| | Class 2 Device Recall Spectral CT 7500 |  |
| Date Initiated by Firm | March 17, 2023 |
| Date Posted | May 12, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1584-2023 |
| Recall Event ID |
92064 |
| 510(K)Number | K203020 |
| Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
| Product | Spectral CT 7500 Computed Tomography X-Ray System, Model Number 728333 |
| Code Information |
UDI-DI: 00884838101111;
Serial Numbers: 10058
10074
10080
10093
10096
10097
10098
10099
10100
10102
10103
10104
10105
10106
10107
10109
10110
10111
10114
10115
10117
10119
10123
396001
396003
10014
10015
10016
10017
10018
10019
10020
10021
10022
10023
10024
10025
10027
10031
10032
10033
10034
10035
10036
10038
10039
10040
10041
10042
10043
10044
10045
10046
10047
10048
10050
10051
10052
10053
10055
10056
10057
10059
10060
10061
10062
10063
10065
10066
10067
10068
10069
10070
10075
10076
10077
10081
10082
10085
10086
10087
10089
10090
10092
10094
10095
10108
10116
10120
10049
10121
10124
|
Recalling Firm/ Manufacturer |
Philips North America Llc 222 Jacobs St Cambridge MA 02141-2289
|
Manufacturer Reason for Recall | Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operators foot. If foot or leg entrapment occurs while pressing Unload Pedal, the risk to operators may include: pain, muscle or ligament sprain, abrasion, laceration, contusion, fracture, loss of function, and joint dislocation. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | An URGENT Medical Device Correction notification letter dated 3/17/23 was sent to customers.
Actions that should be taken by the customer / user in order to prevent risks for patients or users
Immediately stop using the Unload Pedal of the Foot Switch until Philips installs a solution on your system.
Instead of using the Unload Pedal, please use the Unload Patient function on the Gantry Control Panels, or use other Table In/Out, Up/Down functions on the Gantry Control Panels, CT Scan Control Box, or Hardware Interventional Controls to fulfill the normal patient unloading function, as instructed in the IFU.
You may continue to use Load Pedal, and Free-Float Pedal of Foot Switch, as they are not
affected by this Urgent Medical Device Correction Letter.
Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken.
Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed.
Actions planned by Philips to correct the problem
Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the solution to resolve this issue (refer to FCO72800795).
Please be assured that maintaining a high level of safety and quality is our highest priority. If you need any further information or support concerning this issue, please contact your local Philips representative:
For North America, contact the Customer Care Solutions Center (1-800-722-9377). |
| Quantity in Commerce | 92 units |
| Distribution | Worldwide - US Nationwide distribution including in the states of States: AZ, IN, MA, MD, MN, NY, OH, PA, and TX. The countries of Australia, Austria, China, Czech Republic, Denmark, France, Germany, Israel, Japan, Malaysia, Netherlands, Norway, Palestine, Panama, South Korea, Spain, Switzerland, Thailand, United Kingdom.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|