(21)USN19D2127 USN19D2128 (01)00884838056244(21)USN19D2128 USN19D2129 (01)00884838056244(21)USN19D2129 USN20D1688 (01)00884838056244(21)USN20D1688 USO15D0430 N/A USO15D0431 N/A USO15D0974 N/A USO16D0023 (01)00884838056244(21)USO16D0023 USO17D0803 (01)00884838056244(21)USO17D0803 USO17D1054 (01)00884838056244(21)USO17D1054 USO17D1197 (01)00884838056244(21)USO17D1197 USO17D1341 (01)00884838056244(21)USO17D1341 USO18D0424 (01)00884838056244(21)USO18D0424 USO18D0425 (01)00884838056244(21)USO18D0425 USO18D0752 (01)00884838056244(21)USO18D0752 USO18D1609 (01)00884838056244(21)USO18D1609 USO19D0047 (01)00884838056244(21)USO19D0047 USO19D0175 (01)00884838056244(21)USO19D0175 USO19D0183 (01)00884838056244(21)USO19D0183 USO19D0221 (01)00884838056244(21)USO19D0221 USO19D0316 (01)00884838056244(21)USO19D0316 USO19D0418 (01)00884838056244(21)USO19D0418 USO19D0582 (01)00884838056244(21)USO19D0582 USO19D0583 (01)00884838056244(21)USO19D0583 USO19D0584 (01)00884838056244(21)USO19D0584 USO19D0585 (01)00884838056244(21)USO19D0585 USO19D0586 (01)00884838056244(21)USO19D0586 USO19D0811 (01)00884838056244(21)USO19D0811 USO19D0812 (01)00884838056244(21)USO19D0812 USO19D0815 (01)00884838056244(21)USO19D0815 USO19D0912 (01)00884838056244(21)USO19D0912 USO19D0935 (01)00884838056244(21)USO19D0935 USO19D1582 (01)00884838056244(21)USO19D1582 USO19D1585 (01)00884838056244(21)USO19D1585 USO19D1586 (01)00884838056244(21)USO19D1586 USO19D1599 (01)00884838056244(21)USO19D1599 USO19D1600 (01)00884838056244(21)USO19D1600 USO19D1667 (01)00884838056244(21)USO19D1667 USO19D1668 (01)00884838056244(21)USO19D1668 USO19D1669 (01)00884838056244(21)USO19D1669 USO19D1705 (01)00884838056244(21)USO19D1705 USO19D1706 (01)00884838056244(21)USO19D1706 USO19D1707 (01)00884838056244(21)USO19D1707 USO19D1726 (01)00884838056244(21)USO19D1726 USO19D1727 (01)00884838056244(21)USO19D1727 USO19D1728 (01)00884838056244(21)USO19D1728 USO19D1729 (01)00884838056244(21)USO19D1729 USO19D1730 (01)00884838056244(21)USO19D1730 USO19D1803 (01)00884838056244(21)USO19D1803 USO19D1804 (01)00884838056244(21)USO19D1804 USO19D1805 (01)00884838056244(21)USO19D1805 USO19D1826 (01)00884838056244(21)USO19D1826 USO19D1911 (01)00884838056244(21)USO19D1911 USO19D1912 (01)00884838056244(21)USO19D1912 USO19D1913 (01)00884838056244(21)USO19D1913 USO19D2014 (01)00884838056244(21)USO19D2014 USO19D2015 (01)00884838056244(21)USO19D2015 USO19D2016 (01)00884838056244(21)USO19D2016 USO19D2031 (01)00884838056244(21)USO19D2031 USO19D2072 (01)00884838056244(21)USO19D2072 USO19D2120 (01)00884838056244(21)USO19D2120 USO19D2199 (01)00884838056244(21)USO19D2199 USO19D2358 (01)00884838056244(21)USO19D2358 USO19D2359 (01)00884838056244(21)USO19D2359 USO19D2360 (01)00884838056244(21)USO19D2360 USO19D2376 (01)00884838056244(21)USO19D2376 USO19D2377 (01)00884838056244(21)USO19D2377 USO19D2378 (01)00884838056244(21)USO19D2378 USO19D2505 (01)00884838056244(21)USO19D2505 USO19D2506 (01)00884838056244(21)USO19D2506 USO19D2547 (01)00884838056244(21)USO19D2547 USO19D2548 (01)00884838056244(21)USO19D2548 USO19D2549 (01)00884838056244(21)USO19D2549 USO19D2550 (01)00884838056244(21)USO19D2550 USO19D2621 (01)00884838056244(21)USO19D2621 USO19D2622 (01)00884838056244(21)USO19D2622 USO19D2623 (01)00884838056244(21)USO19D2623 USO19D2766 (01)00884838056244(21)USO19D2766 USO19D2767 (01)00884838056244(21)USO19D2767 USO19D2799 (01)00884838056244(21)USO19D2799 USO19D2867 (01)00884838056244(21)USO19D2867 USO19D2913 (01)00884838056244(21)USO19D2913 USO19D2914 (01)00884838056244(21)USO19D2914 USO19D2915 (01)00884838056244(21)USO19D2915 USO20D2118 (01)00884838056244(21)USO20D2118 Affiniti 50 Model Number: 795118 Serial Numbers/UDI Codes US319D0387 (01)00884838056244(21)US319D0387 US519D1329 (01)00884838056244(21)US519D1329 USN18D0258 (01)00884838056244(21)USN18D0258 USN18D1294 (01)00884838056244(21)USN18D1294 USN19D0472 (01)00884838056244(21)USN19D0472 USN19D0777 (01)00884838056244(21)USN19D0777 USO19D0318 (01)00884838056244(21)USO19D0318 USO19D0569 (01)00884838056244(21)USO19D0569 Affiniti 70 Model Number: 795210 Serial Numbers/UDI Codes BZ119F0375 N/A BZ119F0864 N/A BZ120F0638 N/A BZ120F0867 N/A BZ120F0868 N/A BZ120F0911 N/A BZ120F0913 N/A BZ120F0989 N/A BZ120F0990 N/A BZ120F1061 N/A BZ120F1078 N/A BZ120F1081 N/A BZ120F1154 N/A BZ120F1155 N/A BZ120F1157 N/A BZ120F1209 N/A BZ120F1212 N/A BZ120F1213 N/A BZ120F1284 N/A BZ120F1895 N/A BZ120F1948 (01)00884838056251 BZ120F2068 (01)00884838056251 BZ218F1111 (01)00884838056251 BZ220F1621 N/A BZ220F1622 N/A BZ318F1335 N/A BZ320F0234 N/A BZ320F1758 N/A BZ320F1759 N/A BZ320F1990 N/A BZ416F0036 N/A BZ420F0123 N/A BZ420F0124 N/A BZ720F1656 (01)00884838097766 BZ720F1658 (01)00884838056251 BZ720F1659 N/A BZ720F1660 N/A BZ720F1662 N/A BZ720F1663 N/A BZ720F1664 (01)00884838056251 BZ720F1665 N/A BZ720F1736 N/A BZ720F1737 (01)00884838097766 BZ720F1738 N/A BZ720F1739 (01)00884838097766 BZ720F1740 (01)00884838056251 BZ720F1741 (01)00884838097766 BZ720F1743 N/A BZ720F1774 N/A BZ720F1775 (01)00884838056251 BZ720F1776 (01)00884838097766 BZ720F1808 N/A BZ720F1809 N/A BZ720F1810 (01)00884838097766 BZ720F1811 N/A BZ720F1812 (01)00884838097766 BZ720F1813 N/A BZ720F1814 (01)00884838097766 BZ720F1815 (01)00884838097766 BZ720F1828 N/A BZ720F1829 (01)00884838097766 BZ720F1830 N/A BZ720F1831 (01)00884838097766 BZ720F1832 (01)00884838097766 BZ820F0059 (01)00884838097766 BZ820F1003 (01)00884838097766 BZ820F1004 N/A BZ820F1005 (01)00884838097766 BZ820F1006 (01)00884838097766 BZ820F1007 (01)00884838097766 BZ820F1065 N/A BZ820F1066 (01)00884838097766 BZ820F1067 N/A BZ820F1068 N/A BZ820F1145 (01)00884838097766 BZ820F1146 N/A BZ820F1147 N/A BZ820F1222 (01)00884838097766 BZ820F1224 (01)00884838097766 BZ820F1225 (01)00884838097766 BZ820F1304 N/A BZ820F1305 (01)00884838097766 BZ820F1306 N/A BZ820F1307 (01)00884838056251 BZ820F1427 N/A BZ820F1428 (01)00884838056251 BZ918F1610 N/A BZD18F0127 N/A BZN18F0482 N/A BZN18F0588 N/A BZN18F0590 N/A BZN18F0801 N/A BZN18F0902 N/A BZN19F0810 N/A BZN19F1897 N/A BZO20F1898 (01)00884838097766 BZO20F1899 (01)00884838097766 BZO20F1900 (01)00884838097766 BZO20F1901 (01)00884838097766 BZO20F1902 (01)00884838097766 BZO20F2000 (01)00884838097766 BZO20F2001 (01)00884838097766 BZO20F2002 (01)00884838097766 BZO20F2003 (01)00884838097766 BZO20F2004 (01)00884838097766 BZO20F2005 (01)00884838097766 PU120F1540 N/A PU120F1541 N/A PU120F1542 N/A PU120F1543 N/A PU120F1544 N/A PU120F1545 N/A PU120F1546 N/A PU120F1547 (01)00884838056251 PU120F1747 N/A PU120F1748 N/A PU120F1749 N/A PU120F1750 N/A PU120F1751 N/A PU120F1752 N/A PU120F1753 (01)00884838056251 PU220F1271 (01)00884838056251 PU220F1272 N/A PU220F1273 N/A PU220F1274 N/A PU220F1275 (01)00884838056251 PU220F1277 N/A PU220F1278 N/A PU220F1279 N/A PU220F1280 N/A PU220F1281 N/A PU220F1305 (01)00884838056251 PU420F0334 N/A PU420F0335 (01)00884838056251 PU420F0439 N/A PU420F0661 N/A PU420F0662 (01)00884838097766 PU420F0663 N/A PU420F0754 N/A PU420F0755 N/A PU420F0756 (01)00884838097766 PU420F0757 N/A PU420F0758 N/A PU520F0709 (01)00884838056251 PU520F0710 N/A PU520F0711 (01)00884838056251 PU520F0712 (01)00884838056251 PU520F0862 N/A PU520F0863 N/A PU520F0865 N/A PU520F0866 (01)00884838056251 PU620F0016 N/A PU620F0019 N/A PU620F0021 N/A PU620F0253 (01)00884838056251 PU620F0254 (01)00884838056251 PU620F0255 N/A PU620F0256 (01)00884838056251 PU620F0257 N/A PU620F0308 (01)00884838056251 PU620F0309 (01)00884838056251 PU620F0310 (01)00884838056251 PU620F0312 (01)00884838056251 PU720F0450 N/A PU720F0474 (01)00884838056251 PU720F0476 (01)00884838056251 PU720F0612 (01)00884838097766 PU720F0613 (01)00884838056251 PU720F0614 N/A PU720F0618 N/A PU820F0717 (01)00884838056251 PU820F0718 (01)00884838097766 PU820F0720 (01)00884838056251 PU820F0723 N/A PU820F0724 N/A PU820F0746 (01)00884838056251 PU820F0800 N/A PU820F0801 (01)00884838097766 PU820F0910 N/A PU820F0913 (01)00884838056251 PU820F0916 (01)00884838097766 PU820F0919 N/A PUD19F0774 N/A PUD19F0776 N/A PUD19F0830 N/A RU519F1031 N/A RU620F1384 N/A US115F0516 N/A US116F0745 N/A US116F0750 N/A US116F1057 N/A US116F1105 N/A US117F0094 (01)00884838056251(21)US117F0094 US117F0297 (01)00884838056251(21)US117F0297 US117F0362 (01)00884838056251(21)US117F0362 US117F0420 (01)00884838056251(21)US117F0420 US117F0542 (01)00884838056251(21)US117F0542 US117F0598 (01)00884838056251(21)US117F0598 US117F0667 (01)00884838056251(21)US117F0667 US117F1243 (01)00884838056251(21)US117F1243 US118F0267 (01)00884838056251(21)US118F0267 US118F0268 (01)00884838056251(21)US118F0268 US118F0269 (01)00884838056251(21)US118F0269 US118F0507 (01)00884838056251(21)US118F0507 US118F0508 (01)00884838056251(21)US118F0508 US118F1068 (01)00884838056251(21)US118F1068 US118F1069 (01)00884838056251(21)US118F1069 US118F1382 (01)00884838056251(21)US118F1382 US118F1541 (01)00884838056251(21)US118F1541 US118F1566 (01)00884838056251(21)US118F1566 US119F0097 (01)00884838056251(21)US119F0097 US119F0111 (01)00884838056251(21)US119F0111 US119F0146 (01)00884838056251(21)US119F0146 US119F0223 (01)00884838056251(21)US119F0223 US119F0232 (01)00884838056251(21)US119F0232 US119F0618 (01)00884838056251(21)US119F0618 US119F0624 (01)00884838056251(21)US119F0624 US119F0625 (01)00884838056251(21)US119F0625 US119F0626 (01)00884838056251(21)US119F0626 US119F0627 (01)00884838056251(21)US119F0627 US119F0628 (01)00884838056251(21)US119F0628 US119F0629 (01)00884838056251(21)US119F0629 US119F0630 (01)00884838056251(21)US119F0630 US119F0680 (01)00884838056251(21)US119F0680 US119F0706 (01)00884838056251(21)US119F0706 US119F0709 (01)00884838056251(21)US119F0709 US119F1129 (01)00884838056251(21)US119F1129 US119F1130 (01)00884838056251(21)US119F1130 US119F1131 (01)00884838056251(21)US119F1131 US119F1282 (01)00884838056251(21)US119F1282 US119F1290 (01)00884838056251(21)US119F1290 US119F1681 (01)00884838056251(21)US119F1681 US119F1682 (01)00884838056251(21)US119F1682 US120F0017 (01)00884838056251(21)US120F0017 US120F0018 (01)00884838056251(21)US120F0018 US120F0138 (01)00884838056251(21)US120F0138 US120F0139 (01)00884838056251(21)US120F0139 US120F0140 (01)00884838056251(21)US120F0140 US120F0167 (01)00884838056251(21)US120F0167 US120F0169 (01)00884838056251(21)US120F0169 US120F0171 (01)00884838056251(21)US120F0171 US120F0172 (01)00884838056251(21)US120F0172 US120F0173 (01)00884838056251(21)US120F0173 US120F0174 (01)00884838056251(21)US120F0174 US120F0175 (01)00884838056251(21)US120F0175 US120F0316 (01)00884838056251(21)US120F0316 US120F0317 (01)00884838056251(21)US120F0317 US120F0321 (01)00884838056251(21)US120F0321 US120F0322 (01)00884838056251(21)US120F0322 US120F0323 (01)00884838056251(21)US120F0323 US120F0324 (01)00884838056251(21)US120F0324 US120F0325 (01)00884838056251(21)US120F0325 US120F0326 (01)00884838056251(21)US120F0326 US120F0327 (01)00884838056251(21)US120F0327 US120F0328 (01)00884838056251(21)US120F0328 US120F0329 (01)00884838056251(21)US120F0329 US120F0330 (01)00884838056251(21)US120F0330 US120F0331 (01)00884838056251(21)US120F0331 US120F0332 (01)00884838056251(21)US120F0332 US120F0333 (01)00884838056251(21)US120F0333 US120F0334 (01)00884838056251(21)US120F0334 US120F0335 (01)00884838056251(21)US120F0335 US120F0361 (01)00884838056251(21)US120F0361 US120F0439 (01)00884838056251(21)US120F0439 US120F0440 (01)00884838056251(21)US120F0440 US120F0441 (01)00884838056251(21)US120F0441 US120F0442 (01)00884838056251(21)US120F0442 US120F0443 (01)00884838056251(21)US120F0443 US120F0444 (01)00884838056251(21)US120F0444 US120F0509 (01)00884838056251(21)US120F0509 US120F0510 (01)00884838056251(21)US120F0510 US120F0643 (01)00884838056251(21)US120F0643 US120F0644 (01)00884838056251(21)US120F0644 US120F0645 (01)00884838056251(21)US120F0645 US120F0646 (01)00884838056251(21)US120F0646 US120F0647 (01)00884838056251(21)US120F0647 US120F0648 (01)00884838056251(21)US120F0648 US120F0649 (01)00884838056251(21)US120F0649 US120F0650 (01)00884838056251(21)US120F0650 US120F0750 (01)00884838056251(21)US120F0750 US120F0751 (01)00884838056251(21)US120F0751 US120F0753 (01)00884838056251(21)US120F0753 US120F0754 (01)00884838056251(21)US120F0754 US120F0755 (01)00884838056251(21)US120F0755 US120F0757 (01)00884838056251(21)US120F0757 US120F0835 (01)00884838056251(21)US120F0835 US120F0836 (01)00884838056251(21)US120F0836 US120F0837 (01)00884838056251(21)US120F0837 US120F0838 (01)00884838056251(21)US120F0838 US120F0840 (01)00884838056251(21)US120F0840 US120F0841 (01)00884838056251(21)US120F0841 US120F0873 (01)00884838056251(21)US120F0873 US120F0888 (01)00884838056251(21)US120F0888 US120F0889 (01)00884838056251(21)US120F0889 US120F0890 (01)00884838056251(21)US120F0890 US120F0891 (01)00884838056251(21)US120F0891 US120F0956 (01)00884838056251(21)US120F0956 US120F0994 (01)00884838056251(21)US120F0994 US120F0995 (01)00884838056251(21)US120F0995 US120F0996 (01)00884838056251(21)US120F0996 US120F0997 (01)00884838056251(21)US120F0997 US120F0998 (01)00884838056251(21)US120F0998 US120F0999 (01)00884838056251(21)US120F0999 US120F1000 (01)00884838056251(21)US120F1000 US120F1085 (01)00884838056251(21)US120F1085 US120F1086 (01)00884838056251(21)US120F1086 US120F1087 (01)00884838056251(21)US120F1087 US120F1088 (01)00884838056251(21)US120F1088 US120F1090 (01)00884838056251(21)US120F1090 US120F1091 (01)00884838056251(21)US120F1091 US120F1185 (01)00884838056251(21)US120F1185 US120F1186 (01)00884838056251(21)US120F1186 US120F1187 (01)00884838056251(21)US120F1187 US120F1217 (01)00884838056251(21)US120F1217 US120F1294 (01)00884838056251(21)US120F1294 US120F1295 (01)00884838056251(21)US120F1295 US120F1296 (01)00884838056251(21)US120F1296 US120F1297 (01)00884838056251(21)US120F1297 US120F1354 (01)00884838056251(21)US120F1354 US120F1406 (01)00884838056251(21)US120F1406 US120F1408 (01)00884838056251(21)US120F1408 US120F1460 (01)00884838056251(21)US120F1460 US120F1461 (01)00884838056251(21)US120F1461 US120F1462 (01)00884838056251(21)US120F1462 US120F1463 (01)00884838056251(21)US120F1463 US120F1464 (01)00884838056251(21)US120F1464 US120F1465 (01)00884838056251(21)US120F1465 US120F1522 (01)00884838056251(21)US120F1522 US120F1523 (01)00884838056251(21)US120F1523 US120F1566 (01)00884838056251(21)US120F1566 US120F1569 (01)00884838056251(21)US120F1569 US120F1573 (01)00884838056251(21)US120F1573 US120F1574 (01)00884838056251(21)US120F1574 US120F1619 (01)00884838056251(21)US120F1619 US120F1620 (01)00884838056251(21)US120F1620 US120F1621 (01)00884838056251(21)US120F1621 US120F1645 (01)00884838056251(21)US120F1645 US120F1646 (01)00884838056251(21)US120F1646 US120F1647 (01)00884838056251(21)US120F1647 US120F1648 (01)00884838056251(21)US120F1648 US120F1649 (01)00884838056251(21)US120F1649 US120F1714 (01)00884838056251(21)US120F1714 US120F1715 (01)00884838056251(21)US120F1715 US120F1722 (01)00884838056251(21)US120F1722 US120F1810 (01)00884838056251(21)US120F1810 US120F1811 (01)00884838056251(21)US120F1811 US120F1812 (01)00884838056251(21)US120F1812 US120F1813 (01)00884838056251(21)US120F1813 US120F1814 (01)00884838056251(21)US120F1814 US120F1815 (01)00884838056251(21)US120F1815 US120F1816 (01)00884838056251(21)US120F1816 US120F1817 (01)00884838056251(21)US120F1817 US120F1818 (01)00884838056251(21)US120F1818 US120F1819 (01)00884838056251(21)US120F1819 US120F1820 (01)00884838056251(21)US120F1820 US120F1821 (01)00884838056251(21)US120F1821 US120F1881 (01)00884838056251(21)US120F1881 US120F1882 (01)00884838056251(21)US120F1882 US120F1883 (01)00884838056251(21)US120F1883 US120F1884 (01)00884838056251(21)US120F1884 US120F1885 (01)00884838056251(21)US120F1885 US120F1886 (01)00884838056251(21)US120F1886 US120F1887 (01)00884838056251(21)US120F1887 US120F1888 (01)00884838056251(21)US120F1888 US120F1889 (01)00884838056251(21)US120F1889 US120F1891 (01)00884838056251(21)US120F1891 US120F1894 (01)00884838056251(21)US120F1894 US120F1950 (01)00884838056251(21)US120F1950 US120F1951 (01)00884838056251(21)US120F1951 US120F1952 (01)00884838056251(21)US120F1952 US120F1953 (01)00884838056251(21)US120F1953 US120F1969 (01)00884838056251(21)US120F1969 US120F1970 (01)00884838056251(21)US120F1970 US120F1971 (01)00884838056251(21)US120F1971 US121F0604 (01)00884838056251(21)US121F0604 US121F0898 (01)00884838056251(21)US121F0898 US121F0990 (01)00884838056251(21)US121F0990 US121F0991 (01)00884838056251(21)US121F0991 US121F0992 (01)00884838056251(21)US121F0992 US121F0993 (01)00884838056251(21)US121F0993 US121F0994 (01)00884838056251(21)US121F0994 US121F0995 (01)00884838056251(21)US121F0995 US215F0239 N/A US215F0748 N/A US216F0226 N/A US217F0579 (01)00884838056251(21)US217F0579 US217F0580 (01)00884838056251(21)US217F0580 US217F0626 (01)00884838056251(21)US217F0626 US217F0852 (01)00884838056251(21)US217F0852 US218F0108 (01)00884838056251(21)US218F0108 US218F0275 (01)00884838056251(21)US218F0275 US218F1420 (01)00884838056251(21)US218F1420 US219F0466 (01)00884838056251(21)US219F0466 US219F0467 (01)00884838056251(21)US219F0467 US219F0543 (01)00884838056251(21)US219F0543 US219F0546 (01)00884838056251(21)US219F0546 US219F0921 (01)00884838056251(21)US219F0921 US219F1321 (01)00884838056251(21)US219F1321 US219F1322 (01)00884838056251(21)US219F1322 US219F1323 (01)00884838056251(21)US219F1323 US219F1402 (01)00884838056251(21)US219F1402 US220F0045 (01)00884838056251(21)US220F0045 US220F0046 (01)00884838056251(21)US220F0046 US220F0047 (01)00884838056251(21)US220F0047 US220F0114 (01)00884838056251(21)US220F0114 US220F0116 (01)00884838056251(21)US220F0116 US220F0117 (01)00884838056251(21)US220F0117 US220F0118 (01)00884838056251(21)US220F0118 US220F0119 (01)00884838056251(21)US220F0119 US220F0129 (01)00884838056251(21)US220F0129 US220F0132 (01)00884838056251(21)US220F0132 US220F0133 (01)00884838056251(21)US220F0133 US220F0134 (01)00884838056251(21)US220F0134 US220F0135 (01)00884838056251(21)US220F0135 US220F0136 (01)00884838056251(21)US220F0136 US220F0290 (01)00884838056251(21)US220F0290 US220F0311 (01)00884838056251(21)US220F0311 US220F0312 (01)00884838056251(21)US220F0312 US220F0313 (01)00884838056251(21)US220F0313 US220F0314 (01)00884838056251(21)US220F0314 US220F0315 (01)00884838056251(21)US220F0315 US220F0360 (01)00884838056251(21)US220F0360 US220F0361 (01)00884838056251(21)US220F0361 US220F0362 (01)00884838056251(21)US220F0362 US220F0363 (01)00884838056251(21)US220F0363 US220F0364 (01)00884838056251(21)US220F0364 US220F0365 (01)00884838056251(21)US220F0365 US220F0422 (01)00884838056251(21)US220F0422 US220F0457 (01)00884838056251(21)US220F0457 US220F0462 (01)00884838056251(21)US220F0462 US220F0463 (01)00884838056251(21)US220F0463 US220F0464 (01)00884838056251(21)US220F0464 US220F0465 (01)00884838056251(21)US220F0465 US220F0466 (01)00884838056251(21)US220F0466 US220F0467 (01)00884838056251(21)US220F0467 US220F0471 (01)00884838056251(21)US220F0471 US220F0589 (01)00884838056251(21)US220F0589 US220F0590 (01)00884838056251(21)US220F0590 US220F0591 (01)00884838056251(21)US220F0591 US220F0705 (01)00884838056251(21)US220F0705 US220F0764 (01)00884838056251(21)US220F0764 US220F0765 (01)00884838056251(21)US220F0765 US220F0766 (01)00884838056251(21)US220F0766 US220F0767 (01)00884838056251(21)US220F0767 US220F0768 (01)00884838056251(21)US220F0768 US220F0769 (01)00884838056251(21)US220F0769 US220F0777 (01)00884838056251(21)US220F0777 US220F0783 (01)00884838056251(21)US220F0783 US220F0784 (01)00884838056251(21)US220F0784 US220F0785 (01)00884838056251(21)US220F0785 US220F0786 (01)00884838056251(21)US220F0786 US220F0792 (01)00884838056251(21)US220F0792 US220F0793 (01)00884838056251(21)US220F0793 US220F0794 (01)00884838056251(21)US220F0794 US220F0826 (01)00884838056251(21)US220F0826 US220F0827 (01)00884838056251(21)US220F0827 US220F0835 (01)00884838056251(21)US220F0835 US220F0902 (01)00884838056251(21)US220F0902 US220F0903 (01)00884838056251(21)US220F0903 US220F0904 (01)00884838056251(21)US220F0904 US220F0905 (01)00884838056251(21)US220F0905 US220F0906 (01)00884838056251(21)US220F0906 US220F0907 (01)00884838056251(21)US220F0907 US220F0908 (01)00884838056251(21)US220F0908 US220F0909 (01)00884838056251(21)US220F0909 US220F0910 (01)00884838056251(21)US220F0910 US220F0911 (01)00884838056251(21)US220F0911 US220F0912 (01)00884838056251(21)US220F0912 US220F0958 (01)00884838056251(21)US220F0958 US220F1001 (01)00884838056251(21)US220F1001 US220F1005 (01)00884838056251(21)US220F1005 US220F1006 (01)00884838056251(21)US220F1006 US220F1007 (01)00884838056251(21)US220F1007 US220F1008 (01)00884838056251(21)US220F1008 US220F1009 (01)00884838056251(21)US220F1009 US220F1010 (01)00884838056251(21)US220F1010 US220F1011 (01)00884838056251(21)US220F1011 US220F1012 (01)00884838056251(21)US220F1012 US220F1057 (01)00884838056251(21)US220F1057 US220F1059 (01)00884838056251(21)US220F1059 US220F1060 (01)00884838056251(21)US220F1060 US220F1066 (01)00884838056251(21)US220F1066 US220F1068 (01)00884838056251(21)US220F1068 US220F1099 (01)00884838056251(21)US220F1099 US220F1210 (01)00884838056251(21)US220F1210 US220F1211 (01)00884838056251(21)US220F1211 US220F1212 (01)00884838056251(21)US220F1212 US220F1213 (01)00884838056251(21)US220F1213 US220F1214 (01)00884838056251(21)US220F1214 US220F1215 (01)00884838056251(21)US220F1215 US220F1216 (01)00884838056251(21)US220F1216 US220F1371 (01)00884838056251(21)US220F1371 US220F1372 (01)00884838056251(21)US220F1372 US220F1373 (01)00884838056251(21)US220F1373 US220F1407 (01)00884838056251(21)US220F1407 US220F1409 (01)00884838056251(21)US220F1409 US220F1410 (01)00884838056251(21)US220F1410 US220F1411 (01)00884838056251(21)US220F1411 US220F1412 (01)00884838056251(21)US220F1412 US220F1442 (01)00884838056251(21)US220F1442 US220F1485 (01)00884838056251(21)US220F1485 US220F1486 (01)00884838056251(21)US220F1486 US220F1549 (01)00884838056251(21)US220F1549 US220F1610 (01)00884838056251(21)US220F1610 US220F1611 (01)00884838056251(21)US220F1611 US220F1614 (01)00884838056251(21)US220F1614 US220F1615 (01)00884838056251(21)US220F1615 US315F0013 N/A US315F0617 N/A US315F0989 N/A US315F1101 N/A US316F0454 N/A US316F0703 N/A US316F0844 N/A US316F0873 N/A US316F1063 N/A US317F0379 (01)00884838056251(21)US317F0379 US317F1268 (01)00884838056251(21)US317F1268 US318F0446 (01)00884838056251(21)US318F0446 US318F0505 (01)00884838056251(21)US318F0505 US318F1646 (01)00884838056251(21)US318F1646 US318F1758 (01)00884838056251(21)US318F1758 US318F1912 (01)00884838056251(21)US318F1912 US318F1913 (01)00884838056251(21)US318F1913 US318F1914 (01)00884838056251(21)US318F1914 US318F2592 (01)00884838056251(21)US318F2592 US319F0056 (01)00884838056251(21)US319F0056 US319F0592 (01)00884838056251(21)US319F0592 US319F0593 (01)00884838056251(21)US319F0593 US319F0594 (01)00884838056251(21)US319F0594 US319F0595 (01)00884838056251(21)US319F0595 US319F0596 (01)00884838056251(21)US319F0596 US319F0597 (01)00884838056251(21)US319F0597 US319F0617 (01)00884838056251(21)US319F0617 US319F0618 (01)00884838056251(21)US319F0618 US319F0619 (01)00884838056251(21)US319F0619 US319F0620 (01)00884838056251(21)US319F0620 US319F0621 (01)00884838056251(21)US319F0621 US319F0922 (01)00884838056251(21)US319F0922 US319F0923 (01)00884838056251(21)US319F0923 US319F0924 (01)00884838056251(21)US319F0924 US319F0926 (01)00884838056251(21)US319F0926 US319F1037 (01)00884838056251(21)US319F1037 US319F1200 (01)00884838056251(21)US319F1200 US319F1312 (01)00884838056251(21)US319F1312 US319F1313 (01)00884838056251(21)US319F1313 US319F1314 (01)00884838056251(21)US319F1314 US319F1315 (01)00884838056251(21)US319F1315 US320F0026 (01)00884838056251(21)US320F0026 US320F0027 (01)00884838056251(21)US320F0027 US320F0028 (01)00884838056251(21)US320F0028 US320F0125 (01)00884838056251(21)US320F0125 US320F0126 (01)00884838056251(21)US320F0126 US320F0127 (01)00884838056251(21)US320F0127 US320F0128 (01)00884838056251(21)US320F0128 US320F0129 (01)00884838056251(21)US320F0129 US320F0130 (01)00884838056251(21)US320F0130 US320F0131 (01)00884838056251(21)US320F0131 US320F0132 (01)00884838056251(21)US320F0132 US320F0133 (01)00884838056251(21)US320F0133 