Date Initiated by Firm | September 27, 2022 |
Date Posted | October 18, 2022 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-0099-2023 |
Recall Event ID |
90918 |
510(K)Number | K142292 |
Product Classification |
Staple, fixation, bone - Product Code JDR
|
Product | SPEED 15X12X12mm Continuous Compression Implant, Product No. SE-1512 |
Code Information |
UDI-DI (GTIN): 00810633020166
Lot/Expiry
MSE220445 7/1/2027
MSE210184 8/1/2026 |
Recalling Firm/ Manufacturer |
Synthes (USA) Products LLC 1301 Goshen Pkwy West Chester PA 19380-5986
|
Manufacturer Reason for Recall | The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit. |
FDA Determined Cause 2 | Error in labeling |
Action | On September 27, 2022, the firm notified affected customers via urgent medical Device Recall letters. Customers were instructed to examine their inventory to determine if they have affected product and to quarantine them immediately - Customers should not use the recalled products.
Customers should contact their DePuy Synthes Sales Consultant or call the customer support services at ra-dpyus-csspecial@its.jnj.com to coordinate the return/credits of the subject products. |
Quantity in Commerce | 94 US; 30 OUS |
Distribution | US Nationwide distribution. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = JDR
|