| Class 2 Device Recall epoc Host 2 | |
Date Initiated by Firm | December 12, 2022 |
Date Posted | March 03, 2023 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-1214-2023 |
Recall Event ID |
91639 |
510(K)Number | K061597 |
Product Classification |
Electrode measurement, blood-gases (pco2, po2) and blood ph - Product Code CHL
|
Product | epoc Host 2 as follows:
Product SMN
epoc Host 10736387
epoc Host 2, Canada 10736388
epoc Host 2, China 10736389
epoc Host 2, Japan 10736390
epoc Host 2, Refurbished 10736393
epoc Host 2, Canada, Refurbished 10736394
epoc Host 2, Japanese, Refurbished 10736395
epoc Host 2, Chinese, Refurbished 10736433
epoc Host 2, US (MC55X) 11413524
epoc Host 2, ROW (MC55X) 11413528
epoc Host 2, India (MC55X) 11413541
epoc Host 2, China (MC55X) 11413542
epoc Host 2, Japan (MC55X) 11413543
epoc Host 2 (MC55X) - US - Refurbished 11413601
"epoc Host 2 (MC55X) - ROW -Refurbished" 11413602
epoc Host 2 (MC55X) - India - Refurbished 11413603
epoc Host 2 (MC55X) - Japan - Refurbished 11413604 |
Code Information |
Host Software Version /Sensor Configuration: 3.37.3/41.1
Product Description Siemens Material Number (SMN) Unique Device Identification (UDI-DI)
epoc Host 2 10736387 00809708052898;
epoc Host 2, Canada 10736388 00809708069117;
epoc Host 2, China 10736389 00809708097219;
epoc Host 2, Japan 10736390 00809708075576;
epoc Host 2, Refurbished 10736393 00809708053901;
epoc Host 2, Canada, Refurbished 10736394 00809708069124;
epoc Host 2, Japanese, Refurbished 10736395 00809708075583;
epoc Host 2, Chinese, Refurbished 10736433 00809708097226;
epoc Host 2, US (MC55X) 11413524 00630414606460;
epoc Host 2, ROW (MC55X) 11413528 00630414606378;
epoc Host 2, India (MC55X) 11413541 00630414608617;
epoc Host 2, China (MC55X) 11413542 00630414608624;
epoc Host 2, Japan (MC55X) 11413543 00630414608532;
epoc Host 2 (MC55X) - US - Refurbished 11413601 00630414615004;
"epoc Host 2 (MC55X) - ROW -
Refurbished" 11413602 00630414614885;
epoc Host 2 (MC55X) - India - Refurbished 1413603 00630414614762;
epoc Host 2 (MC55X) - Japan - Refurbished 11413604 00630414614892.
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics Inc 2 Edgewater Dr Norwood MA 02062-4637
|
For Additional Information Contact | SAME 781-269-3000 |
Manufacturer Reason for Recall | pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure |
FDA Determined Cause 2 | Under Investigation by firm |
Action | Siemens has issued an Urgent Medical Device Correction POC 23-004.A.US on December 12, 2022 and Urgent Field Safety Notice POC 23-004.A.OUS to customers informing them of the issue and providing workaround for customers to use until the affected sensor configuration/software version expires on June 12, 2023. The next sensor configuration with the fix and its corresponding sw version will be available to customers in April 2023.
Letter states reason for recall, health risk and action to take:
" Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days.
" If you are a distributor, please ensure your customers receive this UMDC letter.
" If you perform QC tests at barometric pressures equal to or above 730 mmHg (97 kPa), no additional action is required.
" If you perform QC tests at barometric pressures less than 730 mmHg (97 kPa), apply barometric correction factors before comparing the gas readings with the published value assignments.
***Important Note*** The above barometric correction factor should be applied only to epoc System Software sensor configuration 41.1, epoc Host 2 SW Version 3.37.3 and epoc NXS Host SW Version 4.10.6, which expires June 12, 2023. Following the expiration of this sensor configuration and the software version refer to the epoc System Manual Section 9.5.1 for the appropriate Barometric Correction factor to apply." Please review this letter with your Medical Director." Please retain this letter with your laboratory records and forward this letter to those who may have received the affected product.
Questions, please contact your Siemens Healthineers Technical Solution Center or your local Siemens Healthineers technical support representative. |
Quantity in Commerce | 22,156 units |
Distribution | Nationwide
Foreign:
Algeria
Argentina
Australia
Austria
Bahrain
Belarus
Belgium
Bolivia
Brazil
Cambodia
Canada
Chile
China
Colombia
Croatia
Czech Republic
Denmark
Ecuador
Egypt
Estonia
Finland
France
Fren.Polynesia
Georgia
Germany
Greece
Guadeloupe
Hong Kong
Hungary
Iceland
India
Indonesia
Iran
Ireland
Israel
Italy
Japan
Jordan
Kazakhstan
Kenya
Kuwait
Laos
Latvia
Lithuania
Luxembourg
Malaysia
Mali
Martinique
Mexico
Myanmar
Netherlands
New Caledonia
New Zealand
Norway
Papua Nw Guinea
Peru
Philippines
Poland
Portugal
Qatar
Republic Korea
Reunion
Romania
Russian Fed.
San Marino
Singapore
Slovakia
Slovenia
Solomon Islands
South Africa
Spain
Sweden
Switzerland
Taiwan
Thailand
Tunisia
Turkey
U.A.E.
Ukraine
United Kingdom
Uruguay
Vietnam
|
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = CHL
|
|
|
|