US320F0134 (01)00884838056251(21)US320F0134 US320F0135 (01)00884838056251(21)US320F0135 US320F0136 (01)00884838056251(21)US320F0136 US320F0137 (01)00884838056251(21)US320F0137 US320F0138 (01)00884838056251(21)US320F0138 US320F0178 (01)00884838056251(21)US320F0178 US320F0179 (01)00884838056251(21)US320F0179 US320F0180 (01)00884838056251(21)US320F0180 US320F0181 (01)00884838056251(21)US320F0181 US320F0182 (01)00884838056251(21)US320F0182 US320F0183 (01)00884838056251(21)US320F0183 US320F0184 (01)00884838056251(21)US320F0184 US320F0185 (01)00884838056251(21)US320F0185 US320F0186 (01)00884838056251(21)US320F0186 US320F0187 (01)00884838056251(21)US320F0187 US320F0188 (01)00884838056251(21)US320F0188 US320F0189 (01)00884838056251(21)US320F0189 US320F0190 (01)00884838056251(21)US320F0190 US320F0191 (01)00884838056251(21)US320F0191 US320F0239 (01)00884838056251(21)US320F0239 US320F0240 (01)00884838056251(21)US320F0240 US320F0241 (01)00884838056251(21)US320F0241 US320F0265 (01)00884838056251(21)US320F0265 US320F0298 (01)00884838056251(21)US320F0298 US320F0299 (01)00884838056251(21)US320F0299 US320F0300 (01)00884838056251(21)US320F0300 US320F0301 (01)00884838056251(21)US320F0301 US320F0302 (01)00884838056251(21)US320F0302 US320F0303 (01)00884838056251(21)US320F0303 US320F0304 (01)00884838056251(21)US320F0304 US320F0305 (01)00884838056251(21)US320F0305 US320F0306 (01)00884838056251(21)US320F0306 US320F0307 (01)00884838056251(21)US320F0307 US320F0308 (01)00884838056251(21)US320F0308 US320F0309 (01)00884838056251(21)US320F0309 US320F0310 (01)00884838056251(21)US320F0310 US320F0336 (01)00884838056251(21)US320F0336 US320F0337 (01)00884838056251(21)US320F0337 US320F0338 (01)00884838056251(21)US320F0338 US320F0339 (01)00884838056251(21)US320F0339 US320F0340 (01)00884838056251(21)US320F0340 US320F0341 (01)00884838056251(21)US320F0341 US320F0342 (01)00884838056251(21)US320F0342 US320F0368 (01)00884838056251(21)US320F0368 US320F0370 (01)00884838056251(21)US320F0370 US320F0371 (01)00884838056251(21)US320F0371 US320F0402 (01)00884838056251(21)US320F0402 US320F0403 (01)00884838056251(21)US320F0403 US320F0405 (01)00884838056251(21)US320F0405 US320F0407 (01)00884838056251(21)US320F0407 US320F0408 (01)00884838056251(21)US320F0408 US320F0409 (01)00884838056251(21)US320F0409 US320F0410 (01)00884838056251(21)US320F0410 US320F0443 (01)00884838056251(21)US320F0443 US320F0445 (01)00884838056251(21)US320F0445 US320F0468 (01)00884838056251(21)US320F0468 US320F0470 (01)00884838056251(21)US320F0470 US320F0620 (01)00884838056251(21)US320F0620 US320F0621 (01)00884838056251(21)US320F0621 US320F0622 (01)00884838056251(21)US320F0622 US320F0649 (01)00884838056251(21)US320F0649 US320F0650 (01)00884838056251(21)US320F0650 US320F0658 (01)00884838056251(21)US320F0658 US320F0659 (01)00884838056251(21)US320F0659 US320F0660 (01)00884838056251(21)US320F0660 US320F0661 (01)00884838056251(21)US320F0661 US320F0662 (01)00884838056251(21)US320F0662 US320F0725 (01)00884838056251(21)US320F0725 US320F0726 (01)00884838056251(21)US320F0726 US320F0727 (01)00884838056251(21)US320F0727 US320F0728 (01)00884838056251(21)US320F0728 US320F0729 (01)00884838056251(21)US320F0729 US320F0730 (01)00884838056251(21)US320F0730 US320F0731 (01)00884838056251(21)US320F0731 US320F0732 (01)00884838056251(21)US320F0732 US320F0733 (01)00884838056251(21)US320F0733 US320F0775 (01)00884838056251(21)US320F0775 US320F0777 (01)00884838056251(21)US320F0777 US320F0778 (01)00884838056251(21)US320F0778 US320F0779 (01)00884838056251(21)US320F0779 US320F0780 (01)00884838056251(21)US320F0780 US320F0781 (01)00884838056251(21)US320F0781 US320F0782 (01)00884838056251(21)US320F0782 US320F0783 (01)00884838056251(21)US320F0783 US320F0784 (01)00884838056251(21)US320F0784 US320F0787 (01)00884838056251(21)US320F0787 US320F0788 (01)00884838056251(21)US320F0788 US320F0816 (01)00884838056251(21)US320F0816 US320F0817 (01)00884838056251(21)US320F0817 US320F0818 (01)00884838056251(21)US320F0818 US320F0819 (01)00884838056251(21)US320F0819 US320F0829 (01)00884838056251(21)US320F0829 US320F0830 (01)00884838056251(21)US320F0830 US320F0831 (01)00884838056251(21)US320F0831 US320F0832 (01)00884838056251(21)US320F0832 US320F0833 (01)00884838056251(21)US320F0833 US320F0834 (01)00884838056251(21)US320F0834 US320F0835 (01)00884838056251(21)US320F0835 US320F0836 (01)00884838056251(21)US320F0836 US320F0837 (01)00884838056251(21)US320F0837 US320F0874 (01)00884838056251(21)US320F0874 US320F0875 (01)00884838056251(21)US320F0875 US320F0929 (01)00884838056251(21)US320F0929 US320F0930 (01)00884838056251(21)US320F0930 US320F0931 (01)00884838056251(21)US320F0931 US320F0932 (01)00884838056251(21)US320F0932 US320F0949 (01)00884838056251(21)US320F0949 US320F0950 (01)00884838056251(21)US320F0950 US320F0953 (01)00884838056251(21)US320F0953 US320F0954 (01)00884838056251(21)US320F0954 US320F0955 (01)00884838056251(21)US320F0955 US320F0956 (01)00884838056251(21)US320F0956 US320F0958 (01)00884838056251(21)US320F0958 US320F0959 (01)00884838056251(21)US320F0959 US320F0960 (01)00884838056251(21)US320F0960 US320F0962 (01)00884838056251(21)US320F0962 US320F0963 (01)00884838056251(21)US320F0963 US320F0964 (01)00884838056251(21)US320F0964 US320F1010 (01)00884838056251(21)US320F1010 US320F1015 (01)00884838056251(21)US320F1015 US320F1075 (01)00884838056251(21)US320F1075 US320F1076 (01)00884838056251(21)US320F1076 US320F1080 (01)00884838056251(21)US320F1080 US320F1081 (01)00884838056251(21)US320F1081 US320F1082 (01)00884838056251(21)US320F1082 US320F1100 (01)00884838056251(21)US320F1100 US320F1198 (01)00884838056251(21)US320F1198 US320F1446 (01)00884838056251(21)US320F1446 US320F1447 (01)00884838056251(21)US320F1447 US320F1448 (01)00884838056251(21)US320F1448 US320F1449 (01)00884838056251(21)US320F1449 US320F1450 (01)00884838056251(21)US320F1450 US320F1452 (01)00884838056251(21)US320F1452 US320F1453 (01)00884838056251(21)US320F1453 US320F1454 (01)00884838056251(21)US320F1454 US320F1455 (01)00884838056251(21)US320F1455 US320F1456 (01)00884838056251(21)US320F1456 US320F1457 (01)00884838056251(21)US320F1457 US320F1461 (01)00884838056251(21)US320F1461 US320F1485 (01)00884838056251(21)US320F1485 US320F1561 (01)00884838056251(21)US320F1561 US320F1563 (01)00884838056251(21)US320F1563 US320F1564 (01)00884838056251(21)US320F1564 US320F1754 (01)00884838056251(21)US320F1754 US320F1765 (01)00884838056251(21)US320F1765 US320F1876 (01)00884838056251(21)US320F1876 US320F1877 (01)00884838056251(21)US320F1877 US320F1889 (01)00884838056251(21)US320F1889 US320F1890 (01)00884838056251(21)US320F1890 US320F1929 (01)00884838056251(21)US320F1929 US320F1930 (01)00884838056251(21)US320F1930 US320F1931 (01)00884838056251(21)US320F1931 US320F1932 (01)00884838056251(21)US320F1932 US320F1933 (01)00884838056251(21)US320F1933 US320F1934 (01)00884838056251(21)US320F1934 US320F1980 (01)00884838056251(21)US320F1980 US320F1982 (01)00884838056251(21)US320F1982 US320F1983 (01)00884838056251(21)US320F1983 US320F1984 (01)00884838056251(21)US320F1984 US415F0801 N/A US415F0854 N/A US416F1050 N/A US417F0031 (01)00884838056251(21)US417F0031 US417F0638 (01)00884838056251(21)US417F0638 US417F0639 (01)00884838056251(21)US417F0639 US417F0640 (01)00884838056251(21)US417F0640 US417F0641 (01)00884838056251(21)US417F0641 US417F0897 (01)00884838056251(21)US417F0897 US418F0231 (01)00884838056251(21)US418F0231 US418F0326 (01)00884838056251(21)US418F0326 US418F0979 (01)00884838056251(21)US418F0979 US418F1100 (01)00884838056251(21)US418F1100 US418F1384 (01)00884838056251(21)US418F1384 US419F1286 (01)00884838056251(21)US419F1286 US419F1352 (01)00884838056251(21)US419F1352 US419F1938 (01)00884838056251(21)US419F1938 US420F0170 (01)00884838056251(21)US420F0170 US420F0171 (01)00884838056251(21)US420F0171 US420F0172 (01)00884838056251(21)US420F0172 US420F0260 (01)00884838056251(21)US420F0260 US420F0508 (01)00884838056251(21)US420F0508 US420F0509 (01)00884838056251(21)US420F0509 US420F0667 (01)00884838056251(21)US420F0667 US420F0668 (01)00884838056251(21)US420F0668 US420F0669 (01)00884838056251(21)US420F0669 US420F0670 (01)00884838056251(21)US420F0670 US420F2990 (01)00884838056251(21)US420F2990 US515F0161 N/A US516F0264 N/A US516F0331 N/A US516F0404 N/A US516F0637 N/A US516F0941 N/A US516F0942 N/A US516F1215 N/A US517F0334 (01)00884838056251(21)US517F0334 US517F0445 (01)00884838056251(21)US517F0445 US517F0622 (01)00884838056251(21)US517F0622 US517F0721 (01)00884838056251(21)US517F0721 US517F0761 (01)00884838056251(21)US517F0761 US517F0895 (01)00884838056251(21)US517F0895 US517F1383 (01)00884838056251(21)US517F1383 US517F1384 (01)00884838056251(21)US517F1384 US517F1385 (01)00884838056251(21)US517F1385 US518F0397 (01)00884838056251(21)US518F0397 US518F0522 (01)00884838056251(21)US518F0522 US518F0536 (01)00884838056251(21)US518F0536 US518F0537 (01)00884838056251(21)US518F0537 US518F0538 (01)00884838056251(21)US518F0538 US518F0586 (01)00884838056251(21)US518F0586 US518F0909 (01)00884838056251(21)US518F0909 US518F1114 (01)00884838056251(21)US518F1114 US519F0401 (01)00884838056251(21)US519F0401 US519F0603 (01)00884838056251(21)US519F0603 US519F0636 (01)00884838056251(21)US519F0636 US519F0838 (01)00884838056251(21)US519F0838 US519F0846 (01)00884838056251(21)US519F0846 US519F1508 (01)00884838056251(21)US519F1508 US519F1541 (01)00884838056251(21)US519F1541 US519F1543 (01)00884838056251(21)US519F1543 US519F1551 N/A US519F1552 N/A US519F1553 (01)00884838056251(21)US519F1553 US519F1554 N/A US519F1634 N/A US520F0354 (01)00884838056251(21)US520F0354 US520F0355 (01)00884838056251(21)US520F0355 US520F2333 (01)00884838056251(21)US520F2333 US615F0112 N/A US615F0557 N/A US615F0791 N/A US615F1206 N/A US615F1255 N/A US615F1256 N/A US616F0403 N/A US616F0476 N/A US616F0683 N/A US616F0684 N/A US616F0685 N/A US617F0153 (01)00884838056251(21)US617F0153 US617F0309 (01)00884838056251(21)US617F0309 US617F0312 (01)00884838056251(21)US617F0312 US617F0314 (01)00884838056251(21)US617F0314 US617F0384 (01)00884838056251(21)US617F0384 US617F0439 (01)00884838056251(21)US617F0439 US617F0481 (01)00884838056251(21)US617F0481 US617F0605 (01)00884838056251(21)US617F0605 US617F0606 (01)00884838056251(21)US617F0606 US617F0610 (01)00884838056251(21)US617F0610 US617F0934 (01)00884838056251(21)US617F0934 US617F1017 (01)00884838056251(21)US617F1017 US617F1018 (01)00884838056251(21)US617F1018 US618F1246 (01)00884838056251(21)US618F1246 US618F1287 (01)00884838056251(21)US618F1287 US618F1796 (01)00884838056251(21)US618F1796 US618F1797 (01)00884838056251(21)US618F1797 US618F1859 (01)00884838056251(21)US618F1859 US618F1862 (01)00884838056251(21)US618F1862 US618F1867 (01)00884838056251(21)US618F1867 US618F1998 (01)00884838056251(21)US618F1998 US619F0672 (01)00884838056251(21)US619F0672 US619F0673 (01)00884838056251(21)US619F0673 US619F0674 (01)00884838056251(21)US619F0674 US619F0675 (01)00884838056251(21)US619F0675 US619F0676 (01)00884838056251(21)US619F0676 US619F0677 (01)00884838056251(21)US619F0677 US619F0678 (01)00884838056251(21)US619F0678 US619F0679 (01)00884838056251(21)US619F0679 US619F0680 (01)00884838056251(21)US619F0680 US619F0681 (01)00884838056251(21)US619F0681 US619F0682 (01)00884838056251(21)US619F0682 US619F0683 (01)00884838056251(21)US619F0683 US619F0871 (01)00884838056251(21)US619F0871 US619F0993 (01)00884838056251(21)US619F0993 US619F1112 (01)00884838056251(21)US619F1112 US619F1319 (01)00884838056251(21)US619F1319 US619F1344 (01)00884838056251(21)US619F1344 US619F1399 (01)00884838056251(21)US619F1399 US619F1794 (01)00884838056251(21)US619F1794 US619F1857 (01)00884838056251(21)US619F1857 US620F0304 (01)00884838056251(21)US620F0304 US620F0463 (01)00884838056251(21)US620F0463 US620F0464 (01)00884838056251(21)US620F0464 US620F1672 (01)00884838056251(21)US620F1672 US620F1682 (01)00884838056251(21)US620F1682 US715F0457 N/A US715F0510 N/A US715F0971 N/A US716F0127 N/A US716F0220 N/A US716F0348 N/A US717F0454 (01)00884838056251(21)US717F0454 US717F0532 (01)00884838056251(21)US717F0532 US717F0652 (01)00884838056251(21)US717F0652 US717F0665 (01)00884838056251(21)US717F0665 US717F0884 (01)00884838056251(21)US717F0884 US717F0923 (01)00884838056251(21)US717F0923 US717F1046 (01)00884838056251(21)US717F1046 US717F1178 (01)00884838056251(21)US717F1178 US717F1179 (01)00884838056251(21)US717F1179 US717F1180 (01)00884838056251(21)US717F1180 US717F1181 (01)00884838056251(21)US717F1181 US717F1182 (01)00884838056251(21)US717F1182 US717F1183 (01)00884838056251(21)US717F1183 US718F0225 (01)00884838056251(21)US718F0225 US718F0226 (01)00884838056251(21)US718F0226 US718F0436 (01)00884838056251(21)US718F0436 US718F0651 (01)00884838056251(21)US718F0651 US718F1236 (01)00884838056251(21)US718F1236 US719F0202 (01)00884838056251(21)US719F0202 US719F0310 (01)00884838056251(21)US719F0310 US719F0554 (01)00884838056251(21)US719F0554 US719F0650 (01)00884838056251(21)US719F0650 US719F0736 (01)00884838056251(21)US719F0736 US719F0891 (01)00884838056251(21)US719F0891 US719F1185 (01)00884838056251(21)US719F1185 US719F1188 (01)00884838056251(21)US719F1188 US719F1189 (01)00884838056251(21)US719F1189 US719F1250 (01)00884838056251(21)US719F1250 US719F1251 (01)00884838056251(21)US719F1251 US719F1308 (01)00884838056251(21)US719F1308 US719F1391 (01)00884838056251(21)US719F1391 US719F1393 (01)00884838056251(21)US719F1393 US719F1394 (01)00884838056251(21)US719F1394 US719F1396 (01)00884838056251(21)US719F1396 US719F1397 (01)00884838056251(21)US719F1397 US719F1398 (01)00884838056251(21)US719F1398 US719F1399 (01)00884838056251(21)US719F1399 US719F1400 (01)00884838056251(21)US719F1400 US719F1436 (01)00884838056251(21)US719F1436 US719F1437 (01)00884838056251(21)US719F1437 US719F1474 (01)00884838056251(21)US719F1474 US719F1475 (01)00884838056251(21)US719F1475 US719F1717 (01)00884838056251(21)US719F1717 US719F1719 (01)00884838056251(21)US719F1719 US719F1720 (01)00884838056251(21)US719F1720 US719F1721 (01)00884838056251(21)US719F1721 US719F1722 (01)00884838056251(21)US719F1722 US719F1723 (01)00884838056251(21)US719F1723 US719F1724 (01)00884838056251(21)US719F1724 US719F1725 (01)00884838056251(21)US719F1725 US719F1727 (01)00884838056251(21)US719F1727 US719F1728 (01)00884838056251(21)US719F1728 US719F1729 (01)00884838056251(21)US719F1729 US719F1730 (01)00884838056251(21)US719F1730 US719F1731 (01)00884838056251(21)US719F1731 US719F1732 (01)00884838056251(21)US719F1732 US719F1733 (01)00884838056251(21)US719F1733 US719F1734 (01)00884838056251(21)US719F1734 US719F1762 (01)00884838056251(21)US719F1762 US719F1820 (01)00884838056251(21)US719F1820 US719F1821 (01)00884838056251(21)US719F1821 US719F1822 (01)00884838056251(21)US719F1822 US719F1823 (01)00884838056251(21)US719F1823 US719F1824 (01)00884838056251(21)US719F1824 US719F1825 (01)00884838056251(21)US719F1825 US719F1826 (01)00884838056251(21)US719F1826 US719F1827 (01)00884838056251(21)US719F1827 US719F1828 (01)00884838056251(21)US719F1828 US719F1829 (01)00884838056251(21)US719F1829 US719F1830 (01)00884838056251(21)US719F1830 US719F1831 (01)00884838056251(21)US719F1831 US719F1832 (01)00884838056251(21)US719F1832 US719F1833 (01)00884838056251(21)US719F1833 US719F1861 (01)00884838056251(21)US719F1861 US719F1862 (01)00884838056251(21)US719F1862 US719F1863 (01)00884838056251(21)US719F1863 US719F1911 (01)00884838056251(21)US719F1911 US719F1940 (01)00884838056251(21)US719F1940 US719F1941 (01)00884838056251(21)US719F1941 US719F2036 (01)00884838056251(21)US719F2036 US719F2038 (01)00884838056251(21)US719F2038 US719F2039 (01)00884838056251(21)US719F2039 US719F2040 (01)00884838056251(21)US719F2040 US719F2041 (01)00884838056251(21)US719F2041 US719F2118 (01)00884838056251(21)US719F2118 US719F2119 (01)00884838056251(21)US719F2119 US719F2150 (01)00884838056251(21)US719F2150 US719F2151 (01)00884838056251(21)US719F2151 US719F2152 (01)00884838056251(21)US719F2152 US719F2153 (01)00884838056251(21)US719F2153 US719F2154 (01)00884838056251(21)US719F2154 US720F0815 (01)00884838056251(21)US720F0815 US720F1017 (01)00884838056251(21)US720F1017 US720F1018 (01)00884838056251(21)US720F1018 US720F1194 N/A US720F1480 (01)00884838056251(21)US720F1480 US720F1666 (01)00884838056251(21)US720F1666 US720F1835 (01)00884838056251(21)US720F1835 US720F1836 (01)00884838056251(21)US720F1836 US814F0006 N/A US814F0345 N/A US814F0612 N/A US815F0175 N/A US815F0857 N/A US815F0858 N/A US816F0064 (01)00884838056251(21)US816F0064 US816F0161 N/A US816F0926 (01)00884838056251(21)US816F0926 US816F0927 (01)00884838056251(21)US816F0927 US816F0997 (01)00884838056251(21)US816F0997 US816F1058 (01)00884838056251(21)US816F1058 US817F0048 (01)00884838056251(21)US817F0048 US817F0049 (01)00884838056251(21)US817F0049 US817F0187 (01)00884838056251(21)US817F0187 US817F0553 (01)00884838056251(21)US817F0553 US817F0746 (01)00884838056251(21)US817F0746 US817F0847 (01)00884838056251(21)US817F0847 US817F1010 (01)00884838056251(21)US817F1010 US817F1197 (01)00884838056251(21)US817F1197 US818F0015 (01)00884838056251(21)US818F0015 US818F0097 (01)00884838056251(21)US818F0097 US818F0137 (01)00884838056251(21)US818F0137 US818F0942 (01)00884838056251(21)US818F0942 US818F1327 (01)00884838056251(21)US818F1327 US818F1502 (01)00884838056251(21)US818F1502 US818F1504 (01)00884838056251(21)US818F1504 US818F1505 (01)00884838056251(21)US818F1505 US818F1506 (01)00884838056251(21)US818F1506 US818F1961 (01)00884838056251(21)US818F1961 US818F2186 (01)00884838056251(21)US818F2186 US819F0067 (01)00884838056251(21)US819F0067 US819F0087 (01)00884838056251(21)US819F0087 US819F0088 (01)00884838056251(21)US819F0088 US819F0089 (01)00884838056251(21)US819F0089 US819F0090 (01)00884838056251(21)US819F0090 US819F0091 (01)00884838056251(21)US819F0091 US819F0092 (01)00884838056251(21)US819F0092 US819F0093 (01)00884838056251(21)US819F0093 US819F0156 (01)00884838056251(21)US819F0156 US819F0157 (01)00884838056251(21)US819F0157 US819F0158 (01)00884838056251(21)US819F0158 US819F0168 (01)00884838056251(21)US819F0168 US819F0169 (01)00884838056251(21)US819F0169 US819F0170 (01)00884838056251(21)US819F0170 US819F0171 (01)00884838056251(21)US819F0171 US819F0172 (01)00884838056251(21)US819F0172 US819F0173 (01)00884838056251(21)US819F0173 US819F0174 (01)00884838056251(21)US819F0174 US819F0175 (01)00884838056251(21)US819F0175 US819F0290 (01)00884838056251(21)US819F0290 US819F0291 (01)00884838056251(21)US819F0291 US819F0292 (01)00884838056251(21)US819F0292 US819F0366 (01)00884838056251(21)US819F0366 US819F0367 (01)00884838056251(21)US819F0367 US819F0373 (01)00884838056251(21)US819F0373 US819F0374 (01)00884838056251(21)US819F0374 US819F0375 (01)00884838056251(21)US819F0375 US819F0377 (01)00884838056251(21)US819F0377 US819F0378 (01)00884838056251(21)US819F0378 US819F0379 (01)00884838056251(21)US819F0379 US819F0380 (01)00884838056251(21)US819F0380 US819F0381 (01)00884838056251(21)US819F0381 US819F0382 (01)00884838056251(21)US819F0382 US819F0383 (01)00884838056251(21)US819F0383 US819F0386 (01)00884838056251(21)US819F0386 US819F0387 (01)00884838056251(21)US819F0387 US819F0430 (01)00884838056251(21)US819F0430 US819F0431 (01)00884838056251(21)US819F0431 US819F0432 (01)00884838056251(21)US819F0432 US819F0475 (01)00884838056251(21)US819F0475 US819F0476 (01)00884838056251(21)US819F0476 US819F0477 (01)00884838056251(21)US819F0477 US819F0478 (01)00884838056251(21)US819F0478 US819F0479 (01)00884838056251(21)US819F0479 US819F0480 (01)00884838056251(21)US819F0480 US819F0481 (01)00884838056251(21)US819F0481 US819F0482 (01)00884838056251(21)US819F0482 US819F0483 (01)00884838056251(21)US819F0483 US819F0484 (01)00884838056251(21)US819F0484 US819F0485 (01)00884838056251(21)US819F0485 US819F0486 (01)00884838056251(21)US819F0486 US819F0487 (01)00884838056251(21)US819F0487 US819F0488 (01)00884838056251(21)US819F0488 US819F0496 (01)00884838056251(21)US819F0496 US819F0497 (01)00884838056251(21)US819F0497 US819F0498 (01)00884838056251(21)US819F0498 US819F0499 (01)00884838056251(21)US819F0499 US819F0500 (01)00884838056251(21)US819F0500 US819F0621 (01)00884838056251(21)US819F0621 US819F0622 (01)00884838056251(21)US819F0622 US819F0623 (01)00884838056251(21)US819F0623 US819F0624 (01)00884838056251(21)US819F0624 US819F0625 (01)00884838056251(21)US819F0625 US819F0626 (01)00884838056251(21)US819F0626 US819F0627 (01)00884838056251(21)US819F0627 US819F0628 (01)00884838056251(21)US819F0628 US819F0632 (01)00884838056251(21)US819F0632 US819F0633 (01)00884838056251(21)US819F0633 US819F0634 (01)00884838056251(21)US819F0634 US819F0643 (01)00884838056251(21)US819F0643 US819F0644 (01)00884838056251(21)US819F0644 US819F0645 (01)00884838056251(21)US819F0645 US819F0646 (01)00884838056251(21)US819F0646 US819F0702 (01)00884838056251(21)US819F0702 US819F0703 (01)00884838056251(21)US819F0703 US819F0704 (01)00884838056251(21)US819F0704 US819F0705 (01)00884838056251(21)US819F0705 US819F0706 (01)00884838056251(21)US819F0706 US819F0707 (01)00884838056251(21)US819F0707 US819F0717 (01)00884838056251(21)US819F0717 US819F0718 (01)00884838056251(21)US819F0718 US819F0787 (01)00884838056251(21)US819F0787 US819F0788 (01)00884838056251(21)US819F0788 US819F0789 (01)00884838056251(21)US819F0789 US819F0790 (01)00884838056251(21)US819F0790 US819F0791 (01)00884838056251(21)US819F0791 US819F0792 (01)00884838056251(21)US819F0792 US819F0810 (01)00884838056251(21)US819F0810 US819F0812 (01)00884838056251(21)US819F0812 US819F0813 (01)00884838056251(21)US819F0813 US819F0814 (01)00884838056251(21)US819F0814 US819F0822 (01)00884838056251(21)US819F0822 US819F0823 (01)00884838056251(21)US819F0823 US819F0824 (01)00884838056251(21)US819F0824 US819F0933 (01)00884838056251(21)US819F0933 US819F0934 (01)00884838056251(21)US819F0934 US819F0935 (01)00884838056251(21)US819F0935 US819F0936 (01)00884838056251(21)US819F0936 US819F0937 (01)00884838056251(21)US819F0937 US819F0938 (01)00884838056251(21)US819F0938 US819F0939 (01)00884838056251(21)US819F0939 US819F1002 (01)00884838056251(21)US819F1002 US819F1042 (01)00884838056251(21)US819F1042 US819F1043 (01)00884838056251(21)US819F1043 US819F1044 (01)00884838056251(21)US819F1044 US819F1045 (01)00884838056251(21)US819F1045 US819F1046 (01)00884838056251(21)US819F1046 US819F1047 (01)00884838056251(21)US819F1047 US819F1048 (01)00884838056251(21)US819F1048 US819F1049 (01)00884838056251(21)US819F1049 US819F1050 (01)00884838056251(21)US819F1050 US819F1054 (01)00884838056251(21)US819F1054 US819F1132 (01)00884838056251(21)US819F1132 US819F1133 (01)00884838056251(21)US819F1133 US819F1134 (01)00884838056251(21)US819F1134 US819F1139 (01)00884838056251(21)US819F1139 US819F1140 (01)00884838056251(21)US819F1140 US819F1141 (01)00884838056251(21)US819F1141 US819F1164 (01)00884838056251(21)US819F1164 US819F1165 (01)00884838056251(21)US819F1165 US819F1166 (01)00884838056251(21)US819F1166 US819F1168 (01)00884838056251(21)US819F1168 US819F1169 (01)00884838056251(21)US819F1169 US819F1173 (01)00884838056251(21)US819F1173 US819F1199 (01)00884838056251(21)US819F1199 US819F1200 (01)00884838056251(21)US819F1200 US819F1201 (01)00884838056251(21)US819F1201 US819F1202 (01)00884838056251(21)US819F1202 US819F1338 (01)00884838056251(21)US819F1338 US819F1364 (01)00884838056251(21)US819F1364 US819F1365 (01)00884838056251(21)US819F1365 US819F1366 (01)00884838056251(21)US819F1366 US819F1367 (01)00884838056251(21)US819F1367 US819F1368 (01)00884838056251(21)US819F1368 US819F1369 (01)00884838056251(21)US819F1369 US819F1455 (01)00884838056251(21)US819F1455 US819F1456 (01)00884838056251(21)US819F1456 US819F1457 (01)00884838056251(21)US819F1457 US819F1458 (01)00884838056251(21)US819F1458 US819F1459 (01)00884838056251(21)US819F1459 US819F1460 (01)00884838056251(21)US819F1460 US819F1461 (01)00884838056251(21)US819F1461 US819F1462 (01)00884838056251(21)US819F1462 US819F1463 (01)00884838056251(21)US819F1463 US819F1464 (01)00884838056251(21)US819F1464 US819F1465 (01)00884838056251(21)US819F1465 US819F1466 (01)00884838056251(21)US819F1466 US819F1514 (01)00884838056251(21)US819F1514 US819F1518 (01)00884838056251(21)US819F1518 US819F1519 (01)00884838056251(21)US819F1519 US819F1520 (01)00884838056251(21)US819F1520 US819F1571 (01)00884838056251(21)US819F1571 US819F1572 (01)00884838056251(21)US819F1572 US819F1573 (01)00884838056251(21)US819F1573 US819F1683 (01)00884838056251(21)US819F1683 US819F1684 (01)00884838056251(21)US819F1684 US819F1685 (01)00884838056251(21)US819F1685 US819F1686 (01)00884838056251(21)US819F1686 US819F1687 (01)00884838056251(21)US819F1687 US819F1688 (01)00884838056251(21)US819F1688 US819F1689 (01)00884838056251(21)US819F1689 US819F1690 (01)00884838056251(21)US819F1690 US819F1746 (01)00884838056251(21)US819F1746 US819F1747 (01)00884838056251(21)US819F1747 US819F1748 (01)00884838056251(21)US819F1748 US819F1749 (01)00884838056251(21)US819F1749 US819F1750 (01)00884838056251(21)US819F1750 US819F1751 (01)00884838056251(21)US819F1751 US819F1761 (01)00884838056251(21)US819F1761 US819F1762 (01)00884838056251(21)US819F1762 US819F1763 (01)00884838056251(21)US819F1763 US819F1764 (01)00884838056251(21)US819F1764 US819F1765 (01)00884838056251(21)US819F1765 US819F1766 (01)00884838056251(21)US819F1766 US819F1767 (01)00884838056251(21)US819F1767 US819F1768 (01)00884838056251(21)US819F1768 US819F1769 (01)00884838056251(21)US819F1769 US819F1770 (01)00884838056251(21)US819F1770 US819F1771 (01)00884838056251(21)US819F1771 US819F1772 (01)00884838056251(21)US819F1772 US819F1773 (01)00884838056251(21)US819F1773 US819F1824 (01)00884838056251(21)US819F1824 US819F1825 (01)00884838056251(21)US819F1825 US819F1826 (01)00884838056251(21)US819F1826 US819F1827 (01)00884838056251(21)US819F1827 US819F1828 (01)00884838056251(21)US819F1828 US819F1829 (01)00884838056251(21)US819F1829 US819F1830 (01)00884838056251(21)US819F1830 US819F1831 (01)00884838056251(21)US819F1831 US819F1832 (01)00884838056251(21)US819F1832 US819F1863 (01)00884838056251(21)US819F1863 US819F1864 (01)00884838056251(21)US819F1864 US819F1865 (01)00884838056251(21)US819F1865 US819F1866 (01)00884838056251(21)US819F1866 US819F1867 (01)00884838056251(21)US819F1867 US819F1868 (01)00884838056251(21)US819F1868 US819F1923 (01)00884838056251(21)US819F1923 US819F1924 (01)00884838056251(21)US819F1924 US819F1925 (01)00884838056251(21)US819F1925 US819F1926 (01)00884838056251(21)US819F1926 US819F1927 (01)00884838056251(21)US819F1927 US819F1928 (01)00884838056251(21)US819F1928 US819F1929 (01)00884838056251(21)US819F1929 US819F1988 (01)00884838056251(21)US819F1988 US819F1989 (01)00884838056251(21)US819F1989 US819F1992 (01)00884838056251(21)US819F1992 US819F1993 (01)00884838056251(21)US819F1993 US819F1994 (01)00884838056251(21)US819F1994 US819F2061 (01)00884838056251(21)US819F2061 US819F2062 (01)00884838056251(21)US819F2062 US819F2063 (01)00884838056251(21)US819F2063 US819F2064 (01)00884838056251(21)US819F2064 US819F2065 (01)00884838056251(21)US819F2065 US819F2066 (01)00884838056251(21)US819F2066 US819F2067 (01)00884838056251(21)US819F2067 US819F2068 (01)00884838056251(21)US819F2068 US819F2069 (01)00884838056251(21)US819F2069 US820F1640 (01)00884838056251(21)US820F1640 US914F0405 N/A US914F0452 N/A US914F0510 N/A US914F0582 N/A US914F0760 N/A US915F0196 N/A US915F0568 N/A US915F0684 N/A US915F0686 N/A US915F0807 N/A US915F0894 N/A US915F1003 N/A US916F0893 (01)00884838056251(21)US916F0893 US916F0981 (01)00884838056251(21)US916F0981 US916F1064 (01)00884838056251(21)US916F1064 US916F1067 (01)00884838056251(21)US916F1067 US917F0246 (01)00884838056251(21)US917F0246 US917F0871 (01)00884838056251(21)US917F0871 US917F1162 (01)00884838056251(21)US917F1162 US917F1248 (01)00884838056251(21)US917F1248 US917F2024 (01)00884838056251(21)US917F2024 US917F2025 (01)00884838056251(21)US917F2025 US917F2026 (01)00884838056251(21)US917F2026 US918F0163 (01)00884838056251(21)US918F0163 US918F0664 (01)00884838056251(21)US918F0664 US918F1090 (01)00884838056251(21)US918F1090 US918F1142 (01)00884838056251(21)US918F1142 US918F1165 (01)00884838056251(21)US918F1165 US918F1512 (01)00884838056251(21)US918F1512 US918F1707 (01)00884838056251(21)US918F1707 US918F1761 (01)00884838056251(21)US918F1761 US919F0024 (01)00884838056251(21)US919F0024 US919F0025 (01)00884838056251(21)US919F0025 US919F0026 (01)00884838056251(21)US919F0026 US919F0027 (01)00884838056251(21)US919F0027 US919F0028 (01)00884838056251(21)US919F0028 US919F0029 (01)00884838056251(21)US919F0029 US919F0030 (01)00884838056251(21)US919F0030 US919F0149 (01)00884838056251(21)US919F0149 US919F0150 (01)00884838056251(21)US919F0150 US919F0154 (01)00884838056251(21)US919F0154 US919F0195 (01)00884838056251(21)US919F0195 US919F0196 (01)00884838056251(21)US919F0196 US919F0197 (01)00884838056251(21)US919F0197 US919F0198 (01)00884838056251(21)US919F0198 US919F0199 (01)00884838056251(21)US919F0199 US919F0200 (01)00884838056251(21)US919F0200 US919F0224 (01)00884838056251(21)US919F0224 US919F0225 (01)00884838056251(21)US919F0225 US919F0229 (01)00884838056251(21)US919F0229 US919F0230 (01)00884838056251(21)US919F0230 US919F0231 (01)00884838056251(21)US919F0231 US919F0232 (01)00884838056251(21)US919F0232 US919F0233 (01)00884838056251(21)US919F0233 US919F0277 (01)00884838056251(21)US919F0277 US919F0282 (01)00884838056251(21)US919F0282 US919F0283 (01)00884838056251(21)US919F0283 US919F0284 (01)00884838056251(21)US919F0284 US919F0285 (01)00884838056251(21)US919F0285 US919F0286 (01)00884838056251(21)US919F0286 US919F0390 (01)00884838056251(21)US919F0390 US919F0391 (01)00884838056251(21)US919F0391 US919F0400 (01)00884838056251(21)US919F0400 US919F0401 (01)00884838056251(21)US919F0401 US919F0423 (01)00884838056251(21)US919F0423 US919F0424 (01)00884838056251(21)US919F0424 US919F0425 (01)00884838056251(21)US919F0425 US919F0426 (01)00884838056251(21)US919F0426 US919F0427 (01)00884838056251(21)US919F0427 US919F0428 (01)00884838056251(21)US919F0428 US919F0429 (01)00884838056251(21)US919F0429 US919F0502 (01)00884838056251(21)US919F0502 US919F0537 (01)00884838056251(21)US919F0537 US919F0539 (01)00884838056251(21)US919F0539 US919F0545 (01)00884838056251(21)US919F0545 US919F0546 (01)00884838056251(21)US919F0546 US919F0547 (01)00884838056251(21)US919F0547 US919F0548 (01)00884838056251(21)US919F0548 US919F0549 (01)00884838056251(21)US919F0549 US919F0550 (01)00884838056251(21)US919F0550 US919F0551 (01)00884838056251(21)US919F0551 US919F0552 (01)00884838056251(21)US919F0552 US919F0609 (01)00884838056251(21)US919F0609 US919F0612 (01)00884838056251(21)US919F0612 US919F0613 (01)00884838056251(21)US919F0613 US919F0614 (01)00884838056251(21)US919F0614 US919F0615 (01)00884838056251(21)US919F0615 US919F0616 (01)00884838056251(21)US919F0616 US919F0617 (01)00884838056251(21)US919F0617 US919F0618 (01)00884838056251(21)US919F0618 US919F0619 (01)00884838056251(21)US919F0619 US919F0620 (01)00884838056251(21)US919F0620 US919F0621 (01)00884838056251(21)US919F0621 US919F0622 (01)00884838056251(21)US919F0622 US919F0623 (01)00884838056251(21)US919F0623 US919F0624 (01)00884838056251(21)US919F0624 US919F0625 (01)00884838056251(21)US919F0625 US919F0627 (01)00884838056251(21)US919F0627 US919F0653 (01)00884838056251(21)US919F0653 US919F0654 (01)00884838056251(21)US919F0654 US919F0655 (01)00884838056251(21)US919F0655 US919F0656 (01)00884838056251(21)US919F0656 US919F0657 (01)00884838056251(21)US919F0657 US919F0658 (01)00884838056251(21)US919F0658 US919F0790 (01)00884838056251(21)US919F0790 US919F0791 (01)00884838056251(21)US919F0791 US919F0792 (01)00884838056251(21)US919F0792 US919F0793 (01)00884838056251(21)US919F0793 US919F0794 (01)00884838056251(21)US919F0794 US919F0795 (01)00884838056251(21)US919F0795 US919F0796 (01)00884838056251(21)US919F0796 US919F0797 (01)00884838056251(21)US919F0797 US919F0798 (01)00884838056251(21)US919F0798 US919F0802 (01)00884838056251(21)US919F0802 US919F0803 (01)00884838056251(21)US919F0803 US919F0820 (01)00884838056251(21)US919F0820 US919F0821 (01)00884838056251(21)US919F0821 US919F0823 (01)00884838056251(21)US919F0823 US919F0864 (01)00884838056251(21)US919F0864 US919F0865 (01)00884838056251(21)US919F0865 US919F0872 (01)00884838056251(21)US919F0872 US919F0892 (01)00884838056251(21)US919F0892 US919F0893 (01)00884838056251(21)US919F0893 US919F0894 (01)00884838056251(21)US919F0894 US919F0895 (01)00884838056251(21)US919F0895 US919F0896 (01)00884838056251(21)US919F0896 US919F0897 (01)00884838056251(21)US919F0897 US919F0898 (01)00884838056251(21)US919F0898 US919F0976 (01)00884838056251(21)US919F0976 US919F0983 (01)00884838056251(21)US919F0983 US919F0984 (01)00884838056251(21)US919F0984 US919F0985 (01)00884838056251(21)US919F0985 US919F0986 (01)00884838056251(21)US919F0986 US919F0987 (01)00884838056251(21)US919F0987 US919F0988 (01)00884838056251(21)US919F0988 US919F0989 (01)00884838056251(21)US919F0989 US919F0990 (01)00884838056251(21)US919F0990 US919F0991 (01)00884838056251(21)US919F0991 US919F0992 (01)00884838056251(21)US919F0992 US919F0993 (01)00884838056251(21)US919F0993 US919F0994 (01)00884838056251(21)US919F0994 US919F0995 (01)00884838056251(21)US919F0995 US919F0996 (01)00884838056251(21)US919F0996 US919F0997 (01)00884838056251(21)US919F0997 US919F0998 (01)00884838056251(21)US919F0998 US919F0999 (01)00884838056251(21)US919F0999 US919F1059 (01)00884838056251(21)US919F1059 US919F1062 (01)00884838056251(21)US919F1062 US919F1070 (01)00884838056251(21)US919F1070 US919F1096 (01)00884838056251(21)US919F1096 US919F1097 (01)00884838056251(21)US919F1097 US919F1098 (01)00884838056251(21)US919F1098 US919F1099 (01)00884838056251(21)US919F1099 US919F1100 (01)00884838056251(21)US919F1100 US919F1101 (01)00884838056251(21)US919F1101 US919F1102 (01)00884838056251(21)US919F1102 US919F1103 (01)00884838056251(21)US919F1103 US919F1104 (01)00884838056251(21)US919F1104 US919F1105 (01)00884838056251(21)US919F1105 US919F1126 (01)00884838056251(21)US919F1126 US919F1127 (01)00884838056251(21)US919F1127 US919F1128 (01)00884838056251(21)US919F1128 US919F1129 (01)00884838056251(21)US919F1129 US919F1130 (01)00884838056251(21)US919F1130 US919F1131 (01)00884838056251(21)US919F1131 US919F1135 (01)00884838056251(21)US919F1135 US919F1136 (01)00884838056251(21)US919F1136 US919F1137 (01)00884838056251(21)US919F1137 US919F1138 (01)00884838056251(21)US919F1138 US919F1187 (01)00884838056251(21)US919F1187 US919F1189 (01)00884838056251(21)US919F1189 US919F1190 (01)00884838056251(21)US919F1190 US919F1191 (01)00884838056251(21)US919F1191 US919F1193 (01)00884838056251(21)US919F1193 US919F1195 (01)00884838056251(21)US919F1195 US919F1196 (01)00884838056251(21)US919F1196 US919F1250 (01)00884838056251(21)US919F1250 US919F1252 (01)00884838056251(21)US919F1252 US919F1253 (01)00884838056251(21)US919F1253 US919F1266 (01)00884838056251(21)US919F1266 US919F1267 (01)00884838056251(21)US919F1267 US919F1268 (01)00884838056251(21)US919F1268 US919F1279 (01)00884838056251(21)US919F1279 US919F1280 (01)00884838056251(21)US919F1280 US919F1281 (01)00884838056251(21)US919F1281 US919F1282 (01)00884838056251(21)US919F1282 US919F1303 (01)00884838056251(21)US919F1303 US919F1304 (01)00884838056251(21)US919F1304 US919F1305 (01)00884838056251(21)US919F1305 US919F1306 (01)00884838056251(21)US919F1306 US919F1307 (01)00884838056251(21)US919F1307 US919F1309 (01)00884838056251(21)US919F1309 US919F1320 (01)00884838056251(21)US919F1320 US919F1387 (01)00884838056251(21)US919F1387 US919F1394 (01)00884838056251(21)US919F1394 US919F1395 (01)00884838056251(21)US919F1395 US919F1470 (01)00884838056251(21)US919F1470 US919F1471 (01)00884838056251(21)US919F1471 US919F1472 (01)00884838056251(21)US919F1472 US919F1473 (01)00884838056251(21)US919F1473 US919F1474 (01)00884838056251(21)US919F1474 US919F1565 (01)00884838056251(21)US919F1565 US919F1577 (01)00884838056251(21)US919F1577 US919F1622 (01)00884838056251(21)US919F1622 US919F1625 (01)00884838056251(21)US919F1625 US919F1626 (01)00884838056251(21)US919F1626 US919F1627 (01)00884838056251(21)US919F1627 US919F1628 (01)00884838056251(21)US919F1628 US919F1629 (01)00884838056251(21)US919F1629 US919F1630 (01)00884838056251(21)US919F1630 