Class 2 Device Recall AcuTwist Acutrak Compression Screws |
|
Date Initiated by Firm |
November 29, 2022 |
Date Posted |
December 23, 2022 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-0770-2023 |
Recall Event ID |
90996 |
510(K)Number |
K221333
|
Product Classification |
Screw, fixation, bone - Product Code HWC
|
Product |
Dual-Trak Clavicle Screws Part Number/Part Description: 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 40-0136-S 3.0mm x 80mm Dual-Trak Clavicle Screw 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 40-0137-S 3.0mm x 90mm Dual-Trak Clavicle Screw 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 40-0138-S 3.0mm x 100mm Dual-Trak Clavicle Screw 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 40-0139-S 3.0mm x 110mm Dual-Trak Clavicle Screw 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 40-0140-S 3.0mm x 120mm Dual-Trak Clavicle Screw 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 40-0143-S 3.8mm x 80mm Dual-Trak Clavicle Screw 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 40-0144-S 3.8mm x 90mm Dual-Trak Clavicle Screw 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 40-0145-S 3.8mm x 100mm Dual-Trak Clavicle Screw 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 40-0146-S 3.8mm x 110mm Dual-Trak Clavicle Screw 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 40-0147-S 3.8mm x 120mm Dual-Trak Clavicle Screw
|
Code Information |
Dual-Trak Clavicle Screws:
Batch Number/Part Number/Part Description/UDI Code:
237352 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-237352000000
241396 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-241396000000
251499 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-251499000000
273721 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-273721000000
295968 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-295968000000
305419 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-305419000000
320123 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-320123000000
320365 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-320365000000
320998 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-320998000000
353959 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-353959000000
360734 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-360734000000
361151 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-361151000000
376436 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-376436000000
377113 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-377113000000
379009 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-379009000000
379939 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-379939000000
379920 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-379920000000
410141 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-410141170413
410142 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-410142170413
423108 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-423108171009
423212 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-423212171009
441951 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-441951180523
442003 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-442003180524
477282 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-477282190808
477323 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-477323190809
500931 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-500931200601
501581 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-501581200610
525006 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-525006210405
525005 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-525005210405
525007 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-525007210405
525102 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-525102210406
525505 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-525505210412
556268 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-556268220222
556269 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-556269220222
556358 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028410-556358220223
293645 40-0136-S 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028427-293645000000200228
383916 40-0136-S 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028427-383916160608230608
421116 40-0136-S 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028427-421116170906240905
534467 40-0136-S 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028427-534467210709280708
565996 40-0136-S 3.0mm x 80mm Dual-Trak Clavicle Screw 10806378028427-565996220615290614
241398 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-241398000000
251500 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-251500000000
273722 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-273722000000
295969 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-295969000000
305610 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-305610000000
320392 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-320392000000
320999 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-320999000000
354105 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-354105000000
360645 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-360645000000
360733 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-360733000000
376553 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-376553000000
380465 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-380465000000
380466 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-380466000000
410223 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-410223170413
410222 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-410222170413
425329 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-425329171030
441811 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-441811180522
441950 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-441950180522
477378 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-477378190812
477519 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-477519190813
501531 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-501531200609
525199 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-525199210407
525200 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-525200210407
555633 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-555633220215
555695 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-555695220216
555694 40-0137 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028434-555694220216
293646 40-0137-S 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028441-293646000000200228
422954 40-0137-S 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028441-422954171004241003
477838 40-0137-S 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028441-477838190816260815
515467 40-0137-S 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028441-515467201210271210
534468 40-0137-S 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028441-534468210709280708
570282 40-0137-S 3.0mm x 90mm Dual-Trak Clavicle Screw 10806378028441-570282220804290803
237369 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-237369000000
241399 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-241399000000
251501 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-251501000000
295970 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-295970000000
305644 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-305644000000
320393 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-320393000000
361063 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-361063000000
376681 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-376681000000
377214 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-377214000000
380638 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-380638000000
381141 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-381141170717
424465 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-424465171019
424175 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-424175171019
477520 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-477520190813
500956 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-500956200602
525287 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-525287210408
525201 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-525201210407
525288 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-525288210408
556907 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-556907220302
556774 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-556774220228
557543 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-557543220310
556608 40-0138 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028458-556608000000
293647 40-0138-S 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028465-293647000000200228
378213 40-0138-S 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028465-378213160322230322
415996 40-0138-S 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028465-415996170627240626
486375 40-0138-S 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028465-486375191203261202
560151 40-0138-S 3.0mm x 100mm Dual-Trak Clavicle Screw 10806378028465-560151220403290402
241400 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-241400000000
251502 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-251502000000
305645 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-305645000000
320950 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-320950000000
320903 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-320903000000
377330 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-377330000000
320950A 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-320950A170720
423259 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-423259171013
423258 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-423258171012
525104 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-525104210406
525103 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-525103210406
554545 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-554545220203
554669 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-554669220204
554933 40-0139 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028472-554933220207
293663 40-0139-S 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028489-293663000000200228
402890 40-0139-S 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028489-402890170113240113
469205 40-0139-S 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028489-469205190423260422
534470 40-0139-S 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028489-534470210709280708
570283 40-0139-S 3.0mm x 110mm Dual-Trak Clavicle Screw 10806378028489-570283220804290803
237368 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-237368000000
