| | Class 2 Device Recall Giraffe OmniBed Carestation CS1 |  |
| Date Initiated by Firm | May 19, 2023 |
| Date Posted | July 06, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2100-2023 |
| Recall Event ID |
92454 |
| 510(K)Number | K152814 K213551 |
| Product Classification |
Incubator, neonatal - Product Code FMZ
|
| Product | Giraffe OmniBed Carestation CS1,
Model Numbers:
a. 2082844-001-XXX,
b. 2074920-001-XXXXXX,
c. M1050205,
d. M1135822,
e. M1145475,
f. M1145476,
g. M1235957,
h. M1149051,
i. M1146098;
Labeled as the following: GIRAFFE OMNIBED CARESTATION CS1, GIRAFFE OMNIBED CARESTATION, GIRAFFE OMNIBED, GIRAFFE OMNIBED CS1, OMNI G 115/230V SERVO2 CARESTATION, ASSY-LAU, OMNIBED CHINA EBASE S HUMID, Manufacturing assembly - Make,
OBSOLETE-Omnibed Czech Servo O2 Nist/Scale/Re-usable Probe/SW w/port hole/CE Plug,
ASSY-LAU, Omnibed Czech S-PH w/Nist O2 w/Reus Probe CE Power Cord, Finished good - Make, ASSY-LAU, Omnibed Czech S-PH w/Reusable Probe CE Power Cord, Finished good - Make, Omnibed Czech Servo O2 Nist/Scale/Re-usable Probe/SW w/port hole/Ce Plug, ASSY-LAU, Omnibed Indonesian S-GR W/Reuseable Probe CE Plug, Finished good - Make, ASSY-LAU, Omnibed ENG S-Gr w/REUProbe CE Power Cord, Finished good - Make, ASSY-LAU, Omnibed Portuguese S-PH w/NIST O2w/Reusable Probe CE Power Cord, Finished good - Make.
Incubator, Neonatal, Warmer, Infant radiant |
| Code Information |
a. Model Number 2082844-001-XXX; GTIN Numbers: 010084068211686221, 010084068211686221; All Lot/Serial Numbers.
b. Model Number 2074920-001-XXXXXX; No GTIN Number; Serial Numbers: HDGR52137, HDGL50099, HDGT60164, HDGT60143, HDGT60142, HDGT60139, HDGT60138, HDGT60137, HDGU60753, HDGT60652, HDGU60069, HDGT60141, HDGU60156, HDGU60120, HDGU60119, HDGT60163, HDGT60703, HDGT60245, HDGT60160, HDGT60459, HDGT60246, HDGT60235, HDGU67645, HDGU60065, HDGU60063, HDGU60061, HDGU63122, HDGU63121, HDGU60062, HDGU57534, HDGU67533, HDGU67534, HDGT60460, HDGU67236, HDGT60249, HDGU60201, HDGU60570, HDGU60620, HDGU60023, HDGU60020, HDGT60612, HDGU60022, HDGU60571, HDGU60686, HDGU60202, HDGV60190, HDGW60244, HDGU67677, HDGU67676, HDGU67675, HDGU67280, HDGU67416, HDGU67415, HDGU67382, HDGU67377, HDGU67369, HDGU67366, HDGU67362, HDGU67289, HDGU67284, HDGU67282, HDGU63165, HDGU63167, HDGU63166, HDGV61084, HDGU67436, HDGU67423, HDGU67422, HDGU67421, HDGU67420, HDGU67419, HDGU67418, HDGU67414, HDGU67413, HDGU67381, HDGU67380, HDGU67379, HDGU67378, HDGU67370, HDGU67368, HDGU67367, HDGU67365, HDGU67363, HDGU67361, HDGU67360, HDGU67359, HDGU67358, HDGU67357, HDGU67304, HDGU67303, HDGU67302, HDGU67301, HDGU67300, HDGU67299, HDGU67298, HDGU67297, HDGU67296, HDGU67295, HDGU67294, HDGU67293, HDGU67292, HDGU67291, HDGU67290, HDGU67288, HDGU67287, HDGU67286, HDGU67285, HDGU67283, HDGU67281, HDGJ51159, HDGT60613, HDGT60614, HDGU60700, HDGU60703.
c. Model Number M1050205; No GTIN Number; Serial Numbers: HDGV60148, HDGU60656, HDGU60658, HDGT60481, HDGS50047, HDGV60146.
d. Model Number M1135822; No GTIN Number; Serial Numbers: HDGL51737, HDGL51738, HDGL51734, HDGM50736, HDGM50841, HDGL51735.
e. Model Number M1145475; No GTIN Number; Serial Numbers: HDGM51813, HDGM52120, HDGM52119, HDGN50474, HDGN50472, HDGN50471, HDGN50469, HDGN50466, HDGN50464, HDGN50848, HDGN50847, HDGN50846, HDGN50845, HDGN50844, HDGQ50049, HDGR53203, HDGS51654, HDGS51653, HDGS51652, HDGS51651, HDGS51650, HDGS51649, HDGS51648, HDGS51647, HDGS51646, HDGS51645, HDGM51179, HDGQ50425, HDGN50465, HDGN50849, HDGN50467, HDGN50470, HDGN50475, HDGR52324, HDGR52323, HDGR52322, HDGR52321, HDGN50468.
f. Model Number M1145476; No GTIN Number; Serial Numbers: HDGR52896.
g. Model Number M1235957; No GTIN Number; Serial Numbers: HDGT60222.
h. Model Number M1149051; No GTIN Number; Serial Numbers: HDGR53018, HDGR53085, HDGR52927, HDGR53086, HDGR53021.
i. Model Number M1146098; No GTIN Number; Serial Numbers: HDGS52233, HDGS52234. |
| FEI Number |
3014220748
|
Recalling Firm/ Manufacturer |
DATEX--OHMEDA, INC. 9900 W Innovation Dr Wauwatosa WI 53226-4856
|
| For Additional Information Contact | GE Healthcare Service 800-437-1171 |
Manufacturer Reason for Recall | GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees beginning on 05/19/2023 by mail. The notice explained the problem, risk, and requeested the following:
Actions to be taken by Customer/User:
Inspect the device bedside panels; if the device has the correct latches, continue to use the device. If the device does not have the correct latches, quarantine the device. Inspect any field replaceable bedside panel stock and quarantine for the affected product.
GE Healthcare will replace affected latches. |
| Quantity in Commerce | 4,540 units |
| Distribution | Worldwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FMZ 510(K)s with Product Code = FMZ
|
|
|
|