Date Initiated by Firm | July 12, 2022 |
Create Date | August 18, 2022 |
Recall Status1 |
Terminated 3 on November 07, 2024 |
Recall Number | Z-1558-2022 |
Recall Event ID |
90632 |
510(K)Number | K150493 |
Product Classification |
Computer, diagnostic, programmable - Product Code DQK
|
Product | AXIOM Sensis or Sensis Vibe Combo systems as follows:
System Material #
AXIOM Sensis, Combo 64 6623974;
AXIOM Sensis, Hemo LOW 6634633;
AXIOM Sensis, Combo 32 6634641;
AXIOM Sensis, EP 129 6634658;
Sensis 10764561;
Sensis Vibe Combo 11007642;
Sensis recording system is intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology. |
Code Information |
1.Sensis 10764561 UDI-DI:04056869010137 2. Sensis Vibe Combo 11007642 UDI-DI: 04056869010205 Serial Numbers: 2022 2025 2032 2060 2113 2139 2154 2155 2186 2187 2201 2205 2224 2225 2227 2229 2269 2272 2276 2282 2291 2293 2294 2331 2332 2349 2350 2352 2403 2408 2425 2428 2442 2444 2500 2501 2515 2520 2521 2532 2552 2563 2576 2578 2581 2589 2608 2629 2744 3014 3015 3021 3037 3038 3041 3049 3050 3051 3057 3059 3060 3069 3075 3091 3097 3101 3103 3112 3128 3129 3134 3136 3140 3142 3166 3205 3206 3224 3226 3316 3321 3333 3352 3383 3390 3405 3406 3407 3408 3417 3432 3434 3437 3440 3464 3472 3475 3529 3534 3536 3564 3641 3642 3650 3653 3655 3660 3671 3676 4003 6101 6103 6110 6208 6622 6625 6631 6702 6727 8067 8068 20004 20015 20016 20023 20026 20045 20047 20048 20056 20062 20065 20109 20127 20139 20161 20162 20319 20321 20380 20383 20385 20391 20392 20399 20407 20430 20448 20466 20525 20538 20546 20556 20557 20561 20573 20574 20575 20585 20598 20606 20618 20622 20660 21000 21034 21035 21037 21043 21051 21052 21056 21074 21076 21077 21095 21096 21098 21100 21102 21106 21109 21112 21226 21244 21297 21332 21334 21345 21346 21363 21383 21384 21388 21391 21399 21438 21452 21498 22000 22022 22037 22051 22217 22249 22251 22252 22267 22282 22312 22336 22341 22345 22357 22366 22406 22415 22423 22427 22436 22437 22438 22461 22462 22474 22480 22490 22512 22522 22523 22526 22528 22605 22612 22619 22623 22627 22628 22646 22648 22651 22652 22663 22667 22669 22670 22671 22681 22682 22688 22708 22720 40011 40012 40108 41017 41019 41021 41202 42109 42209 42308 60020 60022 60023 60024 60025 60040 60052 60058 60060 60062 60065 60079 60087 60089 60095 60110 60111 60116 60120 60123 60125 60127 60128 60129 60130 60134 60138 60143 60146 60147 60154 60166 60167 60171 60308 60311 60312 60322 60323 60325 60347 60348 60355 60369 60370 60372 60395 60410 60412 60427 60436 60444 60453 60468 60473 60474 60480 60483 61002 61008 61009 61014 61030 61031 61048 61061 61064 61071 61072 61081 61083 61084 61086 61089 61103 61108 61116 61121 61124 61128 61129 61133 61138 61141 61142 61143 61144 61151 61169 61183 61186 61198 61201 61207 61219 61224 61225 61239 61241 61246 61251 61253 61256 61266 61281 61292 61297 61299 61302 61305 61309 61310 61313 61315 61317 61318 61320 61325 61330 61331 61344 61345 61394 61403 121014 122006 122011 122026 122029 122042 123004 123021 123047 123053 123055 123056 123058 |
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc 40 Liberty Blvd Malvern PA 19355-1418
|
For Additional Information Contact | SAME 610-219-4834 |
Manufacturer Reason for Recall | The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting position deviates from the default position, the following are possible: Patient touching the Signal Input Box may result in a touch leakage current to the patient; Spillage of liquid entering the Signal Input Box |
FDA Determined Cause 2 | Device Design |
Action | Siemens Healthineers issued Urgent Medical Device Correction letter dated 7/12/22 to US consignees. The letter states the reason for recall, health risk and action to take:
Please follow the instructions in the attached addendum which are a supplement to the Operator Manual chapter "Positioning the signal input box".
Siemens has updated the affected documentation and is distributing an Addendum to the Operator Manual to all affected
customers via Update Instruction AX033/21/S. Please file this Customer Safety Advisory Notice and Addendum to the Operator Manual with all system documentation. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. |
Quantity in Commerce | 399 units U.S. |
Distribution | US Nationwide Distribution
Foreign:
Albania
Algeria
Angola
Argentina
Armenia
Aruba
Australia
Austria
Azerbaijan
Bahrain
Bangladesh
Belgium
Bolivia
Bosnia and Herzegovina
Brunei Darussalam
Bulgaria
Cambodia
Cameroon
Canada
Chile
China
Colombia
Costa Rica
Croatia
Cuba
Cyprus
Czech Republic
Denmark
Ecuador
Egypt
El Salvador
Estonia
Ethiopia
Finland
France
Georgia
Germany
Ghana
Greece
Guatemala
Hong Kong
Hungary
India
Indonesia
Iran
Iraq
Ireland
Israel
Italy
Japan
Jordan
Kazakhstan
Kenya
Kosovo
Kuwait
Latvia
Lebanon
Libya
Luxembourg
Macao
Malaysia
Maldives
Mali
Mauritius
Mexico
Moldova
Montenegro
Morocco
Mozambique
Myanmar
Namibia
Netherlands
New Caledonia
New Zealand
Nigeria
North Korea
Norway
Oman
Pakistan
Palestine, State of
Panama
Paraguay
Peru
Philippines
Poland
Portugal
Qatar
Romania
Rwanda
Saudi Arabia
Serbia
Singapore
Slovakia
Slovenia
South Africa
South Korea
Spain
Sri Lanka
Sudan
Suriname
Sweden
Switzerland
Syria
Taiwan
Tajikistan
Tanzania
Thailand
Trinidad and Tobago
Tunisia
Turkey
Turkmenistan
Uganda
Ukraine
United Arab Emirates
United Kingdom
Uruguay
Uzbekistan
Venezuela, Bolivarian Republic of
Vietnam
Yemen |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = DQK
|