US919F1631 (01)00884838056251(21)US919F1631 US919F1647 (01)00884838056251(21)US919F1647 US919F1648 (01)00884838056251(21)US919F1648 US919F1649 (01)00884838056251(21)US919F1649 US919F1703 (01)00884838056251(21)US919F1703 US919F1704 (01)00884838056251(21)US919F1704 US919F1705 (01)00884838056251(21)US919F1705 US919F1717 (01)00884838056251(21)US919F1717 US919F1718 (01)00884838056251(21)US919F1718 US919F1719 (01)00884838056251(21)US919F1719 US919F1819 (01)00884838056251(21)US919F1819 US919F1869 (01)00884838056251(21)US919F1869 US919F1870 (01)00884838056251(21)US919F1870 US919F1910 (01)00884838056251(21)US919F1910 US919F1911 (01)00884838056251(21)US919F1911 US919F1912 (01)00884838056251(21)US919F1912 US919F1917 (01)00884838056251(21)US919F1917 US919F1946 (01)00884838056251(21)US919F1946 US919F1947 (01)00884838056251(21)US919F1947 US919F1999 (01)00884838056251(21)US919F1999 US919F2000 (01)00884838056251(21)US919F2000 US919F2001 (01)00884838056251(21)US919F2001 US919F2002 (01)00884838056251(21)US919F2002 US919F2003 (01)00884838056251(21)US919F2003 US919F2004 (01)00884838056251(21)US919F2004 US919F2006 (01)00884838056251(21)US919F2006 US919F2007 (01)00884838056251(21)US919F2007 US919F2008 (01)00884838056251(21)US919F2008 US919F2083 (01)00884838056251(21)US919F2083 US919F2086 (01)00884838056251(21)US919F2086 US919F2087 (01)00884838056251(21)US919F2087 US919F2088 (01)00884838056251(21)US919F2088 US919F2090 (01)00884838056251(21)US919F2090 US920F0398 (01)00884838056251(21)US920F0398 US920F0940 (01)00884838056251(21)US920F0940 US920F1494 (01)00884838056251(21)US920F1494 US920F1495 (01)00884838056251(21)US920F1495 USD14F0630 N/A USD14F0631 N/A USD14F0632 N/A USD15F0363 N/A USD15F0554 N/A USD15F0931 N/A USD16F0183 (01)00884838056251(21)USD16F0183 USD16F0184 (01)00884838056251(21)USD16F0184 USD16F0692 (01)00884838056251(21)USD16F0692 USD16F0880 (01)00884838056251(21)USD16F0880 USD16F0881 (01)00884838056251(21)USD16F0881 USD16F0882 (01)00884838056251(21)USD16F0882 USD16F0883 (01)00884838056251(21)USD16F0883 USD16F0912 (01)00884838056251(21)USD16F0912 USD16F0914 (01)00884838056251(21)USD16F0914 USD16F0915 (01)00884838056251(21)USD16F0915 USD16F0920 (01)00884838056251(21)USD16F0920 USD16F0964 (01)00884838056251(21)USD16F0964 USD16F0971 (01)00884838056251(21)USD16F0971 USD17F0886 (01)00884838056251(21)USD17F0886 USD17F1037 (01)00884838056251(21)USD17F1037 USD17F1221 (01)00884838056251(21)USD17F1221 USD17F1547 (01)00884838056251(21)USD17F1547 USD17F1751 (01)00884838056251(21)USD17F1751 USD17F1752 (01)00884838056251(21)USD17F1752 USD17F1753 (01)00884838056251(21)USD17F1753 USD18F0699 (01)00884838056251(21)USD18F0699 USD18F0941 (01)00884838056251(21)USD18F0941 USD18F1099 (01)00884838056251(21)USD18F1099 USD18F1212 (01)00884838056251(21)USD18F1212 USD19F0077 (01)00884838056251(21)USD19F0077 USD19F0092 (01)00884838056251(21)USD19F0092 USD19F0093 (01)00884838056251(21)USD19F0093 USD19F0094 (01)00884838056251(21)USD19F0094 USD19F0095 (01)00884838056251(21)USD19F0095 USD19F0096 (01)00884838056251(21)USD19F0096 USD19F0103 (01)00884838056251(21)USD19F0103 USD19F0125 (01)00884838056251(21)USD19F0125 USD19F0126 (01)00884838056251(21)USD19F0126 USD19F0127 (01)00884838056251(21)USD19F0127 USD19F0171 (01)00884838056251(21)USD19F0171 USD19F0172 (01)00884838056251(21)USD19F0172 USD19F0173 (01)00884838056251(21)USD19F0173 USD19F0174 (01)00884838056251(21)USD19F0174 USD19F0175 (01)00884838056251(21)USD19F0175 USD19F0176 (01)00884838056251(21)USD19F0176 USD19F0177 (01)00884838056251(21)USD19F0177 USD19F0178 (01)00884838056251(21)USD19F0178 USD19F0182 (01)00884838056251(21)USD19F0182 USD19F0183 (01)00884838056251(21)USD19F0183 USD19F0184 (01)00884838056251(21)USD19F0184 USD19F0186 (01)00884838056251(21)USD19F0186 USD19F0244 (01)00884838056251(21)USD19F0244 USD19F0258 (01)00884838056251(21)USD19F0258 USD19F0259 (01)00884838056251(21)USD19F0259 USD19F0260 (01)00884838056251(21)USD19F0260 USD19F0261 (01)00884838056251(21)USD19F0261 USD19F0262 (01)00884838056251(21)USD19F0262 USD19F0263 (01)00884838056251(21)USD19F0263 USD19F0312 (01)00884838056251(21)USD19F0312 USD19F0313 (01)00884838056251(21)USD19F0313 USD19F0386 (01)00884838056251(21)USD19F0386 USD19F0395 (01)00884838056251(21)USD19F0395 USD19F0396 (01)00884838056251(21)USD19F0396 USD19F0419 (01)00884838056251(21)USD19F0419 USD19F0420 (01)00884838056251(21)USD19F0420 USD19F0421 (01)00884838056251(21)USD19F0421 USD19F0422 (01)00884838056251(21)USD19F0422 USD19F0423 (01)00884838056251(21)USD19F0423 USD19F0426 (01)00884838056251(21)USD19F0426 USD19F0427 (01)00884838056251(21)USD19F0427 USD19F0487 (01)00884838056251(21)USD19F0487 USD19F0488 (01)00884838056251(21)USD19F0488 USD19F0489 (01)00884838056251(21)USD19F0489 USD19F0511 (01)00884838056251(21)USD19F0511 USD19F0512 (01)00884838056251(21)USD19F0512 USD19F0513 (01)00884838056251(21)USD19F0513 USD19F0514 (01)00884838056251(21)USD19F0514 USD19F0549 (01)00884838056251(21)USD19F0549 USD19F0550 (01)00884838056251(21)USD19F0550 USD19F0551 (01)00884838056251(21)USD19F0551 USD19F0553 (01)00884838056251(21)USD19F0553 USD19F0554 (01)00884838056251(21)USD19F0554 USD19F0555 (01)00884838056251(21)USD19F0555 USD19F0556 (01)00884838056251(21)USD19F0556 USD19F0557 (01)00884838056251(21)USD19F0557 USD19F0558 (01)00884838056251(21)USD19F0558 USD19F0559 (01)00884838056251(21)USD19F0559 USD19F0560 (01)00884838056251(21)USD19F0560 USD19F0561 (01)00884838056251(21)USD19F0561 USD19F0562 (01)00884838056251(21)USD19F0562 USD19F0563 (01)00884838056251(21)USD19F0563 USD19F0564 (01)00884838056251(21)USD19F0564 USD19F0565 (01)00884838056251(21)USD19F0565 USD19F0566 (01)00884838056251(21)USD19F0566 USD19F0571 (01)00884838056251(21)USD19F0571 USD19F0644 (01)00884838056251(21)USD19F0644 USD19F0645 (01)00884838056251(21)USD19F0645 USD19F0646 (01)00884838056251(21)USD19F0646 USD19F0647 (01)00884838056251(21)USD19F0647 USD19F0648 (01)00884838056251(21)USD19F0648 USD19F0649 (01)00884838056251(21)USD19F0649 USD19F0650 (01)00884838056251(21)USD19F0650 USD19F0651 (01)00884838056251(21)USD19F0651 USD19F0692 (01)00884838056251(21)USD19F0692 USD19F0693 (01)00884838056251(21)USD19F0693 USD19F0694 (01)00884838056251(21)USD19F0694 USD19F0725 (01)00884838056251(21)USD19F0725 USD19F0726 (01)00884838056251(21)USD19F0726 USD19F0727 (01)00884838056251(21)USD19F0727 USD19F0824 (01)00884838056251(21)USD19F0824 USD19F0825 (01)00884838056251(21)USD19F0825 USD19F0826 (01)00884838056251(21)USD19F0826 USD19F0876 (01)00884838056251(21)USD19F0876 USD19F0877 (01)00884838056251(21)USD19F0877 USD19F0906 (01)00884838056251(21)USD19F0906 USD19F0907 (01)00884838056251(21)USD19F0907 USD19F0908 (01)00884838056251(21)USD19F0908 USD19F0909 (01)00884838056251(21)USD19F0909 USD19F0918 (01)00884838056251(21)USD19F0918 USD19F0919 (01)00884838056251(21)USD19F0919 USD19F0927 (01)00884838056251(21)USD19F0927 USD19F0937 (01)00884838056251(21)USD19F0937 USD19F0938 (01)00884838056251(21)USD19F0938 USD19F0939 (01)00884838056251(21)USD19F0939 USD19F0940 (01)00884838056251(21)USD19F0940 USD19F0941 (01)00884838056251(21)USD19F0941 USD19F0942 (01)00884838056251(21)USD19F0942 USD19F0943 (01)00884838056251(21)USD19F0943 USD19F0944 (01)00884838056251(21)USD19F0944 USD19F1006 (01)00884838056251(21)USD19F1006 USD19F1007 (01)00884838056251(21)USD19F1007 USD19F1008 (01)00884838056251(21)USD19F1008 USD19F1074 (01)00884838056251(21)USD19F1074 USD19F1083 (01)00884838056251(21)USD19F1083 USD19F1084 (01)00884838056251(21)USD19F1084 USD19F1085 (01)00884838056251(21)USD19F1085 USD19F1106 (01)00884838056251(21)USD19F1106 USD19F1107 (01)00884838056251(21)USD19F1107 USD19F1108 (01)00884838056251(21)USD19F1108 USD19F1144 (01)00884838056251(21)USD19F1144 USD19F1162 (01)00884838056251(21)USD19F1162 USD19F1212 (01)00884838056251(21)USD19F1212 USD19F1215 (01)00884838056251(21)USD19F1215 USD19F1216 (01)00884838056251(21)USD19F1216 USD19F1218 (01)00884838056251(21)USD19F1218 USD19F1241 (01)00884838056251(21)USD19F1241 USD19F1260 (01)00884838056251(21)USD19F1260 USD19F1261 (01)00884838056251(21)USD19F1261 USD20F0010 (01)00884838056251(21)USD20F0010 USD20F0011 (01)00884838056251(21)USD20F0011 USD20F0434 (01)00884838056251(21)USD20F0434 USN14F0979 N/A USN15F0018 N/A USN15F0352 N/A USN15F0404 N/A USN15F0510 N/A USN16F0501 (01)00884838056251(21)USN16F0501 USN16F0972 (01)00884838056251(21)USN16F0972 USN16F1003 (01)00884838056251(21)USN16F1003 USN16F1244 (01)00884838056251(21)USN16F1244 USN16F1457 (01)00884838056251(21)USN16F1457 USN17F0380 (01)00884838056251(21)USN17F0380 USN17F0486 (01)00884838056251(21)USN17F0486 USN17F0651 (01)00884838056251(21)USN17F0651 USN17F0766 (01)00884838056251(21)USN17F0766 USN17F1039 (01)00884838056251(21)USN17F1039 USN17F1235 (01)00884838056251(21)USN17F1235 USN17F1236 (01)00884838056251(21)USN17F1236 USN17F1237 (01)00884838056251(21)USN17F1237 USN17F1238 (01)00884838056251(21)USN17F1238 USN17F1239 (01)00884838056251(21)USN17F1239 USN17F1240 (01)00884838056251(21)USN17F1240 USN17F1414 (01)00884838056251(21)USN17F1414 USN17F1582 (01)00884838056251(21)USN17F1582 USN18F0457 (01)00884838056251(21)USN18F0457 USN18F1114 (01)00884838056251(21)USN18F1114 USN18F1236 (01)00884838056251(21)USN18F1236 USN18F1237 (01)00884838056251(21)USN18F1237 USN18F1292 (01)00884838056251(21)USN18F1292 USN18F1526 (01)00884838056251(21)USN18F1526 USN18F1799 (01)00884838056251(21)USN18F1799 USN18F1800 (01)00884838056251(21)USN18F1800 USN18F1801 (01)00884838056251(21)USN18F1801 USN18F1802 (01)00884838056251(21)USN18F1802 USN18F1803 (01)00884838056251(21)USN18F1803 USN18F1804 (01)00884838056251(21)USN18F1804 USN18F1805 (01)00884838056251(21)USN18F1805 USN18F1806 (01)00884838056251(21)USN18F1806 USN18F1807 (01)00884838056251(21)USN18F1807 USN18F1808 (01)00884838056251(21)USN18F1808 USN18F1943 (01)00884838056251(21)USN18F1943 USN18F1944 (01)00884838056251(21)USN18F1944 USN18F1945 (01)00884838056251(21)USN18F1945 USN18F1946 (01)00884838056251(21)USN18F1946 USN18F1947 (01)00884838056251(21)USN18F1947 USN18F1948 (01)00884838056251(21)USN18F1948 USN18F1949 (01)00884838056251(21)USN18F1949 USN18F1950 (01)00884838056251(21)USN18F1950 USN18F1951 (01)00884838056251(21)USN18F1951 USN18F1952 (01)00884838056251(21)USN18F1952 USN18F1953 (01)00884838056251(21)USN18F1953 USN18F1954 (01)00884838056251(21)USN18F1954 USN18F1955 (01)00884838056251(21)USN18F1955 USN18F1956 (01)00884838056251(21)USN18F1956 USN18F1957 (01)00884838056251(21)USN18F1957 USN18F1958 (01)00884838056251(21)USN18F1958 USN18F1959 (01)00884838056251(21)USN18F1959 USN18F1960 (01)00884838056251(21)USN18F1960 USN18F1961 (01)00884838056251(21)USN18F1961 USN18F1962 (01)00884838056251(21)USN18F1962 USN18F1963 (01)00884838056251(21)USN18F1963 USN18F2230 (01)00884838056251(21)USN18F2230 USN19F0042 (01)00884838056251(21)USN19F0042 USN19F0043 (01)00884838056251(21)USN19F0043 USN19F0084 (01)00884838056251(21)USN19F0084 USN19F0087 (01)00884838056251(21)USN19F0087 USN19F0088 (01)00884838056251(21)USN19F0088 USN19F0089 (01)00884838056251(21)USN19F0089 USN19F0090 (01)00884838056251(21)USN19F0090 USN19F0091 (01)00884838056251(21)USN19F0091 USN19F0092 (01)00884838056251(21)USN19F0092 USN19F0093 (01)00884838056251(21)USN19F0093 USN19F0094 (01)00884838056251(21)USN19F0094 USN19F0095 (01)00884838056251(21)USN19F0095 USN19F0096 (01)00884838056251(21)USN19F0096 USN19F0165 (01)00884838056251(21)USN19F0165 USN19F0180 (01)00884838056251(21)USN19F0180 USN19F0181 (01)00884838056251(21)USN19F0181 USN19F0182 (01)00884838056251(21)USN19F0182 USN19F0183 (01)00884838056251(21)USN19F0183 USN19F0184 (01)00884838056251(21)USN19F0184 USN19F0185 (01)00884838056251(21)USN19F0185 USN19F0246 (01)00884838056251(21)USN19F0246 USN19F0247 (01)00884838056251(21)USN19F0247 USN19F0248 (01)00884838056251(21)USN19F0248 USN19F0249 (01)00884838056251(21)USN19F0249 USN19F0250 (01)00884838056251(21)USN19F0250 USN19F0251 (01)00884838056251(21)USN19F0251 USN19F0302 (01)00884838056251(21)USN19F0302 USN19F0303 (01)00884838056251(21)USN19F0303 USN19F0304 (01)00884838056251(21)USN19F0304 USN19F0305 (01)00884838056251(21)USN19F0305 USN19F0306 (01)00884838056251(21)USN19F0306 USN19F0307 (01)00884838056251(21)USN19F0307 USN19F0308 (01)00884838056251(21)USN19F0308 USN19F0309 (01)00884838056251(21)USN19F0309 USN19F0310 (01)00884838056251(21)USN19F0310 USN19F0311 (01)00884838056251(21)USN19F0311 USN19F0312 (01)00884838056251(21)USN19F0312 USN19F0338 (01)00884838056251(21)USN19F0338 USN19F0389 (01)00884838056251(21)USN19F0389 USN19F0390 (01)00884838056251(21)USN19F0390 USN19F0391 (01)00884838056251(21)USN19F0391 USN19F0392 (01)00884838056251(21)USN19F0392 USN19F0393 (01)00884838056251(21)USN19F0393 USN19F0394 (01)00884838056251(21)USN19F0394 USN19F0395 (01)00884838056251(21)USN19F0395 USN19F0396 (01)00884838056251(21)USN19F0396 USN19F0397 (01)00884838056251(21)USN19F0397 USN19F0398 (01)00884838056251(21)USN19F0398 USN19F0399 (01)00884838056251(21)USN19F0399 USN19F0400 (01)00884838056251(21)USN19F0400 USN19F0401 (01)00884838056251(21)USN19F0401 USN19F0402 (01)00884838056251(21)USN19F0402 USN19F0519 (01)00884838056251(21)USN19F0519 USN19F0521 (01)00884838056251(21)USN19F0521 USN19F0522 (01)00884838056251(21)USN19F0522 USN19F0524 (01)00884838056251(21)USN19F0524 USN19F0526 (01)00884838056251(21)USN19F0526 USN19F0527 (01)00884838056251(21)USN19F0527 USN19F0528 (01)00884838056251(21)USN19F0528 USN19F0529 (01)00884838056251(21)USN19F0529 USN19F0530 (01)00884838056251(21)USN19F0530 USN19F0580 (01)00884838056251(21)USN19F0580 USN19F0581 (01)00884838056251(21)USN19F0581 USN19F0597 (01)00884838056251(21)USN19F0597 USN19F0598 (01)00884838056251(21)USN19F0598 USN19F0599 (01)00884838056251(21)USN19F0599 USN19F0600 (01)00884838056251(21)USN19F0600 USN19F0601 (01)00884838056251(21)USN19F0601 USN19F0602 (01)00884838056251(21)USN19F0602 USN19F0603 (01)00884838056251(21)USN19F0603 USN19F0684 (01)00884838056251(21)USN19F0684 USN19F0685 (01)00884838056251(21)USN19F0685 USN19F0686 (01)00884838056251(21)USN19F0686 USN19F0687 (01)00884838056251(21)USN19F0687 USN19F0688 (01)00884838056251(21)USN19F0688 USN19F0689 (01)00884838056251(21)USN19F0689 USN19F0690 (01)00884838056251(21)USN19F0690 USN19F0731 (01)00884838056251(21)USN19F0731 USN19F0732 (01)00884838056251(21)USN19F0732 USN19F0733 (01)00884838056251(21)USN19F0733 USN19F0734 (01)00884838056251(21)USN19F0734 USN19F0735 (01)00884838056251(21)USN19F0735 USN19F0736 (01)00884838056251(21)USN19F0736 USN19F0737 (01)00884838056251(21)USN19F0737 USN19F0783 (01)00884838056251(21)USN19F0783 USN19F0805 (01)00884838056251(21)USN19F0805 USN19F0806 (01)00884838056251(21)USN19F0806 USN19F0807 (01)00884838056251(21)USN19F0807 USN19F0822 (01)00884838056251(21)USN19F0822 USN19F0823 (01)00884838056251(21)USN19F0823 USN19F0824 (01)00884838056251(21)USN19F0824 USN19F0825 (01)00884838056251(21)USN19F0825 USN19F0826 (01)00884838056251(21)USN19F0826 USN19F0827 (01)00884838056251(21)USN19F0827 USN19F0861 (01)00884838056251(21)USN19F0861 USN19F0862 (01)00884838056251(21)USN19F0862 USN19F0866 (01)00884838056251(21)USN19F0866 USN19F0936 (01)00884838056251(21)USN19F0936 USN19F0937 (01)00884838056251(21)USN19F0937 USN19F0938 (01)00884838056251(21)USN19F0938 USN19F0939 (01)00884838056251(21)USN19F0939 USN19F0940 (01)00884838056251(21)USN19F0940 USN19F0999 (01)00884838056251(21)USN19F0999 USN19F1003 (01)00884838056251(21)USN19F1003 USN19F1004 (01)00884838056251(21)USN19F1004 USN19F1005 (01)00884838056251(21)USN19F1005 USN19F1006 (01)00884838056251(21)USN19F1006 USN19F1007 (01)00884838056251(21)USN19F1007 USN19F1008 (01)00884838056251(21)USN19F1008 USN19F1009 (01)00884838056251(21)USN19F1009 USN19F1010 (01)00884838056251(21)USN19F1010 USN19F1011 (01)00884838056251(21)USN19F1011 USN19F1014 (01)00884838056251(21)USN19F1014 USN19F1015 (01)00884838056251(21)USN19F1015 USN19F1016 (01)00884838056251(21)USN19F1016 USN19F1124 (01)00884838056251(21)USN19F1124 USN19F1125 (01)00884838056251(21)USN19F1125 USN19F1126 (01)00884838056251(21)USN19F1126 USN19F1127 (01)00884838056251(21)USN19F1127 USN19F1128 (01)00884838056251(21)USN19F1128 USN19F1129 (01)00884838056251(21)USN19F1129 USN19F1130 (01)00884838056251(21)USN19F1130 USN19F1131 (01)00884838056251(21)USN19F1131 USN19F1132 (01)00884838056251(21)USN19F1132 USN19F1133 (01)00884838056251(21)USN19F1133 USN19F1134 (01)00884838056251(21)USN19F1134 USN19F1135 (01)00884838056251(21)USN19F1135 USN19F1136 (01)00884838056251(21)USN19F1136 USN19F1137 (01)00884838056251(21)USN19F1137 USN19F1138 (01)00884838056251(21)USN19F1138 USN19F1139 (01)00884838056251(21)USN19F1139 USN19F1140 (01)00884838056251(21)USN19F1140 USN19F1141 (01)00884838056251(21)USN19F1141 USN19F1142 (01)00884838056251(21)USN19F1142 USN19F1178 (01)00884838056251(21)USN19F1178 USN19F1291 (01)00884838056251(21)USN19F1291 USN19F1351 (01)00884838056251(21)USN19F1351 USN19F1352 (01)00884838056251(21)USN19F1352 USN19F1353 (01)00884838056251(21)USN19F1353 USN19F1354 (01)00884838056251(21)USN19F1354 USN19F1355 (01)00884838056251(21)USN19F1355 USN19F1356 (01)00884838056251(21)USN19F1356 USN19F1357 (01)00884838056251(21)USN19F1357 USN19F1362 (01)00884838056251(21)USN19F1362 USN19F1363 (01)00884838056251(21)USN19F1363 USN19F1383 (01)00884838056251(21)USN19F1383 USN19F1384 (01)00884838056251(21)USN19F1384 USN19F1385 (01)00884838056251(21)USN19F1385 USN19F1386 (01)00884838056251(21)USN19F1386 USN19F1387 (01)00884838056251(21)USN19F1387 USN19F1388 (01)00884838056251(21)USN19F1388 USN19F1389 (01)00884838056251(21)USN19F1389 USN19F1390 (01)00884838056251(21)USN19F1390 USN19F1391 (01)00884838056251(21)USN19F1391 USN19F1392 (01)00884838056251(21)USN19F1392 USN19F1394 (01)00884838056251(21)USN19F1394 USN19F1395 (01)00884838056251(21)USN19F1395 USN19F1443 (01)00884838056251(21)USN19F1443 USN19F1444 (01)00884838056251(21)USN19F1444 USN19F1445 (01)00884838056251(21)USN19F1445 USN19F1446 (01)00884838056251(21)USN19F1446 USN19F1447 (01)00884838056251(21)USN19F1447 USN19F1448 (01)00884838056251(21)USN19F1448 USN19F1449 (01)00884838056251(21)USN19F1449 USN19F1519 (01)00884838056251(21)USN19F1519 USN19F1520 (01)00884838056251(21)USN19F1520 USN19F1521 (01)00884838056251(21)USN19F1521 USN19F1522 (01)00884838056251(21)USN19F1522 USN19F1523 (01)00884838056251(21)USN19F1523 USN19F1524 (01)00884838056251(21)USN19F1524 USN19F1525 (01)00884838056251(21)USN19F1525 USN19F1526 (01)00884838056251(21)USN19F1526 USN19F1527 (01)00884838056251(21)USN19F1527 USN19F1529 (01)00884838056251(21)USN19F1529 USN19F1571 (01)00884838056251(21)USN19F1571 USN19F1572 (01)00884838056251(21)USN19F1572 USN19F1581 (01)00884838056251(21)USN19F1581 USN19F1582 (01)00884838056251(21)USN19F1582 USN19F1583 (01)00884838056251(21)USN19F1583 USN19F1584 (01)00884838056251(21)USN19F1584 USN19F1585 (01)00884838056251(21)USN19F1585 USN19F1586 (01)00884838056251(21)USN19F1586 USN19F1587 (01)00884838056251(21)USN19F1587 USN19F1635 (01)00884838056251(21)USN19F1635 USN19F1636 (01)00884838056251(21)USN19F1636 USN19F1637 (01)00884838056251(21)USN19F1637 USN19F1642 (01)00884838056251(21)USN19F1642 USN19F1746 (01)00884838056251(21)USN19F1746 USN19F1747 (01)00884838056251(21)USN19F1747 USN19F1748 (01)00884838056251(21)USN19F1748 USN19F1749 (01)00884838056251(21)USN19F1749 USN19F1750 (01)00884838056251(21)USN19F1750 USN19F1751 (01)00884838056251(21)USN19F1751 USN19F1752 (01)00884838056251(21)USN19F1752 USN19F1796 (01)00884838056251(21)USN19F1796 USN19F1797 (01)00884838056251(21)USN19F1797 USN19F1798 (01)00884838056251(21)USN19F1798 USN19F1799 (01)00884838056251(21)USN19F1799 USN19F1800 (01)00884838056251(21)USN19F1800 USN19F1801 (01)00884838056251(21)USN19F1801 USN19F1802 (01)00884838056251(21)USN19F1802 USN19F1803 (01)00884838056251(21)USN19F1803 USN19F1804 (01)00884838056251(21)USN19F1804 USN19F1805 (01)00884838056251(21)USN19F1805 USN19F1806 (01)00884838056251(21)USN19F1806 USN19F1807 (01)00884838056251(21)USN19F1807 USN19F1808 (01)00884838056251(21)USN19F1808 USN19F1809 (01)00884838056251(21)USN19F1809 USN19F1810 (01)00884838056251(21)USN19F1810 USN19F1811 (01)00884838056251(21)USN19F1811 USN19F1812 (01)00884838056251(21)USN19F1812 USN19F1813 (01)00884838056251(21)USN19F1813 USN19F1945 (01)00884838056251(21)USN19F1945 USN19F1946 (01)00884838056251(21)USN19F1946 USN19F1947 (01)00884838056251(21)USN19F1947 USN19F1948 (01)00884838056251(21)USN19F1948 USN19F1949 (01)00884838056251(21)USN19F1949 USN19F1950 (01)00884838056251(21)USN19F1950 USN19F1951 (01)00884838056251(21)USN19F1951 USN19F1952 (01)00884838056251(21)USN19F1952 USN19F1953 (01)00884838056251(21)USN19F1953 USN19F1954 (01)00884838056251(21)USN19F1954 USN19F1955 (01)00884838056251(21)USN19F1955 USN19F1956 (01)00884838056251(21)USN19F1956 USN19F1957 (01)00884838056251(21)USN19F1957 USN19F1966 (01)00884838056251(21)USN19F1966 USN19F2029 (01)00884838056251(21)USN19F2029 USN19F2030 (01)00884838056251(21)USN19F2030 USN19F2031 (01)00884838056251(21)USN19F2031 USN19F2113 (01)00884838056251(21)USN19F2113 USN19F2115 (01)00884838056251(21)USN19F2115 USN19F2116 (01)00884838056251(21)USN19F2116 USN19F2130 (01)00884838056251(21)USN19F2130 USN19F2131 (01)00884838056251(21)USN19F2131 USN19F2132 (01)00884838056251(21)USN19F2132 USN19F2133 (01)00884838056251(21)USN19F2133 USN19F2134 (01)00884838056251(21)USN19F2134 USN19F2135 (01)00884838056251(21)USN19F2135 USN19F2136 (01)00884838056251(21)USN19F2136 USN19F2201 (01)00884838056251(21)USN19F2201 USN19F2202 (01)00884838056251(21)USN19F2202 USN19F2235 (01)00884838056251(21)USN19F2235 USN19F2236 (01)00884838056251(21)USN19F2236 USN19F2237 (01)00884838056251(21)USN19F2237 USN19F2238 (01)00884838056251(21)USN19F2238 USN19F2239 (01)00884838056251(21)USN19F2239 USN19F2240 (01)00884838056251(21)USN19F2240 USN19F2241 (01)00884838056251(21)USN19F2241 USN19F2243 (01)00884838056251(21)USN19F2243 USN20F1487 (01)00884838056251(21)USN20F1487 USN20F1488 (01)00884838056251(21)USN20F1488 USN20F1489 (01)00884838056251(21)USN20F1489 USN20F1604 (01)00884838056251(21)USN20F1604 USN20F1605 (01)00884838056251(21)USN20F1605 USN20F1606 (01)00884838056251(21)USN20F1606 USO14F0538 N/A USO14F1076 N/A USO15F0053 N/A USO15F0403 N/A USO15F0447 N/A USO15F0910 N/A USO15F0984 N/A USO15F1005 N/A USO15F1006 N/A USO15F1159 N/A USO16F0089 (01)00884838056251(21)USO16F0089 USO16F0096 (01)00884838056251(21)USO16F0096 USO16F0097 (01)00884838056251(21)USO16F0097 USO16F0208 (01)00884838056251(21)USO16F0208 USO16F0220 (01)00884838056251(21)USO16F0220 USO16F0227 (01)00884838056251(21)USO16F0227 USO16F0274 (01)00884838056251(21)USO16F0274 USO16F0275 (01)00884838056251(21)USO16F0275 USO16F0773 (01)00884838056251(21)0023129282 USO16F0884 (01)00884838056251(21)USO16F0884 USO16F0970 (01)00884838056251(21)USO16F0970 USO17F0739 (01)00884838056251(21)USO17F0739 USO17F0740 (01)00884838056251(21)USO17F0740 USO17F1061 (01)00884838056251(21)USO17F1061 USO17F1062 (01)00884838056251(21)USO17F1062 USO17F1115 (01)00884838056251(21)USO17F1115 USO17F1116 (01)00884838056251(21)USO17F1116 USO17F1117 (01)00884838056251(21)USO17F1117 USO17F1118 (01)00884838056251(21)USO17F1118 USO17F1119 (01)00884838056251(21)USO17F1119 USO17F1120 (01)00884838056251(21)USO17F1120 USO17F1500 (01)00884838056251(21)USO17F1500 USO17F1576 (01)00884838056251(21)USO17F1576 USO18F0242 (01)00884838056251(21)USO18F0242 USO18F1501 (01)00884838056251(21)USO18F1501 USO18F1578 (01)00884838056251(21)USO18F1578 USO18F1618 (01)00884838056251(21)USO18F1618 USO18F1625 (01)00884838056251(21)USO18F1625 USO18F1767 (01)00884838056251(21)USO18F1767 USO18F1768 (01)00884838056251(21)USO18F1768 USO18F2087 (01)00884838056251(21)USO18F2087 USO18F2235 (01)00884838056251(21)USO18F2235 USO19F0017 (01)00884838056251(21)USO19F0017 USO19F0018 (01)00884838056251(21)USO19F0018 USO19F0023 (01)00884838056251(21)USO19F0023 USO19F0024 (01)00884838056251(21)USO19F0024 USO19F0025 (01)00884838056251(21)USO19F0025 USO19F0026 (01)00884838056251(21)USO19F0026 USO19F0027 (01)00884838056251(21)USO19F0027 USO19F0055 (01)00884838056251(21)USO19F0055 USO19F0056 (01)00884838056251(21)USO19F0056 USO19F0057 (01)00884838056251(21)USO19F0057 