• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall SuperCoreTM SemiAutomatic Biopsy Instrument

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall SuperCoreTM SemiAutomatic Biopsy Instrumentsee related information
Date Initiated by FirmOctober 04, 2021
Create DateDecember 09, 2021
Recall Status1 Open3, Classified
Recall NumberZ-0354-2022
Recall Event ID 88814
510(K)NumberK974814 
Product Classification Instrument, biopsy - Product Code KNW
ProductSuperCore Biopsy Instrument: REF/Product Description: 701114090/SUPERCORE BIOPSY INSTR 14GA X 9CM; 701114150/SUPERCORE BIOPSY INSTR 14GA X 15CM; 701116090/SUPERCORE BIOPSY INSTR 16GA X 9CM; 701118060/SUPERCORE BIOPSY INSTR 18GA X 6CM; 701118090/SUPERCORE BIOPSY INSTR 18GA X 9CM; 701118150/SUPERCORE BIOPSY INSTR 18GA X 15CM; 701118200/SUPERCORE BIOPSY INSTR 18GA X 20CM; 701120090/SUPERCORE BIOPSY INSTR 20GA X 9CM; 701120150/SUPERCORE BIOPSY INSTR 20GA X 15CM; 701120200/SUPERCORE BIOPSY INSTR 20GA X 20CM; 701214090/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 9CM; 701214150/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 15CM; 701216090/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 9CM; 701216150/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 15CM; 701218090/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 9CM; 701218150/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 15CM; 701218200/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 20CM; 701220090/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 9CM; 701220150/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 15CM; 701220200/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 20CM; 701114090M/SUPERCORE BIOPSY INSTR 14GA X 9CM-MAMMOTOME; 701114150M/SUPERCORE BIOPSY INSTR 14GA X 15CM-MAMMOTOME; 701116090M/SUPERCORE BIOPSY INSTR 16GA X 9CM-MAMMOTOME; SC 1409/PGI Supercore 14GA x 9CM; SC 1609/PGI Supercore 16GA x 9CM; SC 1806/PGI Supercore 18GA x 6CM;
Code Information REF/Product Description: 701114090/SUPERCORE BIOPSY INSTR 14GA X 9CM; Lot Numbers: 11364350 11365643 11368388 11371872 11374998 11376772 11378864 11382837 11385962 11389848 701114150/SUPERCORE BIOPSY INSTR 14GA X 15CM; Lot Numbers: 11368389; 11376773; 701116090/SUPERCORE BIOPSY INSTR 16GA X 9CM; Lot Numbers: 11362716; 11363320; 11365935; 11377634; 11381340; 11384192; 11388461; 11388552; 11389139; 701118060/SUPERCORE BIOPSY INSTR 18GA X 6CM; Lot Numbers: 11361702; 11365788; 11371012; 11374235; 11376598; 11379722; 11382768; 11383890; 11388656; 11389138; 701118090/SUPERCORE BIOPSY INSTR 18GA X 9CM; Lot Numbers: 11362759; 11363229; 11364420; 11367439; 11370106; 11372917; 11376074; 11377430; 11378217; 11383193; 11383785; 11384243; 11387238; 11388864; 701118150/SUPERCORE BIOPSY INSTR 18GA X 15CM; Lot Numbers: 11370107; 11375663; 11377819; 11382602; 11385139; 11385965; 11389140; 11390503; 701118200/SUPERCORE BIOPSY INSTR 18GA X 20CM; Lot Numbers: 11372225; 11374410; 11377431; 11384193; 11388505; 701120090/SUPERCORE BIOPSY INSTR 20GA X 9CM; Lot Numbers: 11374411; 11378865; 11383280; 701120150/SUPERCORE BIOPSY INSTR 20GA X 15CM; Lot Numbers: 11366252 11367440; 11376374; 11377820; 11381387; 11385960; 701120200/SUPERCORE BIOPSY INSTR 20GA X 20CM; Lot Numbers: 11366253; 11380018; 11388893; 701214090/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 9CM; Lot Numbers: 11369916; 11378293; 11383587; 11387065; 701214150/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 15CM; Lot Numbers: 11377967; 11382843; 701216090/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 9CM; Lot Numbers: 11364421; 11367443; 11374684; 11379724; 11385966; 11388551; 701216150/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 