Class 2 Device Recall Custom Sterile and NonSterile Procedural Kits containing Turkuaz Ultrasound Gel |
 |
Date Initiated by Firm |
May 15, 2023 |
Date Posted |
August 03, 2023 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-2315-2023 |
Recall Event ID |
92399 |
Product Classification |
Catheter care tray - Product Code OHR
|
Product |
Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following:
a. CENTRAL CATHETER INSERTION TRAY, Model Number: CVI680C;
b. CVC INSERTION ACCESSORY KIT, Model Number: ECVC7880;
c. CVL INSERTION TRAY, Model Number: DYNDC1786A;
d. ULTASOUND GUIDED PIV INSERTION, Model Number: IVS3635B;
e. USG PIV INSERTION KIT, Model Number: DYNDV2480A; |
Code Information |
All lot numbers for the following models and corresponding UPCs distributed from Medline between 04/01/2020 and 04/28/2023:
a. Model Number: CVI680C, UPC Number: 065316099294.
b. Model Number: ECVC7880, UPC Number: 019532712961.
c. Model Number: DYNDC1786A, UPC Number: 019348919911.
d. Model Number: IVS3635B, UPC Number: 019348936055.
e. Model Number: DYNDV2480A, UPC Number: 019532732721. |
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield 3 Lakes Dr Northfield IL 60093-2753
|
For Additional Information Contact |
Karin Johnson 886-359-1704
|
Manufacturer Reason for Recall |
Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
Medline notified customers beginning on 05/15/2023, via mail and email with letter. Customers with Medline Probe Cover Kits were instructed to check inventory for affected kits, destroy any affected kits, and complete and return the response form. Distributors were instructed to notify their customers and have them document and destroy affected kits. Customers with the non-sterile and sterile procedural kits were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits. Distributors were instructed to notify customers of the recall and have customers return any affected product for return to Medline.
Medline identified additional units of Medline Probe Cover Kits and notified customers on 06/22/2023. Customers were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits. |
Quantity in Commerce |
6,939 kits |
Distribution |
Worldwide distribution - US Nationwide and the countries of Canada, China, Costa Rica, India, Lebanon, Panama, Qatar, Saudi Arabia, Singapore, United Arab Emirates. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|