Class 2 Device Recall Medline Probe Cover Kits Centurion |
 |
Date Initiated by Firm |
May 18, 2023 |
Date Posted |
July 20, 2023 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-2195-2023 |
Recall Event ID |
92466 |
Product Classification |
Catheter introducer kit - Product Code OFD
|
Product |
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CATHETERLESS INSERTION KIT, Model Number CVI4510; b) VANTEX 7FR 3L 16CM CVC BUNDLE, Model Number STCVC01145B; c) VANTEX TRIPLE LUMEN BUNDLE, Model Number STCVC2000;
|
Code Information |
All lots within expiry, distributed from Medline between December 2017 - May 2023:
a) Model Number CVI4510, UDI/DI (EA) 10653160321194, UDI/DI (CS) 00653160321197;
b) Model Number STCVC01145B, UDI/DI (EA) 10653160351924, UDI/DI (CS) 00653160351927;
c) Model Number STCVC2000, UDI/DI (EA) 10653160336792, UDI/DI (CS) 00653160336795 |
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield 3 Lakes Dr Northfield IL 60093-2753
|
For Additional Information Contact |
Karin Johnson 886-359-1704
|
Manufacturer Reason for Recall |
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
|
FDA Determined Cause 2 |
Device Design |
Action |
Medline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form.
Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to the distributor for collection and return to Medline.
Questions: please contact the Recall Department at 866-359-1704 |
Quantity in Commerce |
620 units |
Distribution |
Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia. |