USO19F0058 (01)00884838056251(21)USO19F0058 USO19F0059 (01)00884838056251(21)USO19F0059 USO19F0060 (01)00884838056251(21)USO19F0060 USO19F0061 (01)00884838056251(21)USO19F0061 USO19F0067 (01)00884838056251(21)USO19F0067 USO19F0068 (01)00884838056251(21)USO19F0068 USO19F0069 (01)00884838056251(21)USO19F0069 USO19F0109 (01)00884838056251(21)USO19F0109 USO19F0110 (01)00884838056251(21)USO19F0110 USO19F0113 (01)00884838056251(21)USO19F0113 USO19F0114 (01)00884838056251(21)USO19F0114 USO19F0115 (01)00884838056251(21)USO19F0115 USO19F0116 (01)00884838056251(21)USO19F0116 USO19F0117 (01)00884838056251(21)USO19F0117 USO19F0118 (01)00884838056251(21)USO19F0118 USO19F0119 (01)00884838056251(21)USO19F0119 USO19F0120 (01)00884838056251(21)USO19F0120 USO19F0121 (01)00884838056251(21)USO19F0121 USO19F0122 (01)00884838056251(21)USO19F0122 USO19F0126 (01)00884838056251(21)USO19F0126 USO19F0127 (01)00884838056251(21)USO19F0127 USO19F0128 (01)00884838056251(21)USO19F0128 USO19F0129 (01)00884838056251(21)USO19F0129 USO19F0176 (01)00884838056251(21)USO19F0176 USO19F0177 (01)00884838056251(21)USO19F0177 USO19F0186 (01)00884838056251(21)USO19F0186 USO19F0216 (01)00884838056251(21)USO19F0216 USO19F0217 (01)00884838056251(21)USO19F0217 USO19F0218 (01)00884838056251(21)USO19F0218 USO19F0219 (01)00884838056251(21)USO19F0219 USO19F0283 (01)00884838056251(21)USO19F0283 USO19F0321 (01)00884838056251(21)USO19F0321 USO19F0323 (01)00884838056251(21)USO19F0323 USO19F0325 (01)00884838056251(21)USO19F0325 USO19F0564 (01)00884838056251(21)USO19F0564 USO19F0565 (01)00884838056251(21)USO19F0565 USO19F0588 (01)00884838056251(21)USO19F0588 USO19F0589 (01)00884838056251(21)USO19F0589 USO19F0642 (01)00884838056251(21)USO19F0642 USO19F0643 (01)00884838056251(21)USO19F0643 USO19F0644 (01)00884838056251(21)USO19F0644 USO19F0769 (01)00884838056251(21)USO19F0769 USO19F0770 (01)00884838056251(21)USO19F0770 USO19F0771 (01)00884838056251(21)USO19F0771 USO19F0800 (01)00884838056251(21)USO19F0800 USO19F0801 (01)00884838056251(21)USO19F0801 USO19F0802 (01)00884838056251(21)USO19F0802 USO19F0844 (01)00884838056251(21)USO19F0844 USO19F0845 (01)00884838056251(21)USO19F0845 USO19F0846 (01)00884838056251(21)USO19F0846 USO19F0847 (01)00884838056251(21)USO19F0847 USO19F0848 (01)00884838056251(21)USO19F0848 USO19F0913 (01)00884838056251(21)USO19F0913 USO19F0914 (01)00884838056251(21)USO19F0914 USO19F0915 (01)00884838056251(21)USO19F0915 USO19F0916 (01)00884838056251(21)USO19F0916 USO19F0917 (01)00884838056251(21)USO19F0917 USO19F0918 (01)00884838056251(21)USO19F0918 USO19F0919 (01)00884838056251(21)USO19F0919 USO19F0920 (01)00884838056251(21)USO19F0920 USO19F0921 (01)00884838056251(21)USO19F0921 USO19F0922 (01)00884838056251(21)USO19F0922 USO19F1038 (01)00884838056251(21)USO19F1038 USO19F1039 (01)00884838056251(21)USO19F1039 USO19F1040 (01)00884838056251(21)USO19F1040 USO19F1041 (01)00884838056251(21)USO19F1041 USO19F1042 (01)00884838056251(21)USO19F1042 USO19F1043 (01)00884838056251(21)USO19F1043 USO19F1044 (01)00884838056251(21)USO19F1044 USO19F1045 (01)00884838056251(21)USO19F1045 USO19F1046 (01)00884838056251(21)USO19F1046 USO19F1047 (01)00884838056251(21)USO19F1047 USO19F1672 (01)00884838056251(21)USO19F1672 USO19F1673 (01)00884838056251(21)USO19F1673 USO19F1674 (01)00884838056251(21)USO19F1674 USO19F1675 (01)00884838056251(21)USO19F1675 USO19F1676 (01)00884838056251(21)USO19F1676 USO19F1677 (01)00884838056251(21)USO19F1677 USO19F1708 (01)00884838056251(21)USO19F1708 USO19F1709 (01)00884838056251(21)USO19F1709 USO19F1710 (01)00884838056251(21)USO19F1710 USO19F1711 (01)00884838056251(21)USO19F1711 USO19F1712 (01)00884838056251(21)USO19F1712 USO19F1713 (01)00884838056251(21)USO19F1713 USO19F1734 (01)00884838056251(21)USO19F1734 USO19F1798 (01)00884838056251(21)USO19F1798 USO19F1799 (01)00884838056251(21)USO19F1799 USO19F1800 (01)00884838056251(21)USO19F1800 USO19F1807 (01)00884838056251(21)USO19F1807 USO19F1808 (01)00884838056251(21)USO19F1808 USO19F1809 (01)00884838056251(21)USO19F1809 USO19F1810 (01)00884838056251(21)USO19F1810 USO19F1840 (01)00884838056251(21)USO19F1840 USO19F1845 (01)00884838056251(21)USO19F1845 USO19F1846 (01)00884838056251(21)USO19F1846 USO19F1847 (01)00884838056251(21)USO19F1847 USO19F1916 (01)00884838056251(21)USO19F1916 USO19F1992 (01)00884838056251(21)USO19F1992 USO19F1993 (01)00884838056251(21)USO19F1993 USO19F1994 (01)00884838056251(21)USO19F1994 USO19F1995 (01)00884838056251(21)USO19F1995 USO19F1996 (01)00884838056251(21)USO19F1996 USO19F1997 (01)00884838056251(21)USO19F1997 USO19F1998 (01)00884838056251(21)USO19F1998 USO19F1999 (01)00884838056251(21)USO19F1999 USO19F2017 (01)00884838056251(21)USO19F2017 USO19F2018 (01)00884838056251(21)USO19F2018 USO19F2019 (01)00884838056251(21)USO19F2019 USO19F2020 (01)00884838056251(21)USO19F2020 USO19F2021 (01)00884838056251(21)USO19F2021 USO19F2022 (01)00884838056251(21)USO19F2022 USO19F2023 (01)00884838056251(21)USO19F2023 USO19F2033 (01)00884838056251(21)USO19F2033 USO19F2034 (01)00884838056251(21)USO19F2034 USO19F2035 (01)00884838056251(21)USO19F2035 USO19F2036 (01)00884838056251(21)USO19F2036 USO19F2073 (01)00884838056251(21)USO19F2073 USO19F2075 (01)00884838056251(21)USO19F2075 USO19F2121 (01)00884838056251(21)USO19F2121 USO19F2123 (01)00884838056251(21)USO19F2123 USO19F2124 (01)00884838056251(21)USO19F2124 USO19F2125 (01)00884838056251(21)USO19F2125 USO19F2126 (01)00884838056251(21)USO19F2126 USO19F2127 (01)00884838056251(21)USO19F2127 USO19F2154 (01)00884838056251(21)USO19F2154 USO19F2200 (01)00884838056251(21)USO19F2200 USO19F2201 (01)00884838056251(21)USO19F2201 USO19F2202 (01)00884838056251(21)USO19F2202 USO19F2203 (01)00884838056251(21)USO19F2203 USO19F2204 (01)00884838056251(21)USO19F2204 USO19F2205 (01)00884838056251(21)USO19F2205 USO19F2303 (01)00884838056251(21)USO19F2303 USO19F2356 (01)00884838056251(21)USO19F2356 USO19F2379 (01)00884838056251(21)USO19F2379 USO19F2380 (01)00884838056251(21)USO19F2380 USO19F2381 (01)00884838056251(21)USO19F2381 USO19F2382 (01)00884838056251(21)USO19F2382 USO19F2383 (01)00884838056251(21)USO19F2383 USO19F2409 (01)00884838056251(21)USO19F2409 USO19F2410 (01)00884838056251(21)USO19F2410 USO19F2411 (01)00884838056251(21)USO19F2411 USO19F2412 (01)00884838056251(21)USO19F2412 USO19F2415 (01)00884838056251(21)USO19F2415 USO19F2416 (01)00884838056251(21)USO19F2416 USO19F2496 (01)00884838056251(21)USO19F2496 USO19F2497 (01)00884838056251(21)USO19F2497 USO19F2499 (01)00884838056251(21)USO19F2499 USO19F2501 (01)00884838056251(21)USO19F2501 USO19F2502 (01)00884838056251(21)USO19F2502 USO19F2503 (01)00884838056251(21)USO19F2503 USO19F2504 (01)00884838056251(21)USO19F2504 USO19F2545 (01)00884838056251(21)USO19F2545 USO19F2546 (01)00884838056251(21)USO19F2546 USO19F2586 (01)00884838056251(21)USO19F2586 USO19F2587 (01)00884838056251(21)USO19F2587 USO19F2588 (01)00884838056251(21)USO19F2588 USO19F2617 (01)00884838056251(21)USO19F2617 USO19F2618 (01)00884838056251(21)USO19F2618 USO19F2619 (01)00884838056251(21)USO19F2619 USO19F2620 (01)00884838056251(21)USO19F2620 USO19F2689 (01)00884838056251(21)USO19F2689 USO19F2763 (01)00884838056251(21)USO19F2763 USO19F2764 (01)00884838056251(21)USO19F2764 USO19F2765 (01)00884838056251(21)USO19F2765 USO19F2768 (01)00884838056251(21)USO19F2768 USO19F2769 (01)00884838056251(21)USO19F2769 USO19F2784 (01)00884838056251(21)USO19F2784 USO19F2785 (01)00884838056251(21)USO19F2785 USO19F2786 (01)00884838056251(21)USO19F2786 USO19F2787 (01)00884838056251(21)USO19F2787 USO19F2788 (01)00884838056251(21)USO19F2788 USO19F2789 (01)00884838056251(21)USO19F2789 USO19F2790 (01)00884838056251(21)USO19F2790 USO19F2791 (01)00884838056251(21)USO19F2791 USO19F2792 (01)00884838056251(21)USO19F2792 USO19F2793 (01)00884838056251(21)USO19F2793 USO19F2794 (01)00884838056251(21)USO19F2794 USO19F2795 (01)00884838056251(21)USO19F2795 USO19F2796 (01)00884838056251(21)USO19F2796 USO19F2797 (01)00884838056251(21)USO19F2797 USO19F2798 (01)00884838056251(21)USO19F2798 USO19F2916 (01)00884838056251(21)USO19F2916 USO19F2917 (01)00884838056251(21)USO19F2917 USO19F2918 (01)00884838056251(21)USO19F2918 USO19F2919 (01)00884838056251(21)USO19F2919 USO19F2920 (01)00884838056251(21)USO19F2920 USO19F2921 (01)00884838056251(21)USO19F2921 USO19F2924 (01)00884838056251(21)USO19F2924 USO19F2925 (01)00884838056251(21)USO19F2925 USO20F1789 (01)00884838056251(21)USO20F1789 Affiniti 70 Model Number: 795119 Serial Numbers/UDI Codes US120F1622 (01)00884838056251(21)US120F1622 US215F0692 N/A US414F0194 N/A US419F0618 (01)00884838056251(21)US419F0618 US514F0204 N/A US519F0207 (01)00884838056251(21)US519F0207 US714F0716 N/A US719F1248 (01)00884838056251(21)US719F1248 US720F1570 (01)00884838056251(21)US720F1570 US720F1571 (01)00884838056251(21)US720F1571 US720F1572 (01)00884838056251(21)US720F1572 US720F1573 (01)00884838056251(21)US720F1573 US720F1574 (01)00884838056251(21)US720F1574 US814F0056 N/A US918F1096 (01)00884838056251(21)US918F1096 USD20F1270 (01)00884838056251(21)USD20F1270 USO18F2086 (01)00884838056251(21)USO18F2086 USO19F0187 (01)00884838056251(21)USO19F0187 USO19F0188 (01)00884838056251(21)USO19F0188 USO19F0189 (01)00884838056251(21)USO19F0189 USO19F1915 (01)00884838056251(21)USO19F1915
Recalling Firm/
Manufacturer
Philips Ultrasound Inc
22100 Bothell Everett Hwy
Bothell WA 98021-8431
For Additional Information ContactHolly Wright Lee
440-483-2015
Manufacturer Reason
for Recall
Due to a software defect that can intermittently cause the system to lock-up which exiting Review Mode while performing an exam.
FDA Determined
Cause 2
Software design
ActionOn 08/06/2021, the firm sent via certified mail an "URGENT Medical Device Correction" Notification Letter to customer informing them that due to a software defect, when reviewing or adjusting xPlane exam results, there is a potential for the ultrasound systems to become unresponsive (Lock-up) preventing users from continuing clinical use. The ultrasound system provides an error notification including method to restart and bring system back to normal use. The Recalling Firm has received a report of an ultrasound system locking up multiple times during an open cardiac procedure while using a Transesophageal Echocardiogram (TEE) transducer. Each lock-up required at least one system reboot to regain functionality. Customers are instructed to: 1) Determine the software version on their ultrasound system by: - Power up the system and allow it to complete the boot sequence; - Press Support on the right side of the control panel; - Under System Management, click System Information; - The software version is listed in the Software Information Section. 2) Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm. 3) Post this notice near the affected unit(s) for ease of reference. 4) Complete and return the response form to the Recalling Firm promptly and no later than 30 days from receipt via email to: ultrasound.corrections@philips.com or fax to 1-833-512-7756. In addition, because the software defect may intermittently cause the system to lock-up when exiting Review mode, customer are asked to take the following steps to minimize the likelihood of occurrence: 1. Acquire a loop while in xPlane, xPlane Doppler, or Dual mode. 2. Do NOT adjust an imaging control (for example: Rotate, Tilt, Gain) while the loop capture is in progress.* (*If the imaging controls have been adjusted during the loop capture, Philips recommends that the system be rebooted prior to reviewing the loop in Review mode). 3. Allow the loop
Quantity in CommerceTotal = 10,583 systems (EPIQ and Afiniti)
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Albania, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Brazil, Brunei Darussalam, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, French Guiana, Germany, Gibraltar, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Virgin Island, Yemen and Zambia.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = IYO
-
-