15CM; Lot Numbers: 11382845; 11388550; 11390251; 701218090/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 9CM; Lot Numbers: 11361701; 11364247; 11367445; 11372366; 11374807; 11378687; 11383192; 11383784; 11388286; 701218150/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 15CM; Lot Numbers: 11371404; 11373084; 11377636; 11380993; 11383194; 11383195; 11387400; 11387833; 11390504; 701218200/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 20CM; Lot Numbers: 11366315; 11372227; 11377432; 11381193; 11382998; 11384716; 701220090/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 9CM; Lot Numbers: 11367446; 11370574; 11375664; 11380589; 11381572; 11388507; 701220150/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 15CM; Lot Numbers: 11368750; 11371876; 11372228; 11376375; 11377429; 11380104; 11384717; 11387148; 701220200/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 20CM; Lot Numbers: 11362655; 11363516; 11372322; 11378295; 11380103; 11382884; 11387584; 701114090M/SUPERCORE BIOPSY INSTR 14GA X 9CM-MAMMOTOME; Lot Numbers: 11371356; 11382375; 11389615; 701114150M/SUPERCORE BIOPSY INSTR 14GA X 15CM-MAMMOTOME; Lot Number: 11371357; 701116090M/SUPERCORE BIOPSY INSTR 16GA X 9CM-MAMMOTOME; Lot Number: 11371359; SC 1409/PGI Supercore 14GA x 9CM; Lot Numbers: 11370550; 11384881; SC 1609/PGI Supercore 16GA x 9CM; Lot Number: 11370552; SC 1806/PGI Supercore 18GA x 6CM; Lot Numbers: 11375600; 11364469;
FEI Number 1625425
Recalling Firm/
Manufacturer
Argon Medical Devices, Inc
1445 Flat Creek Rd
Athens TX 75751-5002
For Additional Information ContactRebecca Ellis
(469) -731-1410
Manufacturer Reason
for Recall
Due to component (housing and plunger) detaching from the semi-automatic biopsy instruments.
FDA Determined
Cause 2
Device Design
ActionOn 09/30/2021, the firm sent an "Urgent-Product Recall Notice" letter to customers via US Postal Service Registered Mail informing them that the firm has received complaints regarding the semi-automatic Biopsy instrument coming apart during shipping or prior to use. According to the firm's investigation, the plastic housing and plunger can be separated more easily than normal for the lots manufactured during this time frame. Customers are instructed to return all affected product to the Recalling Firm's Athens TX facility using RGA#26683, attention Andrea Wieczor. Mailing address is: RGA# 26683 Argon Medical Devices, Inc. 1445 Flat Creek Road Athens, TX 75751 USA Att: Andrea Wieczor In addition, Customers are instructed to complete the response acknowledgement form included with the letter/email and return by fax, mail, or email as soon as possible. For questions. contact the Director, Post-Market Experience at email address Brian.Rogers@argonmedical.com, or contact Arbee Cummings at Arbee.Cummings@argonmedical.com or Andrea Wieczor at Andrea.Wieczor@argonmedical.com. (Updated information 11/2/2021), on 10/28/2021, the firm sent an updated "Urgent-Product Recall Notice for Additional Lots" letter via mail informing customers that they are expanding the recall to additional lots (list on the Inventory Sheet attached to recall notice). Again the Recalling Firm is instructing customer to return the affected product (listed on the Inventory Sheet).
Quantity in Commerce79,910 devices
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. The countries of ALB, ARE, ARG, AUS, AUT, BEL, BGD, BRA, CAN, CHE, CHL, CHN, COL, CZE, DEU, ECU, ESP, EST, FRA, GBR, GRC, HKG, HUN, IND, IRN, ISR, ITA, JOR, JPN, MAR, MEX, MYS, NLD, NPL, NZL, PAK, PER, PHL, POL, PRT, SAU, SGP, SRB, THA, TUR, and TWN.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = KNW
-
-