241407 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-241407000000
251503 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-251503000000
320970 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-320970000000
320970A 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-320970A170630
423354 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-423354220103
423579 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-423579220103
555104 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-555104220208
555184 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-555184220209
555183 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-555183220209
557309 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-557309220308
557310 40-0140 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028496-557310220308
293664 40-0140-S 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028502-293664000000200228
400322 40-0140-S 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028502-400322161212231212
469203 40-0140-S 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028502-469203190423260422
565999 40-0140-S 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028502-565999220615290614
570284 40-0140-S 3.0mm x 120mm Dual-Trak Clavicle Screw 10806378028502-570284220804290803
237370 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-237370000000
241408 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-241408000000
251504 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-251504000000
305469 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-305469000000
361064 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-361064000000
376799 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-376799000000
380044 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-380044170531
380768 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-380768170531
477625 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-477625190814
501582 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-501582200610
555550 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-555550220214
555575 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-555575220214
555635 40-0143 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028519-555635220215
293665 40-0143-S 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028526-293665000000200228
293665A 40-0143-S 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028526-293665A171003241002
469204 40-0143-S 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028526-469204190423260422
565998 40-0143-S 3.8mm x 80mm Dual-Trak Clavicle Screw 10806378028526-565998220615290614
241409 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-241409000000
251505 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-251505000000
305697 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-305697000000
320832 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-320832000000
321000 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-321000170616
361065 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-361065000000
379515 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-379515000000
380368 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-380368170616
424468 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-424468171019
477626 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-477626190814
501493 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-501493200609
555878 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-555878220217
555877 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-555877220217
557544 40-0144 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028533-557544220310
293666 40-0144-S 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028540-293666000000200228
424385 40-0144-S 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028540-424385171018241017
469983 40-0144-S 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028540-469983190503260502
566000 40-0144-S 3.8mm x 90mm Dual-Trak Clavicle Screw 10806378028540-566000220615290614
237371 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-237371000000
241410 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-241410000000
251506 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-251506000000
320969 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-320969000000
376800 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-376800000000
379875 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-379875000000
380969 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-380969170605
379875A 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-379875A170605
425039 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-425039171025
477695 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-477695190815
525289 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-525289210408
525290 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-525290210408
556160 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-556160220221
556161 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-556161220221
556162 40-0145 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028557-556162220221
293667 40-0145-S 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028564-293667000000200228
421117 40-0145-S 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028564-421117170906240905
486376 40-0145-S 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028564-486376191203261202
560152 40-0145-S 3.8mm x 100mm Dual-Trak Clavicle Screw 10806378028564-560152220403290402
241411 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-241411000000
251507 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-251507000000
376955 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-376955000000
410385 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-410385170418
425114 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-425114171030
442004 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-442004180524
500984 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-500984200602
501044 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-501044200603
555307 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-555307220210
555416 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-555416220211
557494 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-557494220309
555415 40-0146 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028571-555415220211
293668 40-0146-S 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028588-293668000000200228
405572 40-0146-S 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028588-405572170210240210
486377 40-0146-S 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028588-486377191203261202
534471 40-0146-S 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028588-534471210709280708
566001 40-0146-S 3.8mm x 110mm Dual-Trak Clavicle Screw 10806378028588-566001220615290614
237367 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-237367000000
241412 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-241412000000
251508 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-251508000000
304885 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-304885000000
377048 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-377048000000
381142 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-381142170623
425328 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-425328171025
501045 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-501045200603
525506 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-525506000000
553968 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-553968220131
554240 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-554240220201
554241 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-554241220201
554546 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-554546220203
557542 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-557542220310
557495 40-0147 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028595-557495220309
293669 40-0147-S 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028601-293669000000200228
383917 40-0147-S 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028601-383917160608230608
424386 40-0147-S 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028601-424386171018241017
486378 40-0147-S 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028601-486378191203261202
501117 40-0147-S 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028601-501117200603270603
566002 40-0147-S 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028601-566002220615290614
570288 40-0147-S 3.8mm x 120mm Dual-Trak Clavicle Screw 10806378028601-570288220804290803
|
Recalling Firm/ Manufacturer |
Acumed LLC 5885 NW Cornelius Pass Rd Hillsboro OR 97124-9370
|
For Additional Information Contact |
Ellie Wood 214-738-1709
|
Manufacturer Reason for Recall |
Distribution without Pre-Market Clearance
|
FDA Determined Cause 2 |
No Marketing Application |
Action |
On 11/29-30/2022, Acumed LLC sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via USPS First Class with signature required to customers informing them that the Acutrak Headless Compression Screws and the Acutrak 2 Headless Compression Screw Systems were commercialized with a Pre-Market Clearance (510(k)).
Customer are instructed to:
Product within the scope of this correction does not need to be returned to Acumed. No actions are required as this is a compliance issue and not a patient safety issue. This notification is solely for your awareness. However, physicians are urged to notify Acumed of any adverse events or product safety issues.
Additionally, adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
" Complete and submit the report Online
" Regular Mail or Fax: Download form or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178.
Please complete the survey on Page 8 of this letter to indicate your receipt of this notice and to acknowledge you have read the content of this notice. Return the survey via email to RecallNotification@acumed.net.
For questions - contact Acumed Business Services via email at BusinessServices@acumed.net or via phone at (888) 627-9957 Option 4 (US and Canada) or +001 (503) 627-9957 and choose Option 4 (outside of US and Canada)
|
Quantity in Commerce |
7,498 units |
Distribution |
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, and WY. The countries of AE, AR, AU, BE, BR, CA, CH, CL, CN, CO, CR, DE, DK, EC, ES, FI, FR, GB, GT, HK, IE, IL, IT, JP, KR, KW, MX, MY, PA, PE, PR, PT, QA, SA, SE, SG, TH, TR, TW, and ZA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = HWC and Original Applicant = Acumed LLC
|
